Article pubs.acs.org/ac

Highly Sensitive Nanomechanical Immunosensor Using Half Antibody Fragments Shangquan Wu,† Hong Liu,*,§ Xin M. Liang,†,∥ Xiaoping Wu,† Baomin Wang,*,‡ and Qingchuan Zhang*,† †

CAS Key Laboratory of Mechanical Behavior and Design of Material, Department of Modern Mechanics, University of Science and Technology of China, Hefei, Anhui 230027, China ‡ College of Agronomy and Biotechnology, China Agricultural University, Beijing 100193, China § Department of Chemical Physics, University of Science and Technology of China, Hefei, Anhui 230026, China ∥ Center for Biomedical Engineering, Department of Electronic Science and Technology, University of Science and Technology of China, Hefei, Anhui 230026, China S Supporting Information *

ABSTRACT: The improvement of sensitivity is of great significance to the application of biochemical sensor. In this study, we propose a micocantilever-based immunosensor in surface stress mode using half antibody fragments as receptor molecules. The thiol-containing half antibody fragment was obtained with a low loss of antibody biological activity and then was covalently and orientedly immobilized on the gold surface of microcantilevers via two native thiol groups. Such a one-step reaction and immobilization of receptor molecule simplify the preparation process of micocantilever immunosensor. Using shortened and highly oriented half antibody fragments as receptor molecules, the generation of surface stress and the transmission of stress from the interaction region of molecules to the surface of the microcantilever have been elevated significantly. The limit of detection (LOD) of the presented sensor has been significantly lowered to 1 pg/mL, or 1.1 pM in equivalence, which is a 500-fold improvement when compared with intact full antibody coated conventional micocantilever sensors. The results indicate that the half antibody fragment is well suited for the functionalization of the microcantilever surface and is generally applicable to all microcantilever immunosensor development, and this principle will help to design a functional film of devices with significantly lower LOD.

M

the preparation of immunosensor. It is crucial to orient the antigen-binding sites of the antibodies toward the analytes. For sensors with a gold-coated surface, self-assembled monolayers (SAM) of thiolated molecules containing a reactive functional group are often used. However, the orientation of antibodies that are immobilized using these approaches is random, which leads to the loss of binding capability due to steric-hindrance effect.19−21 In order to explore the full potential of the immunosensors, it is necessary to develop a proper antibody immobilization method to ensure a uniformed antibody orientation with the antigen-binding sites toward the analytes. Correctly and uniformly oriented antibodies minimize the steric hindrances22 and allow antigens to bind more easily. The oriented antibodies provide 2−8 times higher binding capacity to antigens and, thus, make the immunosensor more sensitive.23

icrocantilever, as a biochemical sensor, has attracted more and more interest for its highly sensitive, label-free, and real-time detection.1,2 In the past decade, the applications of the microcantilever sensors have been extended to the measurements of various molecular interactions, making it an adaptable and versatile sensor.1,2 The fundamental principle of the microcantilever operated in static or surface stress mode is that the interaction between target and the receptor molecules that are immobilized on only one side of the sensor generates a change of surface stress and, then, the change bends the microcantilever. Owing to the attractive features, such as labelfree detection and high sensitivity, microcantilever sensors have been successfully applied to detect various molecular interactions such as proteins, DNA, lipids, pesticides, and heavy metal ions.2−13 A molecule, as an antigen, was detected using a microcantilever immunosensor, where its specific antibody was immobilized on one side of sensor as a receptor molecule.14−18 It is well-known that the immobilization of antibody (receptor molecule) on the sensor surface is a critical step for © 2014 American Chemical Society

Received: December 16, 2013 Accepted: April 7, 2014 Published: April 7, 2014 4271

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry

Article

Figure 1. Schematic of the microcantilever-based sensing system and oriented immobilization of half antibody fragments on the gold (Au) side of the microcantilever via their own thiol groups (−SH): 1, preparation of half fragment antibody by reducing disulfide bonds using 2mercaptoethylamine (2-MEA); 2, oriented immobilization of half antibody fragments via their own two thiol groups conjugated to the gold with a single-step.

