Tacrolimus Measurement: Building a Better Immunoassay Verena Gounden1 and Steven J. Soldin1,2*

Tacrolimus (formerly known as FK506) is a macrolide immunosuppressant that is used for primary immunosuppression after organ transplantation (1 ). The landmark work of Starzl at the University of Pittsburgh paved the way for the clinical use of tacrolimus as a primary immunosuppressant, leading to a shift away from cyclosporine use (2– 4 ). Today, tacrolimus is used in the majority of transplant patients (5 ). Tacrolimus binds to several cytosolic immunophilins of importance; binding to the FK506 binding protein leads to the formation of a complex that interacts with calcium-dependent calcineurin calmodulin translocation pathways that result in prevention of T-cell activation (6 ). Monitoring blood tacrolimus concentrations is essential in assessing organ rejection and toxicity because of its narrow therapeutic index, wide inter- and intraindividual pharmacokinetic variability, and susceptibility to cytochrome P450 –mediated drug interactions (7 ). Immunoassays play a major role in the analysis of many clinical laboratory analytes and in recent years have become more popular for measuring tacrolimus, replacing the mass spectrometric methods that were initially favored over the earlier immunoassay methods. Measurement of drugs by immunoassay offers the advantages of automation and ease of use (assays available around the clock 7 days a week). Furthermore, this approach does not require specialized trained staff, has the advantage of using commercially available calibrators and reagents, and permits random-access testing. Because of these factors, immunoassays allow for reasonably fast turnaround times, particularly for the afterhours testing often needed for assessment of the clinical status of a transplant patient. One of the main disadvantages of immunoassays is the lack of analytical specificity due to cross-reactivity with drug metabolites (active and inactive) and other analytes (8 ). For ta-

1 Department of Laboratory Medicine, National Institutes of Health Clinical Center, Bethesda, MD; 2 Department of Medicine, Georgetown University, Washington, DC. * Address correspondence to this author at: Department of Laboratory Medicine, National Institutes of Health Clinical Center, Bldg. 10, Rm. 2C-249, Bethesda, MD 20892. Fax 301-402-1885; e-mail [email protected], or [email protected]. Received December 31, 2013; accepted January 3, 2014. Previously published online at DOI: 10.1373/clinchem.2013.220012

crolimus the cross-reactivity of the metabolites in most current immunoassays is still substantial (up to 60%), which can cause an unpredictable overestimation of the true drug concentrations (8, 9 ). In contrast, mass spectrometry can be less robust than immunoassay platforms and requires highly trained staff, but it is capable of measuring several compounds in a single run. However, in hospitals mass spectrometry is often not available for afterhours use. Although major improvements in mass spectrometry methods have simplified sample preparation and provided faster turnaround times, the above-mentioned limitations still persist (10 ). In this issue of Clinical Chemistry, Wei et al. elegantly describe the measurement of tacrolimus using 2 antitacrolimus monoclonal antibodies in a sandwich assay that has minimal metabolite interference (11 ). Tacrolimus is a ring-structured hapten. The structure of such haptens makes interaction with 2 binding proteins (e.g., antibodies) very unlikely because of steric hindrance that occurs following binding of a hapten to one binding protein. Because of this phenomenon, the immunoassay platforms for the detection of such hapten drugs are generally competitive immunoassays. Antibody specificity is a major concern for competitive immunoassays. Various metabolites, including M-I (13-O-demethyl), M-II (31-O-demethyl), M-III (15O-demethyl), and M-V (15,31, di-O-demethyl) crossreact with immunoassays for tacrolimus (8 ). Many of the different immunoassays currently available from various manufacturers have been reported to produce tacrolimus values higher than those for HPLC–tandem mass spectrometry (MS/MS).3 The microparticle enzyme immunoassay (MEIA) produces results that are 15– 20% higher, and EMIT (enzyme multiplied immunoassay) and CEDIA (cloned enzyme donor immunoassay) results are reported to be 17% and 19% higher, respectively (12 ). Low hematocrit values also have been reported to give false-positive tacrolimus concentrations using the MEIA assay (13 ). In addition, competitive immunoassays have a limited dynamic range. Yet Wei et al. dem-

3

Nonstandard abbreviations: MS/MS, tandem mass spectrometry; MEIA, microparticle enzyme immunoassay; LC-MS/MS, liquid chromatography–MS/MS.

