bioengineering Article

The Separation of Blood Components Using Standing Surface Acoustic Waves (SSAWs) Microfluidic Devices: Analysis and Simulation Ahmed M. Soliman 1, *, Mohamed A. Eldosoky 1 and Taha E. Taha 2 1 2

*

Biomedical Engineering Department, Faculty of Engineering, Helwan University, Cairo11792, Egypt; [email protected] Communication Department, Faculty of Electronic Engineering, Menoufia University, Menouf 23952, Egypt; [email protected] Correspondence: [email protected]

Academic Editor: Gou-Jen Wang Received: 4 February 2017; Accepted: 26 March 2017; Published: 29 March 2017

Abstract: The separation of blood components (WBCs, RBCs, and platelets) is important for medical applications. Recently, standing surface acoustic wave (SSAW) microfluidic devices are used for the separation of particles. In this paper, the design analysis of SSAW microfluidics is presented. Also, the analysis of SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force are explained and analyzed. The analyses are provided for selecting the piezoelectric material, width of the main microchannel, working area of SAW, wavelength, minimum input power required for the separation process, and widths of outlet collecting microchannels. The design analysis of SSAW microfluidics is provided for determining the minimum input power required for the separation process with appropriated the displacement contrast of the particles.The analyses are applied for simulation the separation of blood components. The piezoelectric material, width of the main microchannel, working area of SAW, wavelength, and minimum input power required for the separation process are selected as LiNbO3 , 120 µm, 1.08 mm2 , 300 µm, 371 mW. The results are compared to other published results. The results of these simulations achieve minimum power consumption, less complicated setup, and high collecting efficiency. All simulation programs are built by MATLAB. Keywords: surface acoustic waves; microfluidics; separation of blood components; biomedical applications; analysis and simulations

1. Introduction White blood cells (WBCs), red blood cells (RBCs), and platelets are the components of blood. The separation of blood components is important for medical treatment and diagnosis. The standing surface acoustic wave (SSAW) microfluidic devices are positively significant in the separation of particles especially in biomedical applications. The separation of particles is depending on volume, density, and compressibility of the particles. The separation of particles using SSAW microfluidic devices has many advantages such as: miniaturization, high accuracy, high sensitivity, portable, low cost, low time consumption, and low samples consumption. The surface acoustic wave (SAW) is generated by applying alternating electric (AC) signal to interdigital transducers (IDTs) based on the piezoelectricity phenomenon. IDTs are metal electrodes deposited on the piezoelectric substrate. The acoustic waves are launched and propagated away from and perpendicular to IDTs fingers. The wavelength is determined by the IDT geometry as following [1]: λ = 4d Bioengineering 2017, 4, 28; doi:10.3390/bioengineering4020028

(1) www.mdpi.com/journal/bioengineering

Bioengineering 2017, 4, 28

Bioengineering 2017, 4, 28

2 of 16

λ = 4d

2 of 16

(1)

where: d is the finger width of IDTs. The acoustic wave propagation velocity is determined by the where: is the finger widthand of IDTs. Theofacoustic wave determined by the acousticd mode, orientation the type substrate. Thepropagation fundamentalvelocity resonantisfrequency is related acoustic mode, orientation and the type of substrate. The fundamental resonant frequency is related to to the acoustic wave velocity as given as following: the acoustic wave velocity as given as following:

f v /

f 00 = v00/λ

(2) (2)

where: f 00,, vv00, , and wavelength of of where: and λλ are are the the unperturbed unperturbed frequency, the unperturbed velocity and wavelength SAW,respectively. respectively.The Theinteraction interactionbetween betweenSAWs SAWsand andthe themedium mediumcan canbe be used used as as sensors sensors [2,3] [2,3] or or as SAW, principles of of mathematical mathematical biosensors in biological applications [4]. In In our our previous work [5,6], the principles models of of using using SSAW SSAW microfluidic microfluidicdevices devicesin inthe theseparation separationprocess processwere weredescribed describedand andanalyzed. analyzed. models The schematic of SSAW microfluidic device for particles separation is shown in Figure The The schematic of SSAW microfluidic device for particles separation is shown in Figure 1. The 1. figure figure is scale. not to In scale. this schematic, the particles is flowed in an inlet microchannel is not to this In schematic, the particles mixturemixture is flowed in an inlet microchannel among among symmetric side microchannels. These flows produce hydrodynamic force. SSAW symmetric side microchannels. These flows produce hydrodynamic force. Also, the Also, SSAWthe force can force can migrate the particles to the pressure node or antinode depending on its acoustic contrast migrate the particles to the pressure node or antinode depending on its acoustic contrast factor sign as factor as willlater. be explained later.are The particles collected by outlet collectingaccording microchannels will besign explained The particles collected byare outlet collecting microchannels to its according to its physical properties. physical properties.

Figure 1. The Theschematic schematic of SSAW microfluidic device for particles separation. (The figure is not Figure 1. of SSAW microfluidic device for particles separation. (The figure is not to scale.) to scale.)

In Reference [7], the separation of three different sizes (1, 5, and 10 μm) of polystyrene particles withIn the same density compressibility was demonstrated. The 10 applied flow particles rate was Reference [7], theand separation of three different sizes (1, 5, and µm) ofsample polystyrene varied 0.03 μL/min tocompressibility 0.3 μL/min. Thewas ratio of the sampleThe flowapplied rate to sample sheath flow with thefrom same density and demonstrated. flow was rate kept was constant as 1:3. From the analysis of the captured images, it was selected that the optimum flow varied from 0.03 µL/min to 0.3 µL/min. The ratio of the sample flow rate to sheath flow was kept rates as as 0.22 and 0.66 μL/min for the images, sample itflow andthat sheath flow rate;flow andrates the constant 1:3.μL/min From the analysis of the captured was rate selected the optimum optimum inputand power 320 mW, as 0.22 µL/min 0.66as µL/min forrespectively. the sample flow rate and sheath flow rate; and the optimum input InasReference [8], the separation of two 10 μm diameter, different-density particle streams power 320 mW, respectively. 3 and melamine: 1710 kg/m3) without using sheath flow was demonstrated. (polystyrene: 1050 [8], kg/m In Reference the separation of two 10 µm diameter, different-density particle streams 3 and 3 ) without The applied input power wasmelamine: ranged from mW to 1280 using mW and theflow flowwas ratedemonstrated. ranged from (polystyrene: 1050 kg/m 1710320 kg/m sheath The applied input power was ranged from 320 mW to 1280 mW and the flow rate ranged from

