Research

Original Investigation

Improving Cardiopulmonary Resuscitation With a CPR Feedback Device and Refresher Simulations (CPR CARES Study) A Randomized Clinical Trial Adam Cheng, MD; Linda L. Brown, MD, MSCE; Jonathan P. Duff, MD; Jennifer Davidson, RN; Frank Overly, MD; Nancy M. Tofil, MD, MEd; Dawn T. Peterson, PhD; Marjorie L. White, MD, MPPM, MEd; Farhan Bhanji, MD; Ilana Bank, MD; Ronald Gottesman, MD; Mark Adler, MD; John Zhong, MD; Vincent Grant, MD; David J. Grant, MBChB, MRCPCH; Stephanie N. Sudikoff, MD; Kimberly Marohn, MD; Alex Charnovich, MD; Elizabeth A. Hunt, MD, MPH, PhD; David O. Kessler, MD, MSc; Hubert Wong, MD; Nicola Robertson, BSc; Yiqun Lin, MD; Quynh Doan, MD, PhD; Jordan M. Duval-Arnould, MPH, CPH; Vinay M. Nadkarni, MD; for the International Network for Simulation-Based Pediatric Innovation, Research, & Education (INSPIRE) CPR Investigators Editorial page 109 IMPORTANCE The quality of cardiopulmonary resuscitation (CPR) affects hemodynamics,

survival, and neurological outcomes following pediatric cardiopulmonary arrest (CPA). Most health care professionals fail to perform CPR within established American Heart Association guidelines.

Supplemental content at jamapediatrics.com

OBJECTIVE To determine whether “just-in-time” (JIT) CPR training with visual feedback (VisF) before CPA or real-time VisF during CPA improves the quality of chest compressions (CCs) during simulated CPA. DESIGN, SETTING, AND PARTICIPANTS Prospective, randomized, 2 × 2 factorial-design trial with explicit methods (July 1, 2012, to April 15, 2014) at 10 International Network for Simulation-Based Pediatric Innovation, Research, & Education (INSPIRE) institutions running a standardized simulated CPA scenario, including 324 CPR-certified health care professionals assigned to 3-person resuscitation teams (108 teams). INTERVENTIONS Each team was randomized to 1 of 4 permutations, including JIT training vs

no JIT training before CPA and real-time VisF vs no real-time VisF during simulated CPA. MAIN OUTCOMES AND MEASURES The proportion of CCs with depth exceeding 50 mm, the proportion of CPR time with a CC rate of 100 to 120 per minute, and CC fraction (percentage CPR time) during simulated CPA. RESULTS The quality of CPR was poor in the control group, with 12.7% (95% CI, 5.2%-20.1%) mean depth compliance and 27.1% (95% CI, 14.2%-40.1%) mean rate compliance. JIT training compared with no JIT training improved depth compliance by 19.9% (95% CI, 11.1%-28.7%; P < .001) and rate compliance by 12.0% (95% CI, 0.8%-23.2%; P = .037). Visual feedback compared with no VisF improved depth compliance by 15.4% (95% CI, 6.6%-24.2%; P = .001) and rate compliance by 40.1% (95% CI, 28.8%-51.3%; P < .001). Neither intervention had a statistically significant effect on CC fraction, which was excellent (>89.0%) in all groups. Combining both interventions showed the highest compliance with American Heart Association guidelines but was not significantly better than either intervention in isolation. CONCLUSIONS AND RELEVANCE The quality of CPR provided by health care professionals is poor. Using novel and practical technology, JIT training before CPA or real-time VisF during CPA, alone or in combination, improves compliance with American Heart Association guidelines for CPR that are associated with better outcomes. TRIAL REGISTRATION clinicaltrials.gov Identifier: NCT02075450 JAMA Pediatr. 2015;169(2):137-144. doi:10.1001/jamapediatrics.2014.2616 Published online December 22, 2014.

Author Affiliations: Author affiliations are listed at the end of this article. Group Information: The International Network for Simulation-Based Pediatric Innovation, Research, & Education (INSPIRE) CPR Investigators are listed at the end of this article. Corresponding Author: Adam Cheng, MD, KidSIM–Assessing Simulation in Pediatrics: Improving Resuscitation Events (ASPIRE) Simulation Research Program, Section of Emergency Medicine, Department of Pediatrics, University of Calgary, Alberta Children’s Hospital, 2888 Shaganappi Trail NW, Calgary, AB T3B 6A8, Canada ([email protected]).

(Reprinted) 137

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

Research Original Investigation

Improving CPR With a CPR Feedback Device

E

ach year, cardiopulmonary resuscitation (CPR) is provided for thousands of children in North America.1,2 The quality of CPR directly affects hemodynamics, survival, and neurological outcomes following cardiopulmonary arrest (CPA).3-9 Unfortunately, health care professionals struggle to retain effective chest compression (CC) skills after basic life support or advanced cardiac life support training.10-12 Even well-trained health care professionals fail to consistently perform CPR within established American Heart Association (AHA) guidelines during CPA.8,13-17 Video-based instruction can be just as or more effective than traditional life support courses to teach CPR skills.18-24 In these studies, learners watched an educational video and were given opportunity to practice but did not receive feedback on the quality of CPR, and the training was not “just-in-time” (JIT). The JIT bedside CPR practice with audiovisual feedback (but without a video component) targeted to health care professionals with a high likelihood of needing to deliver in-hospital CPR was found to improve compliance with AHA CPR-quality guidelines on manikins and real children during CPA.25-28 What is unknown is whether JIT CPR visual feedback (VisF) training with videobased instruction before CPA or real-time VisF during CPR, alone or in combination, is effective in improving CPR quality. The CPR feedback devices provide visual or auditory feedback based on quantitative CPR metrics during CPA and have been shown to improve the quality of CCs when used during training29 and during real CPA events.8,30 Although these devices show potential, no randomized clinical studies to date have evaluated the training method for using CPR-quality VisF devices in CPA. Furthermore, multicenter clinical trials to assess the effect of JIT CPR VisF training before CPA or realtime VisF during simulated or real in-hospital CPA have not been performed to our knowledge. In this study, we evaluated a novel JIT CPR training intervention that includes a video-based learning component and a practical credit card–sized CPR VisF device. We also evaluated the use of the CPR VisF device during simulated CPA. The VisF device captures and displays depth and rate of CCs during CPR practice and CPA. We used explicit methods to deploy a standardized simulated CPA scenario, carefully trained and scripted confederate actors, and captured quantitative CPR data across 10 International Network for Simulation-Based Pediatric Innovation, Research, & Education (INSPIRE) institutions to describe the quality of CPR. The primary objective was to determine whether the use of JIT CPR training with VisF before CPA or realtime VisF during simulated CPA improves the quality of CPR.