molecule but also can directly detect small molecule with high sensitivity.14,27,28 However, there have not been any related studies about the application of half antibody fragment on microcantilever immunosensor. For microcantilever immunosensor which is based on stress effect, we can speculate that highly oriented immobilization of half fragment provides not only high antigen binding capacity but also a greater resultant force due to the same direction of the stress generated from the binding. More importantly, using half fragment leads to the greatly shortened distance between the antigen-binding Fab regions and microcantilever surface (Figure 1). These factors raise the possibility that the utilization of highly oriented half antibody on microcantilever surface can greatly improve the generation and the transmission of surface stress to the microcantilever structure; subsequently, the sensor’s sensitivity can be enhanced significantly. Here, we report the first attempt of using half antibody fragments as the active sensing molecule aiming to develop the next generation microcantilever-based immunosensor with ultrahigh sensitivity. The half fragment is obtained by reducing antibody with 2-mercaptoethylamine (2-MEA) with a low loss of antibody biological activity and then is covalently and orientedly immobilized on the gold surface of microcantilevers via native thiol groups (Figure 1). Antibody with specificity to Ginsenoside Re (GRe) was used for obtaining the half antibody fragment. An enzyme-linked immunosorbent assay (ELISA) was used to evaluate the activity of half antibody fragment. The sensitivity of the presented method using half antibody fragments as receptor molecules was compared with previous studies using another oriented immobilization method to coat the same full antibodies on the microcantilever surface via goat antimouse IgG. In order to demonstrate the usefulness and adaptability of the presented technique, half antiabscisic acid

As previously reported, the sensitivity of microcantileverbased immunosensors can be improved by both oriented immobilization of antibodies18 and reduced distance between the antigen-binding Fab regions and the microcantilever surface.24 As in other immunosensors, the oriented immobilization can enhance the binding capacity of antibody to antigen in microcantilever immunosensors, and more importantly, the stress generated from the binding where the antibody is orientedly immobilized is in the same direction and thus can lead to a greater resultant force, which would greatly improve the sensitivity of the microcantilever sensor whose principle is transduction of the force into a bend of the microcantilever.18 Shortening the distance between the antibody and the microcantilever surface results in significantly elevating the transmission of stress from the reaction region in antigen− antibody to the surface of the microcantilever, thus increasing the sensitivity of the microcantilever sensor by a factor of 100.24 An antibody is a Y-shaped protein molecule that is composed of two identical halves which are combined together by two disulfide bonds (Figure 1). The antibody can be split into two identical half fragments by reducing these disulfide bonds; meanwhile, reduction of the two disulfide bonds introduces two thiol groups to each half fragment. Thus, oriented immobilization of half fragment on gold surface can be achieved because the thiol groups are in a fixed position at the far end that is away from the Fab region. Consequently, the detection sensitivities of quartz crystal microbalance (QCM) and surface plasmon resonance (SPR) can be greatly increased using this fragment as receptor molecule due to its sufficient antigen binding capacity.19,25,26 However, SPR and QCM cannot directly detect small molecule due to its small volume and small molecular weight. On the basis of the principle of surface stress, microcantilever sensors not only can detect macro4272

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry

Article

incubation at 37 °C for 30 min, the plate was washed with PBST four times, and then, an aliquot of 200 μL per well of goat antimouse IgG-HRP (Fab specific) diluted in PBS was added. After incubation at 37 °C for 30 min, the plate was washed again with PBST four times, and then, 200 μL per well of substrate solution was pipetted. The reaction was stopped after 15 min at room temperature by adding 50 μL of the stop solution. The absorbance was read at 450 nm on a microplate reader. Microcantilever Immunosensor Detection. The microcantilever was mounted in a fluid cell (0.6 mL) (Figure 1). PBS was injected into the cell by a peristaltic pump. The flow maintained a constant rate (4 mL/h) after the cell was filled with PBS. The temperature of the room and the cell were maintained at 301 ± 0.1 K and 310.00 ± 0.01 K, respectively. The microcantilever deflection was measured by monitoring the position of a laser beam reflected from the tip of the microcantilever onto a position sensitive detector (PSD). After microcantilever deflection was stabilized, the analytes dissolved in PBS were added into the fluid cell and microcantilever deflection was monitored in situ.

