IEEE TRANSACTIONS 01\ BIOMEDICAL EI\GINEERI\lG. VOL

Communications

37. NO

II. NOYE:vIBER 1990

_ _______________

Computer-Controlled Dynamic Phantom for Ultrasound Hyperthermia Studies JON ZAERR, ROBERT B. ROEMER, AND KULLERVO HYNYNEN

Abstract-A

It consists of four

in vitro

preserved canine kidneys perfused with an

ethanol preparation by

a sin gle

pump, plus four computer-con­

trolled valves and four flow meters. The flow rate to each kidney is computer-controlled, giving flexibility in the types of spatial and tem­ poral flow

devices [9], for developing and testing inverse algorithms [10], and for reproducin g the large scale physiological hlood perfusion re­ sponses of organs [1 I]. Two general approaches are available for

developing such phantoms: constructing a mechanical system from perfusable matrices such as sponges, andlor systems of tubes [7],

[12]; or using a preserved organ for in vitro studies [9], [13], [14]. dynamic phantom sy.tem for use in evaluating hyper­

thermia heating equipment has been designed, constructed, and tested,

80%

III 'j

patterns possible.

Examples are given for both step and ramp

changes in flow, and for a temperature dependent flow which has been used to simulate physiological responses tn elevated temper atures ,

The latter approach has the advantage of possessing a physiologi­ cally more realistic perfusion pattern-at least but

for the organ itself,

not necessarily for tumors.

This communication describes a dynamic phantom system using this latter approach, which has been developed [151 for system test­ ing as described above, rather than for accurately producing actual treatment temperatures. It is based on the process developed by Holmes et al. [13] to preserve in vitro can ine kidneys in alcohol for repeated use. The present study extends the previous applica­ tions of this approach

to include multiple kidneys, with the alcohol

perfusion to each individual kidney controlled by a computer. (The

INTRODUCTION

During hyperthermia treatments the tissue temperature achieved

flow pattern within each kidney can have considerable spatial vari­ ation, and these are not controllable in the present system.) These

depen ds on the magnitude and distribution of the deposited power, thermal conduction, and blood flow. In order to evaluate thc clin­

added features give flexibility

ical heating characteristics of hyperthermia systems, all of the above

both the spatial and temporal distribution of the flow rate.

factors have

uating the power deposition distributions of electromagnetic appli­ cators [I]. [2]. Similarly, water baths and static phantoms can pro­ useful

information

regardng

the

ultrasound

intensity

distributions needed for hyperthermia applications [3]-[6], These nonperfused phantoms have successfully provided absorbed power patterns for simpl e tissues. However. the use of "dynamic" phan­

toms which attempt to simulate the blood perfusion characteristics of tissues has not been so successfuL The long range goal of dy­ namic phantoms is to combine the above static phantom data with simulated patient blood perfusion responses so that patients' tem­ perature distribution responses can be predicted prio r to a treatment for planning purposes. This goal has not been remotely approached due to the com­ plexity of reproducing both the tumor's and the normal tissues' blood perfusion responses. However. even though this long range goal remains remote, such dynamic phantoms can provide other useful functions

for hyperthermia in situations where approximate

tissue temperature profiles are adequate, and other system testing

and evaluation considerations provide the primary experimental goal.

For example, Baisch et at. [7] have used a simple set of parallel tubes embedded in a solid matrix to experimentally reproduce con­ ditions under which the bioheat transfer equation is valid. A second such goal

is

to pro vide a basis for evaluating feedback control sys­

tems under conditions where blood flow can change. This method for experimentally

gives a testing the stability and robustness of

such controllers under approximate clinical conditions, An exam­ ple of such an application is that

Lin et ai, [8] who have used single

preserved in vitro canine kidneys for hyperthermia controller stud­ ies. Other possihle applications include providing a well instru­ mented experimental base for evaluating and optimizing heating Manusc ript received September I 1.1989; revised March 4.1990. This work was supported in part by NCI Grants CA 36428 and CA 33922. The authors are with the Department of Radiation Oncology and the Department of Aerospace and Mechanical Engineering. University � of Arizona, Tucson, AZ 85721.