Inclusion and exclusion criteria are described in the eMethods in the Supplement. Participants were recruited from 10 tertiary care centers (eTable 1 in the Supplement) into teams of 3, with one individual preassigned to the role of team leader to oversee the resuscitation efforts and 2 others assigned as team members expected to perform CCs. Participants included medical students, resident and fellow physicians, nurses and nurse practitioners.

Interventions VisF Device A prototype practical, CPR-quality VisF device (CPRcard; Laerdal Corporation) measuring the size of a credit card (55 × 85 × 1 mm) and placed on the middle of the chest uses accelerometer technology to record and display depth and rate of CCs (eFigure 1 in the Supplement). The device provides real-time VisF via light-emitting diodes for CC depth and rate. The CC rate is indicated by a set of 3 light-emitting diodes set at rates of less than 100, 100 to 120, and greater than 120 compressions per minute, while compression depth is indicated by separate lights set at depths of less than 40 mm (inadequate), 40 to 50 mm (infant), 50 to 70 mm (child or adult on a hard surface), and greater than 70 mm (child or adult on a soft surface). The CPRcard has undergone verification testing using a mechanical device subjecting it to controlled horizontal movements (with accurate depth and rate measurement), which demonstrated variation between −5.9% and 1.4% of target depth and ±3 compressions per minute of target rate (J. Eilledge, Laerdal Corporation, e-mail communication, August 28, 2014). JIT CPR VisF Training Before CPA The JIT CPR training comprised the following 2 components: a 5-minute standardized CPR training video, followed by 2 minutes of practice (per team member) with VisF on depth and rate of CCs during practice (eMethods in the Supplement). Those participants randomized to the no-JIT CPR training study arms did not watch the video but practiced CPR for 2 minutes with no VisF.

Outcome Measures Data were collected from the VisF device, including CC depth, CC rate, and CC fraction (CCF) averaged over the entire 12minute duration of the simulated CPA event. Primary Outcome Measure The primary outcome measure was CC depth. The proportion of CCs during each 12-minute simulated CPA with depth exceeding 50 mm was measured.32

Methods Research ethics board approval was obtained at all sites. Written informed consent was obtained from all participants. We conducted a multicenter, prospective, randomized, clinical 2 × 2 factorial-design trial with explicit simulation-based research methods31 to assess the quality of CCs performed during simulated CPA. Our 2 interventions were (1) JIT standardized, brief CPR training with video and VisF before simulated CPA and (2) real-time VisF during simulated CPA. 138

Study Participants

Secondary Outcome Measures Secondary outcome measures were CC rate and CCF. The proportion of time that the CC rate met AHA guidelines (100-120 per minute)32 was calculated by the rolling median of the last 5 CC measurements in a series. The proportion of CPA time with CC delivery was calculated as the time during CPA with CC delivery divided by the total arrest time.32,33

JAMA Pediatrics February 2015 Volume 169, Number 2 (Reprinted)

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

jamapediatrics.com

Improving CPR With a CPR Feedback Device

Original Investigation Research

Figure 1. Consolidated Standards of Reporting Trials Diagram 124 Teams (372 participants) assessed for eligibility

0 Initial exclusions

124 Teams (372 participants) randomized

Arm 1: No JIT and no VisF

Arm 2: No JIT and VisF

Arm 3: JIT and no VisF

Arm 4: JIT and VisF

31 Teams (93 participants) received allocated intervention as randomized

31 Teams (93 participants) received allocated intervention as randomized

31 Teams (93 participants) received allocated intervention as randomized

31 Teams (93 participants) received allocated intervention as randomized

0 Lost to follow-up

27 Teams (81 participants) analyzed 1 Team (3 participants) excluded from analysis due to protocol violation (CPR done on mattress) 3 Teams (9 participants) excluded due to missing data from technical issue

0 Lost to follow-up

27 Teams (81 participants) analyzed 1 Team (3 participants) excluded from analysis due to protocol violation (CPR done on mattress) 3 Teams (9 participants) excluded due to missing data from technical issue

0 Lost to follow-up

27 Teams (81 participants) analyzed 1 Team (3 participants) excluded from analysis due to protocol violation (CPR done on mattress) 3 Teams (9 participants) excluded due to missing data from technical issue

0 Lost to follow-up

27 Teams (81 participants) analyzed 4 Teams (12 participants) excluded due to missing data from technical issue

CPR indicates cardiopulmonary resuscitation; JIT, just-in-time; and VisF, visual feedback.