Bioengineering 2017, 4, 28

3 of 16

0.2 µL/min to 2 µL/min. The separation efficiencies obtained were in the range of 98.8%–89.4% for flow rates ranging from 0.2 µL/min to 2.0 µL/min at 1280 mW input power. In Reference [9], the separation of two different-size particle streams (3 µm and 10 µm) as well as the separation of two 10 µm diameter with different-density particle streams (1050 kg/m3 and 1710 kg/m3 ) was demonstrated. The applied input power was ranged from 300 mW to 1200 mW and the flow rate ranged from 0.1 µL/min to 0.5 µL/min. The captured fluorescent images were used to verify the experimental results. In Reference [10], input power ranging from 370 mW to 463 mW was applied to separate the blood components. The flow rates of the sample and sheath flow were 0.25 µL/min and 5 µL/min, respectively. As the applied power increased, the RBC clearance ratio also increased. The RBC clearance ratio was over 99% in region of 416–463 mW. On the other hand, the platelets clearance ratio was 74.1% at 439 mW input power. In Reference [11], the separation two different particle mixtures: fluorescent microspheres of different sizes and human peripheral blood mononuclear cells (PBMCs) mixed with E. coli bacteria were demonstrated. The first sample was a mixture of two types of fluorescent microspheres, 1.2 µm in diameter with green emission and 5.86 µm in diameter with red emission. The applied input powers were 85 mW, 240 mW, and 339 mW. The Flow rates of the particle flow and a single side sheath flow were 0.2 µL/min and 0.8 µL/min, respectively. The input power 85 mW was weak to produce sufficient acoustic radiation force. On the other hand, the input power 339 mW was strong to produce sufficient acoustic radiation force. Therefore, the input power 240 mW was used to separate this mixture. The second sample consisted of two different biological cells, purified human PBMCs and E. coli bacteria. PBMCs had an averaged diameter around 7.23 µm; while E. coli bacteria have an averaged volume similar to a sphere with a diameter of 1.1µm. The applied input power was 468 mW. The flow rates of the cell mixture and sheath flow were 0.5 µL/min and 4 µL/min, respectively. The separation efficiency of E. coli and PBMCs were 95.65% and 91.48%, respectively. As in References [7–12], the random high input power, wide range of input power, wide range of flow rates, more than one stage of SSAW, or had low collecting efficiency was used. Also, the design analysis of SSAW microfluidics or the separation analysis was not included in these published papers. In this paper, The SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force are analyzed. Also, the design analysis of SSAW microfluidics for the separation of particles is presented and discussed in details. The design analysis of SSAW microfluidics is depending on three factors. The first factor is the setup properties of the SSAW microfluidics such as: piezoelectric substrate material, input power, wavelength, length of IDTs, width of the main microchannel, widths of outlet collecting microchannels, working area of SAW, working time of SAW, and flow rate of fluid. The second factor is the properties of the fluid such as: density, compressibility, and viscosity. The third factor is the properties of the particles such as: density, compressibility, volume and radius. The design analysis of SSAW microfluidics is provided for selecting the piezoelectric substrate material, width of the main microchannel, working area of SAW, wavelength, minimum input power, and widths of outlet collecting microchannels required for the separation process. The minimum input power required to shift the maximum volume of the particles to the side wall of the main SSAW microchannel at the end of the working time of SSAW is calculated by simulation programs. The calculated minimum input power required for the separation process is determined with appropriated displacement contrast of the particles. Also, the simulations are provided for minimum power consumption, less complicated setup, and high collecting efficiency compared to other published papers [7–12]. The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the design analysis of SSAW microfluidics is presented. Also, the analysis of SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force are presented and discussed.

Bioengineering 2017, 4, 28

4 of 16

2. The Analysis of Forces and Other Effects Acting in SSAW Microfluidics for the Separation of Particles The Surface acoustic waves (SAWs) are generated by two IDTs. These SAWs are overlapped and produced SSAW to generate the acoustic radiation force. Therefore, pressure node and antinode are generated. These lead the particles to migrate toward the pressure node or antinode depending on the sign of the acoustic contrast factor (ϕ). The acoustic radiation force (FSSAW ) is represented as following [13]: ! π p20 Vβ m FSSAW = − ϕ( β, ρ) sin(2kx ) (3) 2λ The acoustic contrast factor ϕ(β,ρ) can be expressed as following:  ϕ( β, ρ) =

5ρ p − 2ρm βp − 2ρ p + ρm βm

 (4)

where p0 , V, λ, ρp , ρm , βp , βm , k, and x are the acoustic pressure amplitude, the volume of the particle, the wavelength, the density of particle, the density of medium, the compressibility of particle, the compressibility of medium, the wave number, and the distance from the pressure node to the initial particle position, respectively. As shown in Equation (4), the acoustic contrast factor (ϕ) is determined by the density and compressibility of the particles and the medium. The acoustic pressure amplitude (p0 ) can be calculated as following: r p0 =

PZ A

(5)

where: P, Z and A are the AC signal power applied to IDT, the acoustic impedance of the piezoelectric substrate, and working area of SAW, respectively. The acoustic impedance of the piezoelectric substrate (Z) can be calculated as following: Z = ρ s v0 (6) where: ρs is the density of the piezoelectric substrate. The power flow in SAW devices can be detected by high sensitive optical measurement [14]. The acoustic contrast factor ϕ(β,ρ) determines whether the particle will move towards the pressure node or the pressure anti-node. If ϕ(β,ρ) > 0, the particles will be moved to the pressure node. If ϕ(β,ρ) < 0, the particles will be moved to the pressure antinode. The most biological particles in liquid medium are moved toward pressure node [15]. The viscous drag force is the resistive force acting on particles in the opposite direction of acoustic radiation force. The viscous drag force is derived from Stoke’s law which describes the motion of sphere particles in a viscous fluid at a specific velocity. The viscous drag force (Fv ) is given by Stoke’s law as following [16]: Fv = 6πηrv (7) where: η, v, and r are the fluid viscosity, the relative velocity between particles and medium, and the radius of the particle, respectively. The acoustic radiation force is dominant compared to the viscous drag force. Also, the viscous drag force cannot be neglected. When the particles have constant velocity in the SSAW field, the acoustic and viscous forces are balanced. According to Equations (3), (4), and (7), the time required for the particle to migrate from position X1 to position X2 based on SSAW and viscous drag forces only (tSSAW ) can be expressed as following [17]:  2  ln tan 2πx  X2 3λ ηr λ X1 tSSAW = (8) π p20 Vβ m ϕ( β, ρ)

  2x     ln  tan     X  3 r    1  t SSAW      p 2V   ,     m 0 2

(6)

The displacement Bioengineering 2017, 4, 28

of particles (dSSAW) based on the SSAW and viscous drag forces can 5 ofbe 16 approximately expressed as following [15]:

