Magnetic Hong
N. Yeung,
PhD
Alex
#{149}
Magnetization with Periodic
M. Aisen,
during
presaturation.
MTC
imaging
is dependent on repetition time; this dependence arises from the dual but conflicting effect of the Selective saturation on tissues rich in macromolecular constituents. Substantial changes in tissue contrast were observed with MTC presaturation. The effect was most marked in muscle, as expected from the high protein content of this tissue. Index
terms:
experimental
netization nance, Radiology
I
From
Magnetic resonance (MR), #{149} Magnetic resonance (MR), magtransfer contrast #{149} Magnetic reso-
technology 1992;
183:209-214
the Department
of Radiology,
Imaging
MD
Transfer Contrast Pulsed Saturation’
Magnetic resonance imaging with magnetization transfer contrast (MTC) was implemented on a 0.5-T clinical system by using a stream of binomial radio-frequency pulses for the selective saturation of the bound protons in tissues. Images were obtamed of an egg phantom and of the calf, head, and abdomen in healthy human volunteers. Nonspecific saturation resulting in diminished signal intensity was manifested in the plots of the ratio of MTC signal to control signal for egg yolk, raw egg white, cerebrospinal fluid, and bone marrow fat. This signal degradation resulted partly from the cumulative effects of T2 on the observed transverse magnetization
Resonance
Univer-
sity of Michigan Medical Center, Ann Arbor, MI 48109-0553. Received Iuly 18, 1991; revision requested August 12; revision received September 30; accepted October 30. Address reprint requests to H.N.Y. C RSNA, 1992
S
contrast
IGNAL
in magnetic
tons.
reso-
nance (MR) imaging is known to be determined by a number of interrelated phenomena, with a major factor being the Ti and T2 of the hydrogen nuclei, or protons, of water and fat. Hydrogen atoms in molecules other than water and fat also participate in the MR imaging process, but because of their low concentration, short
T2,
or both,
they
are
not
directly
observed on MR images. For example, hydrogen is abundantly present in molecules such as membrane lipid and collagen, often in relatively high concentrations.
However,
because
it is
chemically in a relatively bound or, loosely speaking, “solid state” form, the T2 is generally much shorter than 1 msec and cannot be readily observed with commonly used imaging techniques. The term “macromolecular” can also be used for such bound protons.
The
spectral
response
or line
width on MR spectra is inversely related to T2, and the spectrum of macromolecular hydrogen is said to be homogeneously broadened, in contrast
to the
line
width
of natural
free
water, which is narrow, though in practice the spectrum is unavoidably broadened somewhat by magnetic field inhomogeneity. The hydrogen nuclear relaxation times in tissue-especially Ti, the tongitudinal relaxation time-have in turn a complex dependence on many fundamental parameters, such as the intrinsic relaxation rate of free water protons, r = i/T1 (where r5, = relaxation rate of free water protons and T1 = the longitudinal relaxation time of water), and the rate of magnetization transfer, r, between protons of free and bound water or between hydrogen nuclei of nearby macromolecules. Thus, though the bound hydrogen nuclei of macromolecules cannot be observed directly, their presence can have substantial effect on MR image
contrast
with
the
through
directly
their
observed
interaction
water
pro-
by
This the
fact
is shown
recent
most
development
vividly (1-5)
of
magnetization transfer contrast (MTC) imaging. With this technique, substantial change in tissue contrast at MR imaging is induced by altering the magnetization of the bound protons
with
use
of continuous,
off-reso-
nance radio-frequency (RF) irradiation. We have applied similar methods to imaging with a whole-body imager by using
a variation
on
the
technique
that may be easier to apply on commercial imagers. The efficacy of the method is demonstrated with MR images sisted
obtained of two
cooked volunteer
and
in a phantom chicken eggs,
one raw, subjects.
MATERIALS
that one
and
AND
con-
in healthy
METHODS
The pulse sequence used for MTC and the presaturation pulse train are shown schematically continuous,
in Figure monophasic,
irradiation,
1. Instead of using off-resonance
of binomial
a stream
pulses
was
employed, as suggested by Hu et al (6). A train of pulses can be easier to im-
plement tinuous
on commercial irradiation
ter of most
imagers
than
con-
since
the RF transmit-
does
not operate
imagers
in
the continuous wave mode and since continuous irradiation would likely trip RF power monitoring safety features.
