Impact Nola

of Section

M. Hylton,

Doubling

PhD

Ilya Simovsky, PhD Andrew J. Li, PhD James D. Hale, BA To improve the quality of projection angiograms generated from three-dimensional magnetic resonance (MR) angiography data, the authors applied voxel shifting to create intermediate sections (“section doubling”) prior to maximum

intensity

projection.

To

date,

the authors have processed MR angiography studies with and without section doubling in 20 cases. Section doubling resulted in improved vessel contrast and delineation of continuity (especially of small vessels) in all cases.

on

Radiology

S

1992;

OME

of

185:899-902

the

most

successful

magnetic

resonance

(MR) angiography techniques are based on three-dimensional (3D) Fourier transform (FT) acquisition schemes (1,2). In this form of imaging,

phase

encoding

in-plane

is used

resolution

definition.

D larger although

tions (and imaging time) increases in direct proportion to the amount of section thickness reduction; and (b) signalto-noise ratio (S/N) decreases as the square root of the section thickness changes for constant coverage (FOV in the section direction). In MR angiography, vessel connectivity can be difficult to appreciate through viewing of individual sections. This dif-

fect the conclusion) and a section thickness T. Figure 1 shows two extreme

mum

pixel

value

for projection

high

surface.

contrast

studies,

with

vessel

MR

a

of the

in a projec-

processing

used to reduce contrast registration in imaging lesions (3). We describe

has been due to mis-

loss

of small

brain

the application of voxel shifting to MR angiography and illustrate how the effects of misreg-

istration can be substantially reduced the use of voxel shifting along the section axis prior to vessel reformatting with MIP.

as for section

encoding

yields

Materials

and

Theoretical

by

MR angiogra-

tion of these

to

olution

two

the

vessels,

it is desirable

thickness.

This

has

along

MIP

thickness.

shifting From the Radiologic Imaging Laboratory, University of California, San Francisco, South San Francisco, CA 94080 (N.M.H., J.D.H.) and Toshiba America MRI, Inc, South San Francisco, Calif (IS., A.J.L.). Received March 24, 1992; revision requested May 13; revision received July 13; accepted August 18. Supported in part by Toshiba America MRI, Inc. and by grant no. HL 39171 from the National Heart, Lung, and Blood Institute, U.S. Department of Health and Human Services. Address reprint requests to N.M.H., Department of Radiology, University of California, San Francisco, I Irving St. San Francisco, CA 94143. 2 9* indicates generalized vein and artery involvement. (. RSNA, 1992

Volume

185

Number

#{149}

3

ate

interpolation intermediate pixel direction. If, instead,

is first

section

substantial dicted

of the in-plane axes to the section

used

planes

to create

prior

improvement

in vessel

edge

a

can be predefinition,

particu-

larly for thin vessels traveling in directions oblique to the section axis. Section

ization garding ity-related

thickness

in two

affects

principal

vessel

ways

visual-

(disre-

signal loss resulting from velocdephasing): through partial

volume averaging registration. To illustrate the

and

vessel-section

effects of partial volaveraging on vessel visualization, we present examples for a vessel of lumen diameter D (for convenience, with ume

aligned

in the

second

partial

case

(Fig

lb),

because

of

volume

averaging, the apparent vessel signal V = (D/T)V + ([T DJ/T) B. The contrast C becomes C = Va/ B 1 =C0(D/T)forD < TandC=C0for D T. Figure lc shows the contrast in the 90” case for T = 2D. In MR angiography sequences, attempts are made to have the vessel signal be much larger than the background, so that C may be approximated by C (D . V)/(T . B). Because of this, partial volume averaging is more forgiving in MR angiography than it is in other forms of imaging. Figure 2 shows the dependence of C on the ratio D/T for cases where V = 7 . B and V = 1.3 B. The first relationship is typical of MR angiography and the second is typical -

the vessel considering

example

only),

along for 0

the

also affects angle 0

and section axis the two-dimensional

the maximum

contrast

between C0 900. Furtherwill drop way, so that

vessel

will vary = O and C for 0 = more, the signal intensity from maximum in a linear

the intensity profile along the vessel will be triangular or trapezoidal, depending on the angle and the values of D, T, and R. This profile results from the

intermedi-

to projection,

the vessel

contrast

between (again

of

is used values voxel

Linear

to compute along this

I

one

is equivalent

2D, with

=

threshold is reached. Vessel-section registration vessel visualization. For any

phy sequences that exploit in-flow for contrast, the section axis is usually onented parallel to the vessel. MIP reformatting is commonly performed in a projection direction that is perpendicutar to the section axis, such that the res-

section

for T

size R, not af-

pixel

does

of MR imaging of small brain lesions. Notice that for a certain conspicuity threshold (eg, C = .2), the higher the C0, the lower the D/T value at which this

