between
the needle
sound needle
shaft
beam decreases echogenicity.
diameter
enhances
and
the ultra-
from 90#{176}, so will Increasing needle
needle
shaft
echo-
genicity as does axial rotation of a nonsymmetric needle (3). Physical alterations consisting of surface scratches of both the needle shaft and obturator in-
crease needle echogenicity creased scattering of sound
by inwaves
Injection
of an echogenic
tune into
the needle
to increase sonically
needle echogenicity active needle-stylet
rendering
ultrasound
it a passive
used
(4). A combina-
by coupling crystal,
the
mix-
has also been
tion has been created needle to a piezoclectnic ceiver
(4).
aim-jelly
the thus me-
(6).
Although it is important to visualize shaft of a biopsy or drainage nec-
die,
it is needle
tip localization
that
1.
is
of the utmost importance. However, for the tip to remain within the beam duning its passage, the shaft must be aligned parallel to the beam. Bobbing the needle-stylet combination, that is, moving the tip back and forth quickly
over
a short
quently
distance,
used
tip’s
maneuver
location.
We
maneuver the needle stylet
echogenicity
is thought microbubbles en aspects
find
in and
out.
of
needle
the
fre-
to confirm
now
the
to be less traumatic tip never moves,
moves
hanced
is the most
the pump
shaft
shaft
alignment allowing
and
of the constant
tip allows shaft tip
and
Dedicated
tip
Coil
Charles M. Anderson, MD, David Saloner, PhD Ralph E. Lee, RT Alexandra Fortner, MD
outen-
the
needle
tion.
U
1.
3.
which has previously adjunct
shaft
and
en-
tip
visualiza-
Sonographic guidance
biopsy
and
needle
in the
drainage:
for Carotid
N
for
4.
versus computed percutaneous
Intervent
Radiol
Ultrasound localizing
1985;
Radiol
the
90.1214, surface nology
arteries,
MR studies,
172.1214 #{149} Magnetic resonance (MR), coils #{149} Magnetic resonance (MR), tech-
EARLY
From the Department of Radiology San Francisco Veterans Administration Center, 4150 Clement St. San Francisco,
Radiology
868
1990; 176:868-872
#{149} Radiology
Received
requested
quests
The sonographic enhancement of
J Clin Ultrasound
needles.
1983;
JW, James percutaneous
EM. Sobiopsy
of small
masses.
1988;
(3 cm
or less)
AJR
6.
McDicken
WN,
MacKenzie
Ultrasonic
WE,
identification
in amniocentesis.
Anderson
of needle
Lancet
1984;
tips
2:198-199.
MR Angiography’
60% of all surgically accessistenoses of the carotid and vertebral arteries occur at sites other than the carotid bifurcation (1). Dopplen ultrasound (US) often cannot depict these sites because of overlying bone that obscures the vessel origins and the distal internal carotid artery. Magnetic resonance (MR) angiography could be useful in screening patients for cervical arterial disease because it has the potential to depict the entire extent of the carotid and vertebral arteries, includ-
April
R, Seidel KJ. and contrast
12-25.
ble
94121.
ultrasound.
24:555-560.
biopsy
segment.
Carotid
tip echo-
real-time
Reading C, Charboneau nographically guided
ing
the
origins.
ly available
suited
(114), Medical CA
January 2; accepted
to C.M.A.
c RSNA,
1990
November 25, May
21, 1990;
1989;
revision
8. Address
revision received
reprint
re-
However, imaging
for MR
imal
arteries.
specifically of-flight
commercial-
are not well of the prox-
coils
angiography We
have
developed
designed for MR angiography
a coil
carotid with
timecoronal
sagittal excitation, and which can be used to screen the entire extent of the carotid and vertebral arteries for stenotic disease. The ideal coil for MR angiography of the carotid arteries must have many ator
tributes.
First,
it must
vide
the
best
ond,
the
sensitive
common
terms:
phantip af-
Berlin:
1
Index
the and
15 1:189-192.
guided
image.
above shaft
J. Needle
with
1989;
Heckmann appearance
T.
sonographic
Springer-Verlag,
PhD
5, Kwela
a study
Invest
5.
needle puncture. Semin 1986; 3:254-263. Otto RCH, Wellwer J.