Study Procedures Randomization Each team was randomized into 1 of the following 4 study arms: (1) no JIT CPR VisF training before CPA and no real-time VisF during CPA (control), (2) no JIT CPR VisF training before CPA and real-time VisF during CPA, (3) JIT CPR VisF training before CPA and no real-time VisF during CPA, and (4) JIT CPR VisF training before CPA and real-time VisF during CPA (Figure 1). No participant was a member of more than 1 team. Randomization occurred at the level of the team, was stratified by study site, and was conducted in blocks of 4 to ensure equal numbers of teams and participants across the study arms (eMethods in the Supplement). Standardized Simulated Scenarios, Confederate Actors, and the Simulation Environment After randomization, all participants were given a standardized video orientation to the research study, followed by (1) 2 minutes of CPR practice for each team member, with or without JIT CPR training; (2) a pediatric septic shock simulation scenario; and (3) a pediatric simulated CPA scenario (eTable 2 and eFigure 2 in the Supplement). We standardized all elements of the simulation across sites and verified this by video review to minimize the simulation-specific threats to internal validity.31 All recruitment sites used an identical pediatric manikin (SimJunior; Laerdal Corporation) specifically designed and calibrated for CPR training (spring constant, 4.46 kg/cm; 22.3 kg of force required to press to 5 cm; and maximum compression depth of 7 cm). A 15-minute standardized septic shock scenario was conducted for all participants to minimize the Hawthorne effect of being videotaped, orient the team to the simulation environment, and draw attention away from CPR. The sepsis scenario was followed by a 10-minute debriefing with a focus on aspects of clinical care and teamwork (with no discussion of CPR). Next, a stanjamapediatrics.com

dardized 12-minute pediatric CPA scenario was conducted to collect data, depicting a 5-year-old child with pulseless electrical activity progressing to ventricular fibrillation. The VisF device was secured to the center of the manikin’s chest, but in study arms 1 and 3, the light-emitting diode lights were covered so that participants did not receive VisF during CPR. To ensure standardized delivery of the scenario, a preprogrammed scenario and a detailed scenario script were used at all sites (eMethods in the Supplement). Confederate actors were trained in a standardized fashion (eMethods in the Supplement) to play the scripted roles of respiratory therapist and medication nurse, both of whom were instructed not to perform CPR. We standardized the equipment and the location of the defibrillator and medication cart across sites. The simulator was placed on a hard stretcher with no mattress to eliminate mattress compressibility as a confounding variable.34 We verified compliance with this method by (1) centralized review of one prestudy pilot session from each institution and (2) centralized review of all study videotapes (eMethods in the Supplement).

Sample Size The sample size calculation assumed that the main effects of the 2 different interventions on the proportion of CC depth meeting AHA standards would comprise the 2 primary comparisons. Hence, allowing for Bonferroni adjustment, P < .025 was considered statistically significant for each of these 2 comparisons. For the sample size calculation, additivity of the intervention effects was assumed, with a plan to check this assumption in the analysis. The proportion of CCs meeting AHA standards for depth among health care professionals deprived of CPR feedback and JIT CPR training (control) has been reported to be approximately 60%, with an SD of approximately 25%.16 Assuming similar variability among teams receiving either intervention to detect a 15% difference in CCs meeting AHA standards for (Reprinted) JAMA Pediatrics February 2015 Volume 169, Number 2

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

139

Research Original Investigation

Improving CPR With a CPR Feedback Device

Table 1. Demographic Characteristics of Participants in the Study Arms No. (%) Arm 1: No JIT and No VisF (n = 81)

Arm 2: No JIT and VisF (n = 81)

Arm 3: JIT and No VisF (n = 81)

Arm 4: JIT and VisF (n = 81)

Male

24 (29.6)

30 (37.0)

28 (34.6)

23 (28.4)

Female

57 (70.4)

51 (63.0)

53 (65.4)

58 (71.6)

Characteristic Sex

Occupation, level of training Nurse

25 (30.9)

30 (37.0)

35 (43.2)

30 (37.0)

Attending physician

1 (1.2)

5 (6.2)

5 (6.2)

9 (11.1)

Other, RT, paramedic

2 (2.5)

1 (1.2)

0

0

Medical student

8 (9.9)

8 (9.9)

5 (6.2)

5 (6.2)

45 (55.6)

37 (45.7)

36 (44.4)

37 (45.7)

Resident Last basic life support course taken, mo Never

4 (4.9)

1 (1.2)

1 (1.2)

12

31 (38.3)

29 (35.8)

29 (35.8)

36 (44.4)

Never

31 (38.3)

40 (49.4)

34 (42.0)

37 (45.7)

12

31 (38.3)

19 (23.5)

34 (42.0)

29 (35.8)

Never

18 (22.2)

20 (24.7)

13 (16.0)

19 (23.5)

12

29 (35.8)

23 (28.4)

23 (28.4)

24 (29.6)

Never

42 (51.9)

46 (56.8)

47 (58.0)

38 (46.9)

1-5 Times

34 (42.0)

28 (34.6)

25 (30.9)

33 (40.7)

≥6 Times

5 (6.2)

7 (8.6)

9 (11.1)

10 (12.3)

Never

19 (23.5)

15 (18.5)

11 (13.6)

15 (18.5)

1-5 Times

48 (59.3)

48 (59.3)

52 (64.2)

42 (51.9)

≥6 Times

14 (17.3)

18 (22.2)

18 (22.2)

24 (29.6)

Last advanced cardiac life support course taken, mo

Last pediatric advanced life support course taken, mo

CCs on actual pediatric patient within 2 y

CCs on manikin within 2 y

depth between the control team and the teams receiving each of the interventions with 80% power at α = .025 for each of the 2 primary comparisons, the required sample size would be 108 teams (27 per study arm). Based on prior experience with multicenter, simulation-based research,35 we estimated that 15% of the groups would be excluded owing to technical issues with data acquisition (16 teams), so we aimed to recruit 124 teams in total (31 per study arm).

Statistical Analysis All data analyses were performed using statistical software (SPSS Statistics 21; IBM Corporation). Using team CC performance as the unit of measure, CPR event summaries of CC depth, CC rate, and CCF were calculated for each of 4 study arms. We used multivariable linear regression analysis to assess the effects of JIT CPR training or VisF on the primary and secondary outcome measures by including 2 intervention variables (JIT CPR training and VisF) and the interaction term (JIT CPR training × VisF) in the model. Recognizing that the effects could be nonadditive (whether or not the interaction test is statistically significant), we also report on the effect of each interven140

Abbreviations: CCs, chest compressions; JIT, just-in-time; RT, respiratory therapist; VisF, visual feedback.

tion at each level of the other intervention. All estimates of intervention effects are reported as differences in means and 95% CIs, with no adjustment for multiple comparisons.

