Hourly Awakening vs Continuous Contact Lens Sensor Measurements of 24-Hour Intraocular Pressure Effect on Sleep Macrostructure and Intraocular Pressure Rhythm Florent Aptel, MD, PhD; Renaud Tamisier, MD, PhD; Jean-Louis Pépin, MD, PhD; Benjamin Mottet, MSc; Ralitsa Hubanova, MD; Jean-Paul Romanet, MD; Christophe Chiquet, MD, PhD

IMPORTANCE All studies of 24-hour intraocular pressure (IOP) rhythm conducted to date

Supplemental content at jamaophthalmology.com

have used repeated IOP measurements requiring nocturnal awakenings, potentially disturbing sleep macrostructure. OBJECTIVE To evaluate the effects on sleep architecture and IOP rhythm of hourly awakening vs a contact lens sensor (CLS) to continuously monitor IOP without awakening. DESIGN, SETTING, AND PARTICIPANTS Cross-sectional study at a referral center of chronobiology among 12 young healthy volunteers, with a mean (SD) age of 22.3 (2.3) years. INTERVENTIONS Volunteers underwent two 24-hour IOP measurement sessions during a

2-month period. The eye order and session order were randomized. During one session, the IOP of the first eye was continuously monitored using a CLS, and the IOP of the fellow eye was measured hourly using a portable noncontact tonometer (session with nocturnal hourly awakening). During the other session, the IOP of the first eye was continuously monitored using a CLS, and the IOP of the fellow eye was not measured (session without nocturnal awakening). Overnight polysomnography was performed during the 2 sessions. MAIN OUTCOMES AND MEASURES A nonlinear least squares, dual-harmonic regression analysis was used to model the 24-hour IOP rhythm from the CLS data. Comparisons of acrophase, bathyphase, amplitude, and the midline estimating statistic of rhythm were used to evaluate the effect of hourly awakening on IOP rhythm. To evaluate the effects of hourly awakening on sleep architecture, comparisons of sleep structure were used, including total sleep period, rapid eye movement, wake after sleep onset, absolute and relative total sleep time, and non–rapid eye movement sleep (N1, N2, and N3). RESULTS A 24-hour IOP rhythm was found in all individuals for the sessions with and without awakening (P < .05). Hourly awakening for nocturnal IOP measurements increased wake after sleep onset (P = .04) but did not seem to change total sleep time, total sleep period, sleep efficiency, or slow-wave and rapid eye movement sleep stage duration (P > .30). Hourly awakening during noncontact tonometer IOP measurements did not seem to alter the mean variables of the 24-hour IOP pattern evaluated using CLS, including signal, maximum signal, minimum signal, acrophase, and bathyphase (P > .15). CONCLUSION AND RELEVANCE The 24-hour IOP rhythms seem to be unaffected by hourly nocturnal awakening for IOP measurements in young healthy individuals.

JAMA Ophthalmol. 2014;132(10):1232-1238. doi:10.1001/jamaophthalmol.2014.1761 Published online June 26, 2014. 1232

Author Affiliations: Department of Ophthalmology, University Hospital, Centre Hospitalier Universitaire, Université Grenoble Alpes, Grenoble, France (Aptel, Mottet, Hubanova, Romanet, Chiquet); Laboratory of Hypoxia and Physiopathology, Université Grenoble Alpes, Grenoble, France (Tamisier, Pépin). Corresponding Author: Florent Aptel, MD, PhD, University Hospital, Centre Hospitalier Universitaire, Université Grenoble Alpes, 38043 Grenoble CEDEX 09, France ([email protected]). jamaophthalmology.com

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

Hourly vs Continuous 24-Hour IOP Measurements

T

o date, all intraocular pressure (IOP) measurement methods (whether portable or nonportable) require individuals to be awakened for nocturnal measurements. To our knowledge, all studies of 24-hour IOP rhythm have used repeated IOP measurements requiring nocturnal awakenings, potentially disturbing sleep macrostructure. This process may have biased previous descriptions of physiological IOP rhythm.1-7 The effects of awakening for IOP measurements on sleep macrostructure and IOP rhythm are unknown. A contact lens sensor (CLS) (Triggerfish; SENSIMED AG) was recently developed to continuously monitor IOP noninvasively during 24 hours in an ambulatory setting. This innovative method is based on the assumption that a correlation exists between IOP and corneal curvature.8 Animal studies9,10 have shown that an IOP variation of 1 mm Hg produces a change in the central corneal curvature radius of approximately 3 μm. The device has been approved by European regulatory authorities (class IIa device CE-mark) since 2009 and has been evaluated in healthy individuals and in patients with glaucoma in a few studies.11-14 It has recently been demonstrated that this new device represents an accurate and reproducible method to characterize the 24-hour nyctohemeral IOP rhythm in young healthy individuals.15 In the present study, we evaluated the effects of hourly awakening for IOP measurements on sleep using polysomnography and the effects of hourly awakening on IOP rhythm using a CLS that allowed us to continuously monitor IOP during 24 hours. The underlying issue was the validity of the data provided by the previous studies1-7 conducted with awakening after having found that IOP follows a nyctohemeral rhythm in healthy humans and in patients with glaucoma.

