Downloaded from www.ajronline.org by 50.48.152.100 on 11/11/15 from IP address 50.48.152.100. Copyright ARRS. For personal use only; all rights reserved

1095

Stuart

G. Silverman1

David Steven Glare M. C. Douglass

A. Bloom E. Seltzer Ternpany F. Adams

Needle-Tip CT-Guided

Localization Abdominal

Comparison

of Conventional

OBJECTIVE.

This study

was performed

During Biopsy: and Spiral

to determine

whether

CT

the time required

for

needle-tip localization during biopsy of the abdomen would be reduced if continuousvolume data acquisition, also known as spiral CT, were used for guidance instead of conventional CT.

SUBJECTS

underwent

AND METHODS.

needle-tip

Forty patients had biopsies of an abdominal mass; half with conventional CT and half with spiral CT. The

localizations

times required to localize the needle for 104 needle passes and reconstruction times were included, and the radiologist’s

difficulties

were deliberately

excluded.

tional and spiral CT were compared

The mean needle

were calculated; scanning technique and procedural

localization

for the upper abdominal

times with conven-

regions by using the two-tailed unpaired Student’s t-test. RESULTS. The mean time (±SE) for spiral CT was 35 ± 2 sec compared with 105 ± 18 sec for conventional CT (p < .001). When analyzed by region, times with spiral CT were shorter in both the upper abdomen (means, 37 sec for spiral CT vs 150 sec for conventional CT, p < .001) and pelvis (means, 25 sec for spiral CT vs 74 sec for conventional CT, p = 038); the magnitude of the improvement was greater in the upper abdomen. CONCLUSION. The time required to find the needle tip during guided biopsy of an abdominal mass is reduced with spiral CT compared with conventional CT. This improvement is partly a result of the ability to eliminate respiratory misregistration with spiral CT, which is not possible with conventional multisectional CT; hence the greater advantage in upper abdominal biopsy. AJR

159:1095-1097,

The

usefulness

of CT-guided

‘All authors: Department of Radiology, Brigham and Women’s Hospital, 75 Francis St., Boston, MA 021 1 5. Address reprint requests to S. G. Silverman. 0361 -803X/92/1 0 American

595-1095

Roentgen

Ray Society

and

biopsy

has

been

which occurs

careful

prevents

because

instructions,

level. This results

Presented at the annual meeting of the American Roentgen Ray Society, Orlando, FL, May 1992.

aspiration

misregistration,

the needle after revision

percutaneous

[1 -8]. The principal advantage of CT guidance has been the ability locate the needle tip precisely, an important step in the successful lesions [8-1 1]. Conventional multisectional CT, which consists of obtained during separate breath-holds, can be limited by respiratory

despite

6, 1992; accepted

1992

widely reported to position and sampling of all individual scans registration

Received March May 29, 1992.

November

and pelvic

in “skip”

consistent

of respiratory

some

do not end each

breath-hold, resulting

through

during

precise

requiring rotating

the scanner

reconstructed

are truly contiguous,

that

breath

mis-

cycles,

and

at the same

multisectional CT sequence. of the needle tip can

visualization

multiple detector

such

Respiratory

respiratory

organ or lesion, and consequently

a conventional

misregistration,

and subsequently

images

patients

areas in the imaged

be a time-consuming procedure Spiral CT uses a continuously of the CT table

scanning.

vary between

tip may not be imaged

Because

contiguous

tidal volumes

scans. system data

at selected

free of gaps

and continuous

are acquired

section

or overlapping

during

feeding a single

intervals. sections

The

caused

by respiratory misnegistration [1 2]. We hypothesized that use of spiral CT would reduce the time required to localize the needle tip compared with that needed for

1096

SILVERMAN

conventional

CT

during

CT-guided

biopsy

of

CT (p

abdominal

masses.

and Methods

Downloaded from www.ajronline.org by 50.48.152.100 on 11/11/15 from IP address 50.48.152.100. Copyright ARRS. For personal use only; all rights reserved

conventional

CT-guided

CT was used to localize

biopsy of abdominal

masses,

the needle tip in half the patients

and spiral CT was used in the other half. All biopsies were performed by using the Siemens Somatom Plus CT scanner (Siemens Medical Systems, Iselin, NJ). Our usual routine method (with conventional

was to obtain individual scans, await the result, and then decide

if additional

scans were required.

