Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
297
MR Imaging of the Portal Venous System: Value of Gradient-Echo Imaging as an Adjunct
Paul M. Silverman1 Richard H. Patt Brian S. Garra Steven C. Horii Cirrelda Cooper Wendolyn S. Hayes Robert
K. Zeman
to Spin-Echo
Imaging
We evaluated the use of gradient-echo (GRE) as an adjunct to spin-echo (SE) MR imaging of the portal venous system. GRE imaging was performed in 31 subjects, 15 normal volunteers and 16 patients with documented portal venous disease (15 cases) or suspected
disease
(one
case).
Eight
of 16 patients
had complete compression by tumor. Of the two other the other a falsely positive angiogram, focal
thrombus,
subjects,
GRE
intravascular subjects
signal
and
and
that
identified
scans
had
signal three
excellent
patients
could
visualization
compared
with
had
mimic
clot.
of the
surrounding
an artifact
In three
on GRE but not on SE images.
had
venous
thrombosis,
five
had
occlusion. Six patients had extrinsic venous patients, one had an arteriovenous fistula and suggesting portal vein occlusion. In normal
three
consisting
of five
portal
venous
tissues.
Nine
system
(60%)
of a curvilinear
patients
with
In all three patients
area
focal
with
high
of 15 normal of decreased
thrombus,
with occlusion,
clot
was
SE and GRE
images demonstrated similar findings. In five of the six patients with extrinsic venous compression by tumor, SE and GRE studies showed similar findings. Of the other two patients, an arteriovenous fistula was seen on GRE MR in one, and in the other, patency of the left portal vein was seen on SE and GRE images after angiography had suggested portal vein occlusion. Collateral vessels were seen in nine of 16 patients. In five of nine cases, GRE MR demonstrated more extensive collaterals than did SE MR. In summary,
GRE
MR
provides
a useful
adjunct
to standard
include high contrast between vascular structures and motion artifact, and rapid scanning within a breath-hold. AJR
157:297-302,
August
SE
MR
surrounding
imaging.
Benefits
tissues,
reduced
1991
Noninvasive accomplished
imaging of the portal venous system (PVS) has been successfully with a variety of imaging techniques, including sonography, CT, and MR imaging [1 -8]. Doppler sonography has proved to be highly accurate in determining the presence, direction, and velocity of portal blood flow, but is operator dependent [1 , 9-1 1 ]. Although contrast-enhanced CT with dynamic scanning is sensitive in identifying thrombus in the main portal vein as well as large branch vessels, suboptimal opacification of portal vessels, occasionally may limit this examination.
(SE) MR provides
Spin-echo
without
the
necessity
related artifacts, in the Received December revision March 7, 1991.
14, 1990:
accepted
after
clearly
‘All authors: Department of Radiology, Georgetown University Hospital, 3800 Reservoir Rd., NW., Washington, DC 20007. Address reprint requests
to P. M. Silverman. 0361-803X/91/1572-0297 © American
Roentgen
quences vascular
Ray Society
of using
including
anatomic
information
IV contrast
flow-related
enhancement,
portal
vein that mimic thrombus [1 2-14]. have been used in the abdomen and
structures distinguishes
[2, 1 5, 1 6]. The high-intensity the
intraabdominal
similar to that provided
material.
vasculature
However,
may produce
motion
spurious
Gradient-echo (GRE) peripheral extremities
signal produced from
signals pulse Seto image
by flowing
surrounding
by CT
and flow-
blood
stationary
tissues. Additionally, the acquisition time is significantly reduced when using GRE imaging as compared with conventional SE imaging. This allows slice acquisition during a single breath-hold, minimizing motion-related artifact. We have reviewed our experience with GRE imaging of the portal vein in 31 subjects, 1 5 normal volunteers and 1 6 patients with portal vein abnormalities. The
purpose
of the study
adjunct
to standard
Subjects
GRE imaging
as an
including patients
subjects
underwent
MR
imaging
of the
upper
abdomen
the PVS. Fifteen were normal volunteers and 1 6 were in whom portal venous abnormalities were proved on CT (13
teers, 1 0 men and five women, were 1 5-55 years old. The 16 patients, 1 1 men and five women, were 43-77 years old (mean, 60 years). Eight of the sixteen patients had intrinsic thrombus in the
PVS. Five patients had focal thrombus occlusion.