Results Study Population In total, 124 teams (372 participants) were recruited from July 1, 2012, to April 15, 2014. Sixteen teams (48 participants) were excluded from the study because of technical issues (eg, the VisF device was not turned on at the appropriate time, or the VisF device was inadvertently shut off part way through the scenario [13 teams, 39 participants]) or because of a violation of the study protocol (eg, a mattress was used during simulated CPA [3 teams, 9 participants]). Data from the remaining 108 teams (324 participants) were included in the analysis (Figure 1). No significant difference was found in the demographic characteristics of the participants between the study arms at baseline (Table 1). The results were not significantly different in a sensitivity analysis that adjusted for clustering within centers.

JAMA Pediatrics February 2015 Volume 169, Number 2 (Reprinted)

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

jamapediatrics.com

Improving CPR With a CPR Feedback Device

Original Investigation Research

Table 2. Proportion of Chest Compressions (CCs) Meeting American Heart Association Guidelines for CC Depth Exceeding 50 mm Mean (95% CI), % CC Depth >50 mm

No JIT

JIT

JIT Benefit

JIT Mean Benefit Main Effect

No VisF

12.7 (5.2 to 20.1)

37.9 (27.0 to 48.7)

25.2 (12.7 to 37.6)

VisF

33.4 (23.8 to 43.0)

48.0 (39.8 to 56.2)

14.6 (2.2 to 27.0)

VisF benefit

20.7 (8.3 to 33.2)

10.1 (−2.3 to 22.6)

NA

NA

NA

NA

VisF mean benefit main effect

15.4 (6.6 to 24.2)

19.9 (11.1 to 28.7)

Abbreviations: JIT, just-in-time; NA, not applicable; VisF, visual feedback.

Table 3. Proportion of Chest Compressions (CCs) Meeting American Heart Association Guidelines for CC Rate of 100 to 120 per Minute Mean (95% CI), % CC Rate 100-120/min

No JIT

JIT

JIT Benefit

No VisF

27.1 (14.2 to 40.1)

40.7 (27.4 to 54.1)

13.6 (−2.3 to 29.5)

VisF

68.8 (57.3 to 80.3)

79.2 (71.1 to 87.2)

10.4 (−5.5 to 26.3)

VisF benefit

41.7 (25.8 to 57.6)

38.4 (22.5 to 54.5)

NA

NA

NA

NA

VisF mean benefit main effect

40.1 (28.8 to 51.3)

JIT Mean Benefit Main Effect 12.0 (0.8 to 23.2)

Abbreviations: JIT, just-in-time; NA, not applicable; VisF, visual feedback.

Figure 2. Effects of Just-in-Time (JIT) Cardiopulmonary Resuscitation Training and Visual Feedback (VisF) on Chest Compression (CC) Quality A CC depth >50 mm

B

C

Adequate CC rate

100

100

80

80

CC fraction No VisF VisF

100

CCs, %

60

CCs, %

60

CC Fraction, %

95

40

40

20

20

90

85

0

0 No JIT

JIT

Randomization

80 No JIT

JIT

Randomization

No JIT

JIT

Randomization

A, The mean proportion of CCs with adequate depth exceeding 50 mm. B, The mean proportion of CCs with a rate of 100 to 120 per minute. C, Chest compression fraction. Neither intervention had a statistically significant effect on CC fraction, which was excellent (>89.0%) in all groups.

Quality of CPR The overall quality of CPR was poor. The mean percentages of CCs with adequate depth in each study arm were 12.7% for the control, 37.9% for JIT alone, 33.4% for VisF alone, and 48.0% for JIT plus VisF (Table 2). The mean percentages for a CC rate of 100 to 120 per minute were 27.1% for the control, 40.7% for JIT alone, 68.8% for VisF alone, and 79.2% for JIT plus VisF (Table 3). The CCFs were 90.8% for the control, 89.5% for JIT alone, 91.7% for VisF alone, and 90.5% for JIT plus VisF (eTable 3 in the Supplement).

JIT vs No JIT CPR Training Before CPA The use of JIT CPR training before CPA resulted in a 19.9% (95% CI, 11.1%-28.7%; P < .001) absolute increase (ie, an 86.3% relative increase) in CC depth compliance compared with no JIT (Table 2) and a 12.0% (95% CI, 0.8%-23.2%; P = .04) absolute jamapediatrics.com

increase (ie, a 25.0% relative increase) in compliance with rate guidelines (Table 3). The CCF was uniformly high, and no significant increase in CCF was found with the use of JIT CPR training (eTable 3 in the Supplement).

VisF vs No VisF During CPA The use of real-time VisF during CPA resulted in a 15.4% (95% CI, 6.6%-24.2%; P = .001) absolute increase (ie, a 60.9% relative increase) in CC depth compliance and a 40.1% (95% CI, 28.8%-51.3%; P < .001) absolute increase (ie, a 118.3% relative increase) in CC rate compliance compared with no VisF. The CCF was uniformly high, and no significant increase in CCF was found with the use of real-time VisF (eTable 3 in the Supplement). Figure 2 shows the main effect of JIT CPR training before CPA and VisF during CPA on the quality of CPR. (Reprinted) JAMA Pediatrics February 2015 Volume 169, Number 2

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

141

Research Original Investigation

Improving CPR With a CPR Feedback Device

Interaction Effect Combined JIT CPR training and real-time VisF showed the highest compliance with AHA CPR guidelines for CC depth and rate. However, the interaction effect between JIT and VisF was not statistically significant (P = .24 for CC depth, P = .77 for CC rate, and P = .73 for CCF).

