Low-friction hydrophilicsurface for rne~c~ devices ShojiNagaokaand RyojiroAkashi Basic Research Laboratories, Toray Industries, Inc. 1 I If Tebiro, Kamakura. Kanagawa. 248 Japan {Received 76 October 1989; accepted 20 February ?SSC?J

A hydrophilic polymer surface was developed exhibiting excellent low frictional property, namely slipperiness, when in contact with water or physiological fluid due to the reaction of epoxy-containing poly(vinyl pyrrolidone) with the polyamino compound formed on the surface of the substrate. Epoxycontaining poly(vinyl pyrrolidone) was obtained by the copolymerization of vinyl pyrrolidone as a hydrophilic component, glycidyl acrylate as a binding component to the substrate, and vinyl acetate to preserve the strength of the coating layer. The surface friction coefficient depends on the molecular weight of the coated hydrophilic copolymer. It was demonstrated that a molecular weight of 400 000 or more is essential to achieve excellent low surface friction. Using rabbit models, polyurethane catheters, both with and without the hydrophilic low friction coating, were evaluated for surface friction coefficient and blood compatibility. As a result. in the c8se of coated catheters, no lesions of the intima of the blood vessels and no thrombus formations on the surfaces of the catheters were observed. However, the non-coated catheters injured the intima of the blood vessels and severe thrombus formation was found on their surfaces. Keywords: Hydrophilic coatings. poiy~vinyt pyrroiidoneJ, mechanical prapeflies, blood compatibility

Low frictional property of the surface (so-called slipperiness), is required for medical devices in contact with such devices as catheters or guide wires that are inserted into blood vessels, urethra or other parts of the body which have mucous membranes. If these devices do not have low friction, their introduction into the body is not only accompanied by pain, but there is also a danger of damage to the mucous membranes or the intima of the blood vessels, which may lead to infectious diseases or mural thrombus formation’-4. Several methods of decreasing surface friction are known. A particularly practical method involves coating with a hydrophilic polymer. This method is based mainly on the techniques of binding the water-soluble polymer, including an active hydrogen group (top coat) with polyisocyanate groups, which is coated on the substrate (under coat) by the covalent bonds. However, since such isocyanate groups are highly reactive, they become easily inactivated when exposed to the moisture in air, which is why the stability of the coatings poses many problems. Another problem is the poor antihydrolytic characteristic of the chemical bond formed, which causes low durability of the low frictional property of the surface in the body. This report describes the formation of a surface with excellent low frictional property of the surface when in Correspondence to Dr S. Nagaoka. Medical Devices and Diagnostics Department, Toray Industries. Inc., Head Office, 2-1, Nihonbashi-Muromachi 2-chome, Chuo-ku. Tokyo 103, Japan. Q 1990

Bu~e~o~h-Helnemann

Ltd. 0142-961

Z/90/06041

contact with water or physiological fluid, and its evaluation both in vitro and in vivo. We have developed the epoxy groups containing hydrophilic copolymers that form stable, covalent bonds with amino groups coated on the substrate. The epoxy group is stable in normal atmosphere, unlike the isocyanate group. The hydrophilic component of the epoxy-containing hydrophilic polymer (top coat) is biocompatible vinyl pyrrolidone (NVP). Polyamino compounds, which can be obtained by the hydrolysis of polyisocyanate under special conditions, are used as an under coat.

EXPERIMENTAL Polymer

synthesis

The molecular design of the vinyl pyrrolidone (NVP) copolymer is indicated by (1 ) the introduction of binding sites (epoxy group) which react with the amino group on the substrate, (2) the retention of the complex in water or against physical friction, and (3) high molecular weight. Regarding binding sites (l), we have examined glycidyl acrylate (GA). With respect to retention of the complex (2). which will be described in more detail later, it became clear that sufficient durability against water or physical friction cannot be obtained with only NVP and GA; a hydrophobic component is found to be needed. As the

9-06 Biomaterials

1990, Voi 1 I August

419

Low-friction surface for medical devices: S. Nagaoka and R. Akashi

Table 1 The relationship between polymerization molecular weight of the polymer obtained Run