(half anti-ABA) antibody fragments were used to biofunctionalize the microcantilever.



EXPERIMENTAL SECTION Reagents and Solutions. 2-Mercaptoethylamine (2MEA), Ginsenoside Re (GRe), and abscisic acid (ABA) were bought from Sigma (St. Louis, MO, USA). The anti-GRe and anti-ABA monoclonal antibody (anti-GRe mAb and anti-ABA mAb) were obtained from previous research.29 The rest of the chemical reagents were bought from Beijing Chemical Reagents Co. (Beijing, China). Buffers and solutions used include phosphate buffered saline (PBS) (0.1 M phosphate buffer containing 0.9% sodium chloride, pH 7.5), PBS with 0.1% (v/ v) Tween-20 (PBST), PBST containing 0.5% (w/v) gelatin (PBSTG), coating buffer (0.05 M carbonate buffer, pH 9.6), and citrate-phosphate buffer (0.01 M citric acid and 0.03 M monosodium phosphate, pH 5.5). Apparatus. Silicon nitride microcantilevers with a V-shape (0.6 μm × 20 μm × 200 μm Veeco Instruments, Plainview, NY) were used. One side of the microcantilever has a thin film of chromium (15 nm) covered with a 60 nm layer of gold deposited by e-beam evaporation. The experimental setup used for microcantilever detection was home-built allowing one to read out single cantilevers as described in our previous studies.18,24,27,29 The 96-well polystyrene microtiter plate was bought from Costar (Corning, NY). A microplate reader (Multiskan MK3) and an automated plate-washer (Wellwash 4 MK) were bought from Thermo (Vantaa, Finland). Preparation of Half Antibody Fragment. The half antibody fragment was prepared by reducing the intact antibody with 2-MEA. The reduction protocol and the optimal concentration of 2-MEA were the same as those described by Kausaite-Minkstimiene et al.19 To remove unreacted 2-MEA, the reaction mixture was dialyzed against 20 mM PBS buffer (containing 150 mM NaCl and 1.0 mM EDTA, pH 7.2) for 48 h. The dialyzed half antibody fragment was diluted to 1.0 mg/ mL with PBS and stored at −40 °C. Half Antibody Fragment Immobilization. The immobilization procedure was performed in a microplate well that was replaced by a new one after each step. Each microcantilever was pretreated with a “piranha dip” (H2O2/H2SO4 = 1:3, 100 μL) for 15 min and washed with deionized water three times. (CAUTION: “Piranha” solution reacts violently with organic materials; it must be handled with extreme care.) The microcantilever was dried under a gentle stream of nitrogen gas and then immersed into 200 μL of 4.0 μg/mL half antibody fragment solution diluted by PBSTG and incubated for 2 h at 37 °C. After washing with PBST three times, the microcantilever was blocked with 200 μL of 3% nonfat dry milk coating buffer for 30 min at 37 °C. The microcantilevers were washed with PBST three times and dried; the functionalized microcantilevers were ready to use. Different new microcantilevers were used in each binding experiment. Characterization of Half Antibody Fragment Activity. The binding activities of the intact antibody and half antibody fragment were tested by noncompetitive ELISA as described below. Microplate was coated with 200 μL of 0.5 μg/mL coating antigen in coating buffer for 3 h at 37 °C. After washing with PBST four times, the plate was blocked with 200 μL per well of 3% nonfat milk in PBS for 30 min at 37 °C and then washed with PBST four times. To the microplate was added 100 μL of the same concentration (0.2 μg/mL) of the intact antibody and half antibody fragment dissolved in PBS. After