Methods Study Population The study protocol has been recently described in detail.15 This prospective investigation was conducted in the Laboratory of Ocular Chronobiology, Université Grenoble Alpes, Grenoble, France, in accord with the tenets of the Declaration of Helsinki and was approved by the Comité de Protection des Personnes Sud Est V. All participating individuals provided both verbal and written informed consent. Participants were paid for their time and effort. Twenty-four eyes of 12 healthy volunteers were studied in two 24-hour IOP measurement sessions during a 2-month period. The inclusion criteria were individuals free of sleep disorders, with regular lifestyle habits, with usual sleep duration lasting approximately 8 hours, and without endocrine illness or ocular disease (spherical equivalent between −1 and +1 diopter). The exclusion criteria were shift workers, tobacco smokers, individuals with any medical treatment, and persons who had taken a transmeridian flight less than 2 months before the beginning of the study. At the study inclusion visit, all individuals underwent a complete ophthalmic examination, including refraction, jamaophthalmology.com

Original Investigation Research

gonioscopy, biomicroscopy, fundus examination, Goldmann applanation tonometry, and ultrasonographic pachymetry (Pocket II; Quantel Medical). All individuals also filled out a general health questionnaire and underwent a complete physical examination. The eye order and session order (session with nocturnal hourly awakening and session without nocturnal awakening) were randomized.

IOP Measurements The CLS is an oxygen-permeable soft contact lens whose key elements are 2 sensing resistive-strain gauges that are capable of recording circumferential changes in the area of the corneoscleral junction. The contact lens (with a diameter of 14.1 mm and thicknesses of 585 μm at the center and 260 μm at the edge) is available in 3 different base curves (steep, medium, and flat) with an 8.4-mm, 8.7-mm, and 9-mm curvature radius, respectively. Ten data points per second are acquired during a 30-second measurement period, repeated every 5 minutes. The output of the sensor is expressed in electric arbitrary units (eqVm). The noncontact tonometer (NCT) used herein (Pulsair intelliPuff; Keeler) measures the IOP in patients in the sitting or supine position. A mean of 3 readings was recorded hourly during a 24-hour session (every hour from 9 AM to 9 AM). It was previously found that the use of this NCT agrees well with Goldmann applanation tonometry in normotensive and hypertensive individuals in the sitting position.16

Polysomnography Continuous recordings were obtained with the electrode positions C3/A2-C4/A1-Cz/01 of the International 10-20 System of Electrode Placement measuring eye movements, a chin electromyogram, and an electrocardiogram with a modified V2 lead.17 Sleep was scored manually according to standard criteria.17 Airflow was measured by nasal pressure associated with the sum of buccal and nasal thermistor signals. Respiratory effort was monitored with abdominal and thoracic bands. Pulse transit time, an additional indicator of respiratory effort, was recorded concurrently. Oxygen saturation was measured using a pulse oximeter (Biox-Ohmeda 3700; Ohmeda).

Experimental Sessions Participants maintained a self-selected constant sleep-wake schedule (sleep onset between 10 PM and 12 midnight and wake-up between 7 AM and 8 AM) 2 weeks before and during the study, checked by sleep-wake diaries and by ambulatory actigraphy monitoring using a wrist accelerometer (Actiwatch; CamNtech). During the experimental sessions, they were requested not to drink alcohol-containing or caffeinecontaining beverages. Although the participants were housed in a hospital, they were allowed to have free activities between the NCT hourly IOP measurements. At each session, participants were asked to go to bed after the IOP measurement at 10 PM and were asked to get up after the IOP measurement at 8 AM. The protocol is summarized in eTable 1 in the Supplement. JAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

1233

Research Original Investigation

Hourly vs Continuous 24-Hour IOP Measurements

Table 1. Sleep Duration, Sleep Quality Variables, and Sleep Architecture Mean (SD) [95% CI] {% of Total Sleep Time} Variable