2.
Bondestain
genicity:
(1),
that
of equal echogenicity of the night needle
puncture 11:265-268.
Wittich GR. tomographic
beam,
A magnetic resonance imaging coil was developed to improve contrast in direct coronal and sagittal time-of-flight carotid angiograms. The sensitive volume of the coil extends from the carotid origins to the siphons. Angiographic contrast can be optimized for an arterial segment of interest by repositioning the coil to minimize presaturation of blood before it enters
atnaumatic
hances
of
easier
with the visualization.
maneuver, described
needles echogenicity
References
increased
to be due to the formation along the inner and of the needle shaft. This
The pump been cursorily is a useful
in that while the
The
2.
Figures 1, 2. (1) Sonogram of two 21-gauge tom target (an olive). (2) Distinct increased ten the needle stylet has been pumped.
cluding
the it is
bral can from should
be
small
signal-to-noise
to
ratio.
volume
must
pro-
Seccoven
sites
of vascular disease, inat the aortic arch, bifurcation, and the siphon. Third, useful if the coil includes the cereocciput, so that dunal sinus blood be saturated to eliminate signal jugular veins. Fourth, the coil the
be
origin
capable
of providing
September
direct
1990
Ca
Cb
Figure 4. 1.
2.
Figures
1, 2. (1) Circuit diagram of carotid coil. Ca 10 zF, Cb = 30 MF, and Cc = 0-25 zF. By making Ca smaller than Cc, a wide latitude of tuning and matching is achieved, although some symmetry among the four circuit segments is lost. Conducting elements are made of copper
strips.
(2)
Carotid
imaging
coil,
consisting
of
nearly
elliptical
loop
that
rises
at
caudal
end to pass over sternum or neck. Grooves along sides of coil allow it to slide longitudinally on tracks. Height of tracks is adjustable with wing nuts. Coil is shown positioned to image proximal
MR angiogram
of aortic
arch
in
the right posterior oblique projection. Placement of coil edge oven upper sternum limits excitation of blood in ascending arch. Pixel intensity of ascending arch is 170 (arbitrary scale), while that of transverse arch is 1,120. Axial saturation band (arrow) has been applied to eliminate venous signal. BC brachiocephalic artery, CC = left common carotid artery, SC left subclavian artery, J = jugular vein.
carotids.
Materials
and
Methods
Computer simulations of the transverse magnetic field were performed
with
a program
authors
written
by one of the
(C.M.A.).
The
coil frame
was made
of oak for
strength and ease of construction. Coil circuitry was enclosed in plastic conduit set into the undersurface of the coil. The circuit consisted of a 19 X 40-
cm elliptical loop with high-voltage fixed capacitors (Voltronics, East Hanover, NJ) distributed in four segments, a. Figure
b. 3.
Contour
plot
of transverse
magnetization,
for
and
carotid
imaging
coil,
ed in midsagittal (a) and coronal (b) sections. Contour levels are equally spaced on an arbitrary scale. Coil has heterogeneous excitation and sensitivity in head and neck but adequate sensitivity along course of carotid artery. Note that is zero at the edge of the coil (an-
row). In this area, tissues
variable
Rochelle,
calculat-
the carotid view. Furthermore, tion by the body
tients
raphy
with
wide
variations
in body
habitus.
Coils available for conventional imagdo not meet all of these requirements. The sensitive range of the head coil does not extend proximally to the aortic arch and may not include the carotid bifurcation in patients with high shoulders. Coronal or sagittal excitation with the body coil pnesaturates blood before it enters the carotid arteries, mesuiting in a loss of image contrast. The body coil also has a relatively poor signal-to-noise ratio. Although receiveonly surface coils improve the signal-tonoise ratio, they have a limited field of ing
Volume
176
#{149} Number
3
presaturation
they rely on excitacoil, again resulting in
of thonacic
We have
found
studies
that
of the
carotid
MR angiogarteries
fail
because of inadequacies in coil design. After examining computer simulations of coil sensitivities calculated for a vanety of proposed coil geometries, we designed and built a dedicated transmitreceive carotid imaging coil that improves contrast in MR angiography. The important features of this coil are that the sensitive volume extends as far as the aomtic arch proximally, and that the
coil
cunately
can
excitation
flow interest.