 

 on the The displacement of particles (dSSAW ) based SSAW and viscous drag forces can be d SSAW  tan 1 e c (7) approximately expressed as following [15]: 2 λ −1 c dSSAW (9) p02V=  m2π tan  (e ) 2x    c  tw  ln  tan   (10) 2  3 r /      !   2   p0 Vβ m 2πx (10) c = tw + ln tan λ 3λ2 ηr/π where: tw is the working time of the acoustic radiation force on the particle. The angletime (θR)ofisthethe refraction angle produced due to the changing in SAW where: tw Rayleigh is the working acoustic radiation force on the particle. propagation velocity in piezoelectric substrate and the liquid medium. The Rayleigh angle (θR) can The Rayleigh angle (θ R ) is the refraction angle produced due to the changing in SAW propagation be calculated by Snell’s substrate law as following: velocity in piezoelectric and the liquid medium. The Rayleigh angle (θ R ) can be calculated by Snell’s law as following:

 v    sin 1  vmm  θ RR= sin−1  v0  v0

(11) (11)

inin thethe medium. When SAW meets the the edgeedge of the the energy will where: vvmmisisSAW SAWvelocity velocity medium. When SAW meets of liquid, the liquid, the energy couple into into the the liquid at the Rayleigh angle [18]. The Rayleigh will couple liquid at the Rayleigh angle [18]. The Rayleighangle angleisisdepending dependingon onthe the SAW SAW velocities in the piezoelectric material and the liquid medium. As As in in [7–12], the width of the main microchannel is about λ/2 λ/2 or or more. more. Therefore, Therefore, the the Rayleigh Rayleigh angle angle can not be neglected. As shown in Figure 2, when SAW SAW propagated propagated from from the the piezoelectric piezoelectric substrate to the the liquid liquid medium, medium, the the refracted refracted SAW had had horizontal horizontal and vertical vertical components. The acoustic radiation force produced by the horizontal component and can bebe expressed as Fas SSAW sin (θ R). (θ R ). component of ofSSAW SSAWisisused usedininthe theseparation separationprocess process and can expressed FSSAW sin

Figure Figure 2. 2. The The configuration configuration of of Rayleigh Rayleigh angle angle (θ (θ RR)) affect affect on on the the propagation propagation of of SAW. SAW.

The attenuation in liquid-loaded substrate affects on the horizontal component of SSAW force. The attenuation in liquid-loaded substrate affects on the horizontal component of SSAW force. This attenuation is in x-direction. The attenuation will also vary in the z-direction. The attenuation in This attenuation is in x-direction. The attenuation will also vary in the z-direction. The attenuation in x-direction is the main important to the separation process; therefore, the attenuation in z-direction x-direction is the main important to the separation process; therefore, the attenuation in z-direction is not studied in this paper. The horizontal component of SSAW force with the attenuation can be is not studied in this paper. The horizontal component of SSAW force with the attenuation can be represented as following [19]: represented as following [19]: FSSAW All = FSSAW sin (θR ) e−2αH

(12)

where: α, H are the attenuation coefficient and path length of SAW in in liquid-loaded substrate. The attenuation coefficient (α) can be calculated as following [20]: α=

1 ρm vm = lSAW ρ s v0 λ

(13)

where: α, H are the attenuation coefficient and path length of SAW in in liquid-loaded substrate. The attenuation coefficient (α) can be calculated as following [20]:

 Bioengineering 2017, 4, 28

1 l SAW



 m vm  s v0 

(9) 6 of 16

where: lSAW is the attenuation length of SAW in liquid-loaded substrate. The attenuation coefficient (α) is depending the properties ofof liquid piezoelectric substrate. material substrate; and thecoefficient wavelength where: lSAW is theon attenuation length SAWand in liquid-loaded The attenuation (α) of depending SAW. The attenuation is increased exponentially with the attenuation coefficient. is on the properties of liquid and piezoelectric material substrate; and the wavelength of radiation force exponentially (FSSAW) with with Rayleigh angle (θR)coefficient. effect and its attenuation in SAW.The Theacoustic attenuation is increased the attenuation liquid-loaded substrate can be expressed Equations and (11)–(13) as following: The acoustic radiation force (FSSAWfrom ) with Rayleigh(3)angle (θ R ) effect and its attenuation in liquid-loaded substrate can be expressed from Equations (3) and (11)–(13) as following:  p 2V m  ! FSSAW All   π p02 Vβ   ,   sin 2kx sin ( R ) e 2H (10)   m FSSAW All = −  02 ϕ( β, ρ) sin(2kx ) sin (θR ) e−2αH (14)  2λ where: FSSAW All is the acoustic radiation force with Rayleigh angle effect and its attenuation in where: FSSAW All is the acoustic radiation force with Rayleigh angle effect and its attenuation in liquid-loaded substrate. liquid-loaded substrate. As mentioned previously, Constructive and destructive interferences which result from the two As mentioned previously, Constructive and destructive interferences which result from the two opposing SAWs generate SSAW with the formation of pressure nodes (minimum pressure amplitude) opposing SAWs generate SSAW with the formation of pressure nodes (minimum pressure amplitude) and pressure anti-nodes (maximum pressure amplitude) as shown in Figure 3. These pressure and pressure anti-nodes (maximum pressure amplitude) as shown in Figure 3. These pressure fluctuations result in acoustic radiation forces on the particles in the liquid medium, forcing them fluctuations result in acoustic radiation forces on the particles in the liquid medium, forcing them migrate toward either the pressure nodes or the pressure anti-nodes, depending on the sign of migrate toward either the pressure nodes or the pressure anti-nodes, depending on the sign of acoustic acoustic contrast factor. The locations of pressure nodes and antinodes are important for the design contrast factor. The locations of pressure nodes and antinodes are important for the design of SSAW of SSAW microfluidics to separate the particles. microfluidics to separate the particles.

Figure 3. The nodes. The attenuation effect on the equilibrium force line and the locations of the pressure nodes.

Without the attenuation effect, the equilibrium force line will be at zero force line. Regarding to Without the attenuation effect, the equilibrium force line will be at zero force line. Regarding to the attenuation of acoustic radiation force, the locations of pressure nodes will be affected. As shown the attenuation of acoustic radiation force, the locations of pressure nodes will be affected. As shown in Figure 3, the equilibrium force line is shifted due to the attenuation effect. Therefore, the locations in Figure 3, the equilibrium force line is shifted due to the attenuation effect. Therefore, the locations of pressure nodes are changed. As an example, if the width of the microchannel is λ/2, without the of pressure nodes are changed. As an example, if the width of the microchannel is λ/2, without attenuation effect, the locations of pressure nodes will be exactly on the both sides of the SSAW the attenuation effect, the locations of pressure nodes will be exactly on the both sides of the SSAW microchannel. On the other hand, with the attenuation effect, the locations of pressure nodes will be microchannel. On the other hand, with the attenuation effect, the locations of pressure nodes will be slightly moved toward the center of the SSAW microchannel. Therefore, if the SSAW microfluidics is slightly moved toward the center of the SSAW microchannel. Therefore, if the SSAW microfluidics is designed to migrate the particles completely to the side wall of SSAW microchannel (pressure node), the particles will not move completely to the desired locations due to the effect of the attenuation of acoustic radiation force.