Monomer*

Solvent’

V-65

‘fteld

(9)

(9)

Iv)

(%)

20 20 20 20 20 20 20 20 20 20

20 10 5 4 2 0 4 4 4 4

2 2 2 2 2 2 10 6 4 1

70 90 90 90 90 90 90 90 90 70

conditions and the

MW

x IO5

NC0

Primary amine Hydrolysis

NC0 1

2 3 4 5 6 7 8 9 10

*NVP/VAc/GA Polymerization

1.25 2.60 5.14 5.90 7.61 Gel 2.51 4.50 5.49 4.75

of the polyisocyanate

and its hydrolysis

A 20 cm polyurethane (Tecoflex; Thermedics Inc.) tube, with an outer diameter of 0.8 mm, was soaked in a 4% polyisocyanate (Trimethylol propaneflolyrene diisocyanate adduct: isocyanate content 13.2 wt%, Nippon Polyurethane Co.) solution of methyl ethyl ketone for 1 min. Then, it was pulled slowly up and dried for 10 min in a dry nitrogen stream at 40°C. The polyisocyanate is highly compatible with the substrate and never peels off the substrate. The coating thickness observed by the scanning electron microscopy (HITACHI S-800), was 0.2-0.3 pm. Hydrolysis of isocyanate is commonly used to form amino groups, but there is the possibility of a side reaction of the remaining free isocyanate with the amino group to create urea. To prevent this, it is recommended to achieve the reaction in alkaline water of high concentration at a low temperature6. Consequently, 0. I-3.0 N aq. NaOH was used as the hydrolysis medium. After soaking the polyisocyanatecoated polyurethane tubes for a specific amount of time at 2O”C, they were rinsed with distilled water for 3 h and dried in vacuum. Figure 1 shows the reaction scheme.

Coating of the hydrophilic polymer with the polyamine undercoat

and its reaction

After hydrolysis, the test specimen was dipped in a 4% chloroform solution of the hydrophilic copolymer, slowly

420

Polyisocyanate Figure 1

Reaction scheme of hydrolysis of polyisocyanate.

polylsocyanate polyurethane -)

= 85/l O/5 (wt/wt); ’ isopropanol. period, 3 d: temperature, 40°C.

hydrophobic component, we selected vinyl acetate (VAc) which has a suitable molecular reactivity ratio with NVP. High molecular weight (3) is closely correlated with the low frictional property of the surface. The synthesis of the polymer was based on these design indications. NVP, GA and VAc were distilled in vacuum. The polymerization was conducted in a tube by using azobisdimethyl valeronitrile (V-65) as an initiator. A specified quantity of monomers and initiator were dissolved in isopropanol. The mixture was connected to a vacuum system. Then the system was purged with nitrogen. The optimum polymerization temperature and period were 40°C and 3 d respectively. The polymerization conditions are summarized in Table 1. The products were dissolved with chloroform and precipitated in n-heptane. The purified copolymers were dried in a vacuum oven at 50°C for 3 d. The yields of the polymers were determined by the weight of the polymers obtained and the weight-average molecular weight (MW) was determined by a Waters GPC Model 244 using a polystyrene standard.

Coating

-

Biomaterials

1990, Vol 11 August

NC0

NC0

I

I

hydrolysis

c=o

I

3N NaOH

coating

e

CH3 hydrophlllc copolymer

4 capolymercholn

pro]ectlng lntotheflud

hydrophlllc

Figure 2 Schematic low-friction surface.

slhppery coating

representation

extract

-

of the reaction to form hydrophilic

pulled up, dried and coated. Then it was heated at 90°C in air for 3 h to react the epoxy groups with the amino groups. A relatively high temperature and long reaction time were employed, attributable to the lower reactivity of the aromatic amines formed in comparison with the aliphatic amines. The hydrophilic copolymer which was not bound with the polyamine undercoat was extracted by washing in distilled water at 70°C for 3 h. Figure 2 shows the reaction scheme. Next, the durability test of the coating layer was performed. The test specimen was boiled for 2 h for the acceration test to examine the stability against water and squeezed twice between wet pig skins, which is a model of biological tissue, under a compression of 10 g/cm2, to examine the mechanical strength of the coating. The change of the static friction coefficient was then measured to characterize coating durability. Poly(vinyl pyrrolidone) (PVP) and poly(hydroxyethyl methacrylate) p(HEMA) were coated on the surface of the untreated polyurethane tubes using 4% chloroform solution and dried.