Binomial
pulses
have
an added
virtue,
which is demonstrated in the Appendix: The dual-phase composite pulse will have a selective
saturation
geneously
broadened
effect
on
the
homo-
macromolecular
proton resonances, which leaves the narrow resonances, such as water, relatively unscathed.
Application
of this
whether applied on-resonance viates a second RF frequency and
substantially
improves
plementation
potential
on many
disadvantage
Abbreviations: contrast, RF time.
MTC =
radio
pulse
train,
or not, obchannel (1,2) the
ease
imaging
of im-
systems.
A
of this approach
=
magnetization
frequency,
TR
is
transfer =
repetition
209
that
there
may
specific image wave
be
a greater
signal
loss,
quality,
than
hence
with
implementation
The
MTC
Heights,
sequence
MTC
was
The
presaturation
repetition (except
number
192; signals
the echo
3.8
minutes
quence single 2,000 aged.
TR =
with
lines
=
and or 7.5 time
in our study was = 600 msec. For TR sections can be imimages were
continuous
The
mode
of saturation.
pulse train I 1 binomial
consists pulses
an interval T, that can msec to longer periods.
rad/sec. Details are included
software
be
varied Variation
Three MTC imaging corresponding control collected from imaging abdomen, and head in man volunteers. In each ages were acquired with
pre-
of the binomial in the Appendix.
0.2 W/kg
by
supplied
with
the
The
use
of
imager, set
which
by
the
U.S.
Standard receive coils were supplied by manufacturer and included a head coil the human head and leg studies and a single-turn surface images. MTC images
tamed terpulse
by
using delay
coil for the were ob-
several values for (and hence different
the
inde-
grees of presaturation) without shifting the presaturation frequency off-resonance. Control images were obtained by using identical
imaging
the
presaturation. signal intensities
MTC Tissue
mined
by using
of interest. sues were
tamed egg
white,
and fat,
were
ratios from
without deter-
without and
regions
of various tisimages ob-
MTC.
cerebrospinal
unaffected by MTC signal in these tissues
Radiology
#{149}
sequences
operator-specified
Intensity calculated
with
relatively tion, the
210
pulse
Since fluid presaturaserves
raw are as a
msec)
and
interpulse
tion,
of averwith strength
with
2,000
ferent
from 5 in this
rate for a body use of the sequence cycle at this field
to be
rectangular abdominal
Periodi,
pulsed
b.
naturation
representation
of imaging
sequences
lose that
demonstrates
the two
types
signal. The likely explanation raw egg white has a long
while
of
the
two
data sets and data sets were of the calf, healthy hudata set, imtwo TRs (600
two
or three
intervals.
sets
of MTC
interpulse
Images Quantitative intensity with and are shown phantom
data
shown in Figure 2. comparisons of signal ratios in images obtained without selective saturation in Figure 3 for the egg and and in Figure 4 for the
cumulative
during
effects
the
saturasignal
best
illustrated
raw
chicken
by
egg.
This
the
images
Since
the
to show
presaturation presaturation. show that while for the raw egg
any
compared Yet the
this is essentially white, the yolk
does
intervals.
proteins
On
heat-
denature
with much larger moThese solid compo-
are
more
and
and
solid matrices tecular weight.
effective
form
in causing
and
relaxation
numerical
from 25 to 10 msec, the saturating duty
image
intensities
linearly. Thus, with selective further with
data
in Figure
which signifies a tive intensities or effect on contrast between the raw Further, decrease
thereby cycle,
decreased
image contrast saturation and cooking. As the
strate, there is a differential saturation on egg white
3 true
sig-
reaggregate
almost changes changes
MTC
control in Figure
comparing
subsequently
the MTC
of the
with data
by
interpulse
egg-white
decreased increasing
fact is
with
appreciated
and
MTC effects, and cooked egg white will show a greater effect with MTC than wilt its uncooked counterpart. The effect of MTC on egg yolk, which contains littte or no protein, is minimal and does not change with cooking. As ‘ri, was
protein
change
transfer contrast
cross
molecules in egg white are too small to contribute much to magnetization transfer, the absolute signal intensities from the raw egg white and the yolk (which consists mostly of small motecules of cholesterol) might not be ex-
pected
The effect of magnetization itself on signal intensities
nents
magnetization
presaturation.
