Measurement of nasal airflow and resistance

Ian S Mackay FRCS Brompton Hospital, London SW3 6HP and the Westminster Hospitals

Introduction Early attempts at rhinomanometry date back to the end of the last century, when Zwaardermaker described the first hygrometric method (1889). Since then, numerous ingenious methods of anterior and posterior rhinomanometry have been described (Foxen et al. 1971), but despite this no method has gained common usage in clinical practice in this country. One of the main problems in any single measurement of nasal resistance, is that one is measuring a 'moving target'. First, because the resistance increases with inspiration due to indrawing of the alar cartilages, upper lateral cartilages, nasal and nasopharyngeal mucous membranes, the greater the flow the more these structures are indrawn and the more the resistance increases; whilst on expiration these structures dilate and the resistance falls. Secondly, the nasal resistance is continuously altering from side to side over a variable period of some two to three hours; this 'nasal cycle' was first described by Kayser (1895). Despite these problems, it is important to have some objective method of measuring nasal airflow and resistance, to evaluate the effectiveness of topical and systemic medical treatment (Mygind et al. 1977, Vilsvik et al. 1975); and pre- and post-operative studies are needed to evaluate the effectiveness of surgical treatment which in future is likely to gain increasing medicolegal importance. In 1976, Mercury Electronics introduced the NRI Nasal Resistance Meter (Figure 1) which measures nasal airflow using a special face-mask incorporating a linear pneumotachograph and a mouthpiece which allows the intrapharyngeal pressure to be measured via a small tube passed orally into the pharynx.

Figure 1. Nasal resistance meter NRI

Figure 2. Modified cuff

Whilst this apparatus is quick and convenient to use, its main drawback (as with any other method involving the measurement of the nasopharyngeal pressure) is the difficulty of persuading the subjects to relax the soft palate. Mygind et al. (1977), using this method, found 17% of patients unable to relax the soft palate. Foxen et al. (1971), describing a different method of measuring nasopharyngeal pressure, found 25% of subjects unable to be assessed for the same reason. Cottle (1968) described a method of rhino-sphygmomanometry in which the airflow was measured via a nozzle on one side of the nose whilst the pressure was recorded on the other. I

Accepted 6 March 1979, finalized II June 1979

0141-0768/79/110852--04/$01.00/0

.Iv,'-. 1979 The Royal Society of Medicine

Journal of the Royal Society ofMedicine Volume 72 November 1979

853

This method avoids the practical complication of having to measure pressure via the nasopharynx. However, whilst pressure can be accurately recorded by this method, Tucker (1977) has shown it to be necessary to use a flow nozzle with a bore at least as large as the hydraulic cross-section of the normal adult nasal fossa, i.e. at least 8 mm in diameter. This may cause distortion of the nostril, altering the relationship of the lower lateral and upper lateral cartilages to each other and to the nasal septum, which can affect the resistance to airflow. The disadvantages of these two methods can be overcome by measuring flow through one side of the nose with a face-mask incorporating a linear pneumotacograph, whilst pressure is measured with an adjustable nozzle passed through an air-tight seal in the face-mask (Figure 2). The nozzle is so constructed that it can form a seal around the nostril without being passed into the nose and does not employ an inflatable cuff which might alter the resistance on subsequently testing the other side. The apparatus has been modified further so that it will 'cut out' and record pressure on inspiration and expiration, at any preselected flow rate. The patient is requested to breathe comfortably in and out while the flow rate 'cut out' is increased or decreased until a constant, reproducible pressure recording is obtained. The pressure is then read offthe graph and a note is made of the flow rate. Resistance is calculated as pressure divided by flow, with pressure measured in mmH2O and flow in litres/min. To convert the resistance into terms used by pulmonary physiologists, pressure is expressed in cmH20 and flow in 1/sec by applying the formula P/10 + F/60 or 6P/F = RV cmH2O/l/sec. Using this method, convenient flow rates for normal subjects have been found to be in the region of 10 1/min but when nasal resistance is increased, the flow rate at which pressure is recorded has had to be reduced in some cases to 2 or 3 1/min in order to obtain reproducible results with laminar nasal airflow. At higher rates, the flow becomes turbulent and no longer recordable by this method. Criticisms of other methods employing this principle are the possible distortion of the breathing nostril and the fact that breathing through only one nostril is not physiological (Maran, p 849). However, with this particular technique, the breathing nostril is not impeded as flow is measured through the face-mask and pressure recorded at a preselected physiological flow rate. Results