Bioengineering 2017, 4, 28

7 of 16

The Equation (8) can be modified to express the time required for the particle to migrate from position X1 to position X2 based on SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, and viscous drag force (tSSAW All ) as following:  tSSAW All =

3λ2 ηr π





ln tan

2πx λ

 X2

X1 2 p0 Vβ m ϕ( β, ρ) sin(θ R )e−2αH

(15)

The displacement of particle (dSSAW All ) based on SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, and viscous drag force can be approximately expressed as following: λ dSSAW All = tan−1 (e g ) (16) 2π !    p20 Vβ m sin(θ R )e−2αH 2πx (17) g = tw + ln tan λ 3λ2 ηr/π The hydrodynamic force is generated due to the flow of the core stream in the inlet microchannel among the flow of symmetric side microchannels. The core stream will be focused in the center of the main microchannel. The assumptions which used to determine the width of the focused stream (Wf ) in rectangular microchannelsare explained in details in [5,21]. The width and position of the focused stream can be determined by the relative volumetric flow rates of the three inlets microchannels. The width of the focused stream (Wf ) can be calculated using mass conservation law as following [21]: Wf =

Wo Qi γ( Qi + QS1 + QS2 )

(18)

where: Qi , Qs1 , Qs2 , Wo , γ are the volumetric flow rate of the inlet microchannel, the volumetric flow rate of the side microchannels 1 and 2, the width of the main microchannel, and the velocity ratio of the average velocity of the focused stream to the average velocity of the stream in the outlet microchannel which assumed as a unity [22]. The maximum displacement of particle (dHydrodynamic ) from the center of the main microchannel based on the hydrodynamic force can be calculated according to Equation (18) as following: Wf d Hydrodynamic = (19) 2 The diffusion force can affect on the separation process. The diffusion is the movement of the solute from high to low concentrations of the solute in the solvent [23]. The particles in the fluid can be moved due to the diffusion force. In general, a small particle diffuses faster than a larger one. According to the simple constant-step random walk model of diffusion, the diffusion displacement can be expressed as following [23]: d Di f f usion =

q

Dt Di f f usion

(20)

where: dDiffusion , tDiffusion , D are the diffusion displacement, diffusion time, and diffusivity, respectively. The total displacements of the particles (dTotal ) due to SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force can be approximately expressed as following: d Total = dSSAW All + d Hydrodynamic + d Di f f usion

(21)

The total displacements of the particles (dTotal) due to SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force can be approximately expressed as following: Bioengineering 2017, 4, 28

dTotal  d SSAW All  d Hydrodynam ic  d Diffusion

(17)

8 of 16

3. 3. The The Design Design Analysis Analysis of of SSAW SSAW Microfluidics Microfluidicsand andthe theSimulation Simulationof ofthe theSeparation Separationof ofBlood Components Blood Components The design of of SSAW SSAWmicrofluidics microfluidicsisisthe the main factor successful separation of particles. The design main factor forfor successful separation of particles. The The design analysis of SSAW microfluidics are not included in the most published papers. The design design analysis of SSAW microfluidics are not included in the most published papers. The design of of SSAW microfluidics successfulseparation separationofofparticles particlesisisdepending dependingon on three three factors. factors. The SSAW microfluidics forforsuccessful The first first factor factor is is the the setup setup properties properties of of SSAW SSAW microfluidics microfluidics such such as: as: piezoelectric piezoelectric substrate substrate material, material, input input power, wavelength, length of IDTs, width of the main microchannel, widths of outlet power, wavelength, length of IDTs, width of the main microchannel, widths of outlet collecting collecting microchannels, working time of SAW, and flow rate of fluid. The second factor microchannels, working workingarea areaofofSAW, SAW, working time of SAW, and flow rate of fluid. The second is the properties of the fluid such as: density, compressibility, and viscosity. The third factor is the factor is the properties of the fluid such as: density, compressibility, and viscosity. The third factor properties of the particles such as: density, compressibility, volume, and radius. There are dependant is the properties of the particles such as: density, compressibility, volume, and radius. There are relationships among these three these factors. Allfactors. these factors can factors affect on separation of dependant relationships among three All these canthe affect on the process separation particles as explained previously. process of particles as explained previously. The proposed design analysis of SSAW microfluidics for specific separation process The flow flowchart chartof of proposed design analysis of SSAW microfluidics for specific separation is shownisinshown Figurein4.Figure 4. process

Figure 4. The flow chart of proposed design analysis of SSAW microfluidics for specific Figure 4. The flow chart of proposed design analysis of SSAW microfluidics for specific separation process. separation process.

The piezoelectric substrate material is important to generate the SAW. The piezoelectric substrate material is the main factor of the value of the horizontal component ratio of SSAW force due to the effect of Rayleigh angle as shown in Equation (11). The horizontal component ratios of SSAW force

Bioengineering 2017, 4, 28

9 of 16

in distilled water for LiNbO3 , LiTaO3 , GaAs, and quartz piezoelectric materials are shown in Table 1. As explained previously, the piezoelectric substrate materials are affecting on the locations of pressure nodes and the attenuation coefficient as shown in Equation (13). The LiNbO3 is used as substrate material in the simulations due to its maximum electromechanical coupling coefficient (K2 ) compared to the others. Table 1. The horizontal component ratios of SSAW force in distilled water for LiNbO3 , LiTaO3 , GaAs, and quartz. Piezoelectric Substrate Material

LiNbO3

LiTaO3

GaAs

Quartz

The horizontal component ratios of SSAW force in distilled water

0.3745

0.4410

0.5202

0.4565

This proposed design analysis is provided to separate the particles types have different properties (size, density, and compressibility). In the proposed design analysis, the properties of particles and liquid medium are inserted. Also, the starter value of the wavelength (λ) is inserted. Then, the properties of the setup of SSAW microfluidics device are calculated based on the mathematical models which described in the previous chapter as following: 1. 2. 3. 4.

The width of the main SSAW microchannel is about (2λ/5). The wall thickness of the main microchannel is about (λ/20). The SAW piezoelectric substrate material is LiNbO3 . The mixture of components flows in the middle input inlet microchannel as shown in Figure 1.

The lengths of the main SSAW microchannel and IDTs are selected as 15 mm and 9 mm, respectively. The flow rates in the middle input inlet microchannel and the two-symmetric side microchannels are selected as 0.2 µL/min and 0.8 µL/min, respectively. The working time of SSAW force is 0.9 s. The following variables are also calculated: 1. 2. 3. 4. 5. 6.

The SAW working area. The Rayleigh angle between the piezoelectric material and the liquid medium. The displacements of particles due to the hydrodynamic force. The displacements of particles due to the diffusion force. The attenuation in the wall thickness of the main microchannel. The attenuation in the liquid medium.

The particle type which has the maximum effects of SSAW force from the center to the side wall of SSAW microchannel is determined. For this particle type, the minimum input power signal required to shift completely to the side wall of SSAW microchannelis calculated. By calculating minimum input power, the differences among the separated particles (∆Xs) at the end of the working area of SSAW force are calculated. If the differences among the separated particles (∆Xs) are accepted, the dimensions of output microchannels can be calculated. The separation is accepted under the following conditions: 1. 2. 3.

