Use of MMV as a Single Worst-Case Model Virus in Viral Filter Validation Studies Eva Gefroh, Houman Dehghani, Megan McClure, et al.

PDA J Pharm Sci and Tech 2014, 68 297-311 Access the most recent version at doi:10.5731/pdajpst.2014.00978

Downloaded from journal.pda.org on December 14, 2016

TECHNOLOGY/APPLICATION

Use of MMV as a Single Worst-Case Model Virus in Viral Filter Validation Studies EVA GEFROH*, HOUMAN DEHGHANI, MEGAN MCCLURE, LISA CONNELL-CROWLEY, and GANESH VEDANTHAM Purification Process Development, Amgen Inc., Seattle, WA; and Biosafety Development, Amgen Inc., Seattle, WA ©PDA, Inc. 2014 ABSTRACT: Typical platform processes for biopharmaceutical products derived from animal cell lines include a parvovirus filtration unit operation to provide viral safety assurance of the drug product. The industry has adopted this platform unit operation and gained a wider understanding of its performance attributes, leading to the possibility of streamlined approaches to virus clearance validation. Here, the concept of virus validation on a parvovirus-grade filter with a single worst-case model virus is presented. Several lines of evidence, including published literature and Amgen’s own data, support the use of a parvovirus, such as mouse minute virus (MMV), as a worst-case model virus to assess virus removal by parvovirus filters. The evidence presented includes a discussion of the design and manufacture of virus filters with a size exclusion mechanism for removal. Next, the characteristics of different model viruses are compared and a risk assessment on the selection of the relevant model viruses for clearance studies is presented. Finally, a comprehensive summary of literature and Amgen data is provided, comparing the clearance of larger viruses against MMV. Together, this analysis provides a strong scientific rationale for the use of a single, worst-case model virus for assessing virus removal by parvovirus filters, which will ultimately allow for more efficient and streamlined viral clearance study designs. KEYWORDS: Viral filtration, Parvovirus, MMV-only, Viral clearance LAY ABSTRACT: Demonstrating the virus clearance capability of a purification process is an important aspect of biopharmaceutical process development. A key component of the viral safety of the process is the inclusion of a parvovirus-grade filter as an effective and robust virus removal step. Traditional methodologies for viral clearance studies have been based on a conservative, data-intensive approach, but recent trends in the field of virus clearance and process development show evolution towards streamlined and more efficient study designs that are based on understanding the mechanism of viral clearance by downstream unit operations. The publication of scientific datasets and awareness of the underlying mechanisms involved with these unit operations have fueled this trend. Here, the concept of virus validation on a parvovirus-grade filter using a parvovirus as single, worst-case model virus is presented. Multiple lines of evidence are provided to support this proposal, including a review of published literature and Amgen historical data. The adoption of this approach provides benefits in terms of cost savings for executing viral clearance studies, but it also simplifies the necessary dataset and focuses on only supplying value-added information to demonstrate the viral safety of the process.

Introduction Biopharmaceutical products derived from animal cell lines are at risk of contamination from viruses in the cell line or in constituents of the culture media. To

*Corresponding Author: Eva Gefroh, Mail Stop AW2/ D2042, 1201 Amgen Court West, Seattle, WA 98119. Telephone: (206) 265-7439; email: [email protected] doi: 10.5731/pdajpst.2014.00978

Vol. 68, No. 3, May-June 2014

ensure the safety of the product, manufacturing processes include steps specifically designed to remove and/or inactivate viruses. Three complementary approaches to control viral contamination are typically employed as recommended by the U.S. Food and Drug Administration (FDA) and International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) and include (i) selecting and testing cell lines and other raw materials for the absence of viruses, (ii) testing the product at appropriate steps of production for absence 297

Downloaded from journal.pda.org on December 14, 2016

Table I Typical Viruses Used in Viral Clearance Studies Virus Name

Abbrev.

Family

Murine minute virus Murine leukemia virus Pseudorabies virus Reovirus type 3 Poliovirus Porcine parvovirus Simian vacuolating virus 40 Bovine virus diarrhea

MMV MuLV PRV Reo-3 Pol PPV SV40 BVDV

Parvoviridae Retroviridae Herpesviridae Reoviridae Picornaviridae Parvoviridae Polyomaviridae Flaviviridae

of contaminating infectious viruses, and (iii) assessing the capacity of the downstream process to clear viruses (1–3). Downstream processes used for the purification of monoclonal antibodies (mAbs) and other therapeutic proteins employ a series of steps to inactivate and/or remove viruses from process streams (4). Heat, radiation, or chemical treatment (including low pH, solvent/detergent, and chaotropes) can be used for virus inactivation, while virus removal is typically accomplished by chromatography and membrane filtration (5). Although inactivation methods are generally considered robust, these methods may have a limited operating window due to impact on product quality. In addition, these methods are not necessarily effective at inactivating viruses with a broad range of biological, physical, and chemical resistance properties. Removal of virus by chromatography also has specific operating windows for effective clearance and is dependent on the virus chemical properties, type of resin, and operating conditions for the step. Virus filtration offers a robust, size-based virus removal that is easy to use, scalable, and provides high product recovery with little or no impact to product quality (6 – 8). In contrast to other virus clearance methods, virus filtration can be performed at moderate solution conditions, thus minimizing product damage. Its size-based removal mechanism allows for removal of a broad range of viruses with a single filter type, a benefit in terms of meeting current late-phase regulatory expectations. The ability of the purification process to remove or inactivate viruses is demonstrated using model viruses. These are selected to represent a range of characteristics such as size, enveloped vs non-enveloped, 298

Size (nm) Envelope Genome 18–26 80–110 120–200 60–80 22–30 18–26 40–50 50–70

No Yes Yes No No No No Yes

ssDNA ssRNA dsDNA dsRNA ssRNA ssDNA dsDNA ssRNA

Shape

Chem. Resist.

Icosa-hedral Spherical Spherical Spherical Icosa-hedral Icosa-hedral Icosa-hedral Pleo/Spher

V. high Low Med Med Med V. high V. high Low

genome type, physico-chemical resistance, and relevance as models for viruses likely to be present as endogenous or adventitious contaminants in cell culture (1–3). The virus clearance evaluation is performed on individual purification unit operations by spiking a known quantity of model virus into the load material and measuring the amount of virus present in the product pool. These studies are performed using qualified scaled-down models of the process steps. A summary of the typical viruses used in clearance studies for therapeutic proteins is shown in Table I, along with information about the size of the virus, presence or absence of a lipid envelope, genomic composition, shape, and chemical resistance (1, 8, 9). One of the model viruses used in viral clearance studies is xenotropic murine leukemia virus (xMuLV), a model for non-infectious retroviral-like particles (RVLPs) expressed in rodent cells and for large, enveloped RNA viruses. Minute murine virus (MMV) is a member of the parvovirus family of small, non-enveloped DNA viruses and is a known contaminant of rodent production cell cultures (10). MMV is typically included to represent a worst-case adventitious viral agent due to its small size (20 –25 nm) and high resistance to chemical inactivation. An additional panel of two viruses is used for viral clearance studies for licensing applications (1, 3). As a response to observed parvovirus contamination events in biopharmaceutical manufacturing, the general regulatory expectation has evolved to include a step that can robustly and effectively remove these small, non-enveloped viruses. Parvovirus filtration is therefore routinely used in therapeutic protein platform processes at Amgen and by others in the industry (11). As a dedicated and robust virus removal step, PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