Two
of these
and three had complete
patients
had
cirrhosis.
Six
portal
patients
had
extrinsic compression of the main or proximal right or left portal vein by tumor. Abnormalities included metastases (two patients), pancreatic neoplasm (two patients), lymphoma (one patient), and hepatoma (one patient). One patient had an arteriovenous fistula and the
other afalse-positive
angiogram
suggesting
occlusion
ofthe left portal
vein. All 31 subjects underwent GRE imaging of the portal vein performed at 1 .5 T (Siemens, Magnetom, Erlangen, Germany). Fast imaging with steady-state free precession (FISP) was performed in
all 31 cases. First-order
(velocity)
flow compensation
the slice and frequency-encoding included 50/1 4/45#{176}/i (TR/TE/flip acquisition
matrix,
and
a 5-mm
sec for each breath-hold holding their fast-low-angle
FLASH acquisition
was applied in
directions. FISP scan parameters angle/excitation), a 256 x 256 slice
thickness.
Imaging
time
was
16
image. Five of the patients who had difficulty
breath for this time had additional GRE imaging with shot (FLASH) imaging, requiring only 9 sec per image.
technical
parameters
matrix,
and
included
a 5-mm
slice
30/i 0/30#{176}/i , a 128 x 128 thickness.
Maximum-intensity-
projection (MIP) angiographic MR images in the coronal plane were produced in eight patients to supplement standard GRE images. MIPs were calculated on a 256 x 256 grid from a stack of twodimensional GRE slices. Images were viewed at rotational angles from -30#{176}to +30#{176} from the coronal plane at fixed 6#{176} intervals. Reconstruction time was approximately 5 sec per view for a total time
of approximately
1 mm depending
on the
number
of raw
images
used.
Standard
SE imaging was performed 256 x 256 matrix, and a routine 1 0-mm
coverage Thirteen
by using 500/i 5/4 or 8, a thickness to provide enough
to include the entire liver. Six of the 15 volunteers
of the i 6 patients of the
underwent i
and iS
SE imaging.
6 patients
underwent
with a GE 9800 scanner (General imaging. In all cases, enhanced
CT of the upper abdomen
Electric, Milwaukee, WI) before MR scans were obtained dynamically
with a power injector, i .0-i .2 mI/sec (Mark IV, Medrad, Pittsburgh, PA). Scans were obtained at i -cm intervals as part of the standard examination sonographic
of the
upper
evaluation,
abdomen. including
Eight color
of the Doppler
patients flow
also
had
assessment
(Ultramark 9, Advanced Technology Laboratories, Bothell, WA) and four patients had angiographic assessment of the liver and PVS. GRE and SE scans in the volunteers were reviewed to assess the normal appearance of the main portal vein and its major radicles.
GRE scans were also specifically
assessed
for any potential
artifacts
that might be confused with thrombus. In the i 6 patients with portal vein abnormality, GRE imaging was compared with SE imaging in i 5 of i6 patients and with CT scans in 1 3 of i 6 patients. Scans were compared to assess the presence and extent of intrinsic portal vein thrombosis, extrinsic vascular compression, and portal vein collaterals. Detailed statistical analysis was not feasible because of the small number of patients studied. In the five patients with thrombus, the CT criterion for thrombus was a focal
area of decreased hancement.
In three
to identify
in the other
cases), Doppler sonography (eight cases), and angiography (four cases). Angiography was false-positive in one case. The 1 5 volun-
vein
inability
density patients,
in the PVS surrounded total occlusion
a patent
main
portal
vein
and
numerous
enhancing
periportal collaterals. In the six patients with extrinsic portal venous compression, CT demonstrated porta hepatis adenopathy in four (two with pancreatic neoplasm, one with metastatic disease, and one with lymphoma) and intrahepatic metastases and primary hepatoma
of the PVS.
and Methods
Thirty-one
Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
was to evaluate
SE imaging
by contrast
was established
en-
by the
two.