The calculated difference between two types of particles is about (3R1 + 3R2 ). Where: R1 and R2 are the radiuses of the first the second particle types, respectively. The width of each output microchannel should be at least 5 times the radius of the particle type. The distance between the calculated position of the particle type and the nearest edge of the output microchannel should be at least 3 times of the radius of the particle type. The number of output microchannels (NOutputMicrochannels ) can determined as following: NOutputMicrochannels = 1 + (n − 1) × 2

(22)

Bioengineering 2017, 4, 28

10 of 16

where: n is the number of the particles types. The particle type which has the minimum effect of SSAW force will collected from the middle output microchannel. Two-symmetric output microchannels are designed to collect each other particles type. If the differences among the separated particles (∆Xs) are not accepted, the wavelength (λ) will incremented; then, the previous steps will be repeated. The WBCs, RBCs, and platelets are the main particles in the blood. The values of the density of WBCs, RBCs, and platelets are very closed. Otherwise, the values of the compressibility of WBCs, RBCs, and platelets are the same. Therefore, the volume (radius) is the dominant property affecting on the separation process. The values of radius, density, and compressibility of WBCs, RBCs, and platelets are shown in Table 2 [24,25]. The design analysis of SSAW microfluidics for the separation of blood components is presented. Table 2. The values of radius, density, and compressibility for WBCs, RBCs, and platelets. Particle Type

Radius (µm)

Density (kg/m3 )

Compressibility (T·Pa−1 )

WBC RBC Platelet

5 3.5 1.5

1060 1096 1058

348 348 348

In our simulation programs, the selection of minimum input power required for the separation process, piezoelectric substrate material, wavelength, width of the main microchannel, widths of outlet collecting microchannels, and working area of SAW are the main target in our design analysis of SSAW microfluidics for the separation of blood components. The constant values of the length of IDTs and working time of SAW are 9 mm and 0.9 s, respectively. The constant values of the ratio of the flow rates of input inlet microchannel to the symmetric side microchannels are used to provide 0.9 s working time of SAW. To simplify the simulation in this study; therefore, the distilled water is used as a fluid carrier of blood components and as sheath fluid. The density, compressibility, and viscosity of distilled water are 1000 kg/m3 , 448 T·Pa−1 , and 1 mPa·s, respectively. The simulation programs are built by MATLAB. In the design analysis, the minimum input power required to shift the maximum volume of the particles to the side wall of the main SSAW microchannel at the end of the working time of SSAW is calculated and used to decrease the power consumption in SSAW microfluidics. The calculated minimum input power has relationships with the properties of particles, fluid, and other properties of SSAW microfluidics. The relations of the minimum input power required to shift the particle to the side wall of the main SSAW microchannel at the end of the working time of SSAW, radius of the particle, and the acoustic contrast factor are shown in Figure 5. The wavelength, the SAW working area, and the piezoelectric material which used in this simulation are 300 µm, 1.08 mm2 , and LiNbO3 , respectively. The radius of particle is varying from 1 to 7.5 µm. The acoustic contrast factor is varying from 0.1 to 0.75. The required input power is inversely proportional to radius of particle and acoustic contrast factor. In our design analysis, the width of the main SSAW microchannel is selected as 2λ/5 to avoid the slight moving of the pressure nodes due to the attenuation effect of SAW. Therefore, the relation between the width of the main SSAW microchannel and the wavelength is shown in Figure 6. The width of the main SSAW microchannel is proportional to the wavelength. The working area of SAW is depending on and proportional to the width of the main SSAW microchannel because the length of IDTs is constant as 9 mm. Also, the displacement of particles due to the hydrodynamic force is increased proportional to the width of the main SSAW microchannel as explained previously in Equations (18) and (19) because the ratio of the flow rate of input inlet microchannel to the flow rates of the side microchannels is constant.

Bioengineering 2017, 4, 28 Bioengineering 2017, 4, 28

11 of 16 11 of 16

Figure 5. The relations of the input power required to shift the particle to the side wall of the main SSAW microchannel at the end of the working time of SSAW, radius of the particle, and the acoustic contrast factor.

In our design analysis, the width of the main SSAW microchannel is selected as 2λ/5 to avoid the slight moving of the pressure nodes due to the attenuation effect of SAW. Therefore, the relation between the width of the main SSAW microchannel and the wavelength is shown in Figure 6. The width of the main SSAW microchannel is proportional to the wavelength. The working area of SAW is depending on and proportional to the width of the main SSAW microchannel because the length of IDTs is constant as 9 mm. Also, the displacement of particles due towall the hydrodynamic Figure 5. The relations of the input power required to shift the particle to the side of the main forceSSAW is increased proportional to the width of the main SSAW microchannel as explained microchannel at the end of the working time of SSAW, radius radius of the particle, and the acoustic previously in Equations (18) and (19) because the ratio of the flow rate of input inlet microchannel contrast factor. to the flow rates of the side microchannels is constant. In our design analysis, the width of the main SSAW microchannel is selected as 2λ/5 to avoid the slight moving of the pressure nodes due to the attenuation effect of SAW. Therefore, the relation between the width of the main SSAW microchannel and the wavelength is shown in Figure 6. The width of the main SSAW microchannel is proportional to the wavelength. The working area of SAW is depending on and proportional to the width of the main SSAW microchannel because the length of IDTs is constant as 9 mm. Also, the displacement of particles due to the hydrodynamic force is increased proportional to the width of the main SSAW microchannel as explained previously in Equations (18) and (19) because the ratio of the flow rate of input inlet microchannel to the flow rates of the side microchannels is constant.

Figure Figure 6. 6. The The relation relation between between the the width width of of the the main main SSAW SSAW microchannel and the wavelength.

The relation between the minimum input power required to shift the WBCs (the maximum The relation between the minimum input power required to shift the WBCs (the maximum particle compared to RBCs and platelets) to the side wall of the main SSAW microchannel at the particle compared to RBCs and platelets) to the side wall of the main SSAW microchannel at the end of end of the working time of SAW and the wavelength is shown in Figure 7. the working time of SAW and the wavelength is shown in Figure 7. The selection of the minimum input power required for separating the blood components WBCs, RBCs, and platelets is the main target in the design analysis. The shifting of the maximum particles (WBCs) to the side wall of the main SSAW microchannel at the end of the working time of SSAW is the reference to calculate the minimum input power. This minimum input power is applied to all blood Figure 6. The relation the width of the main SSAW microchannel the wavelength. components. Therefore, thebetween displacement contrast or the difference of theand displacement of particle from the center of the main SSAW microchannel (∆X) is essential when selecting the required input The relation between thetheminimum power required shift the WBCs (the power. The relations between differenceinput of particle radius (∆R) to (the reference particle is maximum WBC) and particle compared to RBCs and platelets) to the side wall of the main SSAW microchannel at the the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) at end of the working time of SAWare and the wavelength shown Figure 7. in Figure 8, the width of selected values of wavelengths shown in Figure 8.isFor eachin wavelength

Bioengineering 2017, 4, 28

12 of 16

main microchannel Bioengineering 2017, 4, 28

and the minimum input power are represented in Figures 6 and 7, respectively. 12 of 16 The SAW working time is constant as 0.9 s.

Figure 7. The relation between the minimum input power required to shift the WBCs to the side wall of the main SSAW microchannel at the end of the working time of SAW and the wavelength.