Table II Proposed Approach for Use of a Single Worst-Case Model Parvovirus for Parvovirus-Grade Filter Validation Virus Tested Phase of Application

MMV

XMuLV

PRV

Reo-3

Clinical Marketing

Test Test

Use MMV LRV Use MMV LRV

N/A Use MMV LRV

N/A Use MMV LRV

parvovirus filtration routinely provides ⬎4 log 10 clearance of parvoviruses, such as MMV, through sized-based sieving under typical, non-overloaded conditions (12). Parvovirus-retentive filters are manufactured by several vendors and have a complex, multilayer structure to provide mechanical stability and ensure there are no paths for virus to flow through the membrane. This technology has been instrumental in providing the high levels of parvovirus removal needed to increase viral safety for bioprocessing applications. There is continued interest from industry and health authorities to better understand the robustness of unit operations to provide virus clearance, and the mechanism behind the clearance (11). This improved understanding has resulted in efforts to update the integrated strategy for virus clearance studies, with a goal towards streamlining and efficiency. Proposals for modular or generic validation have been presented for unit operations with well-defined and understood mechanisms of action, such as low-pH virus inactivation and anion exchange chromatography (13, 14). One streamlining proposal for the virus filtration unit operation is the use of a single, small model virus for testing virus removal, instead of the current paradigm of testing 2– 4 model viruses of various sizes (11, 12). In this approach, the parvovirus-grade filter is tested using only a parvovirus model virus, for example, MMV, because this filter type is designed to provide robust retention of viruses sized 20 nm and larger. The approach assumes that the clearance validated for the small virus is a conservative reduction factor that can be applied to larger model viruses as well. This concept is illustrated in Table II for both clinical and marketing phase applications. A set of recent meeting reports from the PDA Virus & TSE Safety Forum in 2011 support this approach (15, 16). The summary report from the Workshop on Virus Removal by Filtration states that murine leukemia Vol. 68, No. 3, May-June 2014

virus (MuLV) has not been reported to pass through a parvovirus filter, and therefore it is logical to use a smaller virus to provide a claim for a large virus like MuLV (16). Also, it is noted in PDA Technical Report 41 that this approach may be valid provided that scientific justification can be provided (12). Here, we provide multiple lines of evidence in favor of this worst-case model virus approach. The evidence includes a summary of the design and manufacture of virus filters with a known size exclusion mechanism for removal. Additionally, a discussion of the risk assessment for the selection of the most relevant model virus for clearance studies is presented. Finally, a comprehensive summary of the literature and Amgen data will be presented, showing that the clearance of larger viruses, such as xMuLV, is comparable to or better than a smaller virus, such as MMV. Taken together, these lines of evidence provide a strong scientific rationale supporting the proposal to use a single, worst-case model virus, such as MMV, to assess virus removal by parvovirus filters. Size Exclusion Mechanism of Virus Filter Retention Virus removal filters are designed, developed, and manufactured by vendors to retain particles by size exclusion. While filter products by different vendors can vary widely in materials of construction and format, the retentive capabilities of the membrane are dictated by the complex internal porous structure of the polymeric matrix. A summary of commercially available virus-retentive filters is shown in Table III (8). These membranes are either tangential flow filters (TFFs) or normal flow filters (also known as direct flow filters, DFFs). Membranes can have very fine pore structures and narrow pore size distribution, with particle capture at the membrane surface, as typically found for TFFs. Normal flow filters contain either an open symmetric or asymmetric pore structure that allows for entrapment of the virus particles within the depth of the membrane structure. Multiple membrane 299

Downloaded from journal.pda.org on December 14, 2016

Table III Summary of Commercially Available Virus Filters Filter Vendor Millipore

Filter Brand

Virus Target Classa

Membrane Pore Symmetry

Modeb

Layers

Membrane Chemistry

Viresolve NFR

Retro

DFF

Asymmetric

Triple layer

Hydrophilic polyethersulfone (PES)

Viresolve NFP

Parvo

DFF

Asymmetric

Triple layer

Hydrophilic polyvinylidene fluoride (PVDF)

Viresolve 70

Parvo

TFF

Asymmetric

Single layer

Hydrophilic PVDF

Viresolve 180

Retro

TFF

Asymmetric

Single layer

Hydrophilic PVDF

Viresolve Pro

Parvo

DFF

Asymmetric

Double layer

Hydrophilic PES

Sartorius Stedim Biotech

Virosart CPV

Parvo

DFF

Symmetric

Double layer

Hydrophilic PES

Pall

Ultipor VF grade DV50

Retro

DFF

Symmetric

Triple layer

Hydrophilic acrylate-modified PVDF

Ultipor VF grade DV20

Parvo

DFF

Symmetric

Double layer

Hydrophilic acrylate-modified PVDF

Pegasus grade LV6

Retro

DFF

Symmetric

Double layer

Hydrophilic acrylate-modified PVDF

Pegasus grade SV4

Parvo

DFF

Symmetric

Double layer

Hydrophilic acrylate-modified PVDF

Planova 35N

Retro

DFF/TFFc

Asymmetric

Hollow fiber

Hydrophilic cuprammonium regenerated cellulose

Planova 20N

Parvo

DFF/TFF

Asymmetric

Hollow fiber

Hydrophilic cuprammonium regenerated cellulose

Planova 15N

Parvo

DFF/TFF

Asymmetric

Hollow fiber

Hydrophilic cuprammonium regenerated cellulose

Planova BioEX

Parvo

DFF/TFF

Asymmetric

Hollow fiber

Modified PVDF

Asahi Kasei

a

“Target virus class” can be a subjective concept based on the application and product type (i.e. biotech vs plasma product). Some large virus retentive filters can retain viruses smaller than retroviruses. Consultation with filter vendors is warranted if clearance of a particular virus is a process requirement. b DFF, direct flow filtration; TFF, tangential flow filtration. c Can be configured by end users to run in either mode. This material is reproduced with permission from John Wiley & Sons, Inc. Miesegaes, G.; Lute, S.; Aranha, H.; Brorson, K. Virus Retentive Filters. In Encyclopedia of Industrial Biotechnology, Flickinger, M. C. Ed.; John Wiley & Sons, Inc.: New York, 2009.

layers are often incorporated to boost overall log reduction factors by providing additive clearance across the layers. Examples of virus filter pore structure via scanning electron microscope (SEM) imaging are shown in Figure 1, illustrating the complex internal pore structure that provides a balance between high flux and passage for the product of interest and high retention of viruses (17). An important design criterion for virus filters is the minimization of non-specific

adsorption to the membrane surface by either using a material of construction that is inherently hydrophilic in nature, or by coating the surface with a hydrophilic polymer. This serves to minimize product loss and fouling on the membrane surface, and thus results in virus particle retention characteristics that are dominated by a size-based mechanism. The membrane chemistry and hydrophilic modifications are summarized for the different virus filters in Table III.