cereand
ing,
of T2 on the
transverse
with longer will emerge
also be applied to the data for brospinat fluid, bone marrow, subcutaneous fat.
ferent
intensity is manifested in the plots of the ratio of MTC signal to control signat for egg yolk, raw egg white, cerebrospinal fluid, and bone marrow fat in Figures 3 and 4. Note that this signat degradation results partly from
the
T2
nat intensities in images of the cooked and raw egg white obtained with both long and short TR and with dif-
are
observed
T2. As dema short
A substance as egg white,
is best
intervals.
human images. Nonspecific tion resulting in diminished
a short
Appendix,
relatively unscathed from the binomiat pulse. The same explanation can
dif-
were acquired for the egg phantom, one with a TR of 1,000 msec and the other with a TR of 10 seconds; each data set was acquired for four differ-
ent
has
in the
saturation. T2, such
In addi-
image
yolk
is T2,
component will not fully restore its longitudinal magnetization at the end of a binomial pulse but rather wilt undergo a diminution in signal proportional to the duty cycle of the pre-
RESULTS
=
and
of a string separated
is still within the guidelines Food and Drug Administration. the for
-a
#{231}#{176}#{231}#{231}#{231}
(a) Schematic
onstrated
parameter will result in different “duty cycles” for the presaturation pulse train and thus will result in differing degrees of presaturation. The RF power level of the pulse train, as measured by means of the nutational frequency imparted by the resulting B1 field to the magnetization, is
is estimated
1.
reference with which to evaluate nonspecific saturation or signal loss resulting from the binomial pulse train.
of MR imager atthrough the period precludthe
specific absorption age weight with the highest duty
6w=
MTC. In both cases, the MTC saturation is applied “before” the 90#{176}-180#{176} imaging sequence, during the intervals that would otherwise be occupied by multisection excitations. The differences between continuous and pulsed saturation are also shown. cw = continuous wave. (b) A 1 1 binomial pulse is illustrated. Note that this pulse can be applied on- or off-resonance. In the latter case, the frequency offset has to be alternated coincidentally with the phase alternation.
the
body coil and limits each transmit to no longer than 10 msec, thus ing the possibility of implementing
pulses
L
Figure
= 600 msec and 2,000 msec. The Se-
used in this study. This particular model lows only RF transmission
2,500
Saturation
2; field of view = 25 cm (human head
implemented section for TR msec, up to four Only single-section
saturation of 5-msec
Continuous
=
for TR
for
Controlled Fn.=;.;os
-
a.
TRs were 1,000 time = 20
of phase-encoding
12.8 minutes
1
fol-
set collected
averaged phantom),
(egg
JH
!
of
imaging
and leg), or 33 cm (human abdomen); section thickness = 5.0 mm (eggs) mm (human subjects); total imaging was
I
consisted
component
data
r
in
(model Highland
The imaging parameters used images presented herein were time (TR) = 600 and 2,000 msec
for one
15 cm
u=a
implemented
spin-echo
egg phantom in which msec and 10 seconds); msec;
continuous-
sequence
by a standard
sequence. for all the
events
diminished
the
MR system International,
Ohio).
towed
Imaging
of non-
of MTC.
a 0.5-T whole-body Vista HP; Picker an
and
degree
3 demonand
effect egg
of yolk,
change in the relatissue contrast; the is quite different and cooked egg.
there is a relatively in the signal of the
greater cooked
April
1992
a.
b.
c.
Figure 2. Representative human head images at (a, b) TR = 600 msec and (c, d) TR = 2,000 msec, all obtained with echo time = 20 msec. (e) T2-weighted image (TR = 90 msec) is also shown to illustrate its superficial re-
4
‘
.. .
,4
t:s
-4
semblance
..‘
masked
.
,
.:7
,$.
changes
to some
at imaging, 4. Intensities MTC, but
extent
TR MTC image in signal
shown
intensity,
by the windows
used
are given quantitatively in Figure of all tissues decrease with the effect is greatest for the my-
elinated white matter, resulting in altered gray-white contrast. Note also in b the increased prominence of the lenticular nuclei
.-
in the image),
(a) TR = 600 msec (control TR = 600 msec (MTC image), (c) TR = 2,000 msec (control image), (d) TR 2,000 msec (MTC image), and (e) TR = 2,000 msec (T2-weighted [echo time = 90 msecj control image).
)
‘I
d.
to the long
in d. Absolute
I
‘
brain.