In one
patient,
an arteriovenous
fistula
was
confirmed
by angiography; in the other, a false-positive angiogram suggested occlusion of the left portal vein. The false appearance of occlusion
was related to a portion of the vessel being seen “end on” angiographically and confirmed nine
of the
i
to be patent
6 patients,
by CT and Doppler
collateral
sonography.
In
vessels were seen on at least one
imaging study (CT, sonography, or angiography). In patients with collaterals, the extent of collateralization, including periportal, mesenteric, perisplenic, retroperitoneal, and pencholecystic collaterals, was evaluated on SE MA, GRE MA, and CT. When Doppler sonography
(eight
patients)
or angiography
(four
patients)
was
performed,
the information from these studies agreed with CT findings. In all patients, MA, CT, and Doppler sonographic examinations were performed within 2 weeks of one another. All studies were reviewed by two
observers,
with
tion consisted parison with
performed
consensus
achieved
in each
case.
Initial
evalua-
of direct assessment of SE MA and GRE MR. ComCT, Doppler sonography, and angiography was also
when these examinations
were available.
Results
GRE imaging was performed in all normal volunteers and showed high signal intensity in all the major intraabdominal vascular structures, including the PVS. In all cases, the main, left, and right
portal
veins
were
subjects,
a linear
or curvilinear
intensity
was noted
in the portal
imaged
successfully.
In nine
area of decreased
signal
vein (Fig. 1). In four cases,
it
was isolated to the main portal vein, and in five cases it extended into the right or left branches of the main portal vein.
Standard
SE imaging,
performed
in six of nine
volun-
teers, confirmed this as an artifact by demonstrating a normal PVS. Supplemental scanning in the coronal plane also confirmed the artifactual nature of this decreased signal. Intrinsic PVS thrombus or occlusion was demonstrated in eight of 1 6 patients. When thrombus was seen on SE images,
it appeared as an area of high signal intensity replacing the normal flow void in the portal vein; on GRE images, it was seen as an area of decreased signal these areas of clot were separate
intensity. On SE images, from regions of artifact
generated from pulsatile flow within adjacent vessels. In three of the five patients with focal thrombus, thrombus was demonstrated on GRE imaging but failed to be confidently identified on SE imaging (Figs. 2 and 3). In one case in which CT clearly demonstrated thrombus in the main portal vein and superior mesenteric vein, GRE and SE imaging were not conclusive. In the other case, thrombus was identified on both
SE and GRE MR. In the three cases with complete
occlusion
of the portal vein, SE and GRE imaging showed similar findings, including failure to identify the main portal vein and numerous periportal collaterals (Fig. 4). Two of these patients had findings consistent with a diagnosis of cavernous trans-
formation
of the portal vein.
Six patients had extrinsic compression of the main or proximal right or left portal vein by tumor. In five of six patients with extrinsic portal vein compression, similar findings were
demonstrated
on SE and GRE imaging
with
5). In one case,
identify
CT (Fig.
the extrinsic
on SE imaging.
compression
Although
GRE
that correlated
imaging
of the
right
was portal
GRE and SE imaging
well
unable vein
to
seen
both showed
Fig. 1.-Artifact on axial GRE MR image. A, Axial GRE 50/14/45#{176} MR image of normal volunteer shows curvilinear artifact in main and right portal veins (arrowheads). B, Coronal GRE image does not substantiate any consistent defect in portal venous system.
Subtie lucencies
Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
signal
Intensity
lack sharp margins seen
and low
with thrombus.
Fig. 2.-Portal vein thrombus equivocal on SE but diagnostic on GRE MR images. A, CT scan at level of main portal vein (arrow) shows low attenuation in main portal vein consistent with thrombus. B, SE 500/15 MR image shows spurious signal within aorta (curved arrow). Lack of clear flow void in aorta and main portal vein (straight arrow) does not allow confident diagnosis of portal vein thrombosis. C, GRE 50/14/45#{176} MR image. High signal in aorta (curved arrow) indicates flowing blood, and distinct lack of signal in main portal vein (straight arrow) is consistent with thrombosis. Dramatic contrast between flowing blood and clot allows a more confident diagnosis of portal vein thrombosis.