Methods methods-In

in-plane

shown, in the first case (Fig la), C0 between the vessel and background (C0 = V/B - 1, where V = vessel signal and B = background signal) is preserved in the image, while the

are extracted

anatomically

the

this assumption

example

angiography

structures

and presented tion view. Voxel shifting

onto

Because

than

at O (Fig la) and at 900 (Fig lb) to the section axis. For the two-dimensional

ficulty can be overcome through use of maximum intensity projection (MIP) processing, in which a projection of the anatomic structure of vessels is created by tracing parallel rays through the 3D volume of data and selecting the maxi-

the other axis of in-plane resolution). Although typically it is desirable to orient the section axis along the principal direction of flow, in practice, some yessels of interest will be aligned along directions that are close to orthogonal to the section axis. For reliable identificanarrow

cases

for one axis of

as well

(Frequency

Angiography’

deleterious effects: (a) For any given field of view (FOV), the number of sec-

two-dimensional Index terms: Angiography #{149} Magnetic resonance (MR), experimental, 9*12142 #{149} Magnetic resonance (MR), image processing #{149}Magnetic resonance (MR), three-dimensional, 9*1214 Magnetic resonance (MR), vascular studies, 9*1214

MR

changing sel cross

registration of section D/sinO thickness T. For a vessel the order of R, angulation thick section can produce tensity corresponding to

trapezoid or to one sult is a vessel with ance.

The

effect

for D = 1.2 are generated

the oblique yeswith the section of lumen D on through a pixels of inthe peak of the

of its edges.

fragmented is illustrated

re3

12#{176}. Two cases the example in Figure 3a by sampling with an R x R pixel size (Fig 3b) and R x 3R pixel size (Fig .

R and from

The

appearin Figure

0

3c). Linear interpolation cases to generate square

fectively

double

=

is used in both pixels and ef-

the display

matrix,

Radiology

as is 899

#{149}

00

R -

90#{176}

900

r

Section Axis I

I

Iftft

III III I I #{149}

I I I I I I I I 1I I I I I I I I

IIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIIFFI

I Fl I I I I I I I I

I1111111111IIII-I-i- IIIIIIIII

I I I FI I I I I I I I I I I I I I I I I I I I I I I I I 1I I

I

II ill-..

T

IIIIIIIIII111111

F IIIIIIIIIII IIIIIIIIIIIIIIIIIIIIII

ji

‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘‘

IIIIIIIIIitiiii’ I I I I I I I I I I I I I I I I I I I I I II I I I I I I I I I I I

IJ+IIIIIIIII 11111

FFFIIIIIIIIIII F-I-FII IIIIIIIIIIIIIIIIIIIIIIIIIIIIIllI

IIIII)IIIII11IIIIIIIIIIIIIIIIIIIIIIII

D

-

FlIlll1lttllllllllllllllttJllllttllI

a.

b.

Figure 1. The lion axis, where background

to the

900

effect of partial T = the section signal) is preserved section axis and D =

volume

averaging

thickness, in a but

R varies

display terminal. A in contrast and in-

creased

vessel

width

on contrast

in-plane as

C

=

is demonstrated

pixel (D /T)

resolution, C0 for D

can

can

appreciate

that

the

effects

impractical,

but

the

exact

analogue can be achieved with postprocessing. In 1985, Leifer and Wiffley (4) pointed out that in 3D FT acquisition, if the time-domain data are multiplied by a linear phase shift as a function of time (in this case, pseudotime along the section-encoding direction), the FT will provide a sampling of the image space that is offset in position, depending on the slope of the phase shift. From a mathematical point of view, images generated after a finite 900

Radiology

#{149}

(c) Resulting

contrast

D aligned

C0

=

p

C

C

=

Co

2

0

D/T Figure

2.

The

ness

D/section

dependence

thickness

The higher

C0, the lower

conspicuity

threshold.

of contrast on the ratio T is shown for two values

the D/T value

required

vessel thickof contrast C0.

to obtain

a given

of

partial volume averaging and vesselsection registration are due to inherent sampling problems, irrespective of S/N levels and of the fidelity with which the imaging techniques represent flow. Section doubling by voxel shifting.-Registration errors can be reduced or eliminated by repeated imaging of the patient, with the patient advanced a fraction of the section thickness between procedures. Such a step is admittedly

T.