0,)
=
e.
egg white, which results in a progressive decrease in the intensity ratio of cooked to raw egg white with increasing RF duty cycle. Figure
2 shows
the
representative
MTC images of the head and their control images. In this sequence of images, both Ti-weighted (TR = 600 msec) and proton-density-weighted (TR = 2,000 msec) images were chosen to illustrate the TR dependence of MTC. This dependence arises from the dual but conflicting effect of the selective saturation on tissues rich in macromolecutar constituents. The saturation of the immobile protons from MTC irradiation causes, on the one hand, attenuation of observable signal through magnetization transfer. On the other hand, this same saturating event also brings about a decrease in the apparent longitudinal relaxation time (7). As a result, for pulse seVolume
183
Number
#{149}
1
quences with short fect partially cancels thereby pulse
TR, the latter the former,
compromising sequence with
ef-
the MTC. For tong TR, the
MTC effect (which has a negative impact on signal strength) is not compensated by the Ti shortening (which impacts
therefore
the
signal
positively)
and
is
left unmitigated.
As expected,
in the
human
images
the relative signal drop-off varied from tissue to tissue, being greatest for muscle and smallest for cerebrospinal fluid in the lateral ventricles of the brain. Thus, the images obtained with MTC display dramatically different contrast than do the control images. 3b and
In comparing Figure 4a with
Figure 4b, we
3a with note also
that some tissues, such as cooked egg white and white matter (as discussed above), show moderate to strong TR dependence on signal attenuation,
a
while others, especially those of the abdomen such as back muscle, renal cortex, and liver, show similar amounts of MTC attenuation with varying TR. Figure 2 also demonstrates an interesting effect of MTC irradiation on a Ti-weighted image of the brain. Note the relatively increased signal intensity
in the
lenticular
nuclei
compared
with the signal intensity of the remainder of the brain, the consequence of a lesser degree of saturation in these increased As gray
nuclei. signal matter
The reason intensity structures,
for this is uncertain. lenticular
nuclei have a relative paucity of myelin compared with the surrounding white matter; hence, greater attenuation
of white
matter
could
be ex-
pected. However, this does not explain why the cortical gray matter does not remain equally intense as Radiology
211
#{149}
the
lenticular
trast
nuclei.
mechanisms
clei that
There
in the
are unique.
are connu-
lenticular
For example,
these structures are known to have diminished signal on high-static-field T2*weighted images, an effect that may be a result of iron deposition (8). Perhaps iron or other substances produce a paramagnetic effect that contributes to the overall isointense appearance of the lenticular nuclei on conventional Ti-weighted images, a mechanism not present to the same
degree brain.
elsewhere
in the
visualized
6
0.9
0.9
0.8
0.8
0.7
0.7
0.6
0.6
0.5 0.4
0.3
The
ing.
authors
are grateful
of MTC nomenon.
in demonstrating
this
0.2
10
results
confirm
and
20
‘
Interpulse
Delay,
observed The effect
muscle, protein
with was
10
this
terms
spin-bath
continuous
images.
show signal intensity of MTC-irradiated
(a) Data
for a short
data samples
for cooked to signal
TR (1 second)
factors
2i2
#{149}
affect
Radiology
TR (10 seconds)
are pre-
0.8. pD
0
i
0.8
6
0.6
.
0.4
0
0.6. 0
0.4.
g
#{163} 0.2
10
15
0.2.
10
20
Delay,
15
lnterpulse
msec
Delay,
20
msec
b.
model
to its
is the inherent specifically,
the
properties relative
of the tongi-
tudinal relaxation rates of the freewater protons versus the cross-relaxation rate between the free and bound fractions. The other factor is s, the saturation parameter that, as we show
in the
Appendix,
is dependent
on both tissue properties and machine settings. With the application pulsed saturation, the steady state
of
hence
key
(b) for a longer
as the ratio spin-echo
0
factor tissue,
signal alteration expressed in
saturation
magnetization. ratio from
eggs, expressed of conventional
0
ton
1 and
and
and raw intensities
sented. In both graphs, the ratios for all substances are less than 1, a combination of MTC-specific as well as nonspecific signal loss. Both graphs also demonstrate a dramatic decrease in the ratio of cooked egg white compared with the raw counterpart; this finding is consistent with the effect of increased MTC on the denatured egg-white protein. Note also that the signal loss decreases slightly as the interpulse delay increases, and the amount of MTC irradiation is consequently diminished. The effects of different TRs are modest; the greatest effect is decreased signal in cooked egg white at the longer TR compared with the shorter TR. #{149} = cooked egg yolk, = cooked egg white, 0 = raw egg yolk, = raw egg white.
signal intensity depends not only on the fundamental rate constants of the tissues and the saturation parameter s, but also on a complex dependence on the interpulse interval, T, which effectively controls the duty cycle and
s = the simple It is clear
25
‘
msec
b.