RESULTS AND DISCUSSION Binding Activity of Half Antibody Fragment. The 2mercaptoethylamine (2-MEA) was used to split the antibody into two half fragments by reducing disulfide bridges between the two heavy chains without affecting the antigen binding sites (Figure 1). The antibody we chose was the mouse monoclonal antibody (mAb) of Ginsenoside Re (GRe, MW = 947 Da). The dissociation constant (Kd) of anti-GRe mAb is 6.7 × 10−10 M. GRe is the main component of American ginseng; it not only has anti-ischemic, antioxidant, and antidiabetic effects but also enhances proliferation of CD4+ T cell.29 The 2-MEA solution used was an optimal concentration.19 The unreacted 2-MEA is removed from the mixture by dialyzing. To confirm binding activity of half antibody fragment, both intact antibody and half antibody fragment were tested for their bindings to immobilized coating antigens using noncompetitive ELISA. The results revealed that the reduction of anti-GRe antibody caused about 18.7% loss of activity for binding antigen. The reduction decreased optical density (OD) from 1.766 ± 0.052 to 1.435 ± 0.071; the data was obtained from 16 time detections in the ELISA plate (Figure 2), and the OD of ELISA plate without coating antigen was only 0.069 ± 0.007. These

Figure 2. Characterization of binding activities of the intact anti-GRe antibodies and half anti-GRe fragments by ELISA. The activities of the half anti-GRe fragments (1. OD = 1.435 ± 0.052) are compared with those of intact anti-GRe antibodies (2. OD = 1.7662 ± 0.071) and microplates without coating antigen (3. OD = 0.069 ± 0.007). The colors confirm the low loss of the binding activity of half antibody fragment and a specific binding of half anti-GRe fragment to GRe. 4273

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry

Article

findings confirm the low loss of the binding activity of antibody because of reductions by 2-MEA and a specific binding of half anti-GRe fragment to GRe. Microcantilever Immunosensor Detection. After functionalized with half antibody fragments, the microcantilever was mounted in a fluid cell, and an optical lever was used to measure the deflection of microcantilever in real time (Figure 1). After the baseline of deflection was stable in a steady continuous flow using phosphate buffered saline (PBS) solution, antigen (GRe) solution with various concentrations was circulated into the cell sequentially. The deflection was monitored in situ for approximately 80 min. The deflection curves for different concentrations of antigens were illustrated (Figure 3a). Antigen binding caused the microcantilevers to

the bending is caused by specific binding of half frag-anti-GRe and GRe. The bending occurred because the binding of GRe to half frag-anti-GRe results in a difference in force between the antibody fragments coated surface on the top side and the silicon nitride surface on the backside of the microcantilever. Figure 4 illustrates the equilibrium deflections of microcantilever for various GRe concentrations. Every data point was an

Figure 4. Equilibrium deflections of microcantilever for various concentrations of GRe.

average of 4 experiments (the data of the other 3 experiments are shown in Figure S1 in Supporting Information) using different microcantilevers, and the error bar reflects the range of deflections. The responses of the microcantilever suggested a positive correlation between the equilibrium deflection and the concentration of GRe. The corresponding microcantilever deflection is proportional to the logarithm of the concentration of antigen. The background noise of the measurement was approximately 3 nm, and the limit of detection (LOD) was calculated to be 3 times higher than the background noise; thus, the LOD of the presented technique was in the 1 pg/mL range (1.1 pM). Our results demonstrated that the microcantilevers based on stress effect using half antibody fragments as receptor molecules are sensitive enough to detect GRe and quantify the GRe. In our previous studies,29 anti-GRe antibody, the same sample used for obtaining half antibody fragment in the present study, was immobilized on the gold surface of the microcantilever via another two methods: (1) random immobilization by sulfhydrylated antibody using the sulfhydrylation reagent which reacts with the −NH2 group of the antibody to give a thiol (−SH) group for conjugation onto the gold surface; (2) oriented immobilization using sulfhydrylated goat antimouse IgG which can specifically bind the Fc part of the antibody to achieve oriented immobilization of antibody (Figure 3b). The microcantilever with oriented intact full antibody coated using sulfhydrylated goat antimouse IgG demonstrated better sensitivity (3−8 times) than that randomly immobilized via sulfhydrylated antibody. The LOD of the microcantilever with antibody immobilized using sulfhydrylated goat antimouse IgG for GRe detection was about 0.5 ng/mL. However, because the region in the goat antimouse IgG for binding the Fc parts of antibodies was shielded due to random immobilization of goat antimouse IgG, the surface concentration of immobilized antibodies using sulfhydrylated goat antimouse IgG was lower than the method using sulfhydrylated antibody. Another point

