Cardiac Robert

R.

Edelman,

MD

Coronary

#{149} Warren

acquired

per

segment.

By using

overlapping 4-mm-thick sections, the coronary arteries were routinely depicted from the coronary ostia distally at MR in healthy subjects. Ultrafast MR angiography of the coronary arteries is feasible with use of a standard body coil. This technique offers considerable potential as an investigational tool and, with further development, may become ful imaging application. Index 54.1214 studies,

Manning,

Arteries:

The authors describe a method for performance of ultrafast magnetic resonance (MR) angiography of coronary arteries with a standard clinical MR system and a body coil. Each image was obtained within a single breath hold by using an electrocardiography-gated, segmented, ultrafast, gradient-echo pulse sequence with an incremental excitation flip angle for the eight phase-encoding steps

J.

a clinically

use-

terms: Coronary vessels, MR studies, #{149} Magnetic resonance (MR), vascular 54.1214 1991;

181:641-643

M

raphy

in several

clinical

ing screening rysms (1) and the extracranial

PhD

(MR) being

applications,

system,

teries (3). Because

of the

giography

in these

includ-

and

success

renal

of MR

other

of

aran-

applications,

techniques.

artery spatial

Moreover,

coro-

flow is phasic (4), and resolution is needed to

The high achievement

MD,

PhD

in rapid

sequence,

constituting

one segment (6). One segment quired in diastole during each

was acheartbeat

by using

triggering,

electrocardiographic

which breath

permitted hold.

imaging

within

Initially, findings with several variations of the pulse-sequence parameters were compared. These variations included the absence or use of flow-compensated gradients with the concomitant alteration in echo times, the use of a fixed versus an incremental excitation flip angle (7), and

the use of prepulses background

signal

comparisons

were

for suppression (8). After made,

of

the initial

most

imaging

was performed with a flow-compensated, fast-imaging-with-steady-state-precession sequence, with repetition time of 14 msec, echo time of 8 msec, section thickness of 4

separate the signal intensities of arteries from those of surrounding fat, myocardium, and blood in cardiac

chambers. requires

acquired

seconds, a single

vivo coronary artery imaging. Because respiratory motion causes blurring, cardiac gating is necessary; therefore, image acquisition within a breath hold is impossible with use of nary high

Paulin,

with 16 segments interleaved in order to complete a 128 x 256 matrix in 16 heartbeats. Imaging time was typically 12-16

we sought to image the coronary arteries, which typically are similar in caliber to intracranial vessels. There are several obvious challenges for in

standard

#{149} Sven

The imaging technique consisted of a two-dimensional, flow-compensated, gradient-echo sequence modified in such a way that eight phase-encoding steps were

angiogused

for intracranial aneudetection of stenoses carotid arteries (2),

vertebrobasilar

mm,

spatial resolution of an adequate

field

of 128

of view

x 256

(16

of 270

x 270

segments

mm,

of eight

matrix phase-

gradient-echo acquisitions (5). We have modified this approach in human studies to permit acquisition of electrocardiography-gated coronary artery images in a body coil within a

encoding steps), with an incremental flip angle series of 10#{176}, o 350 50#{176}, 50#{176}, 50#{176}, 50#{176}, 50#{176}. A scout view was first obtained in the coronal plane, after which transverse images were obtained at the level of the coronary ostia. Sequential transverse images were obtained with 1-mm overlap. In some cases, oblique views were obtained from the transverse images once the proximal coronary artery had been identified. Twenty-two healthy adult volunteers were studied (16 men and six women,

single

aged

of signal-to-noise

difficult

with

onary

ratio,

a body

which

coil.

of in vivo

with

breath

rodent

segmented,

hold.

The

hearts

ultrafast,

method

was

applied in a series of healthy subjects to determine the feasibility of imaging human coronary arteries. SUBJECTS

AND

as a radio-frequency

23-68

years).

formed within Israel Hospital

transmitter

gradient amplitude ramp time to peak

and rewas 10 amplitude

All

studies

were

per-

the guidelines of the Beth Committee on Clinical In-

vestigations.

METHODS

Images were obtained with a 1.5-T whole-body imaging system (Magnetom SP; Siemens Medical Systems, Iselin, NJ) with a circularly polarized body coil used ceiver. Peak mT/rn, with of I msec.

is

published articles have that imaging of the cor-

arteries

is feasible

the Departments of Radiology (R.R.E., D.B., S.P.) and Medicine (Cardiovascular Division) (W.J.M.), Charles A. Dana Research Institute and the Harvard Thorndike Laboratory of the Beth Israel Hospital and Harvard Medical School, Boston. Received May 3, 1991; revision requested June 10; revision received July 8; accepted July 16. Address reprint requests to R.R.E., Department of Radiology, Beth Israel Hospital, 330 Brookline Aye, Boston, MA 02215. C RSNA, 1991 See also the editorial by Caputo (pp 629-630) in this issue.

