Laser Alignment System for HighQuality Portable Radiography1 Heber MacMahon, MD NicholasJ. Yasillo Michael Carlin, RT
In portable radiography, image quality is degraded by scattered radiation. Use of an accurately aligned, antiscatter grid can provide consistently high image quality, but the necessary precision for grid alignment is difficult to achieve with conventional techniques. The authors developed a simple laser alignment method for bedside radiography. A compact laser device is mounted on the collimator housing of a mobile radiography machine, so that the laser beam is precisely parallel to the central x-ray beam. A small reflector device, which attaches to the edge of the grid cassette, indicates beam centering and alignment in a clear, intuitive way. In clinical use, the laser alignment technique provided uniformly high image quality, equivalent to that of fixed equipment. The system has the advantages of being simple to use and inexpensive to implement with existing equipment. INTRODUCTION In portable radiography,
U
usually
poor The
misunderstood.
of mobile diation,
x-ray
equipment;
which
fogs
the
image problem in fact,
radiograph,
quality
is widely
is commonly it is caused reducing
accepted, attributed
mainly contrast
but
by uncontrolled and
the
to intrinsic obscuring
causes
are
limitations scattered
diagnostic
rain-
formation (1) (Fig 1). Use of an accurately aligned antiscatter grid can provide consistently high image quality, but the necessary precision is difficult to achieve with conventional “eyeball” technique. A mechanical alignment system has been demonstrated, but currently it is not available commercially (2). For portable radiography, excellent results can be achieved with a 6: 1 or 8: 1 grid, provided that the beam energy is no greater than 90 kV and that the grid is accunately aligned. Use of a higher kilovoltage requires use of a higher grid ratio, with correspondingly more critical alignment requirements, to achieve adequate reduction of scattered radiation. Figure 2 demonstrates accurate beam alignment compared with inaccurate alignment. In the case of a conventional lead strip grid, align-
Index
terms:
Images,
1992;
RadioGraphics I From Laude
July C
quali
#{149} Lasers
23 and
received
bedside
#{149} Radiography,
technology
12:111-120
the Department ofRadiology, award for a scientific exhibit
R.SNA,
#{149} Radiography,
September
University at the 1990 12; accepted
ofChicago, 5841 S Maryland Ave. RSNA scientific assembly. Received September
13. Address
reprint
Chicago, IL 60637. Recipient ofa April 22, 199 1 ; revision requested
requests
Cum
to MM.
1992
111
.
:-
Figure
1. (a) Portable radiograph obtained with conventional nongrid technique (90 kV, OC film/Lanex medium screen [Eastman Kodak, Rochester, NYJ). Note poor contrast and detail, particularly in the mediastinum. (b) Portable radiograph of the same patient obtained the next day with an accurately aligned 6: 1 grid
and
the same
markedly strated.
technique.
improved
In addition
due
ment
is critical
across along
the grid lines. Moderate the direction of the grid
significantly
in only
impair
to differences
to reduced
one
image
scatter.
that
dimension,
angulation lines does
quality.
With
in positioning
Fine
pulmonary
is, not
a fo-
cused grid, it is also necessary for the beam to be centered accurately. Variations in focusfilm distance affect mainly film density, provided that the x-ray beam is properly centered and aligned. In this article, we illustrate the high image quality that can be achieved by using an accuratehy aligned, antiscatter grid in combination with
a moderately
wide-latitude
screen-film
system for portable radiography. We describe the systems we devised to verify and achieve accurate x-ray beam alignment, and we compare the actual alignment accuracy achieved with conventional techniques and a new laser system
in a series
ofchinical
cases.
and
detail,
U
lung
MATERIALS
inations.
RadioGraphics
U
MacMahon
et a!
lines,
is clearly
AND
METHODS
quality
is
demon-
Two
pairs
of small
radiopaque
tanta-
lum markers (1 x 3 x 0.7 mm) are inserted into the front and rear surfaces of several grid cassettes (Fig 3). Each pair of markers is precisely
positioned
(0. 1-mm
tolerance)
so that
when the x-ray beam is perfectly aligned and centered, the markers are exactly superimposed on the film. Misalignment on decentering of the beam results in misnegistration of the markers, which can be accurately quantitated with a calibrated magnifying loupe (Fig 4). Thus, alignment errors can be determined in both vertical and horizontal dimensions 1#{176}. A simple
been developed alignment and relative positions graph.