Fig. 3.-Portal vein thrombus identified on GRE but not SE MR images. A, CT scan shows clot in portal vein (solid arrows) with patent lumen medially (arrowhead). Subtle evidence of perisplenic arrows). B, SE 500/15 MR image shows artifact. Clot could not be diagnosed confidently. Note high-signal artifact (arrowhead) in area CT and Doppler sonography and a relative signal void in area of documented clot (arrows). Perisplenic collaterals are not seen. C, GRE 50/14/45#{176}image shows findings similar to those on CT with clot peripherally (solid arrows) and a patent lumen medially collaterals are best seen on GRE image (open arrows).
collaterals with
is seen
documented
(arrowhead).
(open flow
on
Perisplenic
SILVERMAN
Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
300
ET
AL.
AJR:157,
Fig. 4.-Occluded main portal vein and extensive collaterals seen on SE and GRE MR images. A, CT scan shows extensive periportal collaterals (solid arrows) due to occlusion of main portal vein. Low-attenuation areas disease. Retroperitoneal, mesenteric, and perisplenic collaterals are present (open arrows). B and C, SE 500/15 (B) and GRE 50/14/45#{176} (C) MR images show findings similar to those seen on CT, with numerous arrows). Additional collaterals appear as on CT (open arrows).
in liver represent periportal
August
1991
metastatic
collaterals
(solid
Fig. 5-Extrinsic compression of portal vein by lymphoma. GRE compared with SE MR shows poorer contrast between tumor and liver. A, SE 500/15 MR image shows lymphomatous tumor (1) to port. hepatis with compression of main portal vein (arrow). Tumor (T) is of low signal intensity. B, GRE 50/14/45#{176} image shows findings similarto those seen on SE MR with very mild portal vein compression (arrow). However, contrast between tumor (T) and adjacent liver is not as good.
extrinsic displayed
portal vein compression the tumor responsible
in five of six patients, GRE for the venous compression
less well than SE imaging did because of the poorer contrast between tumor and surrounding normal tissue on GRE images. In one patient, an arteriovenous fistula seen on GRE imaging
was confirmed
false-positive vein was
angiographically.
angiogram
shown
The patient
suggested
to have a patent
occlusion
in whom
a
of the left portal
vein on both
SE and GRE
MR.
Incidentally noted in three of 16 patients was focal signal within the main portal vein on standard SE imaging unrelated to areas of thrombus or vascular compression that potentially could have given the false impression of thrombus. In each
case, the GRE study, CT, or color Doppler flow sonography was normal in this segment of portal vein. The linear or curvilinear low-intensity area encountered on GRE images in normal volunteers was seen similarly in the patient group but was readily recognized as artifactual by its typical appear-
ance, correlation
In nine of the 1 6 patients,
at least one imaging
collateral
technique
vessels
were
(CT, sonography,
shown
by
In four cases, GRE and SE images were judged to be in showing these vessels. In five cases, GRE images
better
defined
presence
of collaterals
and
showed
better.
MIP imaging
did not provide
information
beyond that provided by GRE imaging, but coronal images did enhance the display of abnormalities, especially in patients with extensive
collaterals.
with other imaging normal
studies,
and experience
volunteers.
Discussion
more
extensive collateralization than was appreciated on standard SE MR (Fig. 6). CT scans were obtained in six of nine patients. In those patients, CT and GRE showed collaterals to a similar extent in four patients, whereas in two cases GRE showed the collaterals
in examining
or angiog-
raphy). similar
the
gained
The application of MR imaging in evaluating the abdominal vasculature provides a noninvasive imaging technique additional to CT and sonography. Standard assessment with SE imaging produces a typical signal void, “black-blood” phenomenon. The presence of high-intensity signal within the main portal vein is suggestive of thrombus, and the absence of a main
portal
suggestive
vein and demonstration
of complete
of tortuous
portal vein thrombosis
collaterals
is
[1 7, 1 8). The
MR
Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
AJA:157, August 1991
Fig. 6.-Improved
depiction
of collateral
vessels
OF
PORTAL
with GRE MR images
VENOUS
and maximum-intensity
MR images. A, SE 500/15
301
SYSTEM
MR image shows portal vein occlusion (solid arrow) and collaterals (open portal vein mimics ascites but represents a confluence of distinct collaterals. B, GRE 50/14/45#{176} image shows portal vein thrombosis (solid arrow) and more extensive C, MIP angiographic image, coronal plane, shows extensive collateralization (arrows).