Burstein,

MR Anglography’

resonance is currently

AGNETIC

Previously demonstrated

1From

#{149} Deborah

Breath-hold

level Radiology

MD

Radiology

RESULTS Initial

studies

consisted

of a com-

parison of findings with use of various pulse-sequence parameters. In all cases, a fast-imaging-with-steadystate-precession sequence was used

rather quence

than

a fast-low-angle-shot

to avoid

the

occurrence

seof

artifacts

at high

excitation

flip

angles

(9). Use of an 8-msec echo flow compensation yielded

time with better im-

age

echo

quality

than

of 6 msec In general,

use

of an

time

without flow compensation. use of the incremental

ex-

citation flip angle series gave better signal-to-noise ratio of coronary artery flow than did use of a fixed flip angle of 15#{176} or 25#{176} and thereby per-

mitted Use

use

of a smaller

of larger

fixed

field

flip

of view.

angles

up

to

tended to suppress the signal from in-plane flow. Use of a prepulse to 500

suppress necessary

stationary to obtain

of coronary not

artery

generally

spins was not sufficient contrast

flow,

and

it was

By using

the

segmented

fast-imag-

ing-with-steady-state-precession quence and the incremental

seexcitation

flip angle series, portions mal left and right coronary were

seen

routinely

The resting to 84 beats imaging left and

well

varied from 46 resulting in an

of 11-21 coronary

delineated

of the proxiarteries

seconds. arteries

by using

The were

a transverse

plane of section (Figs 1, 2). Oblique images were also obtained in several cases (Fig 3) once the proximal coronary artery had been identified.

DISCUSSION We have

demonstrated

the feasibility

of breath-hold MR angiography of the coronary arteries in healthy subjects with use of a standard MR imaging system and body coil. The method proved robust, allowing portions of the proxi-

mal coronary

arteries

to be seen

in all

subjects. The method allowed for acquisition of overlapping 4-mm-thick sections within a brief period of 10-20 mmutes. A major challenge for coronary-flow imaging is to obtain sufficient signal-tonoise ratio to support high-resolution imaging within a body coil. Most highresolution MR angiography has been performed with use of specialized head or surface coils that give higher sensitivity than a body coil. Use of an incremental flip angle has been proposed previously to improve signal-to-noise ratio in fast gradient-echo imaging (7). As applied to flow imaging with our approach, use of an incremental flip angle series permitted application of a large flip angle at the central phase-encoding step without much saturation of inflowing spins, which greatly increased signal-to-noise ratio and flow contrast. This was not the case with use of a constant large flip angle, although centric reordering of the phase-encoding steps might improve the results. Use of an

642

#{149} Radiology

Figure

1.

tamed image, arrow),

with with left

arrow).

Four-millimeter-thick,

breath-hold axial MR images of a healthy subject a 270 x 270-mm field of view, a 12-second imaging

a 128 x 256 matrix, 16 segments of eight phase-encoding anterior descending artery (white

(b) Right

coronary

artery

were

ob-

time per steps each. (a) Left main artery (thick black arrow), and left circumflex artery (thin black

(arrow).

in all subjects.

heart rates per minute,

time right

b.

a.

implemented.

incremental flip angle and k-space segmentatiom has the potential to cause loss of spatial resolution as a result of distortions in the point-spread function. Nonetheless, the resolution was adequate for showing the major coronary arteries. Images obtained with use of this technique were free of ghost artifacts and of blurring from respiratory motion. Substantial improvements in image quality can be expected with faster data acquisition (eg, with use of more efficient pulse sequences and faster gradients). A problem that is less easily overcome is variation in cardiac position from one heartbeat to the next, which causes blurring (10). Potentially, performance of a single-shot technique, such as echo-planar imaging, could overcome this problem, but further development is needed before adequate spatial resolution and signal-to-noise ratio can be obtained by this means. Recently, interleaved spiral scanning, a variation on echo-planar imaging, has shown promise for highresolution imaging of the heart (11). Because interleaved spiral scanning results in fewer saturation effects than gradient-echo methods and is potentially faster, it might prove superior for coronary artery imaging, and further study is needed. Several other techniques have been used to image the coronary arteries. Standard electrocardiography-gated, spin-echo, and cine images have only occasionally depicted portions of the coronary arteries, and these images are not adequate for detailed evaluation (12,13). Three-dimensional acquisition methods also have potential for imaging of the coronary arteries (14). Both the two- and three-dimensional approaches have potential advantages and disad-

vantages. Three-dimensional acquisitions permit use of thinner sections and shorter echo time, thereby minimizing partial volume averaging and flow-related dephasing. With three-dimensional approaches, saturation effects are more severe, resulting in worse flow contrast. The imaging time necessary for three-dimensional acquisitions of more than 1 minute with cardiac gating precludes use of breath holding; therefore, significant blurring from respiratory motion may occur. Further work is needed to determine the relative benefits of the two- and three-dimensional approaches, but one might expect that the methods will be complementary, as they are for MR angiography of the head and neck. Subtraction methods also have potential value for coronary artery imaging. One approach has achieved good background suppression by subtraction of a pair of images, one of which has a spatially localized inversion pulse applied to the aortic root to tag the coronary inflow (15). A major advantage of this method is that a projection image of the coronary arteries is obtained. However, despite initial enthusiasm (16), clinical results with projection angiography have proved disappointing when the technique is applied to other small yessels, such as the carotid arteries. Potential drawbacks to use of the method for coronary artery imaging are the need for excessively long breath-holding periods and a surface coil to achieve adequate signal-to-noise ratio and the possibility of misregistration of small vessels resulting in imperfect subtraction. We have demonstrated the feasibility of performance of breath-hold MR angiography for rapid imaging of the corDecember

