Journal of Medical Imaging and Radiation Oncology •• (2014) ••–•• bs_bs_banner

R ADIOLO GY—OR I G I N AL A RT I C L E

Effects of electrocardiogram gating on CT pulmonary angiography image quality Nicholas D Ardley,1 Ken K Lau,1 Kevin Buchan,2 Eldho Paul3 and John M Troupis1 1 Department of Diagnostic Imaging, Monash Medical Centre, Monash Health, 2Clinical Science, Philips Healthcare, and 3School of Public Health and Preventive Medicine, Monash University, Melbourne, Victoria, Australia

ND Ardley BRadMedImag (HONS); KK Lau MBBS, FRANZCR; K Buchan DCR(R) GDVET; E Paul MSc; JM Troupis MBBS, FRANZCR. Correspondence BRadMedImag (HONS) Nicholas D Ardley, Department of Diagnostic Imaging, Monash Health, 246 Clayton Road, Clayton, Vic. 3168, Australia. Email: [email protected] Conflict of interest: This research was not funded in anyway. One of the authors, Kevin Buchan, is an employee of Philips Healthcare in the field of Clinical Science. The remaining authors have no conflict of interest. Submitted 5 October 2013; accepted 27 November 2013. doi:10.1111/1754-9485.12147

Abstract Introduction: Pulmonary embolism (PE) is the third most common cause of death from cardiovascular disease. Computed-tomographic pulmonary angiography (CTPA) is an accurate and safe test for diagnosing PE. The aim of this retrospective analysis was to evaluate the effects on image quality (IQ) of electrocardiogram (ECG) gating during CTPA. Methods: Fifty consecutive patients presenting for CTPA were included in the study. A single acquisition was performed, resulting in two reconstructions: one at 75% of the R–R interval and the other without ECG influence. IQ evaluation was undertaken by two radiologists, focusing on respiratory and cardiac motion, image noise, low-contrast resolution, vessel and lung clarity, contrast media opacification and artefacts. Various regions of the lungs and vasculature were evaluated, and IQ scores were statistically compared. Results: For the ECG-tagged reconstructions, IQ was noted to be better overall with regard to vessel clarity (P < 0.05) and cardiac motion (P < 0.05), while lung clarity was better only in the left lower zone (P < 0.05). IQ was better with regard to image noise (P < 0.05) and low-contrast resolution (P < 0.05) in the non-ECG-tagged reconstructions. No statistical IQ difference between the two types of reconstruction was noted with regard to respiratory motion, contrast media opacification or presence of artefacts. Conclusion: The two types of reconstruction provide complementary information for evaluating CTPA results. Key words: cardiopulmonary anatomy; computed tomography; ECG gating; pulmonary embolism; vascular.

Introduction Pulmonary embolism (PE) is the third most common cause of death from cardiovascular disease after myocardial infarction and stroke.1 In the United States it has an incidence of 69 per 100 000 patients, with 175 000 new cases occurring every year2 and a mortality rate as high as 30–60%.1,3,4 In the recent past, ventilation–perfusion lung scans, performed in nuclear medicine departments, were the most common investigative tool for suspected pulmonary embolism; however, 60–70% of these scans are deemed non-diagnostic.2,5 Computed-tomographic pulmonary angiography (CTPA) is the imaging modality of choice when investigating for the presence of PE. This is largely due to multidetector CT (MDCT) scanning technology, which enables routine use of thin-slice data with improved low-contrast resolution, improved spatial reso© 2014 The Royal Australian and New Zealand College of Radiologists

lution, and shorter scan times due to better temporal resolution as well as to its use of tools for optimising the timing of the contrast bolus. Today CTPA is regarded as an accurate and safe test for diagnosing PE, as it has improved visualisation of segmental and subsegmental pulmonary arteries6 such that over 90% of subsegmental pulmonary arteries can now be visualised.7 CTPA has a sensitivity of 62–78% and a specificity of 86–95%, according to a large multicentre research study.8 Among clinical centres the protocol for acquiring a CTPA study varies significantly9; however, it would be advantageous to find a more stable and uniform protocol for the implementation of CTPA studies to increase the sensitivity and specificity of CTPA examinations. Pulmonary arterial structures are influenced by cardiac motion, which can induce artefacts such as blurring of vessel edges, dark shading in the otherwise bright column 1