Hourly Awakening

No Awakening

P Value

Total sleep time, min

381.5 (74.5) [339.3-423.7]

365.3 (48.6) [337.8-392.8]

Total sleep period, min

481.0 (65.9) [443.7-518.3]

459.9 (29.4) [443.2-476.5]

.31

79.1 (14.3) [70.8-87.3]

79.4 (11.5) [72.8-86.1]

.92

151.4 (56.7) [119.3-183.5]

110.1 (36.1) [92.2-128.0]

.04a

Sleep efficiency, %

.53

Sleep stage, min W N1

36.8 (8.3) [32.1-41.5] {10.0%}

40.7 (20.1) [29.3-52.1] {11.1%}

N2

195.3 (55.4) [163.9-226.7] {50.6%}

179.0 (35.4) [158.9-199] {49.3%}

N3

66.7 (26.0) [52.0-81.4] {17.3%}

66.1 (17.1) [56.7-76.1] {18.1%}

R

83.5 (19.6) [72.4-94.6] {22.0%}

79.6 (27.8) [63.8-95.3] {21.5%}

Cycles per night, No.

5.1 (0.6) [4.8-5.4]

5.5 (0.5) [5.2-5.8]

.09

Microarousal Index

7.7 (4.4) [5.2-10.2]

6.4 (3.2) [4.6-8.2]

.42

Abbreviations: N, non–rapid eye movement; R, rapid eye movement; W, wake after sleep onset. a

Statistical Analysis From raw IOP data during 24 hours, a nonlinear least squares, dual-harmonic regression analysis was used to model the 24hour IOP rhythm with the following equation:

IOPt = M + A1 cos

2π 2π t + φ2 , t + φ1 + A2 cos 2 τ τ

where A1 is the amplitude of the fundamental cosine fit, A2 is the amplitude of the first harmonic cosine fit, Φ1 is the acrophase of the fundamental cosine fit, Φ2 is the acrophase of the first harmonic cosine fit, τ is the endogenous circadian period (set at 24 hours because of entrained conditions), M is the midline estimating statistic of rhythm, and t is the time. Variables obtained from the modeling of each IOP curve included acrophase (time of the highest IOP value in a 24-hour cycle), bathyphase (time of the lowest IOP value in a 24-hour cycle), the midline estimating statistic of rhythm (mean IOP values in a 24-hour cycle), and unbiased estimates and confidence limits of amplitude (half the difference between the highest and lowest IOP values in a 24-hour cycle).18,19 The distribution of the acrophase and bathyphase over time was analyzed using the Rayleigh test of uniformity and the Watson-Williams test for homogeneity of means.20 The skewness-kurtosis normality test was used to assess the normality of the measurements. The paired t test and analysis of variance were used to compare the means and percentages. McNemar χ2 test was used for the analysis of dichotomous variables. Within-patient across-session relationships were obtained by correlating parallel recording time points using each of the 5-minute recordings with Pearson product moment correlations. To take into account a potential time lag or delay between the 2 measurement series, we also studied the within-patient acrosssession relationships using the cross-correlation function. The cross-correlation function of 2 time series is the product moment correlation as a function of lag (or time offset) between the series. The cross-correlation function identifies the lag that leads to the stronger correlation between the 2 time series. The 2-way random mean agreement intraclass correlation coefficient from the convention by Shrout and Fleiss21 was used to assess the IOP agreement at the 2 sessions and at each time point during the nocturnal period (10 PM to 8 AM) (eg, IOP 1234

By Wilcoxon signed rank test. All other P values are by paired t test.

at 11 PM compared for 2 sessions). The following interpretation scheme for the intraclass correlation coefficient has been described: less than 0.40 represents poor agreement beyond chance, 0.40 to 0.75 represents fair to good agreement beyond chance, and greater than 0.75 represents excellent agreement beyond chance.22 Data analyses were performed using statistical software. These included SPSS (version 17.0; SPSS Inc) and R software (version 2.14; R Foundation for Statistical Computing).23

Results Participant Characteristics Twelve young healthy volunteers (24 eyes) of white race/ ethnicity were included in the study (8 women and 4 men), with a mean (SD) age of 22.3 (2.3) years and a mean (SD) body mass index (calculated as weight in kilograms divided by height in meters squared) of 20.8 (2.0). All participants were easily fitted with the CLS (9 medium lenses, 2 steep lenses, and 1 flat lens depending on previous keratometry) on one randomly selected eye (6 left eyes and 6 right eyes). Six participants had the session with nocturnal hourly awakening first, and 6 participants had the session without nocturnal awakening first. At study inclusion, the mean (SD) IOP values using Goldmann applanation tonometry were 13.8 (2.1) mm Hg in the right eye and 13.7 (1.9) mm Hg in the left eye (P = .86). The 24-hour and nocturnal NCT IOP measurements of the contralateral eye at session 1 are summarized in eTable 2 in the Supplement.