1991

onary minimal

arteries. The technique operator intervention

be initiated

with

requires and can

use of a standard

MR

imaging system and body coil. Future efforts will be directed toward achievement of further improvements in spatial resolution and signal-to-noise ratio, as

well

as use of shorter

echo

data

acquisition,

flow

tion.

With

method

sis.

further could

in patients

and

development,

have

with

time,

clinical

coronary

faster

quantifica-

the usefulness

artery

steno-

U

References 1.

Ruggieri PM, Laub GA, Masaryk TJ, Modic MT. Intracranial circulation: pulse-sequence considerations in three-dimensional (volume) MR angiography. Radiology 1989; 171:785-

2.

Masaryk

TJ, Modic

MT.

Ruggieri

PM,

et al.

Three-dimensional (volume) gradient-echo imaging of the carotid bifurcation: preliminary clinical experience. Radiology 1989; 171 :8013.

4. 5.

6.

7.

Kent KC, Edelman RR, Kim D, Steinman TI, Porter DH, Skiliman JJ. Magnetic resonance imaging: a reliable test for the evaluation of proximal atherosclerotic renal arterial stenosis. J Vasc Surg 1991; 13:311-318. Berne RM, Levy MN. Cardiovascular physiology. 5th ed. St Louis: Mosby, 1986; 200. Burstein

Society

8.

d.

C.

Figure left

main

cumflex

2.

Series

of breath-hold

coronary

artery

(arrow).

coronary

artery

(arrow).

axial MR images (b) Left

anterior

show

left coronary

descending

artery

circulation. (arrow).

(a) Ostium (c, d) Left

cir-

of 9.

10.

11. 12.

13. 14.

15.

D.

MR

imaging

of coronary

artery

flow in isolated and in vivo hearts. JMRI 1991; 1:337-346. Atkinson DJ, Edelman RR. Cineangiography of the heart in a single breath hold with a segmented turboFLASH sequence. Radiology 1991; 178:357-360. Stehling MK. Optimised (incremented) rfangle gradient echo imaging: ORANGE (abstr). In: Book of abstracts: Society of Magnetic Resonance in Medicine, 1990. Berkeley, Calif: of Magnetic

Resonance

in Medicine,

1990; 459. Edelman RR, Chien D, Atkinson DJ, Sandstrom J. Fast time-of-flight MR angiography with improved background suppression. Radiology 1991; 179:867-870. Crawley AP, Wood ML, Henkelman RM. Elimination of transverse coherences in FLASH MRI. Magn Reson Med 1988; 8:248Pearlman JD, Weisskoff RM, Hunter MS. Cohen MS. Brady TJ. Cardiac variance images from single-shot MR imaging (abstr). JMRI 1991; 1:181. Macovski A, Meyer CH, Noll DC. Highspeed imaging with time-varying gradients (abstr). JMRI 1991; 1:138. Paulin 5, von Schulthess GK, Fossel E, Krayenbuehl HP. MR imagine of the aortic root and proximal coronary artenes. AJR 1987; 148:665670. Alfidi RJ, Masaryk TJ, Haacke EM, et al. MR angiography of peripheral, carotid, and coronary arteries. AIR 1987; 149:1097-1109. Paschal CB, Haacke EM, Adler LP, et al. High resolution 3D cardiac imaging (abstr). In: Book of abstracts: Society ofMagnetic Resonance in Medicine 1990. Berkeley, Calif: Sodety of Magnetic Resonance in Medicine, 1990; 278. Wang SJ, Hu BS, Macovski A, Nishimura DC. Coronary angiography using fast selective inversion recovery. Magn Reson Med 1991; 18: 417-423.

16.

Figure 3. Oblique views in another healthy subject were obtained with 15-second time per image. (a) View through axis of the left main coronary artery (arrow) and orthogonal to view in a show cross section of left main artery (arrow).

Volume

181

#{149} Number

3

Wedeen VJ, Meuli RA, Edelman RR, et al. Projective imaging of pulsatile flow with magnetic resonance. Science 1985; 230:946-948.

imaging

(b) view

Railinltwv

M

#{149}

Coronary arteries: breath-hold MR angiography.

The authors describe a method for performance of ultrafast magnetic resonance (MR) angiography of coronary arteries with a standard clinical MR system...
680KB Sizes 0 Downloads 0 Views