Measurement

of the surface

friction

coefficient

After the durability test, the catheter was cut to a length of 5 cm and fixed on a glass plate as shown in Figure 3. Then, the catheter was wetted with physiological saline solution and a weight (100 g, cross-linked collagen film with thickness of about 100 pm, which is a kind of biological tissue model, was attached on the bottom of it using adhesive (cyanoacrylate), and wetted with the same fluid), was placed on the end of the catheter. One end of the glass plate was gradually inclined to obtain the initial inclination angle required for the weight to begin slipping. The static

Low-frxtion

surface

for medical

5.9 X 1 05. The polymer and it was difficult

plate

devices:

obtained

were :ollagen codted weight

used for the synthesis

various compositions. in chloroform

for

were

with

obtained

VAc were using

JEOL

with Test materials Figure

3

the

Evaluatfon

method

of surface

frictional

coefficient

coefficient

formula:

(,u) was

calculated

friction

according

to

Four

of

coefficient.

general

weighing

anaesthesia

urethane inserted

kg

(pentobarbital

catheter

hydrophilic

2-3

and

low frictional

bilaterally

were

surface

into the

used.

sodium),

a polyurethane

Under

a control

catheter

the

arteries

of the

were

and pulled back to the original repeated

three

After

times

longitudinal

inserted

polymer

an additional

position while

5 cm

then

and spattered

performed

microscope. from

lightly rubbing

with

AND

Tab/e

were

initiator

the

125

carbon

scanning

change

which

after

by the

conditions.

static

friction

of the hydrophilic

MW

O/5,

formed

various

of the

consists

= 85/l

The surface

friction

after the durability

= 590 000) and

hydrolysis

of poly-

The concentration

of

surface with

possible

friction

formation

coefficients

to obtain

with

N

0.1

measured

sufficient

aq.

NaOH

low frictional the

the friction

the optimal property,

even

surface

polymer with

coefficient

was

to

in the

1. But

to achieve

the durability conditions hydrolyse

for the

for 3 to 5 min at

the durability

increased

of

in Figure

after

3 N aq. NaOH

However,

reduction

conditions

One of the optimal amination

hydrophilic

polyisocyanate

before and

because

of urea as a side reaction as shown

1 to 3 N aq. NaOH,

binding

were

test.

room temperature.

electron

increased

decreased

sharply under conditions

and of

were

excised

were

and

prepared.

the sections

were

polymer

000

to

between

depends

000,

As

shown

upon

of the polymer

weights

obtained

depending

the molecular

mainly

polymerization

weight

761

in the

weight

on

the

kinetics

of

of the polymer

the

concentration

molecular

weight

of

the

hydrophilic

higher, both with a low concentration

of the

occurred molecular

increased

and the polymer weight

Nevertheless, too much,

10

gel

3200

‘Z&O0

polymer

conditions

obtained

was

FIgwe

4

hydrolysis

Change

of

FT-IR

1200

603

lriuvenumber

could not be dissolved.

polymerization

of the

polymer

of the initiator

of the monomers.

of monomers

Run 4 is one of the most desirable the

with

The examination

surfaces

molecular

conditions.

if the concentration

and

dried

of ethanol

cm-‘)

3 min).

completely

5 cm of the blood vessels

relationship

and with high concentrations formation

series

20°C.

and monomer.

The became

under

excellent

solution.

4200)

microscope.

radical polymerization, obtained

and

FT-IR

by polyisocyanate

disappeared

and 3367 the

surface

test, werefound.