Effects of Hourly Awakening for IOP Measurements on Sleep Table 1 lists the Microarousal Index,17 the number of cycles per night, rapid eye movement (REM) and non-REM (N1, N2, and N3) sleep, sleep efficiency (total sleep time divided by total sleep period), total sleep period (period measured from sleep onset to final awakening), sleep architecture with absolute and percentage wake after sleep onset time, and total sleep time (total of REM and non-REM sleep in a period). Hourly awakening for nocturnal IOP measurements increased wake after sleep onset (P = .04) but did not seem to alter sleep macrostructure, with no differences in total sleep time (P = .53), sleep efficiency (P = .92), the proportions of slow-wave and REM sleep stages in

JAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

jamaophthalmology.com

Hourly vs Continuous 24-Hour IOP Measurements

Original Investigation Research

Figure 1. Two Polysomnography Sessions in an Individual

Sleep Stage

No awakening W R N1 N2 N3

Sleep Stage

Hourly awakening

Sleep period duration is increased, but total sleep time and N1, N2, N3, and rapid eye movement sleep times are not significantly changed. Arrowheads show the hourly awakening times for intraocular pressure measurements. N indicates non–rapid eye movement; R, rapid eye movement; and W, wake after sleep onset.

W R N1 N2 N3

11 PM

6 AM

Time

Table 2. Characteristics of 24-Hour Contact Lens Sensor Signal Measurement Mean (SD) [95% CI] Variable

Hourly Awakening

P Valuea

No Awakening

Signal, eqVm Mean

10.1 (12.1) [3.2 to 16.9]

3.4 (10.9) [−2.8 to 9.6]

.18

Maximum

40.8 (33.2) [22.0 to 59.6]

28.0 (10.2) [22.2 to 33.8]

.23

Minimum

−10.5 (7.1) [−14.5 to −6.5]

−20.7 (21.1) [−32.7 to −8.7]

.17

51.3 (34.9) [31.5 to 71.1]

48.1 (23.3) [34.9 to 61.3]

.59

Amplitude Acrophase, h

2:47 AM (33 min) [2:30 AM to 3:01 AM]

3:51 AM (43 min) [3:26 AM to 4:12 AM]

.28

Bathyphase, h

2:48 PM (50 min) [2:21 AM to 3:12 AM]

2:30 PM (34 min) [2:12 AM to 2:48 AM]

.66

relation to total sleep time (P = .38), and the number of cycles per night (P = .09). Sleep fragmentation also remained unchanged, with no difference in the Microarousal Index (P = .42). Figure 1 shows the polysomnographs from the 2 sessions for 1 participant. It illustrates how participants seemed to compensate for the multiple awakenings imposed and maintained unchanged total sleep time, sleep efficiency, and the duration of each sleep stage (N1, N2, N3, REM, and absolute or percentage relative to total sleep time).

Effects of Hourly Awakening for IOP Measurements on 24-Hour IOP Rhythm The CLS signal showed a significant nyctohemeral rhythm for all 12 participants at sessions 1 and 2 (P < .05). The 24-hour minimum, maximum, and mean signal values and amplitude, acrophase, and bathyphase characteristics of the population are summarized in Table 2. The amplitude of the 95% CIs and residuals (root-mean-square errors) of the nonlinear least squares, dual-harmonic regression analyses of the 24 sessions are summarized in eTable 3 in the Supplement. Nocturnal (10 PM to 8 AM) minimum, maximum, and mean signal values and amplitude of the population are summarized in eTable 4 in the Supplement. In all participants and for both sessions, the mean nocturnal IOP was higher than the mean diurnal IOP: the mean (SD) values at session 1 were 22.1 (15.6) vs −0.4 (8.9) (P = .004), and the mean (SD) values at session 2 were 12.9 (15.5) vs −5.1 (6.7) (P = .008). jamaophthalmology.com

Abbreviation: eqVm, electric arbitrary units. a

By paired t test.