ND Ardley et al.

of contrast within vessels and double-line artefacts along the long axis of a vessel.10 Electrocardiogram (ECG) gating was initially developed for imaging of the coronary arteries to reduce artefacts produced by the moving heart.11 Electrocardiogram gating during CTPA has not been widely adopted in scanners with limited z-axis detector coverage due to the resultant longer scan times and greater radiation exposure.12 The latest CT scanners, with faster rotation times, additional detector rows and more powerful X-ray tubes that allow fast volume coverage, enable better ECG-gated CT scanning of the entire chest within a single breath-hold.13 Applying ECG gating during CTPA examinations can reduce cardiac motion and aortic pulsation artefact and may improve pulmonary vessel clarity, particularly in regions adjacent to the heart borders and around the lung hila, thus potentially leading to better accuracy in diagnosis of pulmonary embolism and better overall image quality (IQ). The aim of this prospective study was to evaluate motion reduction and IQ changes with ECG gating during CTPA.

Methods

radiation dose.14 In addition, ECG-correlated tube current modulation and multisector reconstruction were utilised for the varying heart rates,15 with the dose peak centred at 75% of the R–R interval (to obtain a still phase in end diastole) and lower doses for the remainder of the cardiac cycle. No ECG padding (phase tolerance) was applied to the protocol. Each patient was irradiated only once. This scanning protocol had default parameters that matched the radiation dose delivered during an ungated CTPA examination. The use of ECG gating did not increase the radiation dose.

Intravenous contrast media injection protocol One hundred millilitres of intravenous contrast media (iohexol 350 mg I/mL; Omnipaque 350 (GE Healthcare, Milwaukee, WI, USA)) was injected at a rate of 4 mL/ sec through a 20-gauge intravenous cannula in the antecubital fossa. Each examination was triggered using a bolus-tracking technique whereby the scan began automatically when the region of interest placed in the pulmonary trunk had reached a radiodensity of 130 Hounsfield units.

Patients

CT image reconstruction

Fifty-two consecutive patients of all ages and both sexes who presented to the emergency department of a tertiary referral centre for suspected PE were referred for CTPA between March and September 2011 and included in this study. Two patients who were uncooperative in that they were confused and unable to breath-hold were excluded, leaving 50 data sets available for analysis. The study was approved by the institutional ethics committee.

Reconstruction with retrospective ECG gating produces images that are centred at a selected percentage of the R–R time interval of the ECG trace.16 All percentages of the R–R time interval from any ECG cycle during the acquisition may be reconstructed. Although each patient underwent only one CT acquisition, two fine-slice data sets were reconstructed from the available raw data. The fine-slice data sets were reconstructed with 0.9-mm axial slices overlapped every 0.45 mm on a B (soft) filter with the same level of iterative reconstruction (iDose4, Philips Healthcare). The difference between the two data sets was that one was reconstructed with ECG tagging centred at 75% of the R–R interval (referred to as ‘tagged’ data), whereas the other was reconstructed from the raw data without any consideration of the ECG trace information influencing the reconstruction (referred to as ‘untagged’ data). Traditional helical raw data acquisition utilises spatial interpolation, and an image is produced from at least 180 degrees of data. Reconstruction to an ECG trace still requires 180 degrees of raw data; however, the 180 degrees can be made up of data from multiple sectors comprising different portions of the arc that sum to 180 degrees or more. For example, four sectors could comprise 45 degrees each to sum to 180 degrees.17 Each of these sectors can be taken from different cardiac cycles, although they must be taken at the same phase point (75%). This means that the resultant image has improved temporal resolution, as it is made up of sectors, which inherently have shorter time periods, contained within the data.18 Untagged data differ because the ECG trace that

Heart rate All examinations were acquired without modification of heart rate. No beta-blocking agents were administered for the purposes of this study; however, no investigation into patients’ medication was undertaken to ascertain use of beta blockers prior to CTPA examination. The heart rate of each patient during CT scanning was recorded.