Visual Feedback

This prospective, randomized clinical study evaluated the effect of JIT CPR training before CPA and real-time VisF during CPA on the quality of CPR across multiple institutions. Our results showed that certified health care professionals struggle to provide guideline-compliant CCs during CPA in a simulated, team-based environment. We demonstrated that JIT CPR training before CPA or real-time VisF during CPR, alone or in combination, improves compliance with AHA guidelines for CPR quality. Two previous simulator-based studies with single rescuers demonstrated poor quality of CPR,36,37 with deterioration of CC quality following training.37 Pediatric health care professionals in a tertiary care center struggled to achieve AHA guidelines when performing CPR on real patients.8,15-17 While the performance of CPR in our control group was poor for CC depth and rate, the JIT and VisF–trained group had better CC quality than those who had received bedside JIT CPR training with no video component25-27 and those who had real-time VisF during the performance of CPR on real pediatric patients.8

The use of CPR feedback technology has demonstrated benefits for improving the quality of CPR in the context of CPR by 1 or 2 rescuers in the simulated environment.29,30 Human studies with adult patients in CPA demonstrate that the use of CPR feedback improves CPR quality40-43 and accelerates the rate of return of spontaneous circulation when combined with postevent debriefing.43 In our study, we found a significant improvement in CC depth and a more substantial improvement in CC rate with the use of VisF during simulated CPA. The lack of a more substantial improvement in CC depth suggests that additional focused training may be necessary, particularly given the association of increased CC depth with improved survival.42,44 In contrast, the pronounced improvement in CC rate, as well as recent single-center evidence demonstrating similar improvements in CPR quality in real patients,17 provides compelling evidence to support the widespread implementation of real-time VisF during CPA. Various factors likely influence the effectiveness of VisF during CPA, including health care professional responsiveness to VisF, the duration of CPA, patient age, and the number of health care professionals performing CPR. Future work should describe how health care professionals interact with VisF devices, demonstrate how teams can optimally integrate VisF, and assess whether VisF during CPR improves patient outcomes from CPA across multiple institutions.

JIT CPR Training

Limitations

The JIT CPR training before CPA, or training defined as “a training session conducted directly prior to … and at/near the site of the potential intervention,”25(p910) has emerged as a potential solution to address the poor retention of CPR skills after basic life support education.10-12 The provision of frequent JIT CPR training with integrated feedback (and no video-based learning) has been shown to improve the time to achieve CPR skills success and retention of CPR skills.25-28 In a study of adults, Bobrow et al38 showed that the use of real-time feedback during resuscitation training improved CPR quality and survival in patients with out-of-hospital CPA. Recently, the AHA has announced the launch of the Resuscitation Quality Improvement program,39 consisting of JIT CPR training with integrated feedback and video-based learning. The widespread uptake of this program may eventually replace traditional basic life support training for health care professionals. To our knowledge, our study is to first to date to evaluate JIT CPR training with integrated video-based learning and VisF before CPA in a manner mirroring the design of the AHA’s Resuscitation Quality Improvement program.39 We found significant improvements in CC depth and rate with a brief, focused JIT training session, establishing proof of concept supportive of this design. This evidence supports the implementation of JIT training in clinical environments where CPA is more common (eg, the intensive care unit). The timing, intensity, and frequency of JIT CPR training, as well as the minimally required cutoff for CPR competency, are variables that

Despite the attention to multicenter standardization,31 our study had several limitations. We had one simulated CPA scenario, with 2 team members providing CPR. As such, the results of our study reflect this specific simulated context, making it difficult to predict if our results are directly generalizable to patient care. We had 45 minutes between JIT CPR training and the CPA scenario, while in a real clinical environment, the delay between JIT CPR training and the provision of CPR is unpredictable. Furthermore, participants in the JIT CPR training groups received only one JIT training session. The eventual quality of CPR provided was likely influenced by these factors (ie, the time delay between JIT training and a CPR event and the frequency of JIT training). We excluded participants who had previous experience with CPR feedback devices from our study to remove this as a confounding variable. As such, the effect of the identical interventions on health care professionals with previous experience using CPR feedback devices was not assessed. We believe that the results of our study are generalizable to similar health care professionals naive to CPR feedback devices across a broad spectrum of geographic locations. Because of limitations in VisF technology, we did not collect data for several variables such as ventilation rate, tidal volume, and residual leaning force. Last, we executed an intentto-treat analysis. Variable compliance with VisF may have affected the results but would have imparted selection and ascertainment bias on the analysis.

Discussion

142

should be explored in future research to guide the design of JIT CPR training programs.

JAMA Pediatrics February 2015 Volume 169, Number 2 (Reprinted)

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

jamapediatrics.com

Improving CPR With a CPR Feedback Device

Original Investigation Research

Conclusions The quality of CPR provided by health care professionals is poor. It is feasible to improve CPR quality using a simple

