Sensors 2015, 15, 23205-23217; doi:10.3390/s150923205 OPEN ACCESS

sensors ISSN 1424-8220 www.mdpi.com/journal/sensors Article

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Underwater Imaging Applications Jinlong Song 1,2, Chenyang Xue 1,2,*, Changde He 1,2, Rui Zhang 1,2, Linfeng Mu 1,2, Juan Cui 1,2, Jing Miao 1,2, Yuan Liu 1,2 and Wendong Zhang 1,2,* 1

2

Key Laboratory of Instrumentation Science & Dynamic Measurement, Ministry of Education, North University of China, Taiyuan 030051, China; E-Mails: [email protected] (J.S.); [email protected] (C.H.); [email protected] (R.Z.); [email protected] (L.M.); [email protected] (J.C.); [email protected] (J.M.); [email protected] (Y.L.) Key Laboratory of Science and Technology on Electronic Test & Measurement, North University of China, Taiyuan 030051, China

* Authors to whom correspondence should be addressed; E-Mails: [email protected] (C.X.); [email protected] (W.Z.); Tel.: +86-351-3921-756 (C.X.); +86-351-3920-399 (W.Z.). Academic Editor: Bernhard Tittmann Received: 16 May 2015 / Accepted: 8 July 2015 / Published: 15 September 2015

Abstract: A capacitive micromachined ultrasonic transducer structure for use in underwater imaging is designed, fabricated and tested in this paper. In this structure, a silicon dioxide insulation layer is inserted between the top electrodes and the vibration membrane to prevent ohmic contact. The capacitance-voltage (C-V) characteristic curve shows that the transducer offers suitable levels of hysteresis and repeatability performance. The −6 dB center frequency is 540 kHz and the transducer has a bandwidth of 840 kHz for a relative bandwidth of 155%. Underwater pressure of 143.43 Pa is achieved 1 m away from the capacitive micromachined ultrasonic transducer under 20 excitation. Two-dimensional underwater ultrasonic imaging, which is able to prove that a rectangular object is present underwater, is achieved. The results presented here indicate that our work will be highly beneficial for the establishment of an underwater ultrasonic imaging system. Keywords: CMUT; C-V characteristic curve; sound pressure; bandwidth; ultrasonic imaging

Sensors 2015, 15

23206

1. Introduction Underwater imaging has extensive applications, including underwater observation, differentiation between different objects and recreational underwater activities [1]. Optical visibility in very clear water ranges from about 30 to 60 m, however, the waters within harbors, estuaries and ocean environments are turbid, and this leads to low optical visibility, therefore, the capabilities of underwater optical imaging are limited. Fortunately, acoustic waves can penetrate turbid water and even mud [2]. Ultrasonic waves are a type of sound wave that occurs at frequencies of more than 20 kHz, which is the upper limit of human hearing. Ultrasonic waves offer linear propagation as a result of their small wavelengths, high power (because of the high operating frequency), and strong energy transfer abilities. Ultrasonic devices have actually been used for underwater imaging since World War I and in medicine since the 1930s [3]. They are widely used in a variety of applications, including guiding of interventions in blood vessels and in the heart [4,5], and microelectromechanical systems (MEMS)-based microphones and speakers [6]. Piezoelectric ultrasonic transducers based on lead zirconate titanate (PZT) currently dominate transducer technology. However, low operational efficiency, difficulty in processing two-dimensional arrays and narrow operating bandwidths have been drawbacks when using PZT to meet the demands of modern acoustic transducers [7]. Fortunately, advances in micro-fabrication technology have enabled fabrication of the capacitive micromachined ultrasonic transducer (CMUT), which with its wide bandwidth and design flexibility has emerged as a strong candidate to replace the aging PZT transducer technology [8–10]. Additionally, other advantages such as their compatibility with integrated circuit (IC) fabrication technology, array configuration capabilities, good acoustic impedance matching with liquids and high operational efficiency have brought CMUTs one step closer to being the next generation transducer technology [11–13] and as a result, underwater acoustic imaging using CMUTs is a technology that requires further study. In 1994, Khuri-Yakub led his team to create the first CMUT with a sealed cavity, meaning that it could be used underwater [14]. Over the past few decades, CMUTs have attracted increasingly intensive research interest [9,15–18]. A bandwidth of approximately 40% and pressure of 300 Pa underwater when 1 m away from a CMUT under 100 excitation was achieved at an operating frequency of around 25 kHz using only a single underwater CMUT [19]. Wygant researched the use of transducers working under immersion conditions at around 2.5 MHz for therapeutic applications [20]. Logan fabricated a CMUT with a −6 dB relative bandwidth of 120% [21]. Oralkan has realized 2D imaging using a CMUT with a central frequency at 3 MHz, which only limits the range of action of the transducer [1]. In this work a CMUT with a bandwidth of 155% that requires 20 excitation and has a 540 kHz center frequency for use in underwater imaging has been designed, fabricated and tested. 2. Structural Design The frequency, which is closely related to the radius and the thickness of a vibration, is a vital transducer parameter. Because ultrasonic waves can propagate over greater distances at lower frequencies [22] and because the frequency of the underwater acoustic imaging system is in the 100 kHz–2 MHz range [2], the frequency of a transducer designed for underwater performance is