projection
arrows)
(MIP)
in periportal
collateralization
images
area. Low-Intensity (open
arrows)
area around
than SE Images
occluded
does.
signal
[20]. The sequential two-dimensional acquisition method is favored over three-dimensional techniques, which are particularly sensitive to motion, as they require minutes of aCquisi-
be undetected.
Respiratory
and motion
strating
a persistently
high
signal
on late echoes
in throm-
between
with SE
create
and therefore
high contrast
compared
major pitfall in this type of imaging is caused by high signal from flow-related enhancement occurring within patent vasculature that mimics thrombus. Clot may rarely lack increased artifacts may also superimpose high signal intensity artifacts onto vascular structures, making it difficult to assess their patency. In one study, 14 patients with portal vein thrombosis were evaluated by CT, SE MR, and sonography [1 9]. Acute or subacute thrombi were hyperintense on both Ti - and T2weighted images. Chronic thrombi (greater than 5 weeks’ duration) were hyperintense in eight of 1 1 cases on T2weighted images. Although all patients had thrombosed portal vessels, MR more effectively showed the extent of thrombosis and periportal collateralization than did CT or sonography. Heavily T2-weighted, triple-echo imaging enhanced the ability to differentiate flow-related artifacts from thrombi by demon-
extremely
angiographic
flowing
blood
and
surrounding tissues. They are occasionally subject to prominent high-intensity artifacts from these vascular structures
tion rather than being accomplished within a single breathhold. In standard SE imaging, spins must remain within the slice long enough to be exposed to both 90#{176} and 180#{176} RF
pulses. With relatively rapid blood flow, washout effects dommate so that high-velocity moving spins flow out of the plane of section, yielding characteristic black-blood images. In contrast,
GRE imaging
does
not require
a 1 80#{176} refocusing
pulse,
and echoes can be received sooner after each excitation. Blood entering a slice is more fully magnetized, with each slice acting as an entry slice with flow-related enhancement. Washout effects are minimized and flowing blood generates a stronger signal. When flow compensation techniques are applied, the phase shifts that would normally result in subsequent dephasing and signal loss are minimized. In most cases, correcting for first-order “velocity” phase shifts is adequate
bosed vessels. These images, however, were significantly compromised by their poor signal-to-noise ratio. In the present study, portal venous thrombosis was seen in eight patients. In three of five patients, GRE MR showed focal thrombus not appreciated on the SE MR image. In the three patients with portal vein occlusion and in five of six patients with extrinsic portal vein compression, SE and GRE images were comparable in showing the abnormality. However, in five of nine patients with collateral vessels caused by portal vein compromise, GRE imaging more completely dem-
spins,
dimin-
ishing the bright-blood effect [15]. The high-contrast signal produced by flowing blood GRE imaging well suited to construction of MIP images. produce an angiographic effect that can enhance the of vascular disease (Fig. 6). The MIP program used
makes These display selects
onstrated
the maximum-intensity
the extent
of collateralization.
A significant
benefit
of GRE imaging compared with SE imaging was the ability to image the porta hepatis rapidly and selectively with singleslice breath-hold images, as an adjunct to the much longer SE examination. The
recent
development
of practical
techniques
for
MR
angiography has allowed further clinical application in assessing the abdominal vasculature, especially the PVS [2, 15, 16]. The “bright-blood” images produced by GRE imaging
because
additional
the TE and accentuate
compensatory
washout
voxel
gradients
would
of high-intensity
in the projected
prolong
ray, and there-
fore background signal is not additive as would occur with standard summation imaging [21]. Projectional images have the advantage
offacilitating
viewing
by displaying,
on a single
image,
the path of tortuous vessels that would otherwise need to be traced on numerous sequential images. In evaluating portal venous disease, MIPs can be a useful method of display without tual distortions
requiring invasive angiography. characteristic of these images
Some should
artifacbe rec-
]. A consistent distortion is an artifactual decrease width since the edges of vessels are less intense
ognized
[21
in vessel than
their
central
areas.