CT scanning The CTPA examinations were undertaken using a 128slice MDCT scanner (Ingenuity, Philips Healthcare, Cleveland, OH, USA). Each examination was acquired using the same protocol. Retrospective ECG-gated helical CTPA was undertaken in the craniocaudal direction, covering the lung fields from apices to bases. The scanning parameters were as follows: detector collimation of 64 × 0.625 mm, 400 ms per gantry rotation, pitch of 0.299 and tube voltage of 100 kVp. Each study utilised an automated dose suggestion, which varied with patient size. Tube current was modulated in order to reduce 2

© 2014 The Royal Australian and New Zealand College of Radiologists

ECG gating and CTPA image quality

Table 1. Image quality ranking system Indicator Clarity of vessels Clarity of lungs Respiratory motion Cardiac motion Presence of other artefacts Low-contrast resolution Contrast media opacification Image noise

1

2

3

4

5

Not assessable Not assessable Excessive Excessive Severe None None Excessive

Poor Poor Significant Significant Mild Poor Poor Significant

Fair Fair Moderate Moderate Moderate Fair Fair Moderate

Good Good Minimal Minimal Minimal Good Good Minimal

Excellent Excellent None None None Excellent Excellent None

defines phase points within the data is ignored. These data are reconstructed like a regular CT scan, using 180 degrees or more of spiral interpolated data. The data for where the dose is increased at the 75% phase point and the data for where the dose is dropped away from the 75% phase point are merged into one for reconstruction. Thus, temporal resolution is not optimised in the reconstructed images. These data closely simulate a routine CTPA helical acquisition without ECG gating.

Assessment of CT images Both tagged and untagged data sets for every patient were deidentified and randomised. Two independent thoracic radiologists, each with greater than 15 years’ CT experience, who were blinded to the image type (ECGtagged or untagged) evaluated each data set for IQ with the use of a thin client workstation (Intellispace Portal, Philips Healthcare). A five-point ranking system was used to assess IQ with reference to the assessor’s appraisal of cardiac and respiratory motion, image noise across structures, low-contrast resolution, clarity of vessels and lungs, contrast media density and the presence of other artefacts. Each indicator of IQ mentioned was analysed in each of the following anatomical regions: right and left upper zones, right and left middle zones, right and left lower zones, right and left main pulmonary arteries. A sliding scale of 1–5 was utilised for the IQ assessment, with a definition of each ranking as given in Table 1. All of the IQ scores for all data sets reviewed by both radiologists were collated and compared.

Statistical analysis Data were analysed using SAS software version 9.2 (SAS Institute, Cary, NC, USA). The difference between tagged and untagged IQ scores for various CT IQ indicators were estimated using repeated-measures analysis of variance, with results reported as parameter estimates and standard errors. Agreement between radiologists on IQ scores for various CT IQ indicators was assessed using weighted kappa. A kappa value above 0.8 was considered as almost perfect agreement, 0.6 to 0.8 as good, 0.4 to 0.6 as © 2014 The Royal Australian and New Zealand College of Radiologists

moderate, 0.2 to 0.4 as fair and less than 0.2 as poor. Statistical significance was set at a two-sided P value of 0.05.

Results Study group characteristics The study group consisted of 50 patients (27 male and 23 female) with a mean age of 67.7 ± 13.3 (range 34–89 years). Mean average heart rate for the duration of the scan was 80.32 ± 19.98 bpm (range 30–139). Mean dose length product was 309 ± 188.3 (range 117.9– 1228), which converts to a mean dose of 5.25 mSv (weighting factor = 0.017).

Image quality evaluation The interobserver agreement between the two thoracic radiologists was considered good to almost perfect in each category for the majority of regions, with an average kappa value of 0.79.