Figure 1 SEM cross-sectional images of Ultipor威 DV20 (left) and Viresolve威 Pro (right) membranes. Reprinted from reference 17 with permission of the publisher. 300

PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

In the design of filters with more open pore structures, as typified by the standard normal flow filters used in therapeutic protein processes, the pore size distribution and pore frequency contribute to the overall filter performance and the resultant virus retention. Therefore, it is important for filter manufacturers to control and test for the actual pore size distribution of each membrane lot. With the use of advanced analytical techniques, filter manufacturers provide a high degree of quality assurance of their virus filter product. Examples of non-destructive tests that evaluate the pore size of a filter include liquid–liquid porosimetry (LLP) (18, 19), forward flow tests (e.g. air-water diffusion) (20), and a more recently developed binary gas diffusion test (21). Destructive tests that challenge the filter with a model particle in solution, and thus provide information on the retentive pore size of the membrane, include the gold particle test (22), dextran retention test (23), and bacteriophage challenge test (24). The advantage of non-destructive tests is that the filter can be 100% pre-use tested for every membrane/ device produced, whereas destructive tests can only be performed on a selective lot sampling or post-use basis. Ultimately, filter manufacturers must provide evidence that their integrity test methods and the associated acceptance criteria correlate to retention of the targeted size range of virus at the desired reduction value (12). An example is provided for the development of a correlation between pore size testing and virus particle retention. LLP involves the intrusion of one immiscible fluid into another through the pores of the membrane at a defined transmembrane pressure. Because liquid flow through the membrane pores is proportional to the fourth power of the pore diameter (18), the overall membrane performance may be dictated by a small number of the largest pores. LLP provides an independent physical measure of flow through these largest pores, which can then be correlated to virus retention as measured by log reduction value (LRV). Figure 2 shows an example of the correlation between LLP (or LLDP, liquid–liquid displacement porosimetry) and PP7 (bacteriophage) LRV for Sartorius polyethersulfone (PES) research samples (19). For each membrane sample, the pore size distribution is measured via LLP, and the mean (rp,mean), max (rp,max), and min (rp,min) pore sizes are determined. These same samples are also tested for PP7 retention. In this figure, the maximum pores rp,max in the pore size distribution show the best correlation to PP7 LRV; the trend is as expected, with increasing LRV at smaller Vol. 68, No. 3, May-June 2014

Figure 2 Correlation between LLDP (liquid–liquid displacement porosimetry) and bacteriophage PP7 LRV for Sartorius PES filter. Each membrane sample is tested for pore size distribution by LLDP and PP7 retention, and the resulting PP7 LRV is plotted against the mean (black circles), maximum (gray circles), and minimum (white circles) pore size. Reprinted from reference 19 with permission of the publisher.

pore sizes. The lower coefficient of determination (r2 ⫽ 0.697) seen in this study is likely due to experimental variability in the bacteriophage assay. An additional correlation is later shown between the maximum pore size rp,max and the 90% molecular weight cut-off value for dextran molecules; this later correlation shows a higher coefficient of determination (r2 ⫽ 0.866). Other references also demonstrate the ability to develop a correlation between virus particle retention and pore size testing: Phillips and DiLeo demonstrate a correlation for the Viresolve威 70 and Viresolve威 180 membranes using a LLP test (18), and Giglia and Krisnan demonstrate a correlation for the Viresolve威 Pro membrane using a binary gas test (21). The ability to correlate the retention characteristics of a virus to the pore size of the membrane provides evidence that the predominant mechanism for virus removal is size-based sieving. A paper by Grant and Liu (25) provides some fundamental insight into the particle removal efficiency of membranes in which size-based sieving is the dominant particle capture mechanism. The authors describe a methodology to test the particle capture efficiency of a microfilter using submicron latex particles with well301

Downloaded from journal.pda.org on December 14, 2016

Figure 3 Log reduction value (LRV) as a function of virus size for Viresolve威 70 membranes in buffer or protein solution. The retention coefficient in buffer (closed circles) is compared with that in the presence of protein (open diamonds) as a function of Stokes diameter. Rauscher MuLV was suspended in 10% FCS while the other viruses were in 0.65 mg/mL HSA. Reprinted from reference 26 with permission of the publisher. defined sizes. Their results show that particle capture is a strong function of particle diameter, as well as filter media thickness and filter loading. As expected, particle capture efficiency increases with increasing particle diameter. A sieving model was developed to predict the particle capture efficiency as a function of these variables; this model predicts the concentration of particles exiting the filter matrix relative to the concentration entering the filter. Their analysis showed that the model was able to accurately predict the same trends as the experimental results, thus demonstrating that the predominant mechanism of size-based sieving is understood. There is direct evidence in literature showing the correlation of mammalian virus retention factor to the size of the virus. In a paper by DiLeo et al (26), mammalian viruses Polio Type-1 (32.6 nm), SV-40 (47.9 nm), Sindbis (54.1 nm), Reovirus Type-3 (77.6 nm), and MuLV (85 nm), as well as phage viruses PhiX-174 (28 nm) and Phi-6 (78 nm), were applied to a virus filter membrane, Viresolve威 70 (nominal 70 kDa pore size). While the membrane used in this study was operated in tangential flow mode, the results are still applicable towards the current industry standard normal flow filters, as both filter types operate under a predominantly size-based mechanism of removal. As shown in Figure 3, the virus retention in buffer and protein solution increases constantly with virus diameter, independent of virus class or type, from about 3.5 log10 with polio to greater than 6.8 log10 with MuLV. 302

The authors describe further experiments to demonstrate the absence of adsorption effects through mass balance and filter soaking experiments. Additionally, the trend of monotonically increasing virus LRV with increasing pore size is well predicted by a hard-particle sieving model (solid lines in Figure 3). Altogether, these results support that the mechanism of retention is predominantly size-based sieving. Implicit in this mechanism is the understanding that the retention factor for a larger virus, such as MuLV, should be greater than that of a smaller virus, such as MMV; this is well supported by the experimental data. The arguments presented thus far support the assumed dominant mechanism of size-based sieving: manufacturing of filters designed to capture virus particles within the complex pore structure, minimizing nonspecific adsorption by use of hydrophilic polymers or a hydrophilic base matrix, performing pore-size integrity tests that show correlation to particle retention, and demonstrating the trend of increased virus retention as a function of particle diameter. Other mechanisms such as interception, impaction, diffusive capture, gravitational settling, electrostatic attraction or repulsion, and other non-specific adsorption, may also be present (27); however, the extent of impact of these other mechanisms is likely small given the body of evidence. The proposal of using a single, small virus for parvovirus filtration studies shows some parallels to the current recommendation for a filter consensus rating method for small virus filters. As described in Lute et al (24), a task force was formed to develop a common nomenclature and standardized test method for the classification and identification of virus filters. The final recommendation was a rating method based on a single model particle challenge, bacteriophage PP7 (26 nm), to define the retention capabilities of the virus filter. In another parallel, sterile filters are used in biopharmaceutical processing to produce a sterile product solution through removal of bacteria and other microorganisms. The design of these microfilters is similar to viral filters in that the predominant mechanism for particle removal is through size-based sieving. A sterilizing-grade filter is defined as one that produces a sterile filtrate when challenged by 107 colony-forming units per square centimeter of filter area of Brevundimonas diminuta (previously classified as Pseudomonas diminuta) (28). A standard test procedure for PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