The selection of the minimum input power required for separating the blood components WBCs, RBCs, and platelets is the main target in the design analysis. The shifting of the maximum particles (WBCs) to the side wall of the main SSAW microchannel at the end of the working time of SSAW is the reference to calculate the minimum input power. This minimum input power is applied to all blood components. Therefore, the displacement contrast or the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) is essential when selecting the required input power. The relations between the difference of particle radius (∆R) (the reference particle is WBC) and the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) at selected values of wavelengths are shown in Figure 8. For Figure 7. 7. The Therelation relationbetween betweenthe theminimum minimuminput inputpower powerrequired required to to shift the the WBCs WBCs to to the the side side wall wall eachFigure wavelength in Figure 8, the width of main microchannel andshift the minimum input power are of the the main main SSAW SSAW microchannel microchannel at at the the end end of of the the working working time time of of SAW SAW and the of and the wavelength. wavelength. represented in Figures 6 and 7, respectively. The SAW working time is constant as 0.9 s. The selection of the minimum input power required for separating the blood components WBCs, RBCs, and platelets is the main target in the design analysis. The shifting of the maximum particles (WBCs) to the side wall of the main SSAW microchannel at the end of the working time of SSAW is the reference to calculate the minimum input power. This minimum input power is applied to all blood components. Therefore, the displacement contrast or the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) is essential when selecting the required input power. The relations between the difference of particle radius (∆R) (the reference particle is WBC) and the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) at selected values of wavelengths are shown in Figure 8. For each wavelength in Figure 8, the width of main microchannel and the minimum input power are represented in Figures 6 and 7, respectively. The SAW working time is constant as 0.9 s.

Figure The relations relations between between the the difference difference of of particle and the Figure 8. 8. The particle radius radius (∆R) (R) and the difference difference of of the the displacement of particle particle from from the thecenter centerofofthe themain mainSSAW SSAWmicrochannel microchannel(X) (∆X) selected values displacement of at at selected values of of wavelengths. wavelengths.

∆R between WBCs and RBCs (5 µm–3.5 μm–3.5 µm) μm) is 1.5 1.5 µm, μm, and and the the ∆R ∆R between between WBCs WBCs and and The ∆R platelets (5 µm–1.5 μm–1.5 µm) μm) is is 3.5 3.5 µm. μm. At At 300 300 µm μmwavelength, wavelength, the the displacement displacement contrast contrast (∆X) (∆X) between between platelets WBCs and and RBCs RBCs is is approximately approximately 31 31 µm, μm, and and the the displacement displacement contrast contrast (∆X) (∆X) between between WBCs WBCs and WBCs platelets is approximately 51 µm. From these displacement contrasts, the calculated displacement contrast (∆X) between RBCs and platelets is approximately 20 µm. These displacement contrasts appropriate to optimize the separation process with minimum required input power. Therefore, the 300 µm wavelength is selected for the design analysis. As shown in Figure 7, the minimum input power required for the separation of blood components at 300 µm wavelength is 371 mW. Figure 8. The relations between the difference of particle radius (R) the power, difference of range the Most previous SSAW-based devices as in [7–12] used random highand input wide of displacement of particle from the center of the main SSAW microchannel (X) at selected values of input power, wide range of flow rates, more than one stage of SSAW for the separation of particles, or wavelengths. had low collecting efficiency. This separation process can be followed only with high contrast of the The ∆R between WBCs and RBCs (5 μm–3.5 μm) is 1.5 μm, and the ∆R between WBCs and platelets (5 μm–1.5 μm) is 3.5 μm. At 300 μm wavelength, the displacement contrast (∆X) between WBCs and RBCs is approximately 31 μm, and the displacement contrast (∆X) between WBCs and

contrast (∆X) between RBCs and platelets is approximately 20 μm. These displacement contrasts appropriate to optimize the separation process with minimum required input power. Therefore, the 300 μm wavelength is selected for the design analysis. As shown in Figure 7, the minimum input power required for the separation of blood components at 300 μm wavelength is 371 mW. Most previous SSAW-based devices as in [7–12] used random high input power, wide range of Bioengineering 2017, 4, 28 13 of 16 input power, wide range of flow rates, more than one stage of SSAW for the separation of particles, or had low collecting efficiency. This separation process can be followed only with high contrast of properties of particles. The high complicated setup, or low collecting is the properties of particles. Thepower high consumption, power consumption, complicated setup, or lowefficiency collecting the disadvantage these published papers. Also, thepapers. separation analysis and the design analysis efficiency is the of disadvantage of these published Also, the separation analysis and are the not presented inare these design analysis notpublished presentedpapers. in these published papers. The final SSAW SSAW microfluidic microfluidicdesign designfor forthe theseparation separationof ofblood bloodcomponents components (WBCs, (WBCs, RBCs, RBCs, and platelets) is based on the schematic shown in Figure 1. The most configuration details used in the the schematic shown in Figure 1. simulation are collected from [11]. The width and height of the main SSAW microchannel are 120 (2λ/5) main SSAW microchannel andand IDTs are 15 and (2λ/5) and 25 µm, μm, respectively. respectively.The Thelengths lengthsofofthe the main SSAW microchannel IDTs aremm 15 mm 9and mm, The SAW (λ) is 300 µm. piezoelectric substrate material is 9 respectively. mm, respectively. Thewavelength SAW wavelength (λ) is The 300 SAW μm. The SAW piezoelectric substrate LiNbO flow 3rates the input microchannel and the two symmetric sidesymmetric microchannels material is LiNbO . Theinflow rates inlet in the input inlet microchannel and the two side 3 . The are 0.2 µL/minare and0.2 0.8μL/min µL/min, mixture of blood components flows in the input microchannels andrespectively. 0.8 μL/min, The respectively. The mixture of blood components flows inlet The distilledThe water is usedwater as a fluid. The timeworking of SSAWtime force 0.9 s. in themicrochannel. input inlet microchannel. distilled is used as working a fluid. The of isSSAW The required inputrequired power isinput 371 mW. forceminimum is 0.9 s. The minimum power is 371 mW. The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the design analysis the analysis of of SSAW force with Rayleigh angle effecteffect and analysis of ofSSAW SSAWmicrofluidics microfluidicsand and the analysis SSAW force with Rayleigh angle its in liquid-loaded substrate, viscous drag drag force,force, hydrodynamic force,force, and diffusion force andattenuation its attenuation in liquid-loaded substrate, viscous hydrodynamic and diffusion is shown in Figure 9. The9.displacement of WBCs, RBCs,RBCs, and platelets at theatend theofworking time force is shown in Figure The displacement of WBCs, and platelets theofend the working time of SSAW force dueSSAW to the SSAW withangle Rayleigh and itsinattenuation in of SSAW force due to the force withforce Rayleigh effect angle and itseffect attenuation liquid-loaded liquid-loaded substrate, viscous drag force, force, hydrodynamic force, andare diffusion force are 60, 29, and 9 substrate, viscous drag force, hydrodynamic and diffusion force 60, 29, and 9 µm, respectively. μm, separated respectively. The separated particles can bebysimply collected by five outlets microfluidics The particles can be simply collected five outlets microfluidics as shown in Figureas 1. shown in Figure 1. Therefore, widths of the fivecan outlets microfluidics be determined. The Therefore, the widths of the fivethe outlets microfluidics be determined. Thecan width of middle outlet width of middle outlet microchannel for platelets collection about 30 μm.outlet The microchannels width of two microchannel for platelets collection is about 30 µm. The width of is two mid-lateral mid-lateral outlet microchannels RBCs collection is about 20 μm. The width of lateral outlet for RBCs collection is about 20 µm.for The width of two lateral outlet microchannels fortwo WBCs collection microchannels is about 25 µm. for WBCs collection is about 25 μm.