U
the image
septal
. Verification of Beam Aligmnent We devised a method by which the accuracy ofx-ray beam alignment can be determined in a large number of bedside radiographic exam-
within
112
expansion,
including
computer
program
that determines focus-film distance of the markers
Volume
has
the beam from the on the radio-
12
Number
1
A
B
2.
Correct versus grid alignment.
In position A, the x-ray beam is accurately aligned with the grid (shown sche-
I
I
/
r
Figure incorrect
matically in cross section). With the x-ray tube in position B, the beam strikes
the grid at an angle,
caus-
ing a much greater proportion of the primary x rays (ie, not scattered radialion) tO be absorbed by the
‘I I I
grid.
This
creased and
results image
in decontrast
density.
..
‘1.MM’ ,J ,.
.‘
4/
FRONT Figure tantalum
3.
Schematic markers
were
SIDE illustrates inserted
where into
radiopaque the
front
rear plastic surfaces of 6: i external grid cassettes. The markers were used to accurately determine x-ray beam angle in portable radiographs.
January
1992
and
Figure 4. Enlarged view of a portable chest radiograph shows the tantalum markers, which do not interfere with image interpretation due to their small size and peripheral location. The degree of misnegistration shown here indicates an alignment error of 3.5#{176}.
MacMahon
Ct
a!
U
RadioGraphics
U
113
Figure
5.
Prototype laser device aluminum housing
in a machined
on a mobile radiography be rotated on its pedestal parallel to the grid that the laser beam the
.
reflector
enclosed mounted
unit. The laser can within fixed limits,
lines, in order to ensure coincides with the area
of
device.
Laser
System
Alignment
To facilitate accurate alignment of the central x-ray beam with a focused grid in a clinical setting, a laser alignment system was devised. This
technique
laser
line
employs
projector
reflector on the mobile
device. collimator radiography
Medical
Systems,
and
a single,
low-power
a unique,
compact
The laser device is mounted housing of a conventional machine (AMX 4; GE Milwaukee)
(Fig
5).
The
cassette
[35
x 43-cm]
(which
holds
film
cassette)
a 14
x
when
17-inch the
x-ray
beam is accurately centered. An opaque line on the transparent front surface of the colhimaton housing appears as a dark shadow within limation
the
field light.
projected This
dark
by the line
serves
standard as a dis-
indicator, since it coincides with the reflection at a distance of48 inches (122 which is the focus distance of the grid (Fig 6). Angular beam misalignment across the grid
is indicated
by
proximately cm), which
3 x 2 x 2 inches) temporarily attaches
sible la-
sen is powered from the collimation light cmcuit and is automatically activated when the collimation light is turned on. The device is located so that the laser beam is projected parallel to the central x-ray beam, with a separation of8 inches (20.3 cm). The laser beam is converted from a spot to a line by a cylinder lens. The linear laser reflection aligns with the long side ofthe 16 x 18-inch (40 x 45-cm) grid
tance laser cm),
col-
corner
bright beam
RadioGrapbics
U
MacMahon
et a!
grid
device
(ap-
(7.5 x 5 x 5 to any acces-
cassette
by means
will
reflection on the mesh. is at an angle to the grid,
of
be
at a corresponding
If the x-ray the laser beam
angle
to the
box,
and the reflected laser beam will cause a second reflection on the mesh. The amount of separation of the two reflected lines on the mesh indicates the magnitude of alignment error. A separation between the reflections of 4 mm corresponds to approximately 3#{176} of beam angulation, which we regard as the maximum acceptable angle for optimal image quality ment
with errors
reflector rate, the producing
U
of the
reflector
aluminum brackets and a magnetic strip (Fig 7). The front of the reflector box is covered by a mesh material supported by transparent plastic material. At the bottom of the box is a mirror. The incident laser beam creates a
clearly
114
a compact
indicated
a 6: 1 grid.
that
degrade to the
Thus,
x-ray
image
beam
quality
technologist
align-
are on
the
box. When beam alignment is acculaser reflections are superimposed, a single line (Figs 7, 8).
Volume
12
Number
1
7a.
6.
I
7b.