Figure 3 shows the graph obtained when measuring a normal subject. A flow rate of 10 1/min has been selected and a pressure recording of 5 mmH2O on expiration and 10 mmH2O on

Figure 3. Tracing from normal subject with flow rate set at 10 I/min. The pressure recorded can be seen to be in the region of 10 mmH20 on inspiration (upwards) and 5 mmH20 on expiration (downwards) on both sides

854

Journal of the Royal Society of Medicine Volume 72 November 1979

inspiration is obtained; i.e. in order to produce a flow rate of 10 1/min through each nostril on inspiration, a pressure of 10 mmH20 is required. To express this, as resistance, the formula 6P/F is applied: 6 x 10/10 = 6 cmH2O/l/sec. Figure 4 demonstrates the graph obtained on a patient with the septum grossly displaced to the right. Reading from left to right side of the graph, the pressure is recorded while the flow rate 'cut out' is gradually reduced from 10 1/min down to 3 1/min at which point there is laminar flow and a constant recording is obtained (10 mmH2O) on expiration. Applying the formula 6P/F, the resistance is therefore 6 x 10/3 = 20 cmH2O/l/sec on the right. The left side is then recorded while the flow rate is gradually increased to 10 1/min, at which point the resistance is within the normal range. Examples of other results are shown in Table 1. In one case with a saddle nose and severe deviation of the nasal septum to the right, with hypertrophy of the right inferior turbinate, there was total obstruction on the right side and this method could therefore not be used since it is dependent on having at least a degree of patency on both sides.

Figure 4. Tracing from patient with septum deviated to the right. Reading the graph from left to right, the pressure on the right side of the nose has been recorded, gradually reducing the flow rate sampling point from 10 down to 3 1/min, at which point a constant recording is obtained of 10 mmH20 on expiration (downwards) and 15 mmH20 on inspiration (upwards). The pressure on the left side of the nose has then been recorded, increasing the flow rate setting from 3 to 10 1/min Table 1. Results

Resistance (cmH20/l/sec) Left Right

Age Patient

Sex

(years)

Diagnosis

Insp.

Exp.

Insp.

KP

Male

30

27

12

10

6

LM

Male

35

24

21

12

9

TS

Male

28

27

24

21

12

GS

Male

21

3

3

30

24

MA MC SE

Female Female Male

42 36 15

Anterior deviation of the nasal septum to the right, with inspiratory valvular obstruction Deviation of the nasal septum to the right (midway) Septum dislocated to the right anteriorly and to the left posteriorly Deviated nasal septum to the left. Hypertrophy of left inferior turbinate. Over-patent right airway Atrophic rhinitis; over-patent airway Inspiratory valvular obstruction Nasal septum deviated to right

3 18 30

2 6 14

3 34 4

3 10 4

Exp.

Journal of the Royal Society of Medicine Volume 72 November 1979

855

The modified nasal resistance meter is objective in that it measures pressure required to produce a preselected physiological flow rate on both expiration and inspiration and is therefore not dependent on the effort made by the patient. Left and right sides of the nose are measured separately, which is of particular importance before and after operation in septal surgery. It is not necessary to measure the pressure in the nasopharynx, and this method is therefore not dependent on the patient being able to relax the soft palate. References Cottle M H (1968) International Rhinology 6, 7-26 Foxen E H M, Preston T D & Lack J A (1971) Journal of Laryngology and Otology 85, 811-825 Kayser R (1895) Archives of Laryngology and Rhinology 3, 101 Mygind N, Johnson N J & Thomson J (1977) Clinical Allergy 7, 69-74 Tucker C A (1977) Rhinology 15, 73-79 Vilsvik J S, Jenssen A 0 & Waistad R (1975) Clinical Allergy 5, 291 Zwaardemaker H (1889) Nederlands Tijdschrift voor Geneeskunde 25, 297

Measurement of nasal airflow and resistance.

852 Journal of the Royal Society of Medicine Volume 72 November 1979 Measurement of nasal airflow and resistance Ian S Mackay FRCS Brompton Hospita...
947KB Sizes 0 Downloads 0 Views