performing sterilization validation is outlined under ASTM F838-83 (now F838-05) (29). That procedure recommends the use of a single test microorganism because of its small size, and to produce a consistent standard across the industry. Selection of Model Viruses for Viral Clearance Studies The choice of viruses used for viral clearance studies is based on the cell substrate used for production and the potential risk of introduction of adventitious viruses. Relevant viruses that are known contaminants, and model viruses with a diverse set of physicochemical characteristics are used to evaluate the viral clearance capacity of purification process unit operations. General considerations include the nature of endogenous viruses produced by the cell substrate as well as the susceptibility of the cell substrate to adventitious agents that may be introduced through raw materials (1). Chinese hamster ovary (CHO) cells are widely used cell substrates for production of recombinant protein therapeutics (30). Compared to other less commonly used production cells, such as baby hamster kidney (BHK) cells, CHO cells are less susceptible to infection by viruses based on literature surveys (31). Although viral contaminations in cell culture processes are rare, a few instances have been reported (10, 32–35). In addition to MMV, contamination events in CHO cell culture processes with Cache Valley virus (CVV), Vesivirus 2117, and Reo-3 virus have been reported. Elimination of primary animal-derived raw materials, such as fetal bovine serum and porcine-derived trypsin, and use of chemically defined media for cell culture processes, has significantly reduced the risk of inadvertent introduction of adventitious viral contaminants. However, even in the absence of any primary animal-derived raw material used in cell culture processes, certain viruses such as murine parvoviruses still pose a risk due to their resistance to inactivation and ability to survive in the environment (36). Most of the reported contamination events are hypothesized to have been associated with use of contaminated animal-derived raw materials. However, Moody et al. attributed an MMV contamination of CHO cell culture process to a recombinantly derived raw material (36). Therefore, for mAbs produced in CHO cells without the use of primary animalderived raw materials, MMV is a worst-case relevant virus because it is a small, non-enveloped virus (18 –26 nm), is highly resistant to inactivation, can be a contaminant of non-animal-derived raw materials, Vol. 68, No. 3, May-June 2014

and readily replicates in CHO cells. Contamination events with smaller (⬃17 nm) porcine circovirus (PCV) have been reported (37). However, these events were associated with the use of contaminated porcinederived trypsin in a Vero cell culture system that was capable of supporting PCV replication. Porcine-derived trypsin is not used in the production process for CHO-derived protein therapeutics, and CHO cells have been shown to be not permissive for PCV infection (Amgen unpublished data). Virus morphology is another important consideration in the selection of a worst-case model virus. Most virus particles are spherical in nature. However, there are some viral families that include representative members that are filamentous and are rod- or kidneyshaped. These include filoviridae, rhabdoviridae, and toroviruses in the coronaviridae family. Although filoviridae are rod-shaped and variable in length, they have diameter that is approximately 80 –100 nm. The filoviridae family of viruses contains a single genus Filovirus. The members of this family cause severe hemorrhagic fevers in humans and non-human primates. Rhabdoviridae is a large family of RNA viruses that are also rod-shaped with variable length (100 – 430 nm). The members of this family are more uniform in diameter and range between 45 and 100 nm. Torovirus is a genus of Coronaviriae and is pleomorphic. They may be rod- or kidney-shaped and the diameter of the viral particles ranges from 120 –140 nm (38). Although the morphology of these viruses are different in that they are not spherical, they are generally larger in diameter than parvoviruses and represent viruses with negligible risk with respect to being a contaminant in chemically defined media used in cell culture processes. The quality attributes of the virus stock preparation, such as titer and purity, are also important considerations in the selection of virus used in the spiking study (39). Regulatory guideline document ICH Q5A recommends the use of high-titer stocks to demonstrate adequate viral clearance (1). High-purity stocks are also desirable in virus spiking studies so that the scale-down model system can more accurately represent the actual manufacturing process. This is especially relevant for the viral filtration unit operation for two reasons. First, in the potential event of an actual viral contamination, any virus particles that would reach the viral filter would be of relatively high purity, given the location of the unit operation further downstream in the purification process. Second, there are a 303

Downloaded from journal.pda.org on December 14, 2016

number of published studies that provide evidence for the negative impact of virus spike impurities on filter performance (11, 39 – 42). A review of current literature shows MMV stock preparations to be high in purity and titer (9, 16, 39, 40). It is generally recognized that MMV can be grown to a higher titer and purified more readily as compared to xMuLV (16); further evidence for this assertion is provided in a survey of virus stock preparations from various contract laboratories showing that MMV preparations exhibit relatively low levels of protein impurity and host-cell DNA impurity at high infectious titer (9). Recent work on optimization of MMV stock preparations suggest that virus stock titers and purity can be even further enhanced with modifications to the virus production process (43, 44). Thus, the selection of MMV as a single worst-case model virus is well suited to not only provide high titers to demonstrate the maximum retentive capability of the virus filter, but also high purity so as to minimize non-representative fouling of the filter during challenge studies. Porcine parvovirus (PPV) is another model parvovirus of similar size to MMV and is also commonly used in viral filter challenge studies. In the same survey of virus stock preparations, PPV was similarly observed to exhibit low levels of host-cell impurities at high infectious titer (9). An FDA meta-analysis of viral filtration records from regulatory submissions compared LRVs between MMV and PPV; the mean clearance values were found to be quite similar (MMV 4.6, n ⫽ 99; PPV 4.9, n ⫽ 27; p ⫽ 0.2) (11). While the use of MMV as a single model virus has its merits based on the points described previously, the data reviewed here suggests that PPV could be an acceptable alternative.

discuss viral clearance data in an open forum. The overarching goals of the forum were to consolidate understanding around well understood clearance mechanisms, and to identify areas of improvement for current industry and regulatory practices. In the summary of the proceedings (11), representatives from both FDA and Paul Ehrlich Institut (PEI) regulatory agencies offered their perspectives on virus removal by small virus filters based on a thorough review of their independently generated virus clearance databases. They both concluded that parvoviruses (both MMV and PPV) are effectively removed by parvovirus filters, albeit with variable reduction values, while clearance of much larger retroviruses is consistent and robust. PEI reported that breakthrough of retroviral particles was never observed when an infectivity assay was used (in 74 records), and FDA reported that 95% of their records (in 132 records) demonstrated complete clearance of MuLV. In the small number of MuLV records that showed incomplete clearance, it was noted that substantial removal was seen (over 5 log10) and that these results were likely due to the limitations of retrospective meta-analysis rather than the filter technology itself.

Datasets Comparing Clearance of Different Model Viruses

A couple of datasets were shown for multiple virus filter types that demonstrated comparable virus removal capacity values between MMV and xMuLV. The dataset presented by Genentech on the Viresolve威 Pro filter with proprietary mAbs showed LRVs of ⱖ4.2 for MMV and ⱖ4.8 for xMuLV. Pfizer presented datasets on the Pall DV20, Sartorius CPV, and Planova™ 20N filters for multiple mAbs. The average LRV for MMV was shown to be ⬎4 – 6 log10 and ⬎5–7 log10 for xMuLV. It is worth noting that for these datasets with no virus detected in the permeate, the final achieved LRV number is a function of virus stock titer, spike percentage used in the study, sample dilution requirements due to assay interference/cytotoxicity, and sample volume, all of which impact assay sensitivity. Thus, the values reflect the limit of assay sensitivity rather than the maximum

The evidence presented thus far has focused on the broader aspects of the size-based removal mechanism for virus filters as a class, as well as discussion of the rationale for choosing MMV as a worst-case virus to assess parvovirus filter clearance. This section summarizes datasets for currently available commercial parvovirus filters, and compares the removal efficiency of smaller viruses, such as MMV, to that of larger viruses. Summary of the Proceedings of the 2009 Viral Clearance Symposium A symposium was held in 2009 to allow industry and regulatory representatives to interactively share and 304

A summary was also presented on industry participants’ experience with large virus removal on small virus–retentive filters. Over 100 studies confirmed the capability of these filters to effectively remove retroviruses; this was accepted at the conference proceedings to mean that no virus breakthrough was observed for these studies. The paper proposed that a modular approach to validation of retrovirus removal may be possible for small virus filters.

PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

Figure 4 Summary of Amgen Planova™ 20N virus clearance datasets with MMV and xMuLV. LRV data is shown for 23 mAb processes; some datasets were performed with duplicate runs. Plus signs represent xMuLV LRV with no virus detected, closed and open circles represent MMV LRV with and without virus detected in the permeate, respectively.

retention capability of the filter. Given the similar reported numbers for xMuLV in these datasets as compared to MMV, these datasets support the use of claiming MMV reduction values for a larger virus, such as xMuLV. Amgen Planova™ 20N Dataset A review of Amgen’s historical Planova™ 20N (Asahi Kasei Medical) validation data was performed for viral clearance studies between 2003 and 2011, and includes results from 23 mAb processes (Figure 4). The data include immunoglobulin G1 (IgG1) and IgG2 antibodies, and represent both early-stage and latestage clinical process development with a variety of process and solution conditions. Buffer salts include sodium acetate, sodium phosphate, sodium citrate, sodium sulfate, sodium chloride, and Tris, with the pH of solution ranging from pH 5.0 to 7.5. Filter loadings

were generally around 200 L/m2, but ranged as high as 400 L/m2. Virus spiking was performed with MMV and xMuLV, with either in-house or vendor-sourced virus preparations. In a majority of the cases (39 of 41 runs), MMV was not detected in the filtrate for the chosen process conditions, and for all of the runs, xMuLV was not detected in any of the filtrate samples. No difference in virus retention performance was observed between the sources of virus stock, suggesting that the quality attributes of the virus preparation did not have a large role in determining achievable LRV on the virus filter. Consistent with the data shown by Genetech and Pfizer, these results show that LRV achieved with xMuLV are comparable to MMV, with differences in LRV attributed to virus stock titer and dilution required to eliminate cytotoxicity of the test article in the virus assay. Given the size difference of the viruses, it is generally expected that a greater clearance value should be attainable for xMuLV; however, the achievable LRV is often more limited for xMuLV due to its lower virus stock titers and lower percentage spike used to avoid filter plugging from the virus spike impurities. An example of this scenario is illustrated in Table IV, in which the LRV for the larger xMuLV is demonstrated to be lower than the smaller MMV due to the items listed above and including a higher assay dilution factor due to indicator cell cytotoxicity. In both cases, no virus is detected in the filtrate sample (output titer), but the experimental conditions dictate the value reported for the assay sensitivity limit. Taken together, these results present a compelling case for applying the LRV obtained with MMV, the smallest of the panel of model viruses used for filter validation, to claim LRV across this unit operation for larger viruses such as xMuLV. Amgen Viresolve娂 Pro Dataset An example of the viral clearance capability of the Viresolve威 Pro filter is presented with an in-house viral clearance study for mAb A. Runs were performed

Table IV Example of Experimental Conditions Resulting in Lower LRV for xMuLV as Compared with MMV Example

Load Titer Log10TCID50/mL

Spike %

Input Titer Log10TCID50/mL

Tox Dil

Output Titera Log10TCID50/mL

LRV

MMV 7.50 1.0 5.50 10 ⱕ⫺0.12 ⱖ5.62 xMuLV 7.00 0.5 4.70 30 ⱕ0.36 ⱖ4.34 a Calculations assume a 40 mL large-volume sample and output titer is based on the probability of detecting viruses at low concentrations using the Poisson distribution (1). Vol. 68, No. 3, May-June 2014

305

Downloaded from journal.pda.org on December 14, 2016

Table V Summary of LRV and Percentage Flux Decay Data for mAb A with Viresolve Pro and mAb B with Viresolve NFP Run 1

Molecule/Filter mAb A/Viresolve Pro

Virus

Log10 Reduction Factor

MMV Reo-3 xMuLV PRV mAb B/Viresolve NFP MMV Reo-3 xMuLV PRV a These studies were performed in triplicate. Run 3

Flux Decay (%)

Log10 Reduction Factor

ⱖ4.77 66 ⱖ4.68 ⱖ5.21 92 ⱖ5.38 ⱖ3.72 60 ⱖ4.33 ⱖ3.72 50 ⱖ3.56 4.07 80 4.28 ⱖ4.62a 79 ⱖ4.62a ⱖ4.11 57 ⱖ3.99 ⱖ4.23 64 ⱖ4.48 also showed LRV ⱖ4.62 with 80% flux decay.

in duplicate in 50 mM acetate at pH 5.0 with a panel of four model viruses: MMV, xMuLV, PRV, and Reo-3. The range of throughputs achieved in the study was 904 to 1042 L/m2. The flow decay observed for a 0.1% spike of PRV was around 50 – 60%, for a 0.1% spike of Reo-3 was around 90%, for a 0.1% MMV was as high as 88%, and for a 0.05% spike of xMuLV was around 50 – 60%. The achieved LRVs are shown in Table V. No virus was detected in the pooled permeate samples for all the virus-spiked runs, despite the high level of flow decay observed for some of the runs. The lower LRV numbers for the larger viruses, xMuLV and PRV, as compared with the smaller MMV are reflective of the limitations of the study design due to the chosen spike percentage and starting stock titers. It is believed that the reduction factor for the larger viruses should be theoretically higher than MMV; however, these results further illustrate that the full capabilities of virus retention for larger viruses are often challenging to demonstrate. Amgen Viresolve威 NFP Dataset An example of the viral clearance capability of the Viresolve威 NFP filter is presented with an in-house viral clearance study for mAb B. Runs were performed using the model viruses MMV, xMuLV, PRV, and Reo-3. Virus retention on the NFP viral filter is known to be affected by flux decay, which is defined by the final flow rate normalized to the initial flow rate. As described by Bolton et al (42), flux decay on the filter indicates pore plugging, and the progressive plugging 306

Run 2 Flux Decay (%) 88 91 50 59 80 80 57 57

of the smaller pores of the viral filter leads to a proportional shift in flow through the larger pores, which results in higher virus breakthrough in the permeate. The effect of flux decay on the filter was studied for both the MMV and Reo-3 viruses for this process. Virus spiking studies were performed with MMV in duplicate and Reo-3 in triplicate in which grab samples of the permeate were tested at specific flux decay points (Figure 5). For MMV, virus was observed in most of the permeate grab samples, with increasing virus breakthrough as flux decay increased. This is consistent with the findings reported in Bolton et al (42). The resulting pool also showed MMV breakthrough with LRV of 4.07 and 4.28 out to 80% flux decay for the duplicate runs. In contrast, no Reo-3 was detected in any of the grab or pool samples even after extensive flux decay (up to 80%), with a final pool LRV of ⱖ4.62. The Reo-3 and MMV datasets further provide supporting evidence for the size-based sieving mechanism because the smaller MMV (18 –26 nm) results in higher virus breakthrough as a result of the selective plugging of the smaller membrane pores, whereas the larger Reo-3 virus (60 – 80 nm) does not break through. Viral clearance for xMuLV (80 –110 nm) and PRV (120 –200 nm) was also assessed, and the data is summarized in Table V. As expected, these larger viruses are fully retained by the NFP filter, although as discussed previously, the lower values for the LRV are PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