Figures 6-8. (6) Diagram of the laser alignment The laser (straight arrow) projects a linear beam line)
that
aligns
the x-ray
‘vI
:
Iv]
I,
Alignment Error 8.
Correct Alignment
a long
side
is accurately
of the
centered.
grid
cassette
when
An opaque
line on the
transparent surface row) casts a shadow coincides with the
grid.
of the collimator housing (curved anwithin the collimator light field, which laser beam at the focal distance of the Inaccurate grid alignment causes two bright red (appearing white here) on the front surface of
(7a)
reflections
1
with
beam
system. (dashed
the reflector
device
(arrows)
tion oftube position, the cating accurate alignment.
cides
with
beam
is correctly
device, laser
amount
the edge
(7b)
After
appropriate
in cross Magnitude
of separation
(8)
section.
indicating
Schematic
between
mdicoin-
that the
of the
Dotted
of alignment
conrec-
are superimposed, reflection also
of the grid cassette, centered.
shown beam.
.
reflections The laser
reflector
line nepresents error
is shown
the laser
reflections
the by the
(an-
rows).
For portable chest radiography, the reflecton box is placed at an upper corner of the grid cassette (Fig 9). (Because ofx-ray beam divergence, the grid cassette must project above the patient’s shoulder in every case.) The laser can be tilted, within fixed limits, along the axis of the grid lines so that noncnitical errors in long-axis alignment can be accommodated. For broad, heavy-set patients, the grid and film cassette are placed transversely, and the reflector box is placed ventically. In this situation, the collimator housing, with the attached laser device, is rotated 90#{176}
January
1992
so
that
upper
the laser beam projects parallel edge of the cassette. However,
ticipate sary
that when
near
this a new
maneuver
may
design
is implemented
be
to the we an-
unnecesin the
future.
. Clinical We conducted
Applications
verification
and
three
experiments
laser
alignment
with
the
systems
de-
U
RadioGrapbics
scnibed.
MacMahon
et a!
U
115
Figure
9.
Laser
in operation.
and grid
alignment
In this
cassette
case,
system the
film
are oriented
verti-
cally, and the laser reflection is aligned with the lateral edge of the grid cassette. A shadow caused by
the distance with
the
correct
indicator
laser
coincides
reflection,
focus-film
confirming
distance.
In the first investigation, we sought termine the correlation between grid ment and image quality in a systematic We conducted an observer test using of 161 conventional chest radiographic nations
performed
with
portable
summarized
to dealignway. a series exami-
equipment
and 6: 1 grids with tantalum markers. Three experienced chest radiologists reviewed the radiognaphs, from which the alignment markers had been obscured. The radiologists scored each case for overall image quality, using
the
a 10-point
scale
in which
highest quality and i the In the second experiment,
estimate
focus-film clinical
the
range
of alignment
distance
errors
variations
practice
when
the
gardless
that
tional
to and
occur
in
conventional
eye-
ball technique is used. In a random series of 200 chest radiographs obtained with tantahum-marked grids and conventional portable equipment, we measured the amount of misregistration
of the
alignment
In the last experiment, racy of the laser alignment of 50 clinical cases. The nations
were
performed
markers.
we tested system radiographic with
the
the accuin a series examinew
system
and tantalum-marked grids, and we measured the amount of misnegistration of the alignment markers. In all cases, portable radiography was performed Lanex
at 90 kV, with medium screens
RESULTS The results of the which the accuracy ment was correlated
6: 1 grids, (Eastman
OC film, Kodak).
and
U
116
U
RadioGraphics
U
subjective observer test, in of eyebalhed grid alignwith image quality, are
MacMahon
et a!
10.
The
average
quality
of exposure.
The range of alignment errors in 200 portable chest nadiognaphs obtained with conven-
10 indicated
lowest. we attempted
in Figure
score for each case was plotted against grid alignment error. All cases with the highest quality ratings also had accurate grid alignment. Several cases in which the x-ray beam was accurately aligned received a low quality rating because the radiographs were overexposed. However, no case with a large grid alignment error had a high quality rating, ne-
eyeball
technique
is shown
in Figure
1 1 The accuracy of unaided x-ray beam alignment was quite variable. In many cases, excellent alignment was obtained partly due to intensive technologist training, augmented by feedback provided by the tantalum alignment .
indicators.