IEEE

and simu­

to be considered, Static phantoms have proven their MATERIALS AND METHODS

worth in providing a clinically relevant experimental base for eval­

vide

in designing experiments

lating physiological responses since the user has some control of

Log Number 9038594.

Dynamic Phantom Description

Following is a summary of the system used in this study. The details of all of the system hardware, software, and calibration data are provided by Zaerr [15]. An Apple II P lus computer controlled the flow rates to, and recorded the temperatures in four kidneys mounted in a sonication chamber, as shown in Figs.

sonication chamber

I and 2. The is an 80% alcohol-filled Plexiglas container

with a thin mylar membrane on the bottom to transmit ultrasound. A variable speed gear pump (Micro Pump, In c , ) forced the fluid through the system, and the flow rates to each kidney were set by stepper motor-controlled valves and measured with rotating disk flow meters (Digital Precision Flowmeters, Inc.). The differential pressure across the gear pump head was measured with a pressure transducer (SenSym Model SCX 15DNC). The analog voltage out­ put of the transducer (0,25-9.25 V) was appl ied to one of the channels

of an analog-to-digital convener card on the Apple Com­

puter. An external interface was designed and constructed which: provided power to and multiplexed the drivin g signals to the step­ per motor driver; conditioned the signal from the flow meters; and amplified

and multiplexed the temperature signals. Communication

between the computer and the external interface was provided through a 16-b parallel interface and a 16-channel, 12 b ADC. As

noted earlier, the kidneys were preserved using the alcohol

fixation process developed by Holmes et al. [13], The last stage of

rv ation process involved perfusing the organ with 80%

their prese

ethanol, which was the working fluid of the dynamic phantom pre­ sented here. Using this liquid minimized biological degradation of the phantom while maintaining ease of operation since the normally suggested rehydration

process was eliminated. Thc ultrasound ah­ low compared

sorption in the alcohol-lilled phantom is relatively

to other, normal tissues (e,g" muscle), as is the absorption in nor­ mal kidneys with their high water contents. We have

no reliable

quantitative data on comparative values. The flow control temperatures in the phantom were measured with 16. bare-wire, manganan-constantan welded junction ther­ mocouples (wire diameter 50 I'm) referenced to an AD590JH in­ tegrated circuit temperature sensor. These flow control thermocou-

0018-9294190/1100-1115$0 1.00

@ 1990 IEEE

1116

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 37. NO.

_______________

APPL�

£}!!!�!!.!�J:.t::.�"_!!'?¥__

II PLUS

COMPUTER

I.

NOVEMBER 1990

50.-------�------ --�

___________ _

FLOW RATES

" /

,

, .

·,1'

is X

Fig. 3.

,

,,/',1''''

,@ / ,"KDI NEYl

z 10 o i= U - 10 W a:: -30

DISPLAY OF FLOW RATES. TEMPERATIJRES

,

KIDNEY 3

30

USER INPUT

, ,

.'

-30 -10 10 V-DIRECTION

50

30 [MM]

Thermocouple and scan locations. Kidneys one and three con­

tained thermocouple, as indicated by the X's. The ultrasound focus was scanned alternately on the two octagonal paths.

In each mode, an internal feedback control system ensure that the flow rate called for

by

produced. First, at flow rates above 35 Fig. I. Block diagram of the dyn amic phantom. The Apple computer con­ trolled the flow rates to the four kidneys in the sonication chamber and m easured the temperatures and flow rates via the external interface. The hyperthermia system measured additional temperatures and provided the heating power.

from

was

used to

the software was accurately

mL /min,

the output signals

the flow meters were used directly in a proportional control

algorithm to obtain the specified flow rate. In this mode, the valve position was varied to

force the measured flow rate to track the software. Second, at flow rates optical disks in the flow meters did nol spin

desired flow rate demanded by the

below 35 mL / min the

evenly and thus yielded erroneous measurements. Therefore, the valves were set based on the nearly linear dependence of flow