The
decreased
intensity
of the pe-
ripheral portion of vessels is related in part to partial-volume effect and in part to different flow velocities in vessels. A
Downloaded from www.ajronline.org by 109.161.224.40 on 10/19/15 from IP address 109.161.224.40. Copyright ARRS. For personal use only; all rights reserved
periodic
linear
dark
“striping”
artifact
may
occur
when
vessels
course obliquely between sections owing to the maximum signal intensity being shared between contiguous sections [21}. Thus, although these images provide a high-contrast angiographic
display,
individual
these
cross-sectional
pitfalls
GRE
require
slices
referral
for accurate
to
the
interpreta-
tion. Artifacts may significantly compromise the assessment of portal venous disease on MR. In three of our patients, artifactual signal in the PVS on SE imaging could have been mistaken for thrombus. In these cases, correlation with GRE imaging
and
with
other
imaging
techniques
correctly
attribute
this to artifact.
routinely
available
at the initiation
allowed
Presaturation of this study,
us to
pulses,
hepatic
arterial
flow.
Selective
although provided
presaturation
of portal
less degraded by motion poorer contrast between
artifact tumor
and normal tissues. The appearance of tumor on GRE imaging is variable and complex depending on the specific parameters used, which affect the relative contribution of Ti , T2, and T2* weighting. Increased sensitivity to magnetic susceptibility inherent in GRE imaging also compromises image quality, especially
from
artifacts
related
to curvilinear
artifact
imaging.
artifactual
The
to bowel
gas and clips.
A linear
frequently
on GRE
was encountered
able from previous
nature
of this
was
usually
recogniz-
experience
with normal volunteers and from correlation with other studies. A possible explanation of this effect is referred to as flow separation. Laminar streaming of flowing blood may occur on one side of a vessel, whereas
flow on the opposing side may be relatively slower. The rapidly flowing blood is freely replaced with fresh unsaturated spins that
appear
bright,
whereas
rated
blood
appears
darker.
the slowly A less
flowing,
likely
more
explanation
satuwould
be flow-related displacement. When blood is flowing obliquely through the scanning plane it occupies one location initially when molecules are excited and a different position later when they are encoded for readout. Because GRE images are acquired rapidly, blood would not move a sufficient distance
to make
this
a likely
explanation.
The limitations
of GRE
imaging, including poor contrast with surrounding tissues, difficulty in identifying vascular margins, and flow artifact, do not allow it at the present time to function as a gold standard for assessing
portal
In summary, imaging
of the
venous
GRE imaging portal
venous
thrombosis.
can be a useful adjunct system.
The
nondiagnostic
standard
SE imaging.
This imaging
Because thrombus generally appears as an area of decreased signal rather than increased signal, as with standard SE imaging, these images are less subject to the common pitfall of high-intensity artifacts mimicking clot. Additionally, angiographic MIP images provide a unique display that may complement standard single-slice images in demonstrating vas-
cular impression, encasement, GRE images provide limited with SE imaging
normal tissues, SE imaging.
and collateralization. anatomic resolution
and suboptimal
making
contrast
them unsuitable
between
However, compared tumor
and
as a replacement
for
not
venous flow would require specific presaturation of blood flow perpendicular to the course of the portal vein. GRE vascular images are not as susceptible as SE images to respiratory motion or pulsatile vascular artifact mimicking clot. In the six cases of extrinsic compression of the PVS by tumor, GRE imaging techniques, than SE imaging,
undergo
technique provides significant flexibility compared with standard SE MR because a few slices at a specific anatomic level can be performed to rapidly investigate potential disease.