ARTICLE INFORMATION Accepted for Publication: September 16, 2014. Published Online: December 22, 2014. doi:10.1001/jamapediatrics.2014.2616. Author Affiliations: KidSIM–Assessing Simulation in Pediatrics: Improving Resuscitation Events (ASPIRE) Simulation Research Program, Section of Emergency Medicine, Department of Pediatrics, University of Calgary, Alberta Children’s Hospital, Calgary, Alberta, Canada (Cheng, Davidson, V. Grant, Robertson, Lin); Hasbro Children’s Hospital, Alpert Medical School of Brown University, Providence, Rhode Island (Brown, Overly); Stollery Children’s Hospital, University of Alberta, Calgary, Alberta, Canada (Duff); Children’s of Alabama, University of Alabama at Birmingham (Tofil, Peterson, White); Montreal Children’s Hospital, McGill University, Montreal, Quebec, Canada (Bhanji, Bank, Gottesman); Ann & Robert H. Lurie Children’s Hospital of Chicago, Northwestern University School of Medicine, Chicago, Illinois (Adler); Children’s Medical Center of Dallas, UT Southwestern Medical Center, Dallas, Texas (Zhong); Bristol Royal Hospital for Children, University Hospitals Bristol, Bristol, England (D. J. Grant); Yale New Haven Health, Yale Medical School, New Haven, Connecticut (Sudikoff); Baystate Children’s Hospital, Tufts University School of Medicine, Boston, Massachusetts (Marohn); Departments of Anesthesiology, Critical Care Medicine, and Pediatrics, The Johns Hopkins University School of Medicine, Baltimore, Maryland (Charnovich, Hunt, Duval-Arnould); Division of Pediatric Emergency Medicine, Department of Pediatrics, Columbia University College of Physicians and Surgeons, New York, New York (Kessler); University of British Columbia, Vancouver, British Columbia, Canada (Wong); British Columbia Children’s Hospital, University of British Columbia, Vancouver, British Columbia, Canada (Doan); Critical Care Medicine, The Children’s Hospital of Philadelphia, University of Pennsylvania Perelman School of Medicine, Philadelphia (Nadkarni). Author Contributions: Drs Cheng and Lin had full access to all the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. Study concept and design: Cheng, Brown, Duff, Davidson, Overly, Tofil, Bhanji, Adler, Sudikoff, Hunt, Kessler, Wong, Lin, Doan, Nadkarni. Acquisition, analysis, or interpretation of data: Cheng, Brown, Duff, Davidson, Overly, Tofil, Peterson, White, Bank, Gottesman, Adler, Zhong, V. Grant, D. J. Grant, Marohn, Charnovich, Hunt, Kessler, Wong, Robertson, Lin, Duval-Arnould, Nadkarni. Drafting of the manuscript: Cheng, Brown, Duff, Overly, Kessler, Wong, Robertson, Lin, Doan, Nadkarni. Critical revision of the manuscript for important intellectual content: Cheng, Brown, Duff, Davidson, Overly, Tofil, Peterson, White, Bhanji, Bank, Gottesman, Adler, Zhong, V. Grant, D. J. Grant, Sudikoff, Marohn, Charnovich, Hunt, Kessler, Wong,

jamapediatrics.com

visual CPR-quality feedback device during JIT CPR training (with video) or during a simulated CPA event (with real-time feedback). The JIT CPR training or real-time VisF during CPR can improve compliance with AHA guidelines for CPR quality that are associated with better survival outcomes.

Lin, Doan, Duval-Arnould, Nadkarni. Statistical analysis: Hunt, Wong, Lin, Doan. Obtained funding: Cheng, Gottesman, V. Grant, Hunt. Administrative, technical, or material support: Brown, Davidson, Overly, Tofil, Peterson, Bhanji, Bank, Gottesman, Adler, V. Grant, D. J. Grant, Sudikoff, Marohn, Charnovich, Robertson, Duval-Arnould. Study supervision: Cheng, Brown, Davidson, Peterson, Bank, Gottesman, Zhong, D. J. Grant, Hunt, Kessler. Conflict of Interest Disclosures: Dr Tofil reported speaking at the fall 2013 Laerdal Corporation conference (she used her own created materials and did not receive a speaker stipend but was reimbursed for travel). Dr Nadkarni reported receiving support as the endowed chair, Critical Care Medicine, The Children’s Hospital of Philadelphia, University of Pennsylvania Perelman School of Medicine and from the following research grants: R03HS021583 from the Agency for Healthcare Research and Quality, a Nihon Kohden America research grant, U01HL107681 from the National Institutes of Health, 5R01HL058669-10 from the National Institute of Neurological Disorders and Stroke, and R01HL114484 and 1U01HL094345-01 from the National Heart, Lung, and Blood Institute. No other disclosures were reported. Funding/Support: This study was funded by a research grant from the Heart and Stroke Foundation of Canada and the Canadian Institute for Health Research. Laerdal Corporation donated CPRcards and loaned each recruitment site a SimJunior manikin for use in the study. Data collection for one site (Columbia University College of Physicians and Surgeons) was supported by grant UL1 TR000040 from the National Center for Advancing Translational Sciences, formerly grant UL1 RR024156 from the National Center for Research Resources. Role of the Funder/Sponsor: The research grant funds from the Heart and Stroke Foundation of Canada and the Canadian Institute for Health Research were used for the design and conduct of the study, as well as the collection, management, analysis, and interpretation of data. No funds were received from Laerdal Corporation for conducting this research. Other funding sources had no role in the design and conduct of the study; collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication. Disclaimer: The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. Group Information: The International Network for Simulation-Based Pediatric Innovation, Research, & Education (INSPIRE) CPR Investigators were Aaron Donoghue, MD, MSCE, Robert M. Sutton, MD, MSCE, and Dana Niles, MS, DPhil, The Children’s Hospital of Philadelphia, Philadelphia, Pennsylvania; Shannon D. Scott, PhD, University of Alberta, Calgary, Alberta, Canada; Jenny Chatfield, RN, Alberta Children’s Hospital, Calgary, Alberta, Canada; and Nnenna Chime, MD, MPH, Albert