Evaluation of clarity of vessels There was a significant difference in the clarity of vessels between the tagged and untagged data sets, as shown in Table 2. Each anatomical region was assessed as showing better clarity of vessels in the tagged data set. The left lower zone was the anatomical region that showed the largest difference in clarity of vessels between the tagged and untagged data sets, suggesting that ECG gating allows the influence of cardiac motion on the vessels to be reduced further compared with theother anatomical regions assessed. Figure 1 demonstrates the difference in the clarity of vessels between an ECG-tagged data set and its corresponding untagged data set.

Evaluation of clarity of lungs Table 2 indicates that there was better clarity in the left lower zone only (see Fig. 2), while there was no signifi3

ND Ardley et al.

Table 2. Difference between tagged and untagged scores for each image quality indicator Region

Left lower zone Estimate ± SE P value Left middle zone Estimate ± SE P value Left pulmonary trunk Estimate ± SE P value Left upper zone Estimate ± SE P value Right lower zone Estimate ± SE P value Right middle zone Estimate ± SE P value Right pulmonary trunk Estimate ± SE P value Right upper zone Estimate ± SE P value

Clarity of vessels

Clarity of lungs

Respiratory motion

Cardiac motion

Artefacts

Low-contrast resolution

Contrast media opacification

Image noise

0.41 ± 0.10 ≤0.001*

0.25 ± 0.10 0.02*

−0.16 ± 0.10 0.10

0.93 ± 0.12 ≤0.001*

0.11 ± 0.07 0.13

−0.55 ± 0.09 ≤0.001*

−0.13 ± 0.08 0.09

−0.50 ± 0.10 ≤0.001*

0.34 ± 0.11 ≤0.001*

0.12 ± 0.11 0.28

−0.06 ± 0.10 0.53

0.87 ± 0.11 ≤0.001*

0.07 ± 0.08 0.40

−0.53 ± 0.08 ≤0.001*

−0.11 ± 0.08 0.15

−0.51 ± 0.10 ≤0.001*

−0.07 ± 0.05 0.13

0.56 ± 0.09 ≤0.001*

0.01 ± 0.04 0.80

−0.50 ± 0.09 ≤0.001*

−0.09 ± 0.07 0.19

−0.47 ± 0.09 ≤0.001*

0.21 ± 0.08 0.01* 0.25 ± 0.10 0.02*

0.05 ± 0.09 0.58

−0.03 ± 0.07 0.65

0.42 ± 0.07 ≤0.001*

−0.13 ± 0.08 0.10

−0.54 ± 0.09 ≤0.001*

−0.10 ± 0.07 0.17

−0.55 ± 0.10 ≤0.001*

0.31 ± 0.10 ≤0.001*

0.18 ± 0.10 0.07

−0.16 ± 0.08 0.06

1.15 ± 0.54 0.04*

−0.04 ± 0.06 0.50

−0.52 ± 0.09 ≤0.001*

−0.13 ± 0.08 0.09

−0.49 ± 0.10 ≤0.001*

0.30 ± 0.10 0.01*

0.11 ± 0.10 0.26

−0.05 ± 0.08 0.50

0.62 ± 0.09 ≤0.001*

−0.09 ± 0.07 0.17

−0.53 ± 0.08 ≤0.001*

−0.11 ± 0.08 0.15

−0.52 ± 0.10 ≤0.001*

0.08 ± 0.06 0.15

0.89 ± 0.10 ≤0.001*

0.01 ± 0.04 0.80

−0.50 ± 0.09 ≤0.001*

−0.09 ± 0.07 0.19

−0.48 ± 0.10 ≤0.001*

−0.05 ± 0.06 0.43

0.41 ± 0.07 ≤0.001*

−0.11 ± 0.08 0.15

−0.55 ± 0.09 ≤0.001*

−0.05 ± 0.09 0.57

−0.54 ± 0.09 ≤0.001*

0.22 ± 0.08 0.01* 0.26 ± 0.10 0.01*

0.03 ± 0.09 0.74

*P ≤ 0.05. Positive estimates reflect a swing towards the tagged data series, while negative results reflect a swing towards the untagged data series.

cant difference in the remaining regions of the chest (see Figs. 3 and 4). The left lower zone is the region that is impacted the most by cardiac pulsation.