Figure 3. Graphs of signal intensities
unsaturated counterpart can be shown under some limiting conditions (1-5) to be R = {5, + T(1 (r5, + ri), where s is a fraction (0 s 1) that measures the degree of saturation of the bound fraction of the proration,
20
Delay,
high
(iO-i4) in which the free and bound protons are considered to be two mutually coupled thermodynamic reservoirs, each of which relaxes with its own relaxation mechanism. The ratio of the steady state signal intensity
under
#{149}
Interpulse
MTC presaturamost marked in
The reason for can be quantitatively a binary
15
‘
msec
Figure 4. Plot of intensity ratio of MTC/control in images of human tissues as a function of interpulse delay at (a) TR = 600 msec and (b) TR = 2,000 msec. Error bars on individual data points are omitted from the graphs for clarity. They are approximately 15% for = 10 msec, 10%-15% for = 15 msec, and 10% or less for i, = 20 msec. The larger-than-unity ratios of the bone marrow fat are clearly results of measurement errors. = calf muscle, #{149} = bone marrow, A = gray matter, #{149} = white matter, D = cerebrospinal fluid, 0 = subcutaneous fat, = liver, 0 = renal cortex, = back muscle.
(i-5). contrast
as expected from the content of this tissue.
of
....
25
a.
phe-
extend
of other investigators Substantial changes in tissue
were tion.
15
Interpulse
DISCUSSION Our
0.3
#{163}0.1
a.
those
.
#{163}0.1
to an
anonymous reviewer of the initial version of this manuscript for suggesting the intriguing possibility that paramagnetically produced signal was present in the lenticular nuclei and for suggesting the potential role
0.5 0.4
0.2
control
This paramagnetic effect will not be reduced with MTC irradiation, which produces relatively increased signal as the remainder of the brain loses signal. Interestingly, the lenticular nuclei can appear hyperintense on Ti-weighted images in certain disease states, an observation that has been hypothesized to result from the presence of paramagnetic substances in these nuclei (9). Possibly, MTC may play a role in demonstrating subtle paramagnetic effects in clinical imag-
d
For
complete
R becomes T,/(T,
this
MTC
+
satu-
equal
to
Ti).
relation
signal.
that
One
two
As
the
degree
mentioned
MTC sequences tion of continuous tation
over
of saturation above,
use
requires the or pulsed
a lengthy
duration.
(i4). of the
applicaRF exciAt
least
at lower
field
strengths,
the
RF
power level required is reasonably low. Thus, the RF duty cycle presents problems, perhaps not so much for patient safety but for implementation in most, if not alt, commercial MR imagers manufactured for clinical use. This difficulty represents a major
stumbling block to the development of this potentially powerful technique from an investigational method to use in the clinical arena. We have demonstrated
herein
tend MTC
that
the experimental from in vitro
it is possible
to ex-
technique and restrictive
of in
vivo
settings to a clinical environment a pulse sequence that should be easily implemented on most exist-
by using ing
commercial
mentally,
the
may
eliminate
ing
the MTC
MR imagers. use
Experi-
of a binomial
pulse
the necessity of shiftirradiation frequency April
1992
might
M,
be enhanced
with
MTC
irradia-
tion. Second, tissue characterization may be possible. For example, the contrast effect is particularly pronounced in tissues with a high macromolecular content such as muscle. This phenomenon may permit characterization of myelin in the brain or fibrosis in other somatic tissues. For
example,
the
tissue,
which
identification
of fibrous
is difficult
with
Figure Al. Trajectories netization vectors during the
1 T binomial
the various of the
pulse.
trace the tip of magthe time course of
labels along indicate T2 in milliseconds magnetization compo-
lines
corresponding
Number
nent.