Einstein College of Medicine and The Children’s Hospital at Montefiore, Bronx, New York. Additional Contributions: The following individuals assisted with data collection: Traci Robinson, RN, Elaine Gilfoyle, MD, MEd, Deanna Koot, RN, MN, Wendy Bissett, RN, Helen Catena, RN, and Kevin Murray, KidSIM-ASPIRE Simulation Research Program, Alberta Children’s Hospital, Calgary, Alberta, Canada; Amber Q. Youngblood, BSN, RN, and J. Lynn Zinkan, RN, MPH, Children’s of Alabama, Birmingham; Keven Cabrera, Columbia University College of Physicians and Surgeons, New York, New York; Max G. Dannecker, Lifespan Medical Simulation Center, Providence, Rhode Island; Laura McKenzie, RN, Stollery Children’s Hospital, Edmonton, Alberta, Canada; Jean Francois Trudel, RRT, Montreal Children’s Hospital, Montreal, Quebec, Canada; Linda Crelinsten, RN, MA, The Arnold and Blema Steinberg McGill Medical Simulation Centre, Montreal, Quebec, Canada; Laura Clifford, BSc, Alice Parham, Caroline Box, RGN, RSCN, BSC(Hons), Paul Sampson, BSc, MBChB, Tom Jerrom, Darshana Bhattacharjee, Annie Noble, RN, BSc, Cert Ed FHEA, Rob Kozlowski, BSc(Hons), and Sarah Sibley, Bristol Royal Hospital for Children, Bristol, England; Bonnie Mobley, BSN, Ann & Robert H. Lurie Children’s Hospital of Chicago, Chicago, Illinois; Michele Trinka, MSN, RN, CCRN, PCCN, Dayna Downing, MBA, Roxana Ploski, BS, Kay Martin, BSRC, RRT-NPS, Gail Dutcher, RRT-NPS, AE-C, Michael Williams, BSN, RN, and Marcus Brown, BS, NREMT, Children’s Medical Center of Dallas, Dallas, Texas. No compensation was received by those cited above for their contributions to the study. REFERENCES 1. Parra DA, Totapally BR, Zahn E, et al. Outcome of cardiopulmonary resuscitation in a pediatric cardiac intensive care unit. Crit Care Med. 2000;28(9): 3296-3300. 2. Slonim AD, Patel KM, Ruttimann UE, Pollack MM. Cardiopulmonary resuscitation in pediatric intensive care units. Crit Care Med. 1997;25(12):1951-1955. 3. Cheskes S, Schmicker RH, Christenson J, et al; Resuscitation Outcomes Consortium (ROC) Investigators. Perishock pause: an independent predictor of survival from out-of-hospital shockable cardiac arrest. Circulation. 2011;124(1):58-66. 4. Christenson J, Andrusiek D, Everson-Stewart S, et al; Resuscitation Outcomes Consortium Investigators. Chest compression fraction determines survival in patients with out-of-hospital ventricular fibrillation. Circulation. 2009;120(13):1241-1247. 5. Stiell IG, Brown SP, Christenson J, et al; Resuscitation Outcomes Consortium (ROC) Investigators. What is the role of chest compression depth during out-of-hospital cardiac arrest resuscitation? Crit Care Med. 2012;40(4):1192-1198. 6. Idris AH, Guffey D, Aufderheide TP, et al; Resuscitation Outcomes Consortium (ROC) Investigators. Relationship between chest compression rates and outcomes from cardiac arrest. Circulation. 2012;125(24):3004-3012.

(Reprinted) JAMA Pediatrics February 2015 Volume 169, Number 2

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

143

Research Original Investigation

Improving CPR With a CPR Feedback Device

7. Wik L, Steen PA, Bircher NG. Quality of bystander cardiopulmonary resuscitation influences outcome after prehospital cardiac arrest. Resuscitation. 1994;28(3):195-203.

competence in cardiopulmonary resuscitation through self-directed learning. J Nurses Staff Dev. 2009;25(3):1-13. doi:10.1097/NND .0b013e3181a56f92.

8. Sutton RM, Wolfe H, Nishisaki A, et al. Pushing harder, pushing faster, minimizing interruptions… but falling short of 2010 cardiopulmonary resuscitation targets during in-hospital pediatric and adolescent resuscitation. Resuscitation. 2013; 84(12):1680-1684.

21. Dracup K, Moser DK, Doering LV, Guzy PM. Comparison of cardiopulmonary resuscitation training methods for parents of infants at high risk for cardiopulmonary arrest. Ann Emerg Med. 1998; 32(2):170-177.

9. Wolfe H, Zebuhr C, Topjian AA, et al. Interdisciplinary ICU cardiac arrest debriefing improves survival outcomes. Crit Care Med. 2014; 42(7):1688-1695.

33. Kramer-Johansen J, Edelson DP, Losert H, Köhler K, Abella BS. Uniform reporting of measured quality of cardiopulmonary resuscitation (CPR). Resuscitation. 2007;74(3):406-417. 34. Nishisaki A, Nysaether J, Sutton R, et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-545.

23. Lynch B, Einspruch EL, Nichol G, Becker LB, Aufderheide TP, Idris A. Effectiveness of a 30-min CPR self-instruction program for lay responders: a controlled randomized study. Resuscitation. 2005;67(1):31-43.

35. Cheng A, Hunt EA, Donoghue A, et al; EXPRESS Investigators. Examining Pediatric Resuscitation Education Using Simulation and Scripting (EXPRESS): a multicenter, randomized-controlled trial. JAMA Pediatr. 2013;167(6):528-536.

24. Todd KH, Braslow A, Brennan RT, et al. Randomized, controlled trial of video self-instruction versus traditional CPR training. Ann Emerg Med. 1998;31(3):364-369.

36. Hamrick JT, Fisher B, Quinto KB, Foley J. Quality of external closed-chest compressions in a tertiary pediatric setting: missing the mark. Resuscitation. 2010;81(6):718-723.

12. Berden HJ, Bierens JJ, Willems FF, Hendrick JM, Pijls NH, Knape JT. Resuscitation skills of lay public after recent training. Ann Emerg Med. 1994;23(5): 1003-1008.

25. Niles D, Sutton RM, Donoghue A, et al. “Rolling Refreshers”: a novel approach to maintain CPR psychomotor skill competence. Resuscitation. 2009;80(8):909-912.

37. Badaki-Makun O, Nadel F, Donoghue A, et al. Chest compression quality over time in pediatric resuscitations. Pediatrics. 2013;131(3):e797-e804. doi:10.1542/peds.2012-1892.

13. Abella BS, Alvarado JP, Myklebust H, et al. Quality of cardiopulmonary resuscitation during in-hospital cardiac arrest. JAMA. 2005;293(3):305-310.

26. Sutton RM, Niles D, Meaney PA, et al. “Booster” training: evaluation of instructor-led bedside cardiopulmonary resuscitation skill training and automated corrective feedback to improve cardiopulmonary resuscitation compliance of pediatric basic life support providers during simulated cardiac arrest. Pediatr Crit Care Med. 2011;12(3):e116e121. doi:10.1097/PCC.0b013e3181e91271.