This result suggests that ECG gating has no effect on the amount of respiratory motion seen in a CTPA examination.

Evaluation of respiratory motion

Evaluation of cardiac motion

There was no significant difference in the respiratory motion between the tagged and untagged data sets in any of the anatomical regions, as shown in Table 2.

There was a significant difference in the influence of cardiac motion between the tagged and untagged data sets (shown in Table 2), with the tagged data set

Fig. 1. Oblique image of the left pulmonary arteries showing improved vessel clarity in the ECG-tagged data set (a). The vessel borders are better defined in direct comparison to the nonECG-tagged data set (b).

4

© 2014 The Royal Australian and New Zealand College of Radiologists

ECG gating and CTPA image quality

Fig. 2. Axial image of the left lower lobe of the lung showing superior clarity of lung for the ECGtagged data set (a) in direct comparison to the non-ECG-tagged data set (b).

Fig. 3. Axial image of the lungs taken at the level of the right main pulmonary artery showing equal clarity of lung between the ECG-tagged data set (a) and the non-ECG-tagged data set (b).

Fig. 4. Coronal image of the lungs showing equal clarity of lung between the ECG-tagged data set (a) and the non-ECG-tagged data set (b).

© 2014 The Royal Australian and New Zealand College of Radiologists

5

ND Ardley et al.

Fig. 5. Axial image through level of the left ventricle showing the effects of cardiac motion. The borders of the heart and surrounding vessels in the ECG-tagged data set (a) are better defined with less influence from cardiac motion than in the non-ECG-tagged data set (b). The average heart rate for this acquisition was 80 bpm.

showing less influence of cardiac motion on image quality in each anatomical region. The difference was particularly noticeable in the lower zones (see Fig. 5).

that ECG gating reduces the low-contrast resolution of images.

Evaluation of contrast media opacification Evaluation of the presence of other artefacts There was no significant difference in the presence of artefacts between the tagged and untagged data sets in any of the anatomical regions, as shown in Table 2.

There was no significant difference in the contrast media density between the tagged and untagged data sets (see Table 2).

Evaluation of image noise Evaluation of low-contrast resolution Each anatomical region was assessed as showing better low-contrast resolution in the untagged data set than in the tagged data set (as seen in Table 2), suggesting

There was a significant difference in image noise influencing IQ between the tagged and untagged data sets, as seen in Table 2. More image noise was seen in the tagged data set than in the untagged data set. This is visually represented in Figures 6 and 7.

Fig. 6. Axial image through the pulmonary arteries, showing more image noise in the ECGtagged data set (a) in comparison with the nonECG-tagged data set (b).

6

© 2014 The Royal Australian and New Zealand College of Radiologists

ECG gating and CTPA image quality

Fig. 7. Oblique image of the left pulmonary arteries, showing more image noise in the ECGtagged data set (a) in comparison to the non-ECGtagged data set (b).

Pulmonary embolus detection Pulmonary emboli were detected in 6 of the 50 examinations; each one was detected on both reconstructed data sets. All the emboli detected were noted to be of moderate to large size.

Discussion Electrocardiogram gating played a significant role in influencing the overall image quality obtained from the CTPA examinations. It minimised the cardiac and aortic pulsation artefact. Cardiac motion degrades IQ, not only for the details of the lung adjacent to the heart and aorta, but also for the surrounding pulmonary vessels and lung due to double shadows. The reduction in cardiac motion was complemented by an improvement in vessel clarity. Vessel clarity is of high importance in CTPA examinations, as motion artefact makes vessel borders less distinct and may lead to erroneous diagnosis of pulmonary embolism. Electrocardiogram gating particularly enhanced the clarity of the subsegmental branches of the pulmonary arteries in the mediobasal segments of both lower lobes adjacent to the heart borders. Electrocardiogram gating demonstrated no adverse influences in terms of respiratory motion, contrast media opacification or artefacts. Respiratory motion is primarily influenced by the patient’s ability to breath-hold, and ECG gating did not have any impact. Opacification of iodinated contrast media is altered by the X-ray penetration (i.e. 100 kV and 120 kV) and also by the amount of contrast that reaches the pulmonary vessels at the time of scanning; however, the peak kilovoltage was standardised in this study to 100 kVp. Electrocardiogram gating played no role in contrast media opacification. Image noise and low-contrast resolution were better in non-ECG-tagged data (P < 0.001). Electrocardiogram© 2014 The Royal Australian and New Zealand College of Radiologists