Hourly awakening during NCT IOP measurements did not seem to alter the mean variables of the 24-hour IOP pattern evaluated using CLS, including signal, maximum signal, minimum signal, acrophase, and bathyphase (P > .15). Correlations and cross-correlations of the 24-hour CLS measurements at sessions 1 and 2 are summarized in eTable 5 in the Supplement. Six of 12 participants exhibited a strong correlation (r > 0.7), and 6 of 12 participants exhibited a strong cross-correlation (r > 0.7), with a mean (SD) time lag of 11.7 (9.0) minutes between the 2 sessions. The IOP agreement at the 2 sessions and at each time point during the nocturnal period (10 PM to 8 AM) is summarized in eTable 6 in the Supplement and shown in Figure 2. None of 11 nocturnal hourly IOP values had intraclass correlation coefficients with P < .05, demonstrating poor agreement between the 2 nocturnal period IOP measurements. From Table 2, Figure 2, and Figure 3, it can be seen that the variability of IOP measurements was much greater during the session with nocturnal hourly awakening.

Discussion All previous studies1-7 of the 24-hour IOP conducted to date have found a significant nyctohemeral rhythm in healthy participants and in patients with glaucoma, including those in a constant sitting or supine position, with higher IOP levels usuJAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

1235

Research Original Investigation

Hourly vs Continuous 24-Hour IOP Measurements

Figure 2. Examples of 24-Hour Raw and Modeled Contact Lens Sensor Signal Curves in the Same Individual as in Figure 1

Normalized Change Signal, eqVm

25 20

Session 1 Raw signal Rhythm

15 10 5 0 –5

ΔΦ

Normalized Change Signal, eqVm

–10 25 20 15

Session 2 Raw signal Rhythm

10 5 0 –5 –10 9 PM

10 PM

11 PM

Midnight

1 AM

2 AM

3 AM

4 AM

5 AM

6 AM

7 AM

8 AM

9 AM

Time

The session without nocturnal awakening (session 2) is at the top, and the session with nocturnal hourly awakening (session 1) is at the bottom. eqVm Indicates electric arbitrary units; ΔΦ, the lag between session 1 and session 2 acrophases.

Figure 3. Reproducibility of Nocturnal Mean Contact Lens Sensor (CLS) Values 40

CLS session 1 SD session 1 CLS session 2 SD session 2

Change Signal, eqVm

30

20

10

0

–10

–20 10 PM

Midnight

2 AM

4 AM

6 AM

9 AM

Time

Values were normalized to the first 10 PM measurement among all individuals at the session with nocturnal hourly awakening (session 1) and at the session without nocturnal awakening (session 2). At each session, individuals were asked to go to bed after the intraocular pressure measurement at 10 PM and were asked to get up after the intraocular pressure measurement at 8 AM. eqVm Indicates electric arbitrary units.

ally occurring during the nocturnal period and lower IOP levels usually occurring during the diurnal period. However, when studying the circadian rhythm of a biological variable, a cru1236

cial point is to avoid using a method that could disrupt the normal sleep-wake cycle. Studies24-29 have demonstrated that many biological endogenous circadian rhythms may be perturbed by forced interruption of the sleep-wake cycle, including melatonin secretion, body temperature, plasma cortisol level, blood pressure, heart rate, and others. Until recently, all IOP measurement methods (portable or nonportable) required individuals to be awakened for nocturnal measurements. Therefore, we speculate that the results of several studies conducted with hourly or punctual awakening may have been biased or falsely induced by stress-related artifacts or sleep-wake cycle disorganization. The present study identified few sleep changes related to hourly nocturnal awakening for IOP measurements. Participants seemed to compensate for the multiple awakenings imposed by increasing the total sleep period, so that total sleep time and the proportions of slowwave and REM sleep stages in relation to total sleep time seemed unchanged. Sleep quality or fragmentation was also not altered by nocturnal IOP measurements, with sleep efficiency and the Microarousal Index remaining unchanged. We also found comparable 24-hour IOP rhythms and nocturnal IOP patterns evaluated using the CLS with and without hourly nocturnal awakening for IOP measurements. Nevertheless, an increase in CLS signal value variability throughout the night was observed when individuals were awakened. Few studies30-33 have evaluated the effects of hourly forced awakening on sleep architecture in young healthy individuals. They have evaluated the effects of a few forced awakenings encountered in conditions such as Ramadan fasting or awakenings related to noise disturbances in young healthy individuals, and the results are balanced. In agreement with the

JAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

jamaophthalmology.com

Hourly vs Continuous 24-Hour IOP Measurements

Original Investigation Research

results of our study, these studies found no major disturbances in the different sleep stages, particularly in the number of REM cycles and the proportions of slow-wave and nonREM sleep stages in relation to total sleep time. Forced awakening may have different effects in older individuals according to studies33-37 demonstrating that sleep architecture differs in the elderly, who are also more susceptible to environmental disturbances. No studies to date have evaluated the effects of forced experimental awakening on sleep architecture in the elderly. In older individuals with diseases, such as sleep apnea, that lead to multiple awakenings throughout the night, several sleep architecture disturbances have been demonstrated, including sleep fragmentation, an increase in microarousal, and a decrease in sleep efficiency, slow-wave sleep, and REM sleep.38 Further studies should evaluate the effects of hourly awakening for IOP measurements on sleep architecture and IOP rhythm in patients with glaucoma (most of whom are older than 50 years) in a similar manner as done in the present study. In some sleep disorders such as insomnia and restless legs syndrome that are associated with difficulty in initiating or maintaining sleep, the effect of hourly awakening might be greater. This underscores the value of new IOP measurement methods because insomnia and restless legs syndrome affect 20% and 8% of the general population, respectively.39,40 We found good correlation and cross-correlation of the 24-hour CLS measurements at sessions 1 and 2. Previous studies 7,22,41 conducted with a noncontact or Goldmann applanation tonometer during a range of 12 to 24 hours have demonstrated poor intraindividual reproducibility of IOP values and IOP rhythm variables. In contrast, previous investigations in patients with glaucoma using the same CLS have found good correlation between two 24-hour measurement sessions at a 1-week interval conducted without nocturnal awakening.13 In that study, the greater frequency of

ARTICLE INFORMATION Submitted for Publication: December 21, 2013; final revision received March 15, 2014; accepted March 22, 2014. Published Online: June 26, 2014. doi:10.1001/jamaophthalmol.2014.1761.

Conclusions We found that sleep organization and 24-hour IOP rhythm seem to be unaffected by hourly nocturnal awakening for IOP measurements in young healthy individuals. This finding supports the results of previous studies1-7 conducted with IOP measurement methods requiring participants to be awakened and demonstrating a nyctohemeral rhythm of IOP in humans.

collection, management, analysis, and interpretation of the data; preparation, review, or approval of the manuscript; and decision to submit the manuscript for publication. REFERENCES

Author Contributions: Dr Aptel and Mr Mottet had full access to all the data in the study and take responsibility for the integrity of the data and the accuracy of the data analysis. Study concept and design: Aptel, Tamisier, Pépin, Romanet, Chiquet. Acquisition, analysis, or interpretation of data: Aptel, Tamisier, Mottet, Hubanova, Romanet, Chiquet. Drafting of the manuscript: Aptel, Pépin, Hubanova, Romanet, Chiquet. Critical revision of the manuscript for important intellectual content: All authors. Conflict of Interest Disclosures: None reported. Funding/Support: This study was supported by the Association de Recherche et de Formation en Ophtalmologie and by the Fondation de l’Avenir. Role of the Sponsor: The funding sources had no role in the design and conduct of the study; jamaophthalmology.com

data acquisition with the CLS compared with noncontact or Goldmann applanation tonometry could explain the higher correlation between the 2 measurement sessions (24 time points with hourly noncontact or Goldmann applanation tonometry vs 288 time points with the CLS). The good correlation between the 2 sessions in our study also reinforces the finding that hourly awakening for IOP measurements does not seem to alter the 24-hour IOP pattern. We found an increase in CLS signal value variability throughout the night when participants were awakened. Because sleep architecture is not disorganized by awakening, a technical reason may explain this finding. The CLS provides 10 data acquisitions per second during a 30-second measurement period, repeated every 5 minutes during 24 hours. The device does not indicate when acquisitions are made. Therefore, it is likely that some NCT IOP measurements of the fellow eye were obtained during CLS acquisitions of the first eye. Stress related to awakening, lid opening or contraction, or changes in body position could cause a sudden and unsustainable increase in the CLS signal that leads to greater variability throughout the night. Another limitation of our study may be the few participants, increasing the risk of type I error, which can be explained by the complexity of the experiments (two 24-hour sessions in a sleep laboratory with overnight polysomnography for each individual).