Figure 5 Grab or pool sample LRV for MMV (in duplicate) and Reo-3 (in triplicate with identical results) as a function of flux decay on the Viresolve威 NFP filter. Open circles indicate no virus detected, closed circles indicate virus detected. Three runs performed with Reo-3 (triangles) showed identical results with no virus detected.

ecules and two additional brands of parvovirus filters, Viresolve威 Pro and Virosart威 CPV, was presented with LRV data for MMV/PPV, Reo-3, xMuLV, and PRV. Both filter types achieved full retention of the three larger model viruses. It is worth noting that these latter studies showed parvovirus clearance mostly above 6 LRV, whereas the prior studies on a different filter type obtained lower clearance factors. As the authors note, when parvovirus breakthrough is detected, the demonstrated clearance factor could grossly underestimate the clearance capacity of larger viruses. For this reason, the choice of whether to pursue a worst-case model parvovirus approach or a modular retrovirus database approach will depend on the parvovirus clearance that can be demonstrated, and whether that clearance factor is sufficiently high to claim for larger viruses. This could depend on the filter type used, as noted in the examples provided here, or on process conditions resulting in selective plugging of the smaller membrane pores, as noted in the previous Viresolve威 NFP example. Conclusions and Recommendations

a function of the spike and assay conditions. These two datasets shown for the panel of four viruses demonstrate that a virus challenge with MMV is capable of achieving at least 4 log10 LRV for effective, robust clearance, and that this clearance value is similar to that which can be practically achieved for the larger viruses. Review of Eli Lilly Dataset A recent publication from Eli Lilly (45) presents their approach to achieving modular retrovirus clearance for a parvovirus filter. Their modular database for the Planova™ 20N filter includes xMuLV and PPV LRV results for nine different protein molecules at a range of protein concentrations, pH, and buffer salt solutions. PPV breakthrough was detected in the majority of the runs (LRV ranging from 2.86 to 7.15), whereas xMuLV was not detected in any of the runs (LRV ranging from ⱖ4.67 to ⱖ6.93). While the differential retention behavior between the two sizes of virus supports the size-based sieving mechanism of the filter, the researchers confirmed this outcome in a follow-on study in which the two viruses were co-spiked into the feed solution. This study with six different protein molecules tested in duplicate verifies directly and unequivocally that the Planova™ 20N filter shows complete retention of retrovirus even when parvovirus breakthrough occurs. A further dataset with four molVol. 68, No. 3, May-June 2014

This paper has described the concept of virus validation on a parvovirus-grade filter with a single worstcase model virus and presented multiple supporting arguments in favor of this approach, using both scientific literature and Amgen’s own data. There is a general consensus in the literature that virus filters provide virus removal predominantly through a sizebased sieving mechanism. In the design and production of these filters, non-specific adsorption to the membrane surface is minimized, and the pore size distribution of the membranes is controlled and tested to show correlation to virus retention. Examples were provided from the literature to show a direct correlation between membrane pore size, virus size, and virus retention. Furthermore, the practice of defining the retention capability of a filter with a single model test organism has been well established in the area of sterile filter validation. Because larger viruses are expected to be retained more effectively than smaller viruses, the reduction value obtained for the smallest model virus is a conservative estimate for that of larger viruses. The choice of a parvovirus, MMV, as the worst-case model virus is well supported through literature assessment of virus contamination events applicable to CHO-based cell culture processes with chemically defined media and no animal-derived raw materials. 307

Downloaded from journal.pda.org on December 14, 2016

MMV is a worst-case relevant virus because it is small (18 –26 nm), non-enveloped, highly resistant to physico-chemical inactivation, a known contaminant of non-animal-derived raw materials, and readily replicates in CHO cells. In addition, MMV stock preparations offer the additional benefits of high titer and high purity, resulting in virus spiking studies that better simulate representative manufacturing conditions and demonstrate high log reduction factors. To illustrate the retention capability of different parvovirus-grade filters for a range of model viruses in comparison with MMV, datasets were summarized from the 2009 Viral Clearance Symposium white paper, as well as Amgen data with the Planova™ 20N, Viresolve威 Pro, and Viresolve威 NFP filters. In these studies, the parvovirus filters consistently showed full retention of the larger viruses, with no virus detected in the permeate pool. In general, MMV was also fully retained, except in the case of the NFP filter, which exhibited virus breakthrough as a function of flux decay. This result is expected based on previous literature, and in fact it provides further evidence for the size-based sieving mechanism, as this small model virus exhibits virus breakthrough as a result of the selective plugging of the smaller membrane pores. In contrast, no breakthrough was observed for larger viruses such as Reo-3, xMuLV, and PRV. This differential virus retention behavior was directly demonstrated by researchers at Eli Lilly in virus co-spiking experiments, which showed that retrovirus is fully retained on the virus filter even when parvovirus is exhibiting breakthrough. While these examples bolster the argument for a predominant size-based sieving mechanism, the occurrence of parvovirus breakthrough, and thus lower achieved clearance values on the filter, underscores the need to carefully consider the case for when a single worst-case parvovirus challenge approach can be used. Given that the parvovirus clearance value will be used as a conservative estimate for clearance of larger viruses, it may result in an underestimation of actual clearance capacity for the larger viruses. Higher clearance factors may be necessary for certain large viruses, such as retroviruses, in order to meet an overall clearance target. It is recommended to proceed with this streamlined approach only when the expected parvovirus clearance will provide the desired level to claim for larger viruses. This approach of using a single worst-case model virus has been discussed in a number of recent scientific forums, and published reports from the 2009 Viral 308

Clearance Symposium and the 2011 PDA Virus & TSE Safety Forum Workshop on Virus Removal by Filtration indicate a gaining acceptance (11, 16). Certain regulatory authorities are accepting clinical trial applications with this single worst-case model virus approach (46). Amgen has successfully filed clinical trial applications with this approach. Amgen has also used this approach in support of a parvovirus filter addition to the purification processes of two globally marketed products as a post-approval change. Specifically, the viral clearance data obtained for MMV from the virus filter step was approved as an acceptable surrogate value for the determination of the virus clearance capabilities of the entire purification process for the larger model viruses. Justification was given that for the size-based mechanism of clearance, the larger virus is expected to equal or exceed the level demonstrated for MMV, and therefore use of the MMV value is considered a conservative estimate. These examples of broad acceptance of this scientific justification by regulatory reviewers suggests that use of a single worst-case model parvovirus with accompanying justification based on the mechanism of action could be used to support marketing applications as appropriate. The guideline document ICH Q5A (1), which provides guidance for marketing applications and was published in 1998, suggests that the viral clearance robustness of a step is demonstrated through use of viruses exhibiting a range of biochemical and biophysical properties. The document also describes a need to justify the choice of viruses used in the studies, and that this choice is influenced by the production process. If the purpose of testing multiple viruses is to characterize the robustness of the process, given that the mechanism of removal for these parvovirus filters is known and understood to be size-based, then testing of viruses larger than the filter size rating would not provide additional robustness information about the process step. This is in line with the guideline’s allowance that process step validation testing that does not provide value-added robustness information about the process need not be performed. Given the body of scientific evidence that now exists in support of the size-based mechanism of removal for parvovirusgrade filters, updating the guidelines based on current knowledge and practice should be considered. A recent FDA presentation suggests that “integration of data from multiple sources within the scientific community supporting mechanism of action and reliability of clearance obtained could lead to changes in approach” (47). It is the authors’ desire that the results presented herein are one step in that direction. PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