gated reconstruction is inherently more noisy than nonECG-gated reconstruction because the data are summed from multiple reconstruction angle sectors from multiple cardiac cycles at a predefined phase point of the ECG cycle. This results in improved temporal resolution but limits the inherent signal in the final image, as smaller angles contain fewer data. Ungated data are reconstructed using a traditional spiral interpolation technique commonly employed in CT reconstruction that disregards ECG correlation. The result is a less noisy image. Electrocardiogram gating was not found to provide any benefit in detecting PE via CTPA examination; however, the relatively small sample size of 50 patients must be considered. The majority of patients with PE in our sample presented with moderate to large emboli; detecting differences in sensitivity between the two techniques in the detection of emboli that are both small in size and low in number may require considerably larger sample sizes. One limitation of the study is that the ECG-gated data set was compared with a virtual ungated data set. The virtual ungated data set had a number of differences from actual ungated CTPA examinations, with the main difference being that the pitch used in our study was relatively slow in comparison with routine ungated CTPA. This technique was used to simulate a non-ECG-gated study so that two comparable studies to review could be obtained while only irradiating the patient once. The influence that a slower pitch has on IQ would be highlighted with further comparison between ECG-gated images and routine ungated CTPA images. Due to the slower pitch, a relatively large dose of iodinated contrast media was necessary to accommodate the slower scan time. Another limitation of our study is that only retrospective ECG gating was analysed in assessing how IQ changed across the chest. Retrospective gating allowed us to obtain an untagged data set to which to compare 7

ND Ardley et al.

our tagged data set. Although there was no increase in the radiation doses administered to patients utilising this technique, a prospectively ECG-gated CTPA examination would potentially render lower doses than those achieved in this study. Prospective gating may increase the potential for presence of stair-step artefact in the z direction of the image where the data are not stitched together well. Prospective gating would also have serious consequences for scan time if performed with 4-cm z-axis coverage and would be better achieved on a CT scanner with greater z-axis coverage. Further expansion of a double-blinded randomised study of prospective ECG-gated and routine CTPA examinations with a bigger patient population may show the benefits of ECG gating with regard to pulmonary vessel clarity. The aim for CTPA protocol development going forward is to achieve the benefits of ECG gating while minimising the inherent image noise and optimising vessel clarity. Increasing the dose while maintaining other parameters would result in less image noise but more radiation exposure for the patient. Alternatively, altering the iterative reconstruction settings could achieve reduction in image noise without impacting on radiation dose to the patient. This higher level of iteration would result in the desired reduction of image noise. Another alternative would be to reconstruct the data using novel modelbased reconstruction techniques (which aim to reduce more noise than the statistics-based iterative reconstruction techniques), and this would result in less image noise and hence achieve the desired results. Another limitation of the study is that all observations of the data sets were only rated visually, and a quantitative assessment was not undertaken. Quantitative analysis of factors such as noise and contrast opacification may add further information in the future.

Conclusion We determined that both vessel clarity and cardiac motion improved in the ECG-tagged reconstructions, while image noise and low-contrast resolution were better in non-ECGtagged reconstructions. Both types of reconstruction data set should be assessed when evaluating CTPA examinations, as they provide complementary information without impacting on radiation dose. Future considerations would be to reconstruct ECG-gated CTPA examinations with less image noise from one data set.