Figure 3. Real-time nanomechanical response of microcantilever caused by the specific binding for GRe detection. (a) Microcantilever deflections caused by the binding of GRe and covalently and directionally immobilized half anti-GRe fragment on the gold-coated side of microcantilevers. (b) Microcantilever deflections caused by the binding of GRe and intact anti-GRe antibody that is immobilized using sulfhydrylated goat antimouse IgG which can specifically bind the Fc part of the antibody to achieve oriented immobilization of antibody.29

bend, and the cantilever bent toward the Au side. An equilibrium-bending signal of ≈87, 75, 55, and 32 nm was obtained for 10, 1.0, 0.1, and 0.02 ng/mL antigen solutions, respectively. As a reference, 1000 ng/mL chlorsulfuron (a nonbinder molecule, C12H12ClN5O4S, MW = 357.8 Da) and 10 μg/mL BSA dissolved in PBS were injected in the fluid cell and the microcantilever had no significant bending, indicating that 4274

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry

Article

worth noting is that the film of goat antimouse IgG increases the distance between the antigen binding and the microcantilever surface. This point decreases the stress transmission from antigen−antibody reactions to microcantilever and reduces the microcantilever sensitivity. In the current study, using the half antibody fragment as the receptor molecule, the sensitivity had a 500-fold increase. The improvement of the sensitivity could be mainly attributable to higher density and higher efficiency of stress transmission from antigen−antibody reactions to microcantilever, which resulted from the single-step immobilization of half antibody fragments via two native thiol-groups directly conjugated to the gold and the shortened distance between the antigen-binding Fab regions and the microcantilever surface. In particular, there is only Fab part between the microcantilever surface and the antigen-binding site without linker molecule and Fc part (Figure 1), which greatly shortens the distance between the microcantilever surface and the antigen-binding site, resulting in a more efficient stress transmission. Also, immobilization via two thiol-groups enhances the stability of receptor molecule but, more importantly, increases the stiffness of the connection between the microcantilever surface and the receptor molecule, which in turn further improves stress transmission. Intermolecular interactions governed by the Lennard−Jones potential, van der Waals, electrostatic forces, steric, as well as hydration, and surface hydrophobicity are considered to be the origins of the stresses.8,14,30,31 Antibody conformational change due to antigen binding was proposed as the origins for smallmolecule antigen detection.18 The changes of conformation due to adsorbed molecules are also considered to be the origins in many other intermolecular interactions.32−34 Whatever the origin is, after generating in the antigen-binding Fab regions, the stress transmits through the receptor molecule and the molecule that links microcantilever surface and the receptor molecule and finally reaches and bends the microcantilever. Oriented immobilization and shorter distance between the antigen-binding Fab regions and the microcantilever surface with half antibody fragment as the receptor layer can enhance the generation of surface stress and the stress transmission for every kind of stress, regardless of its origin. As previously reported, the sensitivity of a microcantilever immunosensor in surface stress mode using single-chain Fv (scFv) antibody fragment as receptor molecule was increased to ≈1 nM.17 However, scFv fragments are obtained through genetic engineering rather than a conventional method and may not be easy to acquire since the production process is complex. Attaching a thiol group to the C-terminal end of scFv further increases the complexity and difficulty of the procedure. In the present study, by immobilizing half antibody fragment via two native thiol-groups, which provides better stability of the receptor molecule and a higher stiffness of the connection between receptor molecule and microcantilever, we increased the sensitivity of microcantilevers at limits of detection (LOD) of 1.1 pM (1 pg/mL). Another important advantage is that thiol-containing half antibody fragment was obtained via a simple chemical reaction with only one step. Usefulness and Adaptability of the Presented Technique. To demonstrate the usefulness and adaptability of the presented technique, a microcantilever immunosensor for phytohormone abscisic acid (ABA) was developed, using the procedure described above for the GRe microcantilever (Figure 1). ABA is an important phytohormone that has critical functions for growth and development and for stress resistance,