Flow

Stepper

Meters

Motors

rate

on the pressure rise and valve position, which was approximated

as: V = V,(O/O,) (P/P,.) where V = flow rate in mL / min, II = valve position (zero is closed), P = pressure, and c indicates a

Flow Control Valves

Thermocouples

4

Sonication Chamber

calibration point. Measurements from the /low meters at the lowest measurable rates were used for V", lip and Pc. The !low rate esti­ mates using this technique matched the actual !low rates to within 2 mL /min. Using these two valve positioning schemes, the largest expected error was 4 mL /min at a flow rate of 150 mL / min (the expected upper limit of measurement), the largest expected error

decreased to 1.2 mL / min at a flow rate of 5 mL /min [for tion details see Zaerr (15)]. Hype rt her mia

Differential

r---"'J Pressure Transducer

Thermocouples

calibra­

Hyperthermia System To heat the dynamic phantom the hyperthennia treatment system described by Hynynen et al. [16] was used. The hyperthermia ther­

Return line

mocouples were distributed in the phantom as shown in Fig. 3. The

Variable Speed

Gear Pump

kidney edges were located with ultrasonic B-scans from the hy­

Fig. 2. Schematic of the dynamic phantom. 80% ethanol was pumped with a gear pump through four flow meters and valves into the four kidneys. The valves were operated by the computer controlled stepper motors. The computer also read the flow meters and the pressure rise across the pump.

perthermia system, and the thermocouples were located with the transducers as descrihed hy Hynynen et al. (16). the X's represent thermocouple locations as measured to within 2 mm . A focused ultrasound transducer (frequency 1 MHz, diameter

treatment

=

=

13 em, radius of curvature = 25 em, acoustic efficiency = 70%) was scanned in a translational manner through the phantom in two concetric octagons as shown in Fig. 3, with a total scan time

pies were connected to an analog multiiplexer in the external interface and amplified by a gain of 1000. This signal was directed

to the input of the ADC in the Apple computer. The thermocouples were calibrated to within ±0.25°C, In order to extensively monitor the temperature field in the phantom, additional multijunetion ther­ mocouples (outside diameter 0.7 mm fused silica tuhes), denoted as "hyperthermia thermocouples" in Fig.

I, were inserted into the

kidneys. These hyperthermia thermocouples were connected to an

ical power of 15 W was provided by the heating system. Finally, Hynynen et af. [17] have reported results with four overlapping

beams a t various perfusion levels,

flow rates in

Basic program was written to control the kidney either of two modes: programmed flow, in which a

showing that the heated region,

defined as th e area within which the

temperature

was 75% of the

peak temperature, extends about 5 mm beyond the perimeter of the scan path. Thus, the heated region was approximated as including a 5 mm margin around the outer octagonal scan path.

ex isting scanned, focused ultrasound hyperthermia heating system [16]. An Applesoft

of

9 s. In all tests described in this communication, a constant acoust­

RESULTS-DISCUSSION Several test results are presented here to demonstrate the capa­ bility of this system:

I) programmed flow rates with constant power

list of flow rates and their time durations was preprogrammed, and

input (PF/CP); and 2)

the computer simply stepped through this list; and temperature con­

stant power input (TCF/CP). In the first programmed flow test il­

trolled flow, in which the total flow to each kidney increased or

lustrated, the temperature response of the phantom to step changes in perfusion was investigated. One su ch response is shown in Fig.

decreased as a programmable function of the kidney temperature as measured by the above flow control thermocouples.

temperature

controlled flow rates with con­

4 for kidneys one and three. The th ermocouples shown represent

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 37. NO.

IT

:;... w c:: :::J

44 42

........ .. ._...... ....... _.. .... ...__ .

_

.

U



:;... W a: :J

� \J \

40:r-'\

!;( c::

38'



34

W Q. :::;:

15 MLIMIN...

..��"

H .r"."