It was found that the molecular

polymerization

removed

0~0~

with

(Shimazu coated

cm-‘)

(3250

(NVPNAc/GA

aminated

in 0.2 M phosphate

haematoxylin-eosin.

and the resulting

from

high

aq. NaOH.

5 shows

1%

S-800

time,

of hydrophilic

1 shows

obtained.

Hitachi

of

DISCUSSION

Polymerization conditions

a

in a graded

of the inner

by optical

RESULTS

with

parts of the catheters

sections

staining

observed

were

of the tip of the catheter

critical-point

by the

the inserted

were

(2260

of the surface

It is not

with gold palladium.

At the same

histological After

stained

dehydrated

amylacetate,

dioxide

catheters

1% glutaraldehyde

and

were

The

the

identical

NaOH and the time of hydrolysis were changed independently.

both catheters.

on the surface

the

pH 7.4,

Specimens

was

Figure coefficient the

the blood vessel was excised along

axis.

formation

was fixed with buffer,

with

3 h insertion,

thrombus

and

groups with

isocyanate the catheters

that

because

(3 N, aq. NaOH,

hydrolysis

groups

hydrolysis

were

against the intima of the blood vessels. The procedures

the

and after

same

rabbit. Then,

clear

and reaction

of the polyurethane

and amino poly-

with

(10 cm in length)

femoral

composition

of

measurement

was roughly

Figure 4 shows the FT-IR spectrum

lsocyanate

each

coating

90%).

of polyisocyanate polymer

of the surface

experiment rabbits

monomer

NMR made

obtained

after

The contents

the

before

Animal

spectrometer

with

soluble

of the surface.

,u = tan 8, and this value is used as a parameter

the surface

above

polymer

insoluble

of 5 wt%.

from 0 to 30 wt%.

yield (about

Hydrolysis hydrophilic friction

and

GA content

of the polymer feed

polymer

mentioned

of the hydrophilic

It was found that the polymers

FX-100

composition

R. Akashi

on the substrate.

conditions

purification

changed

and

in Run 5 was too viscous

to coat smoothly

The best polymerization

S. Nagaoka

spectra

of

ROO

Lion

;cm-‘8 the

surface

accompanyrng

the

of polyisocyanate.

Biomatenals

1990,

Voi

1 1 August

421

Low-friction surface for medical devices: S. Nagaoka and R. Akashi

0.3 0.2

0.2

PHEMA coating

0.1

0

0.1

I

1

I

0

10

20 Time

40

60

(min)

Figure 5 Effect of hydrolysis conditions of polyisocyanate on the surface frictional coefficient. Hydrophilic polymer: NVPNAc/GA = 85/l O/5 (wt/wt), MW = 590 000. 0. 0.1; 0, 1.0: 0, 3.0 N NaOH.

higher alkaline concentration or longer incubation time. These conditions may hydrolyse the substrate including the polyamine thin layer. Under scanning electron microscopy, the thickness of the hydrophilic layer was found to be approximately 5 pm after the durability test. This is clearly thicker than the end-toend distance of a polymer with mol. wt 590 000. Consequently, it was suggested that the hydrophilic low friction surface was bound with polyurethane substrate by the reaction between epoxy groups and amino groups at the boundary layer. The upper hydrophilic layer was rendered insoluble by the reaction between the remaining epoxy groups (cross-linking or polymerization) to form a layer with high water content, and polymer chains could project from the outermost surface with relatively free molecular motion as shown in Figure 1.

Effect of the molecular weight of the hydrophilic copolymer on the friction coefficient Figure 6 shows the static friction coefficient (slipperiness) of the polyurethane catheter fixed on the glass plate (as shown in Figure 3). coated and bound with the hydrophilic copolymer (NVP/VAc/GA = 85/l O/5 (wt/wt)) having a different molecular weight. The static friction coefficients were decreased with the increase of the molecular weight of the hydrophilic polymer up to 400 000. The static friction coefficient of an uncoated (control) polyurethane catheter was 0.32 f 0.02 (n = 5). The polyurethane catheter coated with hydrophilic p(HEMA) showed a friction coefficient of 0.18 +_ 0.02 (n = 5, shown by the arrow). This improvement is deemed not enough for protection of mucous membranes of luminal surface and manoeuverability of medical devices. The dotted line (p = 0.035) indicates the static friction coefficient of the polyurethane catheter coated with non-bonded PVP (MW = 400 000), solubilized in water. This low friction surface does not injure mucous membranes or blood vessel. But of course, this slipperiness dissipated in a short while due to the washing off of the PVP from the surface.