at (a) 0#{176} and (b) 90 to the secV/B - 1 (V = vessel signal and B = for the case in which the vessel is aligned at

for a vessel of thickness D = ‘,4T. Contrast C =

be

seen for the larger voxel sampling size. Although these examples have assumed a rectangular rather than sinc point spread function, no important limitations are imposed on the conclusions. The appearance of Figure 3 is a result of the registration of the vessel with respect to the voxel changing from section to section for angles 0 other than O and 900. For 0 = 900, registration is always perfect, and for 0 = O, registration effects can change C by up to a factor of 2, since there are times when the vessel will straddle two sections. The consequence is that the values of C in Figure 2 can be anywhere between the value shown in Figure 3a and half that value. We

and

‘4T.

done on a typical greater reduction apparent

=

C-

phase shift has been applied to the time domain data are as “real” or valid as those reconstructed by application of a zero phase shift (ie, no phase shift). To obtain a true representation of the object, it is necessary that the object have finite support and a bandwidth that meets the Nyquist criterion of the acquisition, meaning that the object does not extend beyond the FOV and that there are no features with edges that are sharper than a pixel width. In practice, MR images do not always meet these conditions, and the consequences are familiar:

In the first

instance,

the

object

is seen to scroll around to the opposite side of the FOV. This is called aliasing. In the latter, “ringing” artifacts associated with the sharp-edged features are seen. The shifted images are subject to the same effects. The scrolling point of the data will move relative to the image, and the ringing artifacts (not object structures) may change character sub-

stantially. It must the “correctness”

obtained

be emphasized of the image

by displacing

that estimates

the subject

by section cause the

shifting is equivalent, same changes would

the object

had

same

value

been

as the

displaced shift.

The

and

beoccur

if

by the resolution

and S/N are maintained (3), as are the section profile characteristics of the 3D FT acquisition and reconstruction processes (5). It should be noted that similar results would be achieved by using sinc interpolation (6), although implementation of this method by using zero padding and performing the Fourier transform would be computationally more intensive and require more computer memory. Experimental

methods.-A

phantom

consisting of a fan of nine oil-filled tubes of 1.2 mm cross-sectional inside diameter was used in the section doubling experiment. The tubes were December

1992

placed

at 7#{176} intervals,

spanning

an angle

allel

and was imaged transaxially FT sequence, with the section

to the OO oil-filled tube. The sehad a section thickness of 3.5 in-plane resolution of 1.1 mm, sections obtained in 9.5 minutes.

quence mm and with 32 Images

of ±28#{176}. This phantom was placed with its plane horizontal to the patient bed with a 3D plane par-

were

obtained

with

permanent-magnet cess;

imaging

Toshiba

America

system

MRI,

(Ac-

South

San

Francisco, Calif). Patients underwent imaging with a O.35-T superconducting magnet imager (MRT 35; Toshiba America MRI); a 3D gradient-echo sequence was used. Sixtyfour 1-mm partitions were acquired

a O.064-T

with mm.

an in-plane resolution of 0.8 x 0.8 A repetition time (TR) of 60 msec, echo time of 7 msec, and flip angle of 30#{176} were employed. Flow compensation

was applied

directions, slab was excited structures. Image

P,

r

to the section-select and and a presaturation

readout

to remove venous reconstruction was

performed with the Access computer and MIP reformatting was performed in a similar computer with this latter software feature. #{149}

Section-doubled

3D image

reconby using intermediate section planes. An increase in reconstruction time of roughly a factor of two is required for section doubling. MIP

structions were generated voxel shifting to compute

projections orthogonal to the section axis were constructed from non-sectiondoubled structions

and section-doubled of the same data

parison.

Because

creases

two, MIP

the

section

number

doubling

of projected

the computation also

reconfor com-

set

time

inlines

by

for a single

doubles.

Results

a. Figure

b. 3.

(a-c)

Simulation

In the phantom comparison, reconstructions using zero, half-voxel, and quarter-voxel shifting were compared.

C-

changes when vessels are angulated with respect to the section axis (vessel thickness D = 1.2 x in-plane pixel size R, angulation (-) is at 12#{176}). The images in b and c are generated with an R x R and R x 3R pixel sampling size, respectively. Linear interpolation is used to create square pixels and double the display matrix. Greater contrast reduction and increased apparent vessel width results with the larger voxel sampling size.

a.

b.

4. Anteroposterior MIP projections created with (a) no voxel shifting, were generated from 3D data with the sections oriented perpendicular to the ness was 3.5 mm. Periodicity in size and contrast along the length of the tubes

the first stage

of voxel

angulation

Volume

185

shifting is apparent

Number

#{149}

3

and

is less appreciable

constructions

Variations described Figure of the

from

each

of the three

re-

are shown in Figure 4. in contrast and vessel width

in the simulated

example

3 are again seen along oil-filled tubes in Figure

of

the length 4. The

c.

Figure

changing

MIPs created

of contrast

between

the half-

and

(b) half-voxel shifting, and (c) quarter-voxel shifting. Projections phantom and parallel to the 0#{176} oil-filled tube. The section thickdecreases from a to c. The improvement is more dramatic with

quarter-pixel-shifted

projections.