can minimize
or eliminate high-signal artifact in blood vessels when applied perpendicularly to the direction of flow. However, even when such pulses are applied, they are routinely used to eliminate flow in the aorta and inferior vena cava and diminish respiratory artifact. Presaturation pulses may also be useful in elimmating
that they can be performed rapidly, as single breath-hold images that can be used in sick patients who might otherwise
images
to SE provide
high contrast between flowing blood, clot, and surrounding tissues. GRE imaging techniques allow marked flexibility in
REFERENCES 1 . RaIls PW. Color Doppler sonography venous system. AJR 1990;155:517-525
of the
hepatic
artery
and
portal
2. Edelman AR, Zhao B, Liu C, et al. MA angiography and dynamic flow evaluation of the portal venous system. AJR 1989:153:755-760 3. Zeman AK, Paushter DM, Scheibler ML, Choyke PL, Clark LA, Jafte MN. CT of the porta hepatis: normal and abnormal (abstr). Radio!ogy 1985;1 57(P): 113
4. Alpem
MB, Rubin JM, Williams
DM, Capek
P. Porta hepatis:
duplex
Doppler US with angiographic correlation. Radiology 1987;162:53-56 5. Weinstein JB, Heiken JP, Lee JKT, et al. High resolution CT of the porta hepatis and hepatoduodenal ligament. RadioGraphics 1986:6(1):55-74
6. Hricak H, Amparo E, Fisher MA, Crooks L, Higgins CB. Abdominal venous system: assessment using MR. Radiology 1985;1 56:415-422 7. Williams DM, Cho KJ, Aisen AM, Eckhauser FE, Portal hypertension evaluated by MA imaging. Radiology i985;1 57:703-706 8. Ohtomo K, Itai Y, Furui 5, Yoshikawa K, Yashiro N, ho M. MA imaging of portal vein thrombus in hepatocellular carcinoma. J Comput Assist Tomogr 1985:9(2):328-329
9. Walter
JP, McGahan JP, Lantz BMT. Absolute flow measurements using pulsed Doppler US. Radiology 1986:159:545-548 10. Miller VE, Berland LL. Pulsed Doppler duplex sonography and CT of portal vein thrombosis. AJR 1985:145:73-76 1 1 . Foley WD, Varma AR, Lawson TL, Borland LL, Smith DF, Thorsen K. Dynamic computed tomography and duplex ultrasonography: adjuncts to arterial portography. J Comput Assist Tomogr 1983:7(1):77-82 12. Axel L. Blood flow effects in magnetic resonance imaging. AJR
1984;143:1157-1166 13.
Bradley WG, Waluch V. Blood flow: magnetic resonance imaging. Radio!ogy 1985;154:443-450 1 4. Bradley WG, Waluch V. Lai KS, Femandez EJ, Spalter C. The appearance of rapidly flowing blood on magnetic resonance images. AiR 1984;143:1
167-1174
1 5. Edelman AR, Mattle HP, Atkinson DJ, Hoogewoud HM. MA angiography. AJR 1990:154:937-946 16. Edelman AR, Wentz KU, Mattle H, et at. Projection arteriography and venography: initial clinical results with MR. Radiology 1989:172:351-357 1 7. Ros PA, Viamonte M Jr. Soila K, Sheldon JJ, Tobias J, Cohen B. Demonstration of cavemomatous transformation of the portal vein by magnetic resonance imaging. Gastrointest Radio! 1986:1 1(1):90-92 1 8. Silverman PM, Feuerstein I, Garra BS, Zeman BZ. STIR imaging as a complement to standard spin echo imaging in the evaluation of the porta hepatis/hepatoduodenal ligament. Magn Reson Imaging 1991;9:73-77 19. Zirinsky K. Markisz JA, Rubenstein WA, et al. MR imaging of portal venous thrombosis: correlation with CT and sonography. AJR 1988:150:283-288 20. Silverman PM, Patt RH, Baum PA, Teitelbaum GP. Ghost artifact on gradient-echo imaging: a potential pitfall in hepatic imaging. AJR 1990;154:633-634 21. Anderson CM, Saloner D, Tsuruda JS, Shapeero LG, Lee RE. Artifacts in maximum-intensity-projection display of MR angiograms. AJR 1990:154:623-629