Figure 9. 9. The The separation separationofofblood bloodcomponents components based design analysis of SSAW microfluidics Figure based onon thethe design analysis of SSAW microfluidics and and the analysis of SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded the analysis of SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, substrate, viscous drag, hydrodynamic, andforces. diffusion forces. viscous drag, hydrodynamic, and diffusion

The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the design analysis of SSAW microfluidics and the SSAW force only is shown in Figure 10. This simulation design analysis of SSAW microfluidics and the SSAW force only is shown in Figure 10. This simulation is done under the same simulation conditions of the previous simulation. As shown, the WBCs and is done under the same simulation conditions of the previous simulation. As shown, the WBCs and RBCs are very closer together (about 7 µm) at the end of the working time of SSAW force. This simulation presents obviously the importance of other forces and effects. The SSAW force only is not enough to provide accurate simulations.

Bioengineering 2017, 4, 28

14 of 16

RBCs are very closer together (about 7 μm) at the end of the working time of SSAW force. This simulation presents is Bioengineering 2017, 4, 28 obviously the importance of other forces and effects. The SSAW force only 14 of 16 not enough to provide accurate simulations.

Figure 10. The Figure 10. The separation separation of of blood blood components components (WBCs, (WBCs, RBCs, RBCs, and and platelets) platelets) based based on on the the design design analysis of SSAW microfluidics and the SSAW force only. analysis of SSAW microfluidics and the SSAW force only.

In Reference [10], the separation of platelets from whole blood components (WBCs, RBCs, and In Reference [10], the separation of platelets from whole blood components (WBCs, RBCs, Platelets) are demonstrated by input power range from 370 mW to 463 mW. On the other hand, our and Platelets) are demonstrated by input power range from 370 mW to 463 mW. On the other hand, results represented not only the separation of platelets but also the full separation of whole blood our results represented not only the separation of platelets but also the full separation of whole components (WBCs, RBCs, and Platelets) using minimum input power 371 mW and only one blood components (WBCs, RBCs, and Platelets) using minimum input power 371 mW and only one separation stage. separation stage. These results of the separation of the blood components (WBCs, RBCs, and platelets) These results of the separation of the blood components (WBCs, RBCs, and platelets) demonstrate demonstrate the design analysis of SSAW microfluidics and the analysis of SSAW force with the design analysis of SSAW microfluidics and the analysis of SSAW force with Rayleigh angle effect Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion hydrodynamic force, and diffusion force. The analysis study is significant and powerful for the force. The analysis study is significant and powerful for the simulation of separation process especially simulation of separation process especially for separation of biological particles. for separation of biological particles. 4. Conclusions 4. Conclusions This paper demonstrated more accurate theoretical simulation of the separation of particles This paper demonstrated more accurate theoretical simulation of the separation of particles using SSAW microfluidic devices. The SSAW force with Rayleigh angle effect and its attenuation in using SSAW microfluidic devices. The SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force were analyzed. liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force were analyzed. The design analysis of SSAW microfluidics for the separation of particles was presented and The design analysis of SSAW microfluidics for the separation of particles was presented and discussed discussed in detail. The design analysis of SSAW microfluidics was depending on three factors. The in detail. The design analysis of SSAW microfluidics was depending on three factors. The first factor first factor was the setup properties of SSAW microfluidics such as: piezoelectric substrate material, was the setup properties of SSAW microfluidics such as: piezoelectric substrate material, input power, input power, wavelength, length of IDTs, width of the main microchannel, widths of outlet wavelength, length of IDTs, width of the main microchannel, widths of outlet collecting microchannels, collecting microchannels, working area of SAW, working time of SAW, and flow rate of fluid. The working area of SAW, working time of SAW, and flow rate of fluid. The second factor was the properties second factor was the properties of the fluid such as: density, compressibility, and viscosity. The of the fluid such as: density, compressibility, and viscosity. The third factor was the properties of the third factor was the properties of the particles such as: density, compressibility, volume, and radius. particles such as: density, compressibility, volume, and radius. The design analysis of SSAW microfluidics provided the selection of the piezoelectric substrate The design analysis of SSAW microfluidics provided the selection of the piezoelectric substrate material, width of the main microchannel, working area of SAW, wavelength, minimum input material, width of the main microchannel, working area of SAW, wavelength, minimum input power, power, and widths of outlet collecting microchannels required for the separation process. The and widths of outlet collecting microchannels required for the separation process. The minimum minimum input power required to shift the maximum volume of the particles to the side wall of the input power required to shift the maximum volume of the particles to the side wall of the main SSAW main SSAW microchannel at the end of the working time of SSAW was calculated by the simulation microchannel at the end of the working time of SSAW was calculated by the simulation programs and programs and used to decrease the power consumption. used to decrease the power consumption. The relations of the minimum input power required to shift the particle to the side wall of the The relations of the minimum input power required to shift the particle to the side wall of the main SSAW microchannel at the end of the working time of SSAW, radius of the particle, and the main SSAW microchannel at the end of the working time of SSAW, radius of the particle, and the acoustic contrast factor were presented. The width of the main SSAW microchannel was selected as acoustic contrast factor were presented. The width of the main SSAW microchannel was selected 2λ/5 to avoid the slightly moving of the pressure nodes due to the attenuation effect of SAW. The as 2λ/5 to avoid the slightly moving of the pressure nodes due to the attenuation effect of SAW. The design analysis of SSAW microfluidics achieved the minimum input power required for the

Bioengineering 2017, 4, 28

15 of 16

separation process with appropriated displacement contrast of the particles. The relations between the difference of particle radius (∆R) and the difference of the displacement of particle from the center of the main SSAW microchannel (∆X) at selected values of wavelengths were presented. The final design of SSAW microfluidics for the separation of blood components (WBCs, RBCs, and platelets) was analyzed. The width and height of the main SSAW microchannel were 120 (2λ/5) and 25 µm, respectively. The lengths of the main SSAW microchannel and IDTs were 15 mm and 9 mm, respectively. The SAW wavelength (λ) was 300 µm. The SAW piezoelectric substrate material was LiNbO3 . The flow rates in the input inlet microchannel and the two symmetric side microchannels were 0.2 µL/min and 0.8 µL/min, respectively. The mixture of blood components will flow in the input inlet microchannel. The distilled water was used as a fluid. The working time of SSAW force was about 0.9 s. The minimum input power required for the separation of blood components was 371 mW. All simulation programs were built by MATLAB. The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the design analysis of SSAW microfluidics was presented. The analysis of SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force was discussed and analyzed. The displacement of WBCs, RBCs, and platelets at the end of the working time of SSAW force due to the SSAW force with Rayleigh angle effect and its attenuation in liquid-loaded substrate, viscous drag force, hydrodynamic force, and diffusion force were 60, 29, and 9 µm, respectively. The separated particles were simply collected by five outlets microfluidics. The widths of the five outlets microfluidics were determined. The width of middle outlet microchannel for platelets collection was about 30 µm. The width of two mid-lateral outlet microchannels for RBCs collection was about 20 µm. The width of two lateral outlet microchannels for WBCs collection was about 25 µm. The simulation of the separation of blood components (WBCs, RBCs, and platelets) based on the design analysis of SSAW microfluidics and the SSAW force only was presented. Under the same simulation conditions, this simulation presented obviously the importance of other forces and effects. The SSAW force only was not enough to provide accurate simulations. This study provided accurate design and simulations for the separation process. The results of these simulations achieved minimum power consumption, less complicated setup, and high collecting efficiency compared to other published papers. This analysis study was significant and powerful for the simulation of the separation process especially for the separation of biological particles. Author Contributions: Ahmed M. Soliman, Mohamed A. Eldosoky, and Taha E. Taha conceived, designed the study, performed the simulations, analyzed the data, and wrote the paper. Conflicts of Interest: The authors declare no conflict of interest.