1. Buguet A, Py P, Romanet JP. 24-Hour (nyctohemeral) and sleep-related variations of intraocular pressure in healthy white individuals. Am J Ophthalmol. 1994;117(3):342-347. 2. Liu JH, Kripke DF, Hoffman RE, et al. Nocturnal elevation of intraocular pressure in young adults. Invest Ophthalmol Vis Sci. 1998;39(13):2707-2712. 3. Noël C, Kabo AM, Romanet JP, Montmayeur A, Buguet A. Twenty-four–hour time course of intraocular pressure in healthy and glaucomatous Africans: relation to sleep patterns. Ophthalmology. 2001;108(1):139-144.

detectable in the sitting position. Invest Ophthalmol Vis Sci. 2003;44(10):4439-4442. 6. Pépin JL, Chiquet C, Tamisier R, Lévy P, Almanjoumi A, Romanet JP. Frequent loss of nyctohemeral rhythm of intraocular pressure restored by nCPAP treatment in patients with severe apnea. Arch Ophthalmol. 2010;128(10):12571263. 7. Mottet B, Chiquet C, Aptel F, et al. 24-Hour intraocular pressure of young healthy humans in supine position: rhythm and reproducibility. Invest Ophthalmol Vis Sci. 2012;53(13):8186-8191. 8. Leonardi M, Leuenberger P, Bertrand D, Bertsch A, Renaud P. First steps toward noninvasive intraocular pressure monitoring with a sensing contact lens. Invest Ophthalmol Vis Sci. 2004;45 (9):3113-3117.

4. Liu JH, Zhang X, Kripke DF, Weinreb RN. Twenty-four–hour intraocular pressure pattern associated with early glaucomatous changes. Invest Ophthalmol Vis Sci. 2003;44(4):1586-1590.

9. Leonardi M, Pitchon EM, Bertsch A, Renaud P, Mermoud A. Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes. Acta Ophthalmol. 2009;87(4): 433-437.

5. Liu JH, Bouligny RP, Kripke DF, Weinreb RN. Nocturnal elevation of intraocular pressure is

10. Hjortdal JO, Jensen PK. In vitro measurement of corneal strain, thickness, and curvature using

JAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

1237

Research Original Investigation

Hourly vs Continuous 24-Hour IOP Measurements

digital image processing. Acta Ophthalmol Scand. 1995;73(1):5-11.

20. Zar JH. Biostatistical Analysis. Upper Saddle River, NJ: Prentice Hall; 1999.

11. Mansouri K, Shaarawy T. Continuous intraocular pressure monitoring with a wireless ocular telemetry sensor: initial clinical experience in patients with open angle glaucoma. Br J Ophthalmol. 2011;95(5):627-629.

21. Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86 (2):420-428.

32. Foret J, Touron N, Clodoré M, Benoit O, Bouard G. Modifications of sleep structure by brief forced awakenings at different times of the night. Electroencephalogr Clin Neurophysiol. 1990;75(3): 141-147.

22. Realini T, Weinreb RN, Wisniewski S. Short-term repeatability of diurnal intraocular pressure patterns in glaucomatous individuals. Ophthalmology. 2011;118(1):47-51.

33. Miyasita A, Fukuda K, Inugami M. Effects of sleep interruption on REM-NREM cycle in nocturnal human sleep. Electroencephalogr Clin Neurophysiol. 1989;73(2):107-116.

23. R Project for Statistical Computing. http://www.r-project.org/. Accessed May 8, 2014.

34. Ohayon MM, Carskadon MA, Guilleminault C, Vitiello MV. Meta-analysis of quantitative sleep parameters from childhood to old age in healthy individuals: developing normative sleep values across the human lifespan. Sleep. 2004;27(7):12551273.

12. Mansouri K, Liu JH, Weinreb RN, Tafreshi A, Medeiros FA. Analysis of continuous 24-hour intraocular pressure patterns in glaucoma. Invest Ophthalmol Vis Sci. 2012;53(13):8050-8056. 13. Mansouri K, Medeiros FA, Tafreshi A, Weinreb RN. Continuous 24-hour monitoring of intraocular pressure patterns with a contact lens sensor: safety, tolerability, and reproducibility in patients with glaucoma. Arch Ophthalmol. 2012;130(12):1534-1539. 14. Pajic B, Pajic-Eggspuchler B, Haefliger I. Continuous IOP fluctuation recording in normal tension glaucoma patients. Curr Eye Res. 2011;36 (12):1129-1138. 15. Mottet B, Aptel F, Romanet JP, Hubanova R, Pépin JL, Chiquet C. 24-Hour intraocular pressure rhythm in young healthy subjects evaluated with continuous monitoring using a contact lens sensor. JAMA Ophthalmol. 2013;131(12):1507-1516. 16. Hubanova R, Aptel F, Zhou T, Arnol N, Romanet JP, Chiquet C. Comparison of intraocular pressure measurements with the Reichert PT100, the Keeler Pulsair intelliPuff portable non-contact tonometers and Goldmann applanation tonometry [published online November 14, 2013]. J Glaucoma. 2013. 17. Berry RB, Budhiraja R, Gottlieb DJ, et al; American Academy of Sleep Medicine, Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. J Clin Sleep Med. 2012;8(5): 597-619. 18. Brown EN, Czeisler CA. The statistical analysis of circadian phase and amplitude in constant-routine core-temperature data. J Biol Rhythms. 1992;7(3):177-202. 19. Gronfier C, Wright KP Jr, Kronauer RE, Czeisler CA. Entrainment of the human circadian pacemaker to longer-than-24-h days. Proc Natl Acad Sci U S A. 2007;104(21):9081-9086.