Frequency offset = 0, w sec, or B1 = 4.5 .tT (the actual strength of the binomial pulses experiment is = 9.3 p.T).
off-resonance. pulses Figure
= 1,200 rad/ RF field used in our
By applying
off-resonance, i, some of the
binomial
as illustrated nonspecific
in satu-
ration problem described above may be alleviated. Investigation of this possible improvement is currently under way in our laboratory. It should
be noted
that
the
2-4
agree
in terms
that we constants.
off-resonance
saturation,
at (7) demonstrated the cross-relaxation measured.
Grad
et
that both r,, and rate, TX could be
In a related
development
reported elsewhere, we have provided a general quantitative solution to the theoretical problem of the application of selective pulsed saturation to a heterogeneous spin system such
as tissues
by
using
the
spin-bath model; we have vided a means to estimate as well
ters
as the
(ie, the
remaining
binary
also TX
two
pro-
and r5,, parame-
longitudinal rethe molar ratio f of the bound to free protons) that are required to describe fully the theoretical model of coupled spin bath (15). The clinical usefulness of MTC imaging requires further investigation. taxation
Tissue
whether agnostic
There MTC First, Volume
intrinsic
rate,
T,
contrast
, and
is clearly
this
altered,
alteration
efficacy
enhances
remains
but
di-
to be seen.
are at least two possible ways techniques might be of benefit. the conspicuity of certain lesions 183
Number
#{149}
1
is
spin-bath
know
BINOMIAL
=
x)”. The
2 (1 2 1). In our
nance, as shown alternation that
in Figure is coincidental
can the
ii
work,
total
To
very
short
vectors
transverse
we
of the
illustrate
how
saturate free ones,
the bound in Figure
lb.
T2 components,
to realign lose strength
themselves in the
ond
lobe
I 1 pulse
of the
and
long
long
on
coherence,
the
hand,
magnetization tively unaffected. part tion short
the
which
quickly
ter in biologic more nonspecific will
will
vectors
the
maintain
tipped
in the
will
continuously
de-
Thus, the action of a has the effect of reducof the very short T2 hydroleaving the
T2 nuclei accounts
a dynamic
relaxation
with
composite
pulse.
favorable
condition
interplay the
of trans-
RF action
On-resonance,
of the the
in bringing
but not plot the travecof
‘
in Figure is slightly reason
Al, -yB1 less than
strength we
actual
used chose
B1 for
B1 field
T20)
rad/sec, of the
in our experiments. this value instead illustration
happens
to yield
0.45
=
msec
(0.2
msec
This
for
a T2 that
sponse
of the
shows
which zation ignores
is long compared composite pulse,
a final
preserves as expected relaxation for proton continues
the
phase T2 is much
by
taking
into
field in the ation. In the
account
I 1 pulse
diagram,
the
and the
solid
effect
transverse line
of the
RF
relaxrepresents
actual
tip angle
near
the
longitudinal for binomial effects. On
fractions to approach
with the
Al,
of a single
the
effect
zero,
with T2 = 0 regardless
of the RF field. Note shorter than T2,,D,
effect
the re-
magnetipulses and the other T2,, of
that when the saturapulse
di-
B1 field
in Figure
The
(21)
condiT2 (or
the
msec.
equations
the
a 360#{176} flu-
is superfluous. critically damped
tion efficiency of the binomial minishes rapidly unless the strength is increased.
of the magfrom the tran-
of
is that
B1 field in use).
hand,
is
the
1,200 one-half
=
tation on the long T2 protons after the first lobe of the pulse. This fortuity gives the misleading impression that the binomial-
M(t)
applied RF field of the I T pulse. In
Bloch
most
“down”
resonance (in the manner described Fig Ib) is not quite the same as that
to the
relain
T2. magnetization
quency. This diagram represents a tracing of the magnetization vectors of the proton components, with T2 spanning three orders of magnitude from 15 sec to 15
solution
op-
tissues, will tend to show signal attenuation than
show
verse
For
case, the RF irradiation proton-resonance fre-
netization,
sec-
affect
differently.
be
tissues with longer When T2 is short, the
RF action
pulses
the illustrated applied at the
sient
the
longitudinally process. The
of the longer This also
ity of the pulse tion makes the
with a sign to the
by duration
time development M(t), is derived
to-
for the differential nonspecific saturathat tissues will undergo. Tissues with T2 values, in the range found in wa-
actual
used
from equilibrium for the whole
The
of
for
(or the immobile macromolecular gen) nuclei to near zero while
the
i (1 1)
jectory of the tip of the magnetization tor M(t) in the rotating reference frame the proton fractions with different T2 times as they are deflected or nutated the
tilted
which will
other
RF field
corn=
binomial protons Al we
cases msec. pulse,
except
T2 fractions
T2 fractions,
dine longitudinally. complete RF pulse ing the magnetization
RF pulses.