CT consisted

The localization

of individual

scans

protocol

for con-

(1 0 x 1 0 mm sequence,

2-sec exposure,

120 kVp, 21 0 mA) beginning at the suspected location of the needle tip within the patient; scans were repeated until the needle tip was visualized. The localization protocol for spiral CT consisted of a 10-mm-thick volumetric data acquisition (1 20 kVp, 165 mA) obtained with a table feed rate of 10 mm/sec and beginning 1 cm above the estimated location of the needle tip within the patient. During a 5-sec exposure a total volume of 5 cm was scanned during a single breath-hold. Images were then reconstructed by using a 10mm slice thickness

and 1 0-mm

intervals;

the first and last slices were

yielding a 3-cm volume ofimaged tissue. The interventional radiology team (including a staff radiologist and a resident or fellow) aligned the scan sections for both groups by using the localizer light and was instructed to repeat the scan or scan sequences until the needle tip, as indicated by the low-attenuation beam-hardening artifact, was detected [8]. The time required to localize the needle, defined as the interval between when the technologist begins the scan and the radiologist discarded,

views the needle tip, was calculated for 104 needle placements in the 40 patients. The time for each conventional

CT sequence

a 3-sec start-up delay, 2 sec for the first scan, the time

included

between

the

first localization image and the image showing the needle tip (which includes scanning times for the remainder of the images and reconstruction times for all but the final image showing the needle tip), and the reconstruction time for the last image (8 sec). The time for each spiral sequence included a 3-sec start-up delay, a 5-sec scan time, and a time determined by the number of images necessary to visualize the needle tip multiplied by 1 1 sec (the reconstruction time, per image, of the software for this spiral CT scanner). The time determination deliberately excludes operator technique and lesion size and location, which generally determine the number of needle passes. It also excludes variables such as the skill of the technologist and the medical

condition “upper

abdomen,”

including

the

liver,

two

spleen,

the two-tailed

The

needle

use

of spinal

localizations

CT

in biopsies

pelvic organs and of netropenitoneal lymph nodes (mean time for spiral CT, 25 sec; mean time for conventional CT, 74 sec). The improvement accorded by using spiral CT was greaten in the upper

abdomen

group

than

in the netnopenitoneum/pelvis

group. When spiral CT was used, upper abdominal

needle-tip

localizations took an average of only 1 2 sec longer than those in the pelvis, but there was still a statistically significant difference between the two regions (p = .001). With conven-

tional CT, the time required for localization of the needle tip was longer for masses in the upper abdomen than for pelvic masses on netnopenitoneal lymph nodes. Discussion Compared

with

sonography

of CT-guided

abdominal

and procedure

length.

on fluonoscopy,

biopsy

include

disadvantages

its cost,

The accessibility

availability,

of the lesion

and the

radiologist’s skill are important determinants of overall procedure times. In addition, biopsy with conventional CT requines intermittent imaging between needle manipulations, a disadvantage not associated with real-time sonography on fluoroscopy. The results of our study suggest that continuous-volume data acquisition, although it does not eliminate the need to image the needle after placement, reduces the time needed

to perform

this step compared

with conventional

average amount of time saved by using spinal CT was approximately 70 sec for each needle placement. There are several reasons for this improvement. CT.

The

In conventional abdomen, where

CT-guided respiratory

cant, the times required than those for biopsies lymph nodes. Elimination

TABLE 1: Needle Abdominal Biopsy

biopsies of organs in the upper misnegistration may be signifi-

for needle localization were longer of pelvic organs or retroperitoneal of respiratory misnegistration pnob-

Localization

Times

During

CT-Guided

unpaired

Student’s

No. of Sec ± SE (No. of Needle Placements)

groups: those adrenal

Mean

gland,

kidney, and pancreas, and those in the “pelvis,” including pelvic masses as well as retroperitoneal lymph nodes. These assignments, although admittedly arbitrary, are based on the hypothesis that respiratory misregistration is greater during biopsies of the upper abdominal organs than during biopsies of pelvic organs and retroperitoneal lymph nodes, where little or no respiratory misregistration should occur. Mean needle localization times with conventional and spiral CT were compared for each organ and abdominal region by using

shortened the time of upper abdom(mean time for spinal CT, 37 sec; mean time for CT, 1 50 sec) and biopsies of masses in the

for

of the patient.

Data were analyzed by dividing the biopsies into in the

.001).

Self-expanding metal stents for palliative treatment of malignant ureteral obstruction.

We studied the immediate and long-term efficacy of the Wallstent device in the treatment of ureteral obstruction caused by malignant disease...
703KB Sizes 0 Downloads 0 Views