422

\

Biomaterials

1990, Vol 11 August

C

I

2 Molecular

I

4

I

I

6

8 x105

Weight

of Copolymer

Figure 6 Effect of molecular weight of the hydrophilic polymer on the surface frictional coefficient of the surface. Mean + SD(n = 5). NVPNAc/GA = 85/10/5 (wt/wtJ.

As shown in Figure 6, a molecular weight of the bound hydrophilic copolymer of at least 400 000 is necessary to obtain excellent low frictional property of the surface when in contact with water or physiological fluid. The dependency of the frictional property on the molecular weight of the copolymer may be explained in connection with so-called Toms effect’. Although this effect has not been fully explained in molecular terms, it is known that very small traces of water-soluble polymers added to water can greatly reduce turbulent friction on the surfaces past which the fluid flows, depending on the molecular weight of the polymers’*. When the water-soluble copolymer is bound at the outermost surface of the substrate and projecting into the water or physiological fluid with relatively free molecular motion as shown in Figure 7, the similar effect could be expected and copolymers with higher molecular weight could reduce the friction coefficient more than the copolymers with lower molecular weight.

Effect of the chemical composition of the hydrophilic polymer on the durability of low surface friction Figure 7 shows the relationship between the content of VAc in the hydrophilic polymer and the friction coefficient after durability test. In these cases, the content of GA was kept at 5 wt%. As the content of NVP was increased, the friction coefficient decreased, but, unless hydrophobic monomer (VAc) was copolymerized with 10 wt% or more, it was not possible to obtain sufficient durability, and the low surface friction deteriorated easily (Cl -+ 0 in Figure 7). These results suggest that the durability of low surface friction is closely correlated with the mechanical strength of the hydrophilic copolymer. It is essential to copolymerize the hydrophobic monomer (VAc) together with NVP and GA to make the hydrophilic low friction layer stable against boiling water or physical friction. However, if the content of VAc was over 10 wt%, the hydrophilicity of the copolymer decreased and the low surface friction property deteriorated.

Low-friction

Almost

z

no friction

vessel

C

surfaces artery

.w .-f U

E

for med/cal

devices:

was felt between

in the case of the coated Figure

3

surface

8 shows

of the

rabbit.

were covered

network,

erythrocyte,

electron

of the

and blood

polyurethane

composed

and

platelet

catheters

no thrombus

of the

in the femoral

non-coated

leucocyte,

slippery surface showed

R. Akashi

micrographs

with thrombus

contrary, all of the polyurethane

8

the catheter

3 h after insertion

All

catheters

and

catheter.

scanning

of two catheters

S. Nagaoka

of a fibrin (a). On the

with the hydrophilic

formation

or adhesion

v

of blood components

C ._0

graphs of the cross-sections

of the femoral

U

rubbed

In the case of the non-coated

I= U ._

polyurethane destroyed.

+I

elastic

3

lesions.

Figure

+I

.-

VI

catheter, There

the

were

anywhere.

the

These

in the

hydrophilic

photomicro-

arteries that were

of the femoral

no endothelial

contrary,

with

histological

the intima

cells

were

case

artery was nor

internal

catheter-induced

of the

slippery

were observed and the endothelial

3b

2b

lb

the typical

the catheters.

membrane On

"0

9 shows

with

catheter

(b).

polyurethane

surface,

no lesions

cell layer was not damaged.