References 1. Goldhaber SZ, Bounameaux H. Pulmonary embolism and deep vein thrombosis. Lancet 2012; 379: 1835–46. 2. Rathbun SW, Raskob GE, Whitsett TL. Sensitivity and specificity of helical computed tomography in the diagnosis of pulmonary embolism: a systematic review. Ann Intern Med 2000; 132: 227–32.

8

3. Naess IA, Christiansen SC, Romundstad P, Cannegieter SC, Rosendaal FR, Hammerstrom J. Incidence and mortality of venous thrombosis: a population-based study. J Thromb Haemost 2007; 5: 692–9. 4. Ocak G, van Stralen KJ, Rosendaal FR et al. Mortality due to pulmonary embolism, myocardial infarction, and stroke among incident dialysis patients. J Thromb Haemost 2012; 10: 2484–93. 5. Anderson DR, Kahn SR, Rodger MA et al. Computed tomographic pulmonary angiography vs ventilation–perfusion lung scanning in patients with suspected pulmonary embolism: a randomized controlled trial. JAMA 2007; 298: 2743–53. 6. Perrier A, Roy PM, Sanchez O et al. Multidetector-row computed tomography in suspected pulmonary embolism. N Engl J Med 2005; 352: 1760–8. 7. Le Gal G, Righini M, Parent F, van Strijen M, Couturaud F. Diagnosis and management of subsegmental pulmonary embolism. J Thromb Haemost 2006; 4: 724–31. 8. Perrier A, Howarth N, Didier D et al. Performance of helical computed tomography in unselected outpatients with suspected pulmonary embolism. Ann Intern Med 2001; 135: 88–97. 9. Perrier A, Roy PM, Aujesky D et al. Diagnosing pulmonary embolism in outpatients with clinical assessment, D-dimer measurement, venous ultrasound, and helical computed tomography: a multicenter management study. Am J Med 2004; 116: 291–9. 10. Shuman WP, Leipsic JA, Busey JM et al. Prospectively ECG gated CT pulmonary angiography versus helical ungated CT pulmonary angiography: impact on cardiac related motion artifacts and patient radiation dose. Eur J Radiol 2012; 81: 2444–9. 11. Schernthaner RE, Stadler A, Beitzke D et al. Dose modulated retrospective ECG-gated versus non-gated 64-row CT angiography of the aorta at the same radiation dose: comparison of motion artifacts, diagnostic confidence and signal-to-noise-ratios. Eur J Radiol 2012; 81: e585–90. 12. d’Agostino AG, Remy-Jardin M, Khalil C et al. Low-dose ECG-gated 64-slices helical CT angiography of the chest: evaluation of image quality in 105 patients. Eur Radiol 2006; 16: 2137–46. 13. Johnson TR, Nikolaou K, Wintersperger BJ et al. ECG-gated 64-MDCT angiography in the differential diagnosis of acute chest pain. AJR Am J Roentgenol 2007; 188: 76–82. 14. Angel E, Yaghmai N, Jude CM et al. Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT. Phys Med Biol 2009; 54: 497–512. 15. Horiguchi J, Kiguchi M, Fujioka C et al. Radiation dose, image quality, stenosis measurement, and CT densitometry using ECG-triggered coronary 64-MDCT angiography: a phantom study. AJR Am J Roentgenol 2008; 190: 315–20.

© 2014 The Royal Australian and New Zealand College of Radiologists

ECG gating and CTPA image quality

16. Shuman WP, Branch KR, May JM et al. Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 2008; 248: 431–7. 17. Grass M, Manzke R, Nielsen T et al. Helical cardiac cone beam reconstruction using retrospective ECG gating. Phys Med Biol 2003; 48: 3069–84.

© 2014 The Royal Australian and New Zealand College of Radiologists

18. Manzke R, Koken P, Hawkes D, Grass M. Helical cardiac cone beam CT reconstruction with large area detectors: a simulation study. Phys Med Biol 2005; 50: 1547–68.

9

Effects of electrocardiogram gating on CT pulmonary angiography image quality.

Pulmonary embolism (PE) is the third most common cause of death from cardiovascular disease. Computed-tomographic pulmonary angiography (CTPA) is an a...