References 1.

2.

3.

4.

Taha, T.; El-Dosoky, M.; El-Sayed, A. On the analysis of the interaction between surface acoustic wave (SAW) and adjacent media. In Proceedings of the 24th National Radio Science Conference (NRSC), Cairo, Egypt, 13–15 March 2007. Tang, Y.; Li, Z.; Ma, J.; Wang, L.; Yang, J.; Du, B.; Yu, Q.; Zu, X. Highly sensitive surface acoustic wave (SAW) humidity sensors based on sol–gel SiO2 films: Investigations on the sensing property and mechanism. Sens. Actuators B Chem. 2015, 215, 283–291. [CrossRef] Taha, T.; El-Dosoky, M.; El-Sayed, A. Estimation of Mixing Ratios of Three Component Mixture Using Shear Horizontal Surface Acoustic Wave Sensor and Neural Network. J. Univ. Chem. Technol. Metall. 2009, 44, 307–312. Taha, T.; El-Dosoky, M.; El-Sayed, A. The Effect of Loss Tangent on the Perturbation of Shear Horizontal Surface Acoustic Wave (SH-SAW) Biomedical Sensors in Case Human Blood and Glucose. J. Univ. Chem. Technol. Metall. 2009, 44, 98–102.

Bioengineering 2017, 4, 28

5.

6. 7. 8. 9.

10. 11. 12. 13. 14.

15. 16. 17. 18. 19. 20. 21.

22. 23. 24.

25.

16 of 16

Soliman, A.; Eldosoky, M.; Taha, T. Modelling and Simulation of Microparticles Separation using Standing Surface Acoustic Waves (SSAWs) Microfluidic Devices for Biomedical Applications. Int. J. Comput. Appl. 2015, 129, 30–38. Soliman, A.; Eldosoky, M.; Taha, T. Analysis Improvement of Standing Surface Acoustic Wave (SSAW) Microfluidic Devices for Bio-Particles Separation. Int. J. Comput. Appl. Technol. 2017, 55. accepted. Nam, J.; Lee, Y.; Shin, S. Size-dependent microparticles separation through standing surface acoustic waves. Microfluidics Nanofluidics 2011, 11, 317–326. [CrossRef] Jo, M.; Guldiken, R. Active density-based separation using standing surface acoustic waves. Sens. Actuators A Phys. 2012, 187, 22–28. [CrossRef] Chan, M.; Guldiken, R. Acoustic microfluidic platform for size and density-based cell separation. In Proceedings of the 2011 IEEE International of Ultrasonics Symposium (IUS), Orlando, FL, USA, 10–21 October 2011. Nam, J.; Lim, H.; Kim, D.; Shin, S. Separation of platelets from whole blood using standing surface acoustic waves in a microchannel. Lab Chip 2011, 11, 3361–3364. [CrossRef] [PubMed] Ai, Y.; Sanders, C.; Marrone, B. Separation of Escherichia coli Bacteria from Peripheral Blood Mononuclear Cells Using Standing Surface Acoustic Waves. Anal. Chem. 2013, 85, 9126–9134. [CrossRef] [PubMed] Jo, M.; Guldiken, R. Two-stage microfluidic device for acoustic particle manipulation. In Proceedings of the Smart Biomedical and Physiological Sensor Technology VIII, Orlando, FL, USA, 25 April 2011. Li, X.; Zhou, Y. Microfluidic Devices for Biomedical Applications; Woodhead Publishing: Cambridge, UK, 2013. Franke, K.; Ross-Messemer, M.; Menck, A.; Hoeller, R.; Schmidt, H.; Weilmacht, M. The highly sensitive optical measurement of absolute SAW amplitudes for power flow analysis. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2003, 50, 77–80. [CrossRef] [PubMed] Guldiken, R.; Jo, M.; Gallant, N.; Demirci, U.; Zhe, J. Sheathless Size-Based Acoustic Particle Separation. Sensors 2012, 12, 905–922. [CrossRef] [PubMed] Vila, R. Waves in Fluids and Solids; In Tech: Rijeka, Croatia, 2011. Shi, J.; Huang, H.; Stratton, Z.; Huang, Y.; Huang, T. Continuous particle separation in a microfluidic channel via standing surface acoustic waves (SSAW). Lab Chip 2009, 9, 3354–3359. [CrossRef] [PubMed] Devendran, C.; Albrecht, T.; Brenker, J.; Alan, T.; Neild, A. The importance of travelling wave components in standing surface acoustic wave (SSAW) systems. Lab Chip 2016, 16, 3756–3766. [CrossRef] [PubMed] Nyborg, W. Acoustic Streaming due to Attenuated Plane Waves. J. Acoust. Soc. Am. 1953, 25, 68–75. [CrossRef] Jiao, Z.; Huang, X.; Nguyen, N. Scattering and attenuation of surface acoustic waves in droplet actuation. J. Phys. A Math. Theor. 2008, 41, 355502. [CrossRef] Rahim, M.; Selamat, N.; Yunas, J.; Ehsan, A. Effect of geometrical shapes on 3D hydrodynamic focusing of a microfluidic flow cytometer. In Proceedings of the 12th IEEE International Conference on Semiconductor Electronics (ICSE), Kuala Lumpur, Malaysia, 17–19 August 2016. Lee, G.; Chang, C.; Huang, S.; Yang, R. The hydrodynamic focusing effect inside rectangular microchannels. J. Micromech. Microeng. 2006, 16, 1024–1032. [CrossRef] Laurell, T.; Lenshof, A. Microscale Acoustofluidics; The Royal Society of Chemistry: London, UK, 2015. Dykes, J.; Lenshof, A.; Åstrand-Grundström, I.; Laurell, T.; Scheding, S. Efficient Removal of Platelets from Peripheral Blood Progenitor Cell Products Using a Novel Micro-Chip Based Acoustophoretic Platform. PLoS ONE 2011, 6, e23074. [CrossRef] [PubMed] Petersson, F.; Åberg, L.; Swärd-Nilsson, A.; Laurell, T. Free Flow Acoustophoresis: Microfluidic-Based Mode of Particle and Cell Separation. Anal. Chem. 2007, 79, 5117–5123. [CrossRef] [PubMed] © 2017 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).