be quickly
translated
set the proximal
contrast
volume,
which
in the arterial
edge
and
mounted translated
to ac-
of the
optimizes
segment
New
matching
artery
in-
of
(Fig 2). The
This
could
be done
out removing the patient magnet by pulling on the calibrated rod (not shown). platform was designed to existing cushions for the Imaging was performed Siemens Magnetom unit
Federal
coil was
on a track so that it could be longitudinally with respect
to the patient.
blood.
many
(Polyflon,
(Fig 1). The coil was bent anteriorly so that the caudal edge of the coil could pass above the upper sternum while the body of the coil was positioned more posteriorly along the course of
are not excited.
coronal and sagittal images of the neck without saturating blood proximal to the arterial segment of interest. Finally, the coil must be easy to position on pa-
capacitors
NY) for tuning
Republic
with-
from the coil with a The coil be used with head coil. with a 1.5-T (Erlangen,
of Germany).
Carotid
angiograms (2) and maximum-intensity projections (3) were obtained with software provided by Gerhard Laub, PhD, of Siemens Medical Systems (Iselin, NJ). MR angiognams were acquired with three-dimensional Fourier transform, FLASH (fast low-angle shot) sequences with velocity compensation in the section-select and frequency-encod-
Radiology
#{149} 869
id coronal two-dimensional FLASH with velocity compensation in the section-select and frequency-encoding directions, a TR of 22 msec, a TE of 12 msec, a flip angle of 30#{176}, one acquisi-
ing directions and with a repetition time (TR) of 30-40 msec, an echo time (TE) of 7-8 mscc, a flip angle of 20#{176}30#{176}, one acquisition, a 256 X 256 X 32 or 64 matrix, and a voxel size of 1 X 1 X 1 .25 mm. A coronal saturation band
was applied venous grams,
to the occiput
signal. For composite the field of view was
corresponding 1.37
icd
to a voxel
tion,
of 1.37
X
from
mm. Acquisition times van4 to 9 minutes. After selection
number the vessel
maximum-intensity
matrix,
and
An axial
Results The
a pixel
presatura-
transmit-receive
carotid
imaging
tion
coil is an elliptical loop placed in the coronal plane around the patient’s
nal
head
a parasagittal plane arch, then combined
X 1.25
of the minimum that contained
X 256
algo-
rithm.
band was applied to eliminate sigfrom the jugular veins. Sections 4 mm thick were obtained every 2 mm in
angio350 mm,
size
a 256
size of 1 X 1 mm.
to eliminate
projection
maximum-intensity
(Fig 2). The
longitudinally
coil can be translated
on an adjustable
track.
When the coil is moved toward the feet, the caudal edge of the coil is sus-
through the aortic by means of the
of sections of interest,
projections
were
calculated on a 256 X 256 grid. A typical examination consisting of two coronal acquisitions and one sagittal acquisition, with translation of the coil and calculation of preliminary projections, could be accomplished in less than 40 minutes.
Images of the aontic arch were obtamed with a sequence written by one of the authors (D.S.), consisting of rap-
Figure 5. MR angiograms of vertebral antery origins. Caudal edge of coil has been placed over clavicles. (a) Sagittal image of left common carotid (CC) and proximal vertebral
(V)
artery
origin
tery (SC). Arteries rower more distally ration
of blood
bilateral
from
appear
subclavian
because
nan-
of partial
(4). (b) Coronal
subclavian
an-
progressively
image
and proximal
satuof
vertebral
arteries.
a.
a.
b.
Figure 6. Lateral mon carotid artery, applied to occipital reaches bifurcation. tions
of projection.
dal edge
just below
ty does not project 870
b.