1117

II. NOVEMBER 1990

�a:

W Q. :::!! W I-

36

-��---����32��o 5 10 15 20 25 30 35 40 45

44 42 40 38 36 34 32 0

10

5

IT :;... w a: :::J I< a: w Q.

:::;:

w I-

.. .. ......�. .�. .. .�.��_��_..__.. ... __..f�·�·". .. ·····r-- r�..... � �

44

_.

42

·· ·

Z

","., . ............. ,

:i :J �

40

w I< a:

38 36

::

!

34 32 0

5

10

15

20

15

20

25

15

20

25

(a)

(a)

25

30

20 18 16 14 12 10

8

0 .... u.

35

40

6

0

10

5

45

TIME [MINUTES]

TIME [MINUTES] (b) Fig. 4. The s im ulta neou s temperature responses to two kidneys to step chang es in p erfusio n under constant power heating. (a) Minimum and maximum temperatures in the heated region of kidney one, and the flow

Fig.

5.

(b) Temperature [e'ponse to a ramp change in perfusion. (a) Maxi­

mum and minimum temperatures in the heated region of kidney three. (b) Flow into kidney three.

rate into kidney one. (b) Minimum and maximum temperatures in the heated region of kidne y three. and the flow rate into kidney three.

30

38

the upper and lower bounds on the temperatures measured inside the heated region. In region one. as the flow was increasell from IS to 25 mL/min, the temperature dropped about 2.5°C. When the flow rate was brought back to its original value the temperature increased to its previous level. Kidney three showed similar results for three-step changes in flow rate. These changes in temperature demonstrate the sensitivity of the phantom temperatures to changes in flow initiated by the computer, and they also show that perfusion changes in one region did not significantly affect the temperatures in the second region. To test repeatability. this series of flow rate changes was repeated three times with very similar results. The temperature changes associated with each change in flow rate for these three repetitions wcre compared, and the maximum differ­ ence between runs was less than 22% for all cases (based on the largest temperature change of the three runs). In a second programmed flow test the temperature response for a ramp change in How was measured (Fig. 5). As the flow rate increased, the temperature decreased. This illustrates the phan­ tom's versatility in providing various programmed temporal How patterns. (The temperature time lag in the hotter thermocouple's response is probably due to it's being located in a region of low perfusion, and is thus cooled mainly by thermal conduction.) The temperature dependent flow gives one the ability to study various algorithms (one can program the flow response as almost any function of temperature) which could possibly reproduce the temperature dependent perfusion observed in vivo (e.g., [II], [18], [19]). For example, we have used this apparatus [20] to system­ atically reproduce the oscillatory response of in vivo canine thighs to hyperthermia [II]. For this TCF ICP Study, the flow rate to each kidney was controlled with the algorithm: Vet) Vb + G*to(t td) where V flow rate in mL/min, Vb basal flow rate in mL/ min, G gain in mL/min per °C, to T temperature ele­ vation in °C, t time, and td response delay. Selected results from that study (to illustrate the use of this apparatus) are shown in Fig. 6 for a test with Vb 8 mL/min, G 4, and a time delay of 4 min. One can see that the highest flow ratc in kidney three is 28.4 mL/ min which occurred at 14.5 min. This is about 4 min after the peak temperature (38.8°C measured by the flow control

� w c:: :J I< II: w Q. :::;: w I-

25

:J

36

20

3:

.

""""""

10

0

(a)

10 0

.... U.

FLOW

20

30

B

5

30 25

38

Z :i :J

� 20 W

36

15

TEMP

34 .__•....•••.•.....•

32

W l-

a:

34

40

w a: :::J I< a: w Do :::;: w l-



15

Computer-controlled dynamic phantom for ultrasound hyperthermia studies.

A dynamic phantom system for use in evaluating hyperthermia heating equipment has been designed, constructed, and tested. It consists of four in vitro...
621KB Sizes 0 Downloads 0 Views