The range consecutive
of alignment portable chest
tamed
the
with
laser
error for the radiographs
device
is also
50 ob-
shown
in
Figure 1 1 With laser alignment, the range of error was markedly reduced. Major alignment errors, which significantly degrade quality, were essentially eliminated. .
DISCUSSION In most hospitals, for portable chest U
a nongrid radiography.
technique is used Because of the large amounts of scattered radiation reaching the film, contrast and detail are reduced, particuharly in heavy patients. To retain adequate contrast in this situation, a relatively high-contrast film is commonly used. This practice can provide acceptable detail in optimally exposed areas of the lungs; however, visibility of mediastinal anatomy and the retrodiaphragmatic portions of the lower lobes is limited. Although the benefits of using an antiscatter
Volume
12
Number
1
10
0. 45
40. 0
35
LI
30.
U CONVENTIONAL LASER ALIGNED
LI
I.
25. 20.
5
10
15
20
ANGULATION ERROR ( DEGREES
ANGULATION
10.
ERROR
(DEGREES)
11.
Figures 10, 11. (10) Subjective image quality ratings correlated ble chest radiographs obtained with 6: 1 grids and conventional alignment with conventional eyeball technique compared with laser-aligned cases are accurate within 3#{176}, which is the maximum image quality.
with grid alignment errors for eyeball technique. (1 1) Accuracy the laser alignment method. Note acceptable error with a 6: 1 grid
161 portaof grid that 95% for optimal
of
a. b. Figure 12. (a) Portable radiograph of a 3 1-year-old woman with mediastinal emphysema and diffuse interstitial infiltrates due to bronchiolitis obhiterans with organizing pneumonia. The radiograph was obtained at 90 kV with an accurately aligned 6: 1 grid (OC film/Lanex medium screen) and shows excellent lung and mediastinal
detail.
the radiology medium
p = pot-table.
department
screen).
The
(b)
with image
Posteroantenior
a dedicated
quality
radiograph
chest
is virtually
obtained
radiography
unit
indistinguishable
with
the
patient
at 125 kV with from
that
erect
a 12: 1 grid
ofthe
portable
2 days
later
in
(OC fllm/Lanex
examination.
s
standard.
grid
in radiography are well established, use in portable radiography frequently produces unsatisfactory results. Such poor results can be attributed to inaccurate grid alignment and to use of inappropriately high kilovoltage levels for a given grid ratio. Also, the additional weight and bulk of an external
of a grid
grid gists,
cassette
may
particularly
ceived
as marginal
January
1992
be
a deterrent
when
the
As demonstrated here, use of a wide-latitude film with a 6: 1 grid and a beam energy of 90 kV or lower can provide consistently high image
quality
in portable
the contrast and ally indistinguishable with
state-of-the-art
detail
radiography.
achieved from that fixed
In fact,
can be virtuproduced
equipment
(Fig
12).
to technolobenefits
are
per-
or inconsistent.
MacMahon
Ct
a!
U
RadioGraphics
U
117
14a.
14b.
Figures patient
13, 14. (13) was radiographed
A chest phantom at 90 kV with
of 0#{176} (a) and 3#{176} (b) are illustrated. these controlled conditions, with
with 1 inch (2.5 cm) ofLucite added to simulate a 6: 1 grid (OC film/Lanex medium screen). Grid
and
However,
relation
endotnacheal
accurate
of the grid in x-ray beam is critically This can be achieved fixed geometry de-
central important (Figs 13-15). either with a mechanical, vice
118
U
RadioGraphics
or with
the
U
laser
technique.
MacMahon
Portable density
with detail,
radiograph obtained with due to grid misalignment.
the grid vertically including
oriented
improved
the
and
visibility
of
tube.
alignment
to the
adult errors
The effect of an alignment error as small as 3#{176} is clearly visible under resulting diminished contrast and mediastinal detail. With a 6#{176} alignment
ennon (not shown), grid cutoffeffects were quite pronounced. (14a) grid and film transversely oriented shows poor contrast and decreased ( 14b) Portable radiograph of the same patient, obtained subsequently accurately aligned, shows improved overall contrast and mediastinal
the catheter
an average alignment
Although
et a!
the
fixed
geometry
advantages and film tive
to the
method
has
has certain unique the ability to support the grid in any desired position rela-
approach
(eg, cassette patient),
the
the
virtues
laser
of low
Volume
alignment
cost
12
and
sim-
Number
1
d.
b.