#{149} Radiology
c.
ulceration of the sinus (arrow). CC = cornIC internal carotid artery, EC external carotid artery. (b) MR angiogram obtained with head coil. Saturation band was blood to eliminate jugular signal. Contrast is poor because weak inflow leads to partial saturation of blood before it Note that proximal extent of common carotid artery in excitation volume is not seen because it is not included in 5cc-
angiograms
(c) Contrast
bifurcation. over artery
of
carotid
is much
bifurcation.
improved
Submandibular in lateral view
(a)
in MR
Conventional
angiogram
angiogram
obtained
shows
with
signal intensity from fat (arrow) and can be removed from projection
same
shallow
sequence
is the result in frontal
but
with
carotid
coil
of proximity of coil to chin. view by choice of sections.
positioned
This
with
signal
September
cau-
intensi-
1990
pended above the patient’s sternum and the cephalic edge of the coil is positioned at the skull vertex. When the coil is moved toward the chin, the caudal edge of the coil is just above the hyoid
The
bone.
coil is translated
to determine
the proximal edge of the excitation ume, which can be used to optimize flow contrast for a chosen segment the artery. Contrast between blood
volinof and
stationary tissues in time-of-flight imaging is due to a short TR, which mcsuits in saturation of stationary tissues (2). Blood entering the volume is not
in the
volume
before
the arterial segment of A calculated contour transmitted field strength coil is shown in Figure zero at the caudal edge miffing precise placement of the saturation volume. is not Figure 7. Sagittal MR angiogram of carotid siphon (arrow), acquired with thin, three-dimensional sagittal slab (40 mm) medial to the common carotid arteries, to prevent presatunation.
homogeneous
volume
arch
only
and
not
to the
in images
of the
origins
with
was performed twice, once with the coil over the sternum to optimize contrast of proximal vessels, then, without moving the patient, with the coil pulled up to the pa-
is obtained
tient’s chin to image distal vessels. Selected sections from each acquisition were projected, after which the two resulting images
were projected Acquisition
Volume
to form the composite time
176
was
less
#{149} Number
than
3
20
image.
minutes.
the
head
coil.
After
utes,
the lower edge of the carotid below the bifurcation, greaten with
the
same
slab
positioned
medial
to identify lesions in if endartenectomy of a lesion is under consid-
projection.
image
required
illustrating
Acquisition
less than
the
feasibility
20 mmof per-
forming screening of the entire length of the carotid in a reasonably short cxamination time. Discussion The carotid coil shown in Figure 2 can be used to image all segments of the carotid artery with direct coronal or
of
as-
of the
adjustment
midline
mum-intensity
vertebral arteries (Fig 5), with strong contrast. If the coil is pulled back farther, so that the caudal edge is near the chin, presaturation of the carotid bifurcation is minimized. In Figure 6, a convcntional angiognam of the bifurcation is compared with MR angiognams obtuned with the head and carotid coils. In this patient, poor inflow results in reduced contrast on images obtained Figure 8. Coronal MR angiogram of carotid arteries with large field of view used to display entire course of vessels. Sequence
An MR angiognam of the carotid siphon (Fig 7) was obtained with a thin
of this
cending arch or the heart. The sharp definition of the saturated volume is achieved by positioning the coil over the upper sternum. The neck vessels are clearly depicted. If the coil is pulled slightly cephalad, so that the caudal edge is over the clayides, one can minimize presaturation of the subclavian arteries. This position results
in
tissue.
two
the aortic arch in the right posterior oblique projection. This image demonstrates that excitation is limited to the transverse
resulted
two
of
imaging. angiogram
have
of stationary
emation. Figure 8 is a composite of coronal images obtained with the carotid coil first over the sternum, then pulled up to the chin. The first image demonstrates strong contrast in the proximal vessels, while the second shows strong contrast distally. The projection images were then combined in a maxi-
the head and neck. Rather, it delivers the largest flip angles and has the greatest sensitivity along the anterior neck. It would not be adequate for spin-echo imaging of the neck, which requires precise 90#{176} and 180#{176} flip angles. It is suitable, however, for gnadient-refocused angiographic Figure 4 shows an MR
would
it is important this distribution more proximal
reaching
the
but this suppression
to the common carotid arteries, so that blood was not excited proximal to the intrapetnous segment. The siphons and intracranial circulation arc generally not amenable to surgical connection, but
interest. plot of the of the carotid 3. Excitation is of the coil, perof the edge The coil field
oven
less
sagittal
yet saturated and appears bright. If the blood must follow a long course in the excited volume or is moving so slowly that it is subjected to many pulses, it will become saturated, resulting in poor contrast. Saturation of blood may be minimized by careful placement of the edge of the excitation volume so that the blood travels only a short distance
angle,
of
coil to just contrast
sequence.