In summary, this paper provides a strong scientific rationale for using a single worst-case model virus, MMV, for parvovirus filter virus clearance studies. The data were presented in a comprehensive manner to comment on filter design and the size-based sieving mechanism, and on supporting evidence from in-house and external datasets of the retentive capabilities of these filters. This new level of understanding for the underlying viral clearance mechanism should support the adoption of this new approach, which will ultimately allow for more efficient and streamlined viral clearance studies. Acknowledgements The authors would like to acknowledge Suresh Vunnum for his contributions to the introduction, Carol Krantz for review of the manuscript, and Herb Lutz for literature review of viral filter size-exclusion mechanism. Declarations The contents of this publication do not represent an Amgen endorsement of any products described herein and are not meant to imply that Amgen uses any of these products for clinical or commercial manufacturing. The authors declare that they have no competing interests related to this manuscript, financial or otherwise. References 1. FDA; ICH. Guidance for Industry. Q5A Viral Safety Evaluation of Biotechnology Products Derived from Cell Lines of Human or Animal Origin. U.S. Department of Health and Human Services, FDA, Center for Drug Evaluation and Research, Center for Biologiscs Evaluation and Research (CBER): Rockville, MD, 1998.

man Use. U.S. Department of Health and Human Services, FDA, CBER: Rockville, MD, 1997. 4. Zhou, J. X.; Solamo, F.; Hong, T.; Shearer, M.; Tressel, T. Viral clearance using disposable systems in monoclonal antibody commercial downstream processing. Biotechnol. Bioeng. 2008, 100 (3), 488 – 496. 5. Kundu, A.; Reindel, K. Evaluation of Viral Clearance in Purification Processes. In Process Scale Bioseparations for the Biopharmaceutical Industry, Shukla, A. A., Etzel, M. R., Gadam, S, Eds.; CRC Press: Boca Raton, FL, 2006; pp 419 – 448. 6. Carter, J.; Lutz, H. An overview of viral filtration in biopharmaceutical manufacturing. Eur. J. Parenter. Sci. 2002, 7(3), 72–78. 7. Levy, R. V.; Phillips, M. W.; Lutz, H. Filtration and the Removal of Viruses from Biopharmaceuticals. In Filtration in the Biopharmaceutical Industry, Meltzer, T. H., Jornitz, M. W., Eds.; Dekker: New York, 1998; pp 619 – 646. 8. Miesegaes, G.; Lute, S.; Aranha, H.; Brorson, K.; Flickinger, M. C. Virus Retentive Filters. In Encyclopedia of Industrial Biotechnology, Flickinger, M. C., Ed.; John Wiley & Sons, Inc.: New York, 2009. 9. Miesegaes, G.; Lute, S.; Dement-Brown, J.; Kaushal, S.; Strauss, D.; Chen, D.; Brorson, K. A survey of quality attributes of virus spike preparations used in clearance studies. PDA J. Pharm. Sci. Technol. 2012, 66(5), 420 – 433. 10. Garnick, R. L. Experience with viral contamination in cell culture. Dev. Biol. Stand. 1996, 88, 49 –56. 11. Miesegaes, G.; Bailey, M.; Willkommen, H.; Chen, Q.; Roush, D.; Blumel, J.; Brorson, K. Proceedings of the 2009 Viral Clearance Symposium. Dev Biol (Basel) 2010, 133, 3–101.

2. The European Agency for the Evaluation of Medicinal Products: Human Medicines Evaluation Unit, Committee for Proprietary Medicinal Products (CPMP). Guideline on Virus Safety Evaluation of Biotechnological Investigational Medicinal Products, 2009. CPMP/BWP/398498/2005.

12. PDA Technical Report No. 41: Virus Filtration. Parenteral Drug Association: Bethesda, MD, 2008.

3. FDA. Points to Consider in the Manufacture and Testing of Monoclonal Antibody Products for Hu-

13. Brorson, K.; Krejci, S.; Lee, K.; Hamilton, E.; Stein, K.; Xu, Y. Bracketed generic inactivation

Vol. 68, No. 3, May-June 2014

309

Downloaded from journal.pda.org on December 14, 2016

of rodent retroviruses by low pH treatment for monoclonal antibodies and recombinant proteins. Biotechnol. Bioeng. 2003, 82(3), 321–329. 14. Curtis, S.; Lee, K.; Blank, G. S.; Brorson, K.; Xu, Y. Generic/matrix evaluation of SV40 clearance by anion exchange chromatography in flowthrough mode. Biotechnol. Bioeng. 2003, 84(2), 179 –186. 15. Willkommen, H.; Blu¨mel, J.; Brorson, K.; Chen, D.; Chen, Q.; Gröner, A.; Kreil, T. R.; Robertson, J. S.; Ruffing, M.; Ruiz, S. Meeting Report: PDA Virus and TSE Safety Forum. PDA J. Pharm. Sci. Technol. 2013, 67(2), 81–97. 16. Willkommen, H.; Blu¨mel, J.; Brorson, K.; Chen, D.; Chen, Q.; Gröner, A.; Kreil, T. R.; Robertson, J. S.; Ruffing, M.; Ruiz, S. Meeting report— workshop on virus removal by filtration: trends and new developments. PDA J. Pharm. Sci. Technol. 2013, 67(2), 98 –104. 17. Bakhshayeshi, M.; Jackson, N.; Kuriyel, R.; Mehta, A.; van Reis, R.; Zydney, A. L. Use of confocal scanning laser microscopy to study virus retention during virus filtration. J. Membr. Sci. 2011, 379(1–2), 260 –267. 18. Phillips, M. W.; DiLeo, A. J. A validatible porosimetric technique for verifying the integrity of virus-retentive membranes. Biologicals 1996, 24(3), 243–253. 19. Peinador, R. I.; Calvo, J. I.; ToVinh, K.; Thom, V.; Pra´danos, P.; Herna´ndez, A. Liquid–liquid displacement porosimetry for the characterization of virus retentive membranes. J. Membr. Sci. 2011, 372(1-2), 366 –372.

particles using cuprammonium regenerated cellulose hollow fiber (i-BMM or BMM) from aqueous solution containing protein. Polym. J. 1994, 26(11), 1244 –1256. 23. Bakhshayeshi, M.; Kanani, D. M.; Mehta, A.; van Reis, R.; Kuriyel, R.; Jackson, N.; Zydney, A. L. Dextran sieving test for characterization of virus filtration membranes. J. Membr. Sci. 2011, 379(1), 239 –248. 24. Lute, S.; Riordan, W.; Pease III, L. F. Tsai, D.-H.; Levy, R.; Haque, M.; Martin, J.; Moroe, I.; Sato, T.; Morgan, M.; Krishnan, M.; Campbell, J.; Genest, P.; Dolan, S.; Tarrach, K.; Meyer, A.; PDA Virus Filter Task Force; Zachariah, M. R.; Tarlov, M. J.; Etzel, M.; Brorson, K. A consensus rating method for small virus-retentive filters. I. Method development. PDA J. Pharm. Sci. Technol. 2008, 62(5), 318 –333. 25. Grant, D. C.; Liu, B. Y. H. Sieving capture of liquidborne particles by microporous membrane filtration media. Particle & Particle Systems Characterization 1991, 8(1– 4), 142–150. 26. DiLeo, A. J.; Vacante, D. A.; Deane, E. F. Size exclusion removal of model mammalian viruses using a unique membrane system, Part I: Membrane qualification. Biologicals 1993, 21(3), 275–286. 27. Grant, D. C.; Liu, B. Y. H.; Fisher, W. G.; Bowling, R. A. Particle capture mechanisms in gases and liquids: an analysis of operative mechanisms in membrane/fibrous filters. J. Environ. Sci. 1989, 32(4), 43–51. 28. FDA, U.S. Department of Health and Human Services. Guidance for Industry. Sterile Drug Products Produced by Aseptic Processing—Current Good Manufacturing Practice, 2004.