these
are
plane,
tend and
The
duration of 5 rnsec (w = 0 in Fig Ib). also be applied off-reso-
alternation
the
trated which
most
probably
the
longitudinal magnetization (ie, makM, = 0) is the “critical damping” condition or yB1 = 1/(2 T2), where -y is the gyromagnetic ratio. In the example illus-
PULSES
1 T pulse with a applied on-resonance These pulses can
phase
funda-
can have different modulation as of x in the bino-
mial expansions (1 monly used cases are
n
all the
magnetization
the ing
The binomial pulse varieties of amplitude given by the coefficients
and
model
A CONSIDERATION
OF
relaxation
ous
of the binary
APPENDIX:
intrinsic
scte-
what
represent
posite direction and will return toward their initial state. The short T2 fractions,
known about MTC in terms of the theoretical spin-bath model. Tissues rich in proteinaceous fibers, such as muscle, tend to demonstrate a larger reduction in signal intensity than do tissues such as cerebrospinat fluid that have only a small protein content, and this signal reduction increases as the interpulse interval in pulsed saturation decreases. It is also possible to interpret these results quantitatively provided mental
rate of the free-water protons, r5, , cannot be determined simply by measuring Ti with conventional techniques. However, by performing Ti measurements with an inversionrecovery sequence by using continu-
with
the ward
phase
in Figures
qualitatively
of brokenness,
a T2 of 2.67, 0.47, 0.084, and 0.015 After the first lobe of the binomial
The
sequences, would be of great in establishing a differential diagnosis (16-19). One preliminary report suggests a rote for MTC imagof multiple
order
short
conven-
tional benefit
ing in the evaluation rosis (20). The results illustrated
the proton fractions with a T2 of 15 msec, while the rest of the lines, in decreasing
pulse
of an
applied
off-
for shown
on-resonance
pulse. pulse
When the B field of the is applied on-resonance,
called verse
effective field, B1,, lies in the transplane. When the binomial pulse is
applied off-resonance, that brings about
the
binomial the so-
the effective field nutation of the spin
Radiology
#{149} 213
vectors
now
xy)
on
but
result,
to the
tions, be
lies not on the transverse the
an
just
meridional long
ability
pulse
of an
therefore if all the
T2,
immo-
bile fractions will be severely impaired. However, if a sufficient number of such pulses is applied in series, this attenuated saturation will accumulate, which gives a stronger net effect. Use of a train of pulses will also allow for the time required for
saturation
in the immobile
fraction
(
or transfer
to the
free
to communicate
1/r,)
=
fractions. One aspect of the difference between application of continuous and pulsed RF in MTC imaging is the trade-off between use of a weak, long, continuous-RF field and a shorter, more intense, amplitudemodulated field. The key factor in determining the saturation efficiency of the two
techniques tion nus
(measured
by degree
achieved of the immobile that of the free fraction
power)
time
is T2 , the of the immobile
the T2,,, turn-on
the more time and
transverse protons. important the more
RF
complex
as well as on machine level, off-resonance cycle). while
It should also the quantitative
tween MTC and tors is complex,
,
r,
2.
3.
4.
5.
of the
6.
damped
the aforementioned the quantitative
on T2),
(RF power and duty
emphasized relationship
7.
that be-
8.
9.
facrelation10.
214
Radiology
#{149}
13.
14.
15.
Wolff SD, EngI, Balaban RS. Magnetizahon transfer contrast: method for improving contrast in gradient-recalled-echo images. Radiology 1991; 179:133-137. Wolff SD, Chesnick 5, FranklA, Lim KO, Balaban RS. Magnetization transfer contrast: MR imaging of the knee. Radiology 1991; 179:623-628. Wolff SD, Balaban RS. Magnetization transfer contrast (MTC) and tissue water proton relaxation in vivo. Magn Reson Med 1989; 10:135-144. Eng J, Ceckler TL, Balaban RS. Quantitalive 1H magnetization transfer imaging in vivo. Magn Reson Med 1991; 17:304-314. Grad I Mendelson D, Hyder F, Bryant RC.
of nuclear
magnetic
cross-re-
la.xation spectroscopy to tissue. Magn Reson Med 1991; 17:452-459. Hu B, Nishimura D, Macovski A. Pulsed
magnetization
comparison
factors frequency, be
1.
the
dependent T20, , and
11.