The change

in periodicity

with

in a.

Radiology

901

#{149}

periodicity

of the artifact

increasing

angulation

with

with

increases respect

to

the 0#{176} (horizontal) axis. A reduction in the stairstep pattern can be appreciated between the unshifted and half-voxelshifted images. A less dramatic but noticeable reduction is evident between the images shifted by one-half and onequarter pixel. Anteroposterior projections created for non-section-doubled and sectiondoubled MR angiography data are compared in Figure 5 for a patient with an arteriovenous malformation. The improvement in vessel definition and contrast is especially noticeable in vessels oriented oblique to the section axis. In 3D time-of-flight MR angiography studies such as these, the persistence signal into the imaging volume

of flow

(in this case, arterial flow entering from the bottom of the slab) is limited by flow velocity, TR, and flip angle. A decrease in flow signal is seen with increasing distance from the entry. Section doubling is particularly beneficial in the more distat branches, allowing better visualization of these low-intensity structures through maximization of their contrast and improvement of their continuity. To date, MR angiographic studies in 20 patients have been processed with and without section doubling improved vessel delineation with section doubling resulted in all cases. Because a sampling

artifact

can

usually

fled by the periodicity trast

changes,

be identi-

of size and con-

no incident

of a false-

positive finding being corrected has occurred in our limited patient data. The improvement in vessel edge definition has allowed a more accurate assessment of the degree of vessel narrowing in the left carotid arteries of two patients. Discussion

902

doubling

Radiology

#{149}

time.

the

number

of sections,

If image

matrix

size

is limited,

gained by the

by voxel geometric

imaging

shifting will relationship

of vascular

features.

a.

a

reduced FOV might also result. Although resolution does not change with voxel shifting, an improvement in yessel-voxel registration can result in a new intensity maximum projected by the MIP process. This detection improvement is appreciated mainly at vessel edges and in imaging of small vessels, of a size on the order of the voxel dimension. Improvement is greatest for smalldiameter vessels and those with lower intrinsic contrast. Registration artifacts that occur for angulated vessels will be more severe for larger voxel dimensions. Clearly, the extent of the benefit be affected created in

b. Figure 5. section-doubled

1.

mation doubling improvement

appreciated at angles

4.

5.

W.

Three-dimensional Magn Reson

giography.

phase contrast anMed 1989; 9:139-

6.

with

and (b) nonsame data set

an artenovenous

demonstrate on projection

for views

U

Alfidi RJ, Masaryk TJ, Haacke EM, et al. MR angiography of peripheral, carotid and coronary arteries. AJR 1987; 149:1097-1109. Dumoulin CL, Souza SP, Walker MF, Wagle

(a) Section-doubled images of the

for a patient

References

2.

The technique of voxel shifting for creating intermediate sections in a 3D time-of-flight MR angiography study, thus

provides a substantial improvement in the visualization of small vessels in MIP projections. Artifactual variations in contrast and vessel width caused by registration changes as vessels cross sections obliquely are substantially reduced. Further improvement with smaller shift distances is limited, however, by the finite voxel sampling size, which this process does not address. Voxel shifting also requires increased data reconstruction and processing times, as well as increased demand for data storage. Given these considerations, we have found voxel shifting by one-half pixel to be most effective. The geometric relationships inherent in MR angiography make the section direction the most desirable for doubling. In-plane voxel shifting would also provide some benefit but will require the additional factor increase in imaging

orthogonal in vessel

the

impact angiograms

malforof section created

to the section delineation

axis. The is best

for the thinner vessels oriented oblique to the section axis.

Leifer MC, Wilfley BP. NMR volume imaging with half slice offsets (abstr). In: Book of abstracts: Society of Magnetic Resonance in Medicine 1985. Berkeley, Calif: Society of Magnetic Resonance in Medicine, 1985; 1013. Carlson JC, Crooks LE, Ortendahl DA, Kramer DM, Kaufman L. Signal-to-noise ratio and section thickness in two-dimensional versus three-dimensional Fourier transform MR imaging. Radiology 1988; 166: 266-270. Bracewell RN. The Fourier transform and its applications. 2nd ed. New York: McGraw-Hill, 1978; 194.

149. 3.

Kramer D, Li A, Simovsky I, Hawryszko C, Hale J, Kaufman L. Applications of voxel shifting in magnetic resonance imaging. Invest Radiol 1990; 25:1305-1310.

December

1992

Impact of section doubling on MR angiography.

To improve the quality of projection angiograms generated from three-dimensional magnetic resonance (MR) angiography data, the authors applied voxel s...
791KB Sizes 0 Downloads 0 Views