The path length that the blood traversed in the common carotid artery before reaching the bifurcation was 9 cm in the head coil but only 4 cm in the carotid coil. Saturation of blood in the head coil could have been reduced by choosing a longer TR or a smaller flip
sagittal cause
time-of-flight of the small
size
sequences. Beof the coil, sig-
nal-to-noise ratio is adequate, yielding images with good contrast. The coil overcomes common reasons for failure of MR angiognaphy encountened with of conventional coils. First, the unique design of the carotid coil allows imaging of artery bifurca-
two
use
tions body
and origins regardless habitus. This is not
of patient always possi-
ble with the transmit-receive head coil, which may not extend proximally enough to include even the bifuncation. Second, coronal on sagittal excitation with the carotid coil does not mesuit in saturation of thoracic blood. Saturation
is a problem
when
the
body
coil is used to excite and a receive-only surface neck coil to detect. This prob1cm is solved with the carotid coil by positioning
the
caudal
edge
mal to the arterial segment Other investigators have specialized pulse sequences tions
to the
problem
just
proxi-
of interest. proposed as solu-
of proximal
satu-
ration. Two of these arc sequential axial thin-section acquisition followed by coronal reconstruction (5) and axial slab
excitation
followed
by coronal
sec-
tion readout (6). The carotid coil described here could be used together with these pulse sequences but does not require them to eliminate presaturation. Direct coronal and sagittal acquisitions en in-plane
have the resolution
advantage than
does
Radiology
of greatse#{149} 871
quential
thin-section
Furthermore,
reformation.
they
have
than both thin-section and excitation-coronal readout thereby
minimizing
ous
shorter
axial slab sequences,
turbulent
This may be disadvantageous if strong subcutaneous fat signal obscures underlying vessels in projection images. Because the coil also transmits, tissues
to the coil are more saturated, offsetting this disadvantage. one may “edit out” subcutane-
partially
Further,
Monorail System of the Greenfield Diana F. Guthaner, MD James 0. Wyatt, MD John Thomas Mehigan, Allan M. Wright, MD Jerome F Breen, MD Lewis Wexler, MD
terms:
plications,
nae cavae,
al-
clipping
projection approach
the coil wider,
calculawould
1.
thereby
in-
obese. The extended range of the carotid coil highlights one of the key advantages of MR angiography. It can be used to visualize carotid origins and phons
and
the
vertebrobasilar
which
are not accessible
US. A screening MR ing bilateral origins,
1990; 176:872-874
angiognam bifurcations,
includsi-
ning.
of a Gncenficld
infe-
malpositioning
of the
filter,
or the
sub-
sequent migration on embolization of the filter. Because the small hooks at the feet of the filter legs are sharp and pierce the wall of the vena cava, mcor repositioning
of a malposi-
tioned filter is difficult. Several recent case reports in the literature have descnibed the percutaneous retrieval of a Greenfield filter (1-7). The technique we have developed is simple and also allows repositioning of the filter in some
cases.
Case
Reports
the
Departments
Radiology,
Mayo 13; revision
Medical
of Diagnostic
Center,
872
RSNA,
1990
#{149} Radiology
Pasteur
(D.F.G.,
LW.)
Dr. Rrn H2321,
pulmonary
presence
and
Surgery
Stanford,
embolus
of a Mobin-Uddin
(J.O.W.,
J.T.M.),
CA 94305; Department
Samaritan Hospital, San Jose, Calif (A.M.W.); and Department of Diagnostic Clinic, Rochester, Minn (J.F.B.). Received January 12, 1990; revision requested received April 23; accepted April 26. J.F.B. supported in part by National Heart
March Lung Blood Institute Cardiovascular print requests to D.F.G. C
Radiology 300
a new the
Radiology
Research
Training
203:961-968. MT, Ruggieni
Three-dimensional aging of the carotid clinical experience.
(volume)
Rossnick
grant
HLO
bifurcation:
Radiology
S, Laub
PM,
C, Braeckle
dimensional
display
Proceedings
of the
of blood IEEE
et al.
gradient-echo impreliminary 1989; 171:801-806. R, et al.