575

Editorials onstrate in this article that their newly developed sandwich immunoassay has a similar dynamic range (1–30 ng/mL) to liquid chromatography–MS/MS (LC-MS/ MS) methods, with negligible bias and markedly less metabolite cross-reactivity compared to other commercially available competitive immunoassays (11 ). It should be noted on review of the Bland–Altman plots presented by the authors that there appears to be an increased deviation from LC-MS/MS values at concentrations greater than or equal to 16 ng/mL, although this is perhaps not of clinical significance. The approach used by Wei and colleagues in this sandwich immunoassay for tacrolimus measurement opens the door to the development of similar assays for other drugs that have potentially cross-reactive metabolites. Examples of such drugs include cyclosporine, sirolimus, digoxin, and phenytoin. Immunoassay technology has evolved from the original assays of 50 years ago. Many new technologies allow for improvement of immunoassays, with the development of more specific antibodies and with better affinity to the target antigen/hapten. The advent of in vitro antibody display technologies provides the ability to engineer new or augment specific attributes of interest, such as affinity, specificity, or stability, into antibody molecules (14 ). Software developments allowing for 3-D modeling of interactions between antibodies and the antigen also serve to improve immunoassay development. However, there remain challenges for immunoassays in areas beyond the quantification of drugs. These challenges include the measurement of total and free thyroid hormones and steroids. In general, immunoassays are unreliable when used for quantifying low concentrations of steroids such as cortisol, testosterone, and estradiol. Improvement in quantification using a similar approach to that employed by Wei et al. is urgently needed. A longitudinal replicate study of immunosuppressive drugs performed by the College of American Pathologists showed substantial imprecision for the immunoassays, which accounted for 77% to 90% of the total variance. The major source of imprecision came from within-laboratory factors, with tacrolimus showing large between-year within-laboratory variation (15 ). This is likely to be a result of lot-to-lot changes in the immunoassay reagent. The development of more specific immunoassays does not ensure the prevention of lot-to-lot variability. The onus remains on the manufacturer to employ strict QC in the manufacture of the reagents as well as testing for minimal variability between lots and monitoring shifts in QC ranges among customers using the product. Nevertheless, the devel-

576 Clinical Chemistry 60:4 (2014)

opment of this novel sandwich assay for the measurement of tacrolimus by Wei and colleagues represents a milestone in immunoassay development that will hopefully assist in improving immunoassay measurement of drugs and other analytes.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article. Authors’ Disclosures or Potential Conflicts of Interest: No authors declared any potential conflicts of interest.

References 1. Hong JC, Kahan BD. Immunosuppressive agents in organ transplantation: past, present and future. Semin Nephrol 2000;20:108 –25. 2. Todo S, Fung JJ, Demetris AJ, Jain A, Venkataramanan R, Starzl TE. Early trials with FK506 as primary treatment in liver transplantation. Transplant Proc 1990;22:13– 6. 3. Starzl TE, Fung JJ, Jordan M, Shapiro R, Tzakis A, McCauley J, et al. Kidney transplantation under FK506. Ann Surg 1990;264:63–7. 4. Jusko WJ, Thomson AW, Fung J, McMaster P, Wong SH, Zylber-Katz E, et al. Consensus document: therapeutic monitoring of tacrolimus (FK-506). Ther Drug Monit 1995;17:606 –14. 5. Port FK, Dykstra DM, Merion RM, Wolfe RA. Trend and results for organ donation and transplantation in the United States, 2004. Am J Transplant 2005;5(4 Pt 2):843–9. 6. Dumont FJ, Saruch MJ, Koprak SL, Melino MR, Sigal NH. The immunosuppressive macrolides FK-506 and rapamycin act as reciprocal antagonists in murine T cells. J Immunol 1990;144:1418 –24. 7. Shaw LM, Holt DW, Keown P, Venkataramanan R, Yatskoff RW. Current opinions on therapeutic drug monitoring of immunosuppressive drugs. Clin Ther 1999;21:1632–52. 8. Murthy JN, Davis DL, Yatskoff RW, Soldin SJ. Tacrolimus metabolite crossreactivity in different tacrolimus assays. Clin Biochem 1998;31:613–7. 9. Salm P, Taylor PJ, Clark A, Balderson GA, Grygotis A, Norris RL, et al. High performance liquid chromatography-tandem mass spectrometry as a reference for analysis of tacrolimus to assess two immunoassays in patients with liver and renal transplantation. Ther Drug Monit 1997;19:694 –700. 10. Volosov A, Napoli KL, Soldin SJ. Simultaneous simple and fast quantification of three major immunosuppressants by liquid chromatography–tandem mass-spectrometry. Clin Biochem 2001;34:285–90. 11. Wei TQ, Zheng YF, Dubowy M, Sharma M. Sandwich assay for tacrolimus using 2 antitacrolimus antibodies. Clin Chem 2014;60:621–30. 12. Butch AW. Immunosuppressive drugs: pharmacokinetics, preanalytic variables, and analytical considerations. In: Dasgupta A, ed. Handbook of drug monitoring methods: therapeutic and drugs of abuse. 1st ed. Clifton (NJ): Humana Press; 2008. p 165–93. 13. Armedariz Y, Garcia S, Lopez RM, Pou L. Hematocrit influences immunoassay performance for the measurement of tacrolimus in whole blood. Ther Drug Monit 2005;27:766 –9. 14. Siegel RW, Baugher W, Rahn T, Drengler S, Tyner J. Affinity maturation of tacrolimus antibody for improved immunoassay performance. Clin Chem 2008;54:1008 –17. 15. Steele BW, Wang E, Soldin SJ, Klee G, Elin RJ, Witte DL. A longitudinal replicate study of immunosuppressive drugs: a College of American Pathologists study. Arch Pathol Lab Med 2003;127:283– 8.

Tacrolimus measurement: building a better immunoassay.

Tacrolimus measurement: building a better immunoassay. - PDF Download Free
60KB Sizes 2 Downloads 0 Views