and it regulates the expression of many genes in plants.35 The sensitivity of the method using half antibody fragment was compared with the microcantilever with antibody immobilized via thiolated anti-ABA antibody using the sulfhydrylation reagent 2-iminothiolane hydrochloride which reacts with the −NH2 group of the antibody to give a thiol (−SH) group for conjugation onto the gold surface. Figure 5a shows the real-

Figure 5. Real-time nanomechanical response of microcantilever caused by the specific binding for ABA detection. (a) Microcantilever deflections caused by the binding of ABA and half anti-ABA fragment that is covalently and orientedly immobilized on the gold surface of microcantilevers via native thiol groups. (b) Microcantilever deflections caused by the binding of ABA and intact anti-ABA antibody using the sulfhydrylation reagent 2-iminothiolane hydrochloride which reacts with the −NH2 group of the antibody to give a thiol (−SH) group for conjugation onto the gold surface.

time deflections of the microcantilever coated with half antiABA fragment with the injection of different concentrations of ABA. As a reference, 1000 ng/mL chlorsulfuron (a nonbinder molecule, C12H12ClN5O4S, MW = 357.8 Da) and 10 μg/mL BSA dissolved in PBS were injected in the fluid cell, and the microcantilever had no significant bending, indicating that the bending is caused by specific binding of half frag-anti-ABA and ABA. Figure 5b shows the real-time deflections of the microcantilever coated with thiolated anti-ABA antibody with the injection of different concentrations of ABA. In the same ABA concentration of 10 ng/mL, the deflections of the microcantilever with antibody immobilized via half anti-ABA fragment and thiolated anti-ABA antibody were ≈125 and ≈65 nm, respectively. The results indicated that the microcantilevers 4275

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry



ACKNOWLEDGMENTS This work was supported by the National Natural Science Foundation of China (Grant Nos. 11127201 and 11072233), Chinese National Key Technology R&D Program (2012BAK08B05), the Fundamental Research Funds for the Central Universities (WK2090050032), and the National Basic Research Program of China (2011CB302105).

coated with half anti-ABA fragment had a higher sensitivity than those with thiolated anti-ABA antibody.



CONCLUSIONS Immobilization of receptor molecules on sensor surfaces is a difficult point, but a key point, especially for further application of the technique. In the current study, we investigated the application of half antibody fragment as receptor molecule for microcantilever sensor in surface stress mode. Half antibody fragment was chosen for its shorter distance between the antigen-binding Fab regions and the microcantilever surface compared with intact antibody molecule and its native thiol group that readily covalently immobilized onto the gold surfaces with oriented immobilization. For microcantilever sensor whose principle is transduction of the force into a bend of microcantilever, maximizing the generation of surface stress and the transmission of stress from the interaction region of molecules to the surface of the microcantilever is significant for improving microcantilever sensitivity. Oriented immobilization of half antibody fragment provides higher binding capacity, and the stress generated from the binding in the receptor molecule is in the same direction; thus, it can lead to a greater resultant force. The shorter distance between the antigen-binding Fab regions and the microcantilever surface significantly increases the stress transmission from the interaction region of molecules to the surface of the microcantilever. We demonstrated that using half antibody fragment as receptor molecule improved the sensitivity of microcantilever significantly. The current LOD of the method for GRe detection is as low as 1 pg/mL, which was improved 500-fold compared with a previous study using intact antibody molecule. One-step reaction and immobilization of receptor molecule are two advantages of the current method. The results indicate that the half antibody fragment is well suited for the functionalization of microcantilever surface and is generally applicable to all antibodies in microcantilever sensor development. Further research should focus on improvement and preserving biological activity of antibody fragment on the gold surface of the microcantilever and reuse of microcantilever sensors.