Three-
vessels
in MRI.
Computers
in Cardiol-
ogy Conference, Boston, 1986. New York: IEEE Publication Services, 1986; 193-196. Anderson CM, Saloner D, Tsuruda JS, et al. Artifacts in maximum intensity projection display of MR angiograms. AJR 1990; 154: 623629. Keller
PJ, Drayer
BP,
Fram
EK, et al.
MR
angi-
ography with two-dimensional acquisition and three-dimensional display. Radiology 1989; 173:527-532. Masaryk TJ, Laub G, Modic T, Ruggieni P. Ross Js. 3DFT MR angiography of the carotid bia comparison
of time-of-flight
tech-
niques (abstr). In: Book of abstracts: Society of Magnetic Resonance in Medicine 1989. Vol 1. Berkeley, Calif: Society of Magnetic Resonance 1989;
164.
Repositioning
vena caval filter (Mcdi-tech/ Boston Scientific, Watertown, Mass) under fluoroscopic guidance is now a standard procedure performed with percutaneous techniques. Occasionally, optimal positioning of the filter is not achieved because of premature ejection of the filter from the carrier device,
despite
tions. JAMA 1968; Masaryk TJ, Modic
in Medicine,
HE placement
tnieval
Hass WK, Fields WS, North RR, et al. Joint study of extracranial arterial occlusion. II. Arteniography, techniques, sites, and complica-
furcation:
U
developed
From
6.
phons, and vertebral arteries can be accomplished in less than an hour, which compares favorably with US scan-
33-year-old woman with pulmonary arterial hypertension secondany to chronic pulmonary emboli
1
5.
Doppler
Case 1.-A
Stanford University of Radiology, Good
3.
si-
nor
MD
2.
4.
system,
with
References
be
creasing the distance between coil and subcutaneous fat. A wider coil would also permit excitation deep to the sternum in patients who arc unusually
T
Interventional procedure, corn949.458 #{149} Venae cavae, filters #{149}Veinterventional procedures, 949.1299
Radiology
a regional
for Percutaneous Vena Caval Filter’
The authors describe a technique for removing or repositioning a malpositioned Greenfield inferior vena caval filter. A “monorail” system was used, in which a wire was passed from the femoral vein through the apical hole in the filter and out the internal jugular vein; the wire was held taut from above and below and thus facilitated repositioning or removal of the filter. The technique was used successfully in two cases. Index
with
before the alternative
to construct
dephas-
ing signal loss. In our experience, axial slab excitation-coronal readout acquisition works well for visualizing the distal carotid arteries but very poorly for the carotid origins, even with a low-lying surface coil. The close-fitting carotid coil enables detection of subcutaneous tissues near the coil elements with great sensitivity.
adjacent
tissues
gonithm tion. An
TEa
7425.
Address
re-
umbrella placed in the infranenal
fenion
vena
many years previously inferior vena cava. In-
cavography
occlusion
of the
demonstrated
inferior
vena
cava
dis-
tal to the umbrella, with development of large lumbar and gonadal collateral veins draining into the cava and renal veins and bypassing the occlusion. In the operating suite, a Greenfield filter was placed at the level of T-12, above the renal veins, in a position considcred satisfactory in this patient. After surgery, the filter was found to have
migrated. Twenty-four hours after placement, the filter had moved to the level of L-1 and had an oblique orientation; at least 2 cm of the filter extended into
the
right
gonadal
vein
Because the filter tive in this oblique tioning
was
(Fig
la).
would be ineffecorientation, neposi-
indicated.
The
right
inter-
nal jugular vein was cannulated and a balloon catheter partially inflated in the inferior vena cava above the level of the renal veins to prevent embolization of the filter into the right atrium. A 0.035-inch Amplatz wire (Cook, Bloomington, md) was introduced from the left femoral vein, and, with the assistance of a Benenstein catheten-with an angle near its tip that facilitates directional control-traversed the lumbar collateral vessel into the infcmion vena cava above the occlusion.
The
wine
was
then
through
the
ten. The
balloon
placed
wine,
with
and
hole
manipulated in the
catheter
apex
was
a doubled-over
the Amplatz
of the
then
fil-
mc-
exchange
wire
protruding
September
1990