38. Bobrow BJ, Vadeboncoeur TF, Stolz U, et al. The influence of scenario-based training and real-time audiovisual feedback on out-of-hospital cardiopulmonary resuscitation quality and survival from out-of-hospital cardiac arrest. Ann Emerg Med. 2013;62(1):47-56.e1. doi:10.1016/j.annemergmed .2012.12.020.

10. Kaye W, Mancini ME. Retention of cardiopulmonary resuscitation skills by physicians, registered nurses, and the general public. Crit Care Med. 1986;14(7):620-622. 11. Broomfield R. A quasi-experimental research to investigate the retention of basic cardiopulmonary resuscitation skills and knowledge by qualified nurses following a course in professional development. J Adv Nurs. 1996;23(5):1016-1023.

14. Wik L, Kramer-Johansen J, Myklebust H, et al. Quality of cardiopulmonary resuscitation during out-of-hospital cardiac arrest. JAMA. 2005;293(3): 299-304. 15. Sutton RM, Niles D, Nysaether J, et al. Quantitative analysis of CPR quality during in-hospital resuscitation of older children and adolescents. Pediatrics. 2009;124(2):494-499. 16. Sutton RM, Maltese MR, Niles D, et al. Quantitative analysis of chest compression interruptions during in-hospital resuscitation of older children and adolescents. Resuscitation. 2009;80(11):1259-1263. 17. Sutton RM, Niles D, French B, et al. First quantitative analysis of cardiopulmonary resuscitation quality during in-hospital cardiac arrests of young children. Resuscitation. 2014;85(1):70-74. 18. Batcheller AM, Brennan RT, Braslow A, Urrutia A, Kaye W. Cardiopulmonary resuscitation performance of subjects over forty is better following half-hour video self-instruction compared to traditional four-hour classroom training. Resuscitation. 2000;43(2):101-110. 19. Braslow A, Brennan RT, Newman MM, Bircher NG, Batcheller AM, Kaye W. CPR training without an instructor: development and evaluation of a video self-instructional system for effective performance of cardiopulmonary resuscitation. Resuscitation. 1997;34(3):207-220. 20. Cason CL, Kardong-Edgren S, Cazzell M, Behan D, Mancini ME. Innovations in basic life support education for healthcare providers: improving

144

22. Einspruch EL, Lynch B, Aufderheide TP, Nichol G, Becker L. Retention of CPR skills learned in a traditional AHA Heartsaver course versus 30-min video self-training: a controlled randomized study. Resuscitation. 2007;74(3):476-486.

cardiac resuscitation outcomes both inside and outside the hospital: a consensus statement from the American Heart Association [published corrections appear in Circulation. 2013;128(20):e408 and 2013;128(8):e120]. Circulation. 2013;128(4):417-435.

27. Sutton RM, Niles D, Meaney PA, et al. Low-dose, high-frequency CPR training improves skill retention of in-hospital pediatric providers. Pediatrics. 2011;128 (1):e145-e151. doi:10.1542/peds.2010-2105.

39. American Heart Association. Resuscitation quality improvement. http://www.heart.org /HEARTORG/General/Resuscitation-Quality -Improvement_UCM_459324_SubHomePage.jsp. Accessed May 16, 2014.

28. Wolfe H, Sutton RM, Nadkarni V, et al. Blood pressure directed booster trainings improve intensive care unit provider retention of excellent CPR skills. Pediatr Emerg Care. In press.

40. Abella BS, Edelson DP, Kim S, et al. CPR quality improvement during in-hospital cardiac arrest using a real-time audiovisual feedback system. Resuscitation. 2007;73(1):54-61.

29. Yeung J, Meeks R, Edelson D, Gao F, Soar J, Perkins GD. The use of CPR feedback/prompt devices during training and CPR performance: a systematic review. Resuscitation. 2009;80(7): 743-751.

41. Hostler D, Everson-Stewart S, Rea TD, et al; Resuscitation Outcomes Consortium Investigators. Effect of real-time feedback during cardiopulmonary resuscitation outside hospital: prospective, cluster-randomised trial. BMJ. 2011;342:d512.

30. Kirkbright S, Finn J, Tohira H, Bremner A, Jacobs I, Celenza A. Audiovisual feedback device use by health care professionals during CPR: a systematic review and meta-analysis of randomised and non-randomised trials. Resuscitation. 2014;85(4): 460-471.

42. Kramer-Johansen J, Myklebust H, Wik L, et al. Quality of out-of-hospital cardiopulmonary resuscitation with real time automated feedback: a prospective interventional study. Resuscitation. 2006;71(3):283-292.

31. Cheng A, Auerbach M, Hunt EA, et al. Designing and conducting simulation-based research. Pediatrics. 2014;133(6):1091-1101. 32. Meaney PA, Bobrow BJ, Mancini ME, et al; CPR Quality Summit Investigators, the American Heart Association Emergency Cardiovascular Care Committee, and the Council on Cardiopulmonary, Critical Care, Perioperative and Resuscitation. Cardiopulmonary resuscitation quality: improving

43. Edelson DP, Litzinger B, Arora V, et al. Improving in-hospital cardiac arrest process and outcomes with performance debriefing. Arch Intern Med. 2008;168(10):1063-1069. 44. Sutton RM, French B, Niles DE, et al. 2010 American Heart Association recommended compression depths during pediatric in-hospital resuscitations are associated with survival. Resuscitation. 2014;85(9):1179-1184.

JAMA Pediatrics February 2015 Volume 169, Number 2 (Reprinted)

Copyright 2015 American Medical Association. All rights reserved.

Downloaded From: http://archpedi.jamanetwork.com/ by a Penn State Milton S Hershey Med Ctr User on 05/26/2015

jamapediatrics.com

Improving cardiopulmonary resuscitation with a CPR feedback device and refresher simulations (CPR CARES Study): a randomized clinical trial.

The quality of cardiopulmonary resuscitation (CPR) affects hemodynamics, survival, and neurological outcomes following pediatric cardiopulmonary arres...