REFERENCES

(1) Huber, F.; Lang, H. P.; Backmann, N.; Rimoldi, D.; Gerber, C. Nat. Nanotechnol. 2013, 8, 125−129. (2) Tamayo, J.; Kosaka, P. M.; Ruz, J. J.; San Paulo, A.; Calleja, M. Chem. Soc. Rev. 2013, 42, 1287−1311. (3) Nugaeva, N.; Gfeller, K. Y.; Backmann, N.; Lang, H. P.; Duggelin, M.; Hegner, M. Biosens. Bioelectron. 2005, 21, 849−856. (4) Hou, H.; Bai, X. J.; Xing, C. Y.; Gu, N. Y.; Zhang, B. L.; Tang, J. L. Anal. Chem. 2013, 85, 2010−2014. (5) Singamaneni, S.; McConney, M. E.; LeMieux, M. C.; Jiang, H.; Enlow, J. O.; Bunning, T. J.; Naik, R. R.; Tsukruk, V. V. Adv. Mater. 2007, 19, 4248−4255. (6) Joo, J.; Kwon, D.; Yim, C.; Jeon, S. ACS Nano 2012, 6, 4375− 4381. (7) Gruber, K.; Horlacher, T.; Castelli, R.; Mader, A.; Seeberger, P. H.; Hermann, B. A. ACS Nano 2011, 5, 3670−3678. (8) Mertens, J.; Rogero, C.; Calleja, M.; Ramos, D.; Martin-Gago, J. A.; Briones, C.; Tamayo, J. Nat. Nanotechnol. 2008, 3, 301−307. (9) Zhang, J.; Lang, H. P.; Huber, F.; Bietsch, A.; Grange, W.; Certa, U.; McKendry, R.; Guntgerodt, H. J.; Hegner, M.; Gerber, C. Nat. Nanotechnol. 2006, 1, 214−220. (10) Pera, I.; Fritz, J. Langmuir 2007, 23, 1543−1547. (11) Alvarez, M.; Calle, A.; Tamayo, J.; Lechuga, L. M.; Abad, A.; Montoya, A. Biosens. Bioelectron. 2003, 18, 649−653. (12) Chen, X.; Chen, W.; Mulchandani, A.; Mohideen, U. Sens. Actuators, B 2012, 161, 203−208. (13) Xu, Y.; Zhang, B.; Wu, S.; Xia, Y. Anal. Chim. Acta 2009, 649, 117−122. (14) Wu, G. H.; Datar, R. H.; Hansen, K. M.; Thundat, T.; Cote, R. J.; Majumdar, A. Nat. Biotechnol. 2001, 19, 856−860. (15) Yen, Y.-K.; Huang, C.-Y.; Chen, C.-H.; Hung, C.-M.; Wu, K.-C.; Lee, C.-K.; Chang, J.-S.; Lin, S.; Huang, L.-S. Sens. Actuators, B 2009, 141, 498−505. (16) Lee, D.; Kwon, D.; Ko, W.; Joo, J.; Seo, H.; Lee, S. S.; Jeon, S. Chem. Commun. 2012, 48, 7182−7184. (17) Backmann, N.; Zahnd, C.; Huber, F.; Bietsch, A.; Pluckthun, A.; Lang, H. P.; Guntherodt, H. J.; Hegner, M.; Gerber, C. Proc. Natl. Acad. Sci. U.S.A. 2005, 102, 14587−14592. (18) Wu, S. Q.; Nan, T. G.; Xue, C. G.; Cheng, T.; Liu, H.; Wang, B. M.; Zhang, Q. C.; Wu, X. P. Biosens. Bioelectron. 2013, 48, 67−74. (19) Kausaite-Minkstimiene, A.; Ramanaviciene, A.; Kirlyte, J.; Ramanavicius, A. Anal. Chem. 2010, 82, 6401−6408. (20) Bonroy, K.; Frederix, F.; Reekmans, G.; Dewolf, E.; De Palma, R.; Borghs, G.; Declerck, P.; Goddeeris, B. J. Immunol. Methods 2006, 312, 167−181. (21) Tang, D. Q.; Zhang, D. J.; Tang, D. Y.; Ai, H. J. Immunol. Methods 2006, 316, 144−152. (22) Oh, B. K.; Kim, Y. K.; Lee, W.; Bae, Y. M.; Lee, W. H.; Choi, J. W. Biosens. Bioelectron. 2003, 18, 605−611. (23) Turkova, J. J. Chromatogr., B 1999, 722, 11−31. (24) Wu, S. Q.; Liu, H.; Cheng, T.; Zhou, X. R.; Wang, B. M.; Zhang, Q. C.; Wu, X. P. Sens. Actuators, B 2013, 186, 353−359. (25) Karyakin, A. A.; Presnova, G. V.; Rubtsova, M. Y.; Egorov, A. M. Anal. Chem. 2000, 72, 3805−3811. (26) Sharma, H.; Mutharasan, R. Anal. Chem. 2013, 85, 2472−2477. (27) Tan, W. M.; Huang, Y.; Nan, T. G.; Xue, C. G.; Li, Z. H.; Zhang, Q. C.; Wang, B. M. Anal. Chem. 2010, 82, 615−620. (28) Dionisio, M.; Oliviero, G.; Menozzi, D.; Federici, S.; Yebeutchou, R. M.; Schmidtchen, F. P.; Dalcanale, E.; Bergese, P. J. Am. Chem. Soc. 2012, 134, 2392−2398.