20. Bolton, G.; Cormier, J.; Krishnan, M.; Lewnard, J.; Lutz, H. Integrity testing of normal flow parvovirus filters using air-liquid based tests. Bioprocessing J. 2006, 5(1), 50.

29. ASTM F838 – 05 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration, 2005.

21. Giglia, S.; Krishnan, M. High sensitivity binary gas integrity test for membrane filters. J. Membr. Sci. 2008, 323(1), 60 – 66.

30. Wurm, F. M. Production of recombinant protein therapeutics in cultivated mammalian cells. Nat. Biotechnol. 2004, 22(11), 1393–1398.

22. Hirasaki, T.; Noda, T.; Nakano, H.; Ishizaki, Y.; Manabe, S.; Yamamoto, N. Mechanism of removing Japanese encephalitis virus (JEV) and gold

31. Berting, A.; Farcet, M. R.; Kreil, T. R. Virus susceptibility of Chinese hamster ovary (CHO) cells and detection of viral contaminations by

310

PDA Journal of Pharmaceutical Science and Technology

Downloaded from journal.pda.org on December 14, 2016

adventitious agent testing. Biotechnol. Bioeng. 2010, 106(4), 598 – 607. 32. Garnick, R. L. Raw materials as a source of contamination in large-scale cell culture. Dev. Biol. Stand. 1998, 93, 21–29. 33. Nims, R. W. Detection of adventitious viruses in biologicals—a rare occurrence. Dev. Biol. (Basel) 2006, 123, 153– 64, discussion 183–197. 34. Oehmig, A.; Buttner, M.; Weiland, F.; Werz, W.; Bergemann, K.; Pfaff, E. Identification of a calicivirus isolate of unknown origin. J. Gen. Virol. 2003, 84(10), 2837–2845. 35. Onions, D. Animal virus contaminants of biotechnology products. Dev. Biol. (Basel) 2004, 118, 155–163. 36. Moody, M.; Alves, W.; Varghese, J.; Khan, F. Mouse minute virus (MMV) contamination—a case study: detection, root cause determination, and corrective actions. PDA J. Pharm. Sci. Technol. 2011, 65(6), 580 –588.

41. Khan, N. Z.; Parrella, J. J.; Genest, P. W.; Colman, M. S. Filter preconditioning enables representative scaled-down modelling of filter capacity and viral clearance by mitigating the impact of virus spike impurities. Biotechnol. Appl. Biochem. 2009, 52(4), 293–301. 42. Bolton, G.; Cabatingan, M.; Rubino, M.; Lute, S.; Brorson, K.; Bailey, M. Normal-flow virus filtration: detection and assessment of the endpoint in bio-processing. Biotechnol. Appl. Biochem. 2005, 42(2), 133–142. 43. Slocum, A.; Burnham, M.; Genest, P.; Venkiteshwaran, A.; Chen, D.; Hughes, J. Impact of virus preparation quality on parvovirus filter performance. Biotechnol Bioeng. 2013, 110(1), 229 – 239. 44. Asher, D.; Slocum, A.; Bergmann, K.; Genest, P.; Katz, A.; Morais, J.; Lawrence, C.; Greenhalgh, P. Predicting virus filtration performance with virus spike characterization. BioProcess Int. 2011, 9(3), 26 –37.

37. Victoria, J. G.; Wang, C.; Jones, M. S.; Jaing, C.; McLoughlin, K.; Gardner, S.; Delwart, E. L. Viral nucleic acids in live-attenuated vaccines: detection of minority variants and an adventitious virus. J. Virol. 2010, 84(12), 6033– 6040.

45. Stuckey, J.; Strauss, D.; Venkiteshwaran, A.; Gao, J.; Luo, W.; Quertinmont, M.; O’Donnell, S.; Chen, D. A novel approach to achieving modular retrovirus clearance for a parvovirus filter. Biotechnol. Prog. 2014, 30(1), 79 – 85.

38. Tidona, C. A.; Darai, G. The Springer Index of Viruses, 2nd ed.; Springer: New York, 2011.

46. Blumel, J., Paul Ehrlich Institut, Germany. Reduced Viral Clearance Program Based on Platform Data for Biotech and Plasma Derived Products. PDA European Virus & TSE Safety Forum, Barcelona, Spain, 2011.

39. PDA Technical Report No. 47: Preparation of Virus Spikes Used for Virus Clearance Studies. Parenteral Drug Association: Bethesda, MD, 2010. 40. Cabatingan, M. Impact of virus stock quality on virus filter validation. BioProcess Int. 2005, 3(10), S39 –S43.

Vol. 68, No. 3, May-June 2014

47. King, K. FDA CDER. Virus Clearance Integration: Critical Considerations and Orthogonality. PDA European Virus & TSE Safety Forum, Berlin, Germany, 2013.

311

Downloaded from journal.pda.org on December 14, 2016

An Authorized User of the electronic PDA Journal of Pharmaceutical Science and Technology (the PDA Journal) is a PDA Member in good standing. Authorized Users are permitted to do the following: ·Search and view the content of the PDA Journal ·Download a single article for the individual use of an Authorized User ·Assemble and distribute links that point to the PDA Journal ·Print individual articles from the PDA Journal for the individual use of an Authorized User ·Make a reasonable number of photocopies of a printed article for the individual use of an Authorized User or for the use by or distribution to other Authorized Users Authorized Users are not permitted to do the following: ·Except as mentioned above, allow anyone other than an Authorized User to use or access the PDA Journal · Display or otherwise make any information from the PDA Journal available to anyone other than an Authorized User ·Post articles from the PDA Journal on Web sites, either available on the Internet or an Intranet, or in any form of online publications ·Transmit electronically, via e-mail or any other file transfer protocols, any portion of the PDA Journal ·Create a searchable archive of any portion of the PDA Journal ·Use robots or intelligent agents to access, search and/or systematically download any portion of the PDA Journal ·Sell, re-sell, rent, lease, license, sublicense, assign or otherwise transfer the use of the PDA Journal or its content ·Use or copy the PDA Journal for document delivery, fee-for-service use, or bulk reproduction or distribution of materials in any form, or any substantially similar commercial purpose ·Alter, modify, repackage or adapt any portion of the PDA Journal ·Make any edits or derivative works with respect to any portion of the PDA Journal including any text or graphics ·Delete or remove in any form or format, including on a printed article or photocopy, any copyright information or notice contained in the PDA Journal

Use of MMV as a Single Worst-Case Model Virus in Viral Filter Validation Studies.

Typical platform processes for biopharmaceutical products derived from animal cell lines include a parvovirus filtration unit operation to provide vir...