Figure 15. Series ofportable radiographs ofa 205-lb (76.5-kg) liven transplant recipient illustrate the effects of grid alignment on image quality. Examples with approximately equivalent density were selected to facilitate visual comparison. The beam alignment angles were determined from the tantalum markers in each case (arrow). Alignment errors of 20#{176} (a), 8#{176} (b), 3#{176} (C), and 0#{176} (d) are shown. Image quality, including lung
and mediastinal detail, improves as the alignment the image quality is high, even though the lungs failure to use a grid results in poor image quality,
phicity, easily
as well
as the
to existing
facility
equipment.
to be adapted The
devices
are
particularly
damage
or
loss
is not
a serious
we have reflector
been boxes
able to reduce substantially
cently, of the sacrificing
alignment
reflector
inexpensive,
accuracy.
error decreases. are poorly inflated.
so that concern.
Re-
the size without
In the
future,
plan to incorporate the reflectors into a specially designed grid cassette. The highly reliable, solid state laser and all optics are enclosed within a machined, solid aluminum housing for maximum strength and durability.
we
January
1992
particularly
When grid alignment Except in children
in the
mediastinum.
In addition from scattered sure is a major phy.
In the
is accurate, as in d, and very small adults,
case
to poor image quality radiation, incorrect problem in portable of conventional
resulting film exporadiogra-
nongrid
radi-
ography performed with high-contrast film, the range of acceptable exposure is quite nannow. Apart from the kilovoltage and milhiamperage settings, variations in the focus-film distance can have a marked effect on exposure. Although a retractable tape measure is supplied
a projected
with
portable
distance
radiography
indicator
MacMahon
machines,
(as provided
et a!
U
RadioGraphics
U
119
with
the
laser
alignment
practical
advantages
variation
in focus-film
system)
and
has
may
result
distance.
some
in less
More
tant, the reduction in scatter means of a laser-aligned grid
impor-
achieved by allows the use
of
a wide-latitude film without sacrificing excellent local contrast throughout the lungs and mediastinum. Thus, diagnostic quality images can be achieved over a relatively wide range of exposures, compared with the nongrid and high-contrast film alternative. Automated exposure control devices have been developed for portable radiography and are likely to be used more widely as the technology improves. However,
accurate
exposure
alone
provide high image quality in the presence of uncontrolled scatter. The prototype laser alignment device has been readily accepted by technologists, as it does not interfere with the normal operation of the equipment and provides distance and alignment information that can be easily assimilated.
Initially,
the
expose grid
alignment
eliminates
this ence,
cutoff of the
Because
sistent
and
grid
predictable
excellent
consistently,
with
laser with
can
device
beam
alignment the
to the
is con-
be obtained
in technically
difficult
(Fig
Laser-aligned,
patient. The tantalum markers confirming accurate alignment Image quality is similar to that achieved dard high-kilovoltage examination.
views tions
RadioGraphics
U
MacMahon
et a!
of
are super(arrow). in a stan-
such as lateral and decubitus projecfor which accurate grid alignment is pan-
ticularly
important
scatter
due
to the
large
amount
of
generated.
situa-
Acknowledgments: lowing server
MD, and U 1. 2.
We are grateful
radiologists tests: John
Steven
who participated J. Fennessy, MD,
M. Montnen,
to the fol-
in the obCarl Vyborny,
MD.
REFERENCES Barnes
and
aging.
GT. Contrast RadioGraphics MacMahon H, Yasillo
1991;
ography Thoracic graduate
U
radiograph
an adult imposed,
experience
120
portable
device, expeni-
16). Although primarily developed for chest radiography, the laser system is potentially adaptable to any other type of bedside radiographic examination, including nonstandard tions
16.
to over-
laser
x-ray
“
_&
F
Figure
accurate
maximum
corrected
results
even
the
allows
primary
is easily
the
with
and
accurate
problem
and
is a tendency
because
achieved
grid
transmission
film.
there
radiograph,
‘
cannot
with
scatter
in x-ray
NJ, LoughhinJJ.
a fixed
geometry
system. Proceedings Radiology, Annual Course, San Diego,
Volume
im-
11:307-323.
Clinical
mobile
of the Meeting 1989.
12
nadi-
Society of and Post-
Number
1