ASSOCIATED CONTENT

S Supporting Information *

Additional information as noted in text. This material is available free of charge via the Internet at http://pubs.acs.org/.



Article

AUTHOR INFORMATION

Corresponding Authors

*Phone: +86-551-63607613. Fax: +86-551-63601248. E-mail: [email protected]. *Phone: +86-10-62731305. Fax: +86-10-62732567. E-mail: [email protected]. *Phone: +86-551-63607613. Fax: +86-551-63601248. E-mail: [email protected]. Author Contributions

The manuscript was written through contributions of all authors. All authors have given approval to the final version of the manuscript. Notes

The authors declare no competing financial interest. 4276

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Analytical Chemistry

Article

(29) Nan, T. G.; Wu, S. Q.; Zhao, H. W.; Tan, W. M.; Li, Z. H.; Zhang, Q. C.; Wang, B. M. Anal. Chem. 2012, 84, 4327−4333. (30) Kang, K.; Sachan, A.; Nilsen-Hamilton, M.; Shrotriya, P. Langmuir 2011, 27, 14696−14702. (31) Watari, M.; Galbraith, J.; Lang, H. P.; Sousa, M.; Hegner, M.; Gerber, C.; Horton, M. A.; McKendry, R. A. J. Am. Chem. Soc. 2007, 129, 601−609. (32) Buchapudi, K.; Xu, X. H.; Ataian, Y.; Ji, H. F.; Schulte, M. Analyst 2012, 137, 263−268. (33) Federici, S.; Oliviero, G.; Hamad-Schifferli, K.; Bergese, P. Nanoscale 2010, 2, 2570−2574. (34) Oliviero, G.; Chiari, M.; De Lorenzi, E.; Colombo, R.; Cretich, M.; Damin, F.; Federici, S.; Depero, L. E.; Bergese, P. Sens. Actuators, B 2013, 176, 1026−1031. (35) Fujii, H.; Chinnusamy, V.; Rodrigues, A.; Rubio, S.; Antoni, R.; Park, S. Y.; Cutler, S. R.; Sheen, J.; Rodriguez, P. L.; Zhu, J. K. Nature 2009, 462, 660−664.

4277

dx.doi.org/10.1021/ac404065m | Anal. Chem. 2014, 86, 4271−4277

Highly sensitive nanomechanical immunosensor using half antibody fragments.

The improvement of sensitivity is of great significance to the application of biochemical sensor. In this study, we propose a micocantilever-based imm...
2MB Sizes 1 Downloads 3 Views