1238

24. Goichot B, Weibel L, Chapotot F, Gronfier C, Piquard F, Brandenberger G. Effect of the shift of the sleep-wake cycle on three robust endocrine markers of the circadian clock. Am J Physiol. 1998; 275(2, pt 1):243-248. http://ajpendo.physiology.org /content/275/2/E243.long. Accessed May 8, 2014. 25. Salín-Pascual RJ, Ortega-Soto H, Huerto-Delgadillo L, Camacho-Arroyo I, Roldán-Roldán G, Tamarkin L. The effect of total sleep deprivation on plasma melatonin and cortisol in healthy human volunteers. Sleep. 1988;11(4):362369. 26. von Treuer K, Norman TR, Armstrong SM. Overnight human plasma melatonin, cortisol, prolactin, TSH, under conditions of normal sleep, sleep deprivation, and sleep recovery. J Pineal Res. 1996;20(1):7-14. 27. Redwine L, Hauger RL, Gillin JC, Irwin M. Effects of sleep and sleep deprivation on interleukin-6, growth hormone, cortisol, and melatonin levels in humans. J Clin Endocrinol Metab. 2000;85(10):3597-3603. 28. Dettoni JL, Consolim-Colombo FM, Drager LF, et al. Cardiovascular effects of partial sleep deprivation in healthy volunteers. J Appl Physiol (1985). 2012;113(2):232-236. 29. Sasaki Y, Miyasita A, Takeuchi T, Inugami M, Fukuda K, Ishihara K. Effects of sleep interruption on body temperature in human subjects. Sleep. 1993;16(5):478-483. 30. Bahammam AS, Almushailhi K, Pandi-Perumal SR, Sharif MM. Intermittent fasting during Ramadan: does it affect sleep? J Sleep Res. 2014;23 (1):35-43.

35. Nivison ME, Endresen IM. An analysis of relationships among environmental noise, annoyance and sensitivity to noise, and the consequences for health and sleep. J Behav Med. 1993;16(3):257-276. 36. Missildine K, Bergstrom N, Meininger J, Richards K, Foreman MD. Sleep in hospitalized elders: a pilot study. Geriatr Nurs. 2010;31(4):263271. 37. Öhrström E, Rylander R, Björkman M. Effects of night time road traffic noise: an overview of laboratory and field studies on noise dose and subjective noise sensitivity. J Sound Vibrat. 1988; 127:441-448. 38. Stradling JR. Sleep-related breathing disorders, 1: obstructive sleep apnoea: definitions, epidemiology, and natural history. Thorax. 1995;50 (6):683-689. 39. Sateia MJ, Nowell PD. Insomnia. Lancet. 2004; 364(9449):1959-1973. 40. Hening W, Allen RP, Tenzer P, Winkelman JW. Restless legs syndrome: demographics, presentation, and differential diagnosis. Geriatrics. 2007;62(9):26-29. 41. Realini T, Weinreb RN, Wisniewski SR. Diurnal intraocular pressure patterns are not repeatable in the short term in healthy individuals. Ophthalmology. 2010;117(9):1700-1704.

31. Basner M, Müller U, Elmenhorst EM. Single and combined effects of air, road, and rail traffic noise on sleep and recuperation. Sleep. 2011;34(1):11-23.

JAMA Ophthalmology October 2014 Volume 132, Number 10

Copyright 2014 American Medical Association. All rights reserved.

Downloaded From: http://archopht.jamanetwork.com/ by a University of California - San Diego User on 06/01/2015

jamaophthalmology.com

Hourly awakening vs continuous contact lens sensor measurements of 24-hour intraocular pressure: effect on sleep macrostructure and intraocular pressure rhythm.

All studies of 24-hour intraocular pressure (IOP) rhythm conducted to date have used repeated IOP measurements requiring nocturnal awakenings, potenti...