CONCLUSION of

Content Figure

7

coefficient

l,

after

Effect of

of hydrophilic

the

durability

surface. test.

polymer

Mean

MW

f

= 50

VAc

composition

so (n = 5). x

(wt%) 0.

on the surface Before

A new

frictional

durability

test;

hydrophilic

been developed to the

104.

polyamino

substrate.

determined

the optimal

to be NVP/VAc/GA

chemical = 85/l

composition O/5

was

(wt/wt).

amino

compound

polyisocyanate

with

obtained

aq. NaOH

In viva experiments The

insertion

catheter

and

weight the

manipulation

with a hydrophilic

with

8

Scanning

hydrophilrc

electron

low-friction

of the

low frictional

easier than those with the non-coated

Figure

or physiological

micrographs

surface

polyurethane

of the surface

was much

have

a

3 h after

insertion:

essential

molecular of

low

(a) polyurethane

formed

on the

was composed

of vinyl

acrylate.

by

the when

depends

polymer.

friction

The

poly-

hydrolysis

of

in contact with

upon the molecular To obtain

hydrophilic

greater

than

coefficient

catheter

use has (top coat)

of high concentration.

for the

weight

polymer

coat)

coefficient,

fluid,

hydrophilic

it was

durability

catheter.

of the catheter

of the

friction,

polyurethane

for medical

and glycidyl

was

The surface frictional water

(under

copolymer

vinyl acetate,

compound

surface

a hydrophilic

The hydrophilic

pyrrolidone, As explained,

low friction

by binding

(non-coated),

was

low surface copolymer

400

000.

related

(bl polyurethane

to The

to the

cathetef

surface.

Biomatenals

1990,

Vol

1 1 August

423

Low-friction surface for medical devices: S. Nagaoka and R. Akashi

/

, lesions

endothelial cell layer

Figure 9 Histological observation of the mural section of the blood vessel after catheter manipulation (stained by haematoxylin-eosineJ. (aJ Polyurethane catheter (non-coatedJ. Catheter-induced lesions were clearly observed and endothelial cell layer was fallen off (bJ Polyurethane catheter with hydrophilic lowfriction surface. No lesions were observed and the endothelial cells formed normal monolayer.

mechanical strength of the hydrophilic copolymer, and the addition of hydrophobic vinyl acetate was essential to prevent the deterioration of the low friction coating. The polyurethane catheter coated with the hydrophilic low friction surface was easy to insert in and did not damage the blood vessel during manipulation and, furthermore, showed improved blood compatibility.

2

3

4

5 6

REFERENCES

7 1

424

Ducatman, B.S., McMichan, J.C. and Edwards, W.D., Catheterinduced lesions of the right side of the heart, JAMA 1985, 253, 791-795

Biomaterials

1990. Vol 11 August

8

Ford, SE. and Manley, P.N.. Indwelling cardiac catheters: an autopsy study of associated endocardial lesions, Arch. Pathol. Lab. Med. 1982,106, 314-317 Lange. H.W.. Galliani, CA. and Edwards, J.E.. Local complications associated with indwelling Swan-Ganz catheters: autopsy study of 36 patients, Am. J. Cardiol. 1983, 52, 1 108- 1 1 1 1 Lazarus, H.M., Lowder, J.N. and Herzig, R.H., Occlusion and infection in Broviac catheters during intensive cancer therapy, Cancer 1983, 52, 23-48 Micklus, J.M.. Coated substrate having low coefficient of friction, United States Patent Number 4, 1 19, 094, 1978 Tanaka, H., Hydrolysis of polyisocyanate. Bull. Chem. Sot. Japan. 1976,49,2821-2823 Gadd, G.E., Turbulence damping and drag reduction produced by certain additives in water, Nature 1965, 206, 463-467 Kohn, M.C., Energy stage in drag-reducing polymer solutions, J. Polym. Sci., Polym. Phys. Ed. 1973, 11, 2339-2356

Low-friction hydrophilic surface for medical devices.

A hydrophilic polymer surface was developed exhibiting excellent low frictional property, namely slipperiness, when in contact with water or physiolog...
820KB Sizes 0 Downloads 0 Views