U
Application
are beyond the scope of this article. It suffices to point out here that the MTC in tissues is intimately related to this saturation efficiency and is therefore many tissue factors (rn,
(15)
References
is the RF advantageous
continuous irradiation over mode, unless the amplitude pulse is raised to the critically
level of 1/(2’yT2,). The details o/this
mi-
relaxation The shorter
is the pulsed
formulated
can be calculated numerifundamental constants are
Acknowledgments: The authors express their sincere appreciation for the technical support provided by Picker International. They also benefited greatly from stimulating discussions with many of their colleagues, in particular, Scott Swanson, PhD, and Thomas Chenevert, PhD, of the Department of Radiology, University of Michigan Medical Center, Ann Arbor, Mich.
of saturafraction unit
per
theoretically
12.
bino-
short
known.
been
z pulse.
off-resonance the
has
and cally
will
the
on-resonance
to saturate
ship
pulse
in preserving as the
(or As a frac-
binomial
as effective
the
mial
plane.
free-proton
off-resonance
magnetization But
T2,
(xz)
transfer
contrast
(abstr).
16.
17.
spin-lattice
18.
19.
In:
Book of abstracts: Society of Magnetic Resonance in Medicine 1990. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1990; 352. Grad I Mendelson D, Hyder F, Bryant RC. Direct measurements of longitudinal relaxation and magnetization transfer in heterogeneous systems. I Magn Reson 1990; 86: 416-419. Drayer B, Burger P. Darwin R, Riedere 5, Herfkens R, Iohnson CA. MRI of brain iron. AIR 1986; 147:103-1 10. BrunberglA, Kanal E, Hirsch W, VanThiel DH. Chronic acquired hepatic failure: MR imaging of the brain at 1.5 T. AINR 1991; 12:909-914. Koenig SH, Bryant RC, Hallenga K, Iacob CA. Magnetic cross-relaxation among protons in protein solutions. Biochemistry 1978; 17:4348-4358.
Edzes HT, Samulski ET. The measurement of cross-relaxation effects in proton NMR spin-lattice relaxation of water in biological systems: hydrated collagen and muscle. I Magn Reson 1978; 31:207-229. Edzes HT, Samulski ET. Cross relaxation and spin diffusion in the proton NMR of hydrated collagen. Nature 1977; 265:521523. Fung BM, McCaughy TW. Cross relaxation in hydrated collagen. I Magn Reson 1980; 39:413-420. Grad I Bryant RC. Nuclear magnetic cross-relaxation spectroscopy. I Magn Reson 1990; 90:1-8. Yeung HN. Proton relaxation in tissue under periodic pulsed saturation (abstr). In: Book of abstracts: Society of Magnetic Resonance in Medicine 1991. Berkeley, Cahf; Society of Magnetic Resonance in Medicine, 1991; 174. Charles HC, Baker ME, Hathorn W, Sostman D. Differentiation of radiation fibrosis from recurrent neoplasia: a role for 31-P MR spectroscopy? AIR 1990; 154:6768. Chamuleau RA, Creyghton JH, De Nie I, Moreland MA, Van der Lende OR, Smidtl. Is the magnetic resonance imaging proton
20.
21.
relaxation
time a reliable
nonin-
vasive parameter of developing liver fibrosis? Hepatology 1988; 8:217-221. Lee fl(, Glazer HS. Controversy in the MR imaging appearance of fibrosis (editorial). Radiology 1990; 177:21-22. Negendank WC, al-Katib AM, Karanes C, Smith MR. Lymphomas: MR imaging contrast characteristics with cinical-pathologic correlations. Radiology 1990; 177:209216. Foust RI, Tanttu II, Sepponen RE, Kinnunen E. Magnetization transfer contrast in multiple sclerosis at 0.1 T. In: Book of abstracts: Society of Magnetic Resonance in Medicine 1990. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1990; 626. Torrey HC. Transient nutations in fluclear magnetic resonance. Phys Rev 1949; 76:1059-1066.
April
1992