Sensors 2015, 15

23207

determined to be about 300 kHz. Based on our previous study and the process conditions [23], the radius and the thickness of the membrane are determined to be 90 μm and 2.8 μm, respectively. A transducer operating in transmitting mode requires a high cavity height to obtain high output pressure, while a transducer working in receiving mode requires only a low cavity height to attain high sensitivity [24]. The optimum cavity height is confirmed to be 0.65 μm after balanced consideration of these requirements. When the size of the top electrode is 40% to 50% of the size of the membrane, the transducer can have a large bandwidth [25]. As a result, the top electrode radius is determined to be 45 μm. A highly-doped silicon wafer is used as the vibration membrane [19]. Unfortunately, this could easily result in ohmic contact between the top electrodes and the membrane, making the membrane become the top electrode. Therefore, a silicon dioxide insulation layer is added between the top electrodes and the membrane. A cross-sectional schematic of a sensitive cell is shown in Figure 1a and the plan form of the element is shown in Figure 1b. The specific parameters of a designed structural element are listed in Table 1.

ti1 tm d0

ti 2

(a)

(b)

Figure 1. (a) The cross-section of a sensitive cell; (b) The planform of an element. Table 1. The parameters of designed CMUT structure. Parameters Membrane radius/μm Membrane thickness/μm Electrode radius/μm Electrode thickness/μm Number of cells Electrode insulation layer thickness/μm Insulation layer thickness/μm Cavity height/μm

Value 90 2.8 45 1 900 0.15 0.15 0.65

The axial resolution (∆x) is defined as the minimum distance that can be distinguished between two echo sources in the direction of propagation. The axial resolution can be expressed as: ∆x =

λ 2

where n is the number of scanning lines, λ is the wavelength and nλ is the pulse width.

(1)

Sensors 2015, 15

23208

Increasing the operating frequency and reducing the number of scanning lines can improve the axial resolution of the transducer. For the CMUTs designed in this paper, the axial resolution is approximately 5.55 mm when the number of scanning lines is 3. The lateral resolution (∆y) is defined as the minimum distance that can be distinguished in the vertical direction of propagation. The lateral resolution can be expressed as: ∆y =

1.22λ ×S D

(2)

where D is the aperture size of the probe and S is the distance between the focus and the CMUT. The lateral resolution at a specific focus can be improved by reducing the operating wavelength and increasing the aperture size of the probe. The probe aperture size for our designed CMUT is 6 mm, and the lateral resolution is approximately 6.78 mm when the distance between the focus and the CMUT is 50 mm. 3. Fabrication Process Like other MEMS transducers, the transducer presented in this paper can be fabricated by the MEMS sacrificial technique [26]. A detailed description of this fabrication process can be found in the literature [8]; however, the required fabrication processes have been summarized briefly in this paper. Before the fabrication process begins, a silicon wafer and a silicon-on-insulator (SOI) wafer should first be prepared. The requirements of the silicon wafer and the SOI wafer are listed in Table 2. Table 2. The detailed parameters of the silicon wafer and SOI wafer. Parameter Size (inches) Conductive type Resistivity (ohm·cm) Orientation Thickness (μm)

SOI Wafer 6 Device layer P Handle wafer N Device layer 0.01–0.08 Handle wafer 0.01–0.02 (100) Device layer 2.8 ± 0.1 Box layer 0.8 ± 0.08 Handle wafer 430 ±15

Silicon Wafer 6 P 0.01–0.02 (100) 400 ± 10

The main fabrication flow-chart is illustrated in Figure 2. The fabricated transducers on a 150 mm wafer are shown in Figure 3. Step 1 Standard RCA cleaning is performed on both the silicon wafer and the SOI wafer to remove organic matter, dust and oxide layers. Step 2 The silicon wafer is then oxidized to form a 0.8-μm oxide layer, which will be part-etched to form cavities, as shown in Figure 2a. Step 3 Gluing and exposure processes are performed on the front area of the silicon wafer; a 0.65-μm oxide layer is etched to form the cavities and a 0.15-μm oxide layer is left behind to prevent membrane contact with the substrate, as shown in Figure 2b.

Sensors 2015, 15

23209

Step 4 The silicon wafer and the SOI wafer are bonded together at low temperature [27,28] and in a vacuum environment [3], as shown in Figure 2c. Step 5 The handle layer of the SOI wafer, the buried oxide (BOX) layer of the SOI wafer and the oxide layer are etched to produce the basic transducer structure, as shown in Figure 2d. Step 6 A silicon dioxide layer is deposited on the vibration membrane to prevent the formation of an ohmic contact between the top electrodes and the vibration membrane, as shown in Figure 2e. Step 7 A metal layer is sputtered on the vibration membrane by the evaporation method, and the top electrodes and pads are formed by the peeling method, as shown in Figure 2f.

(a)

(b)

(c)

(d)

(e)

(f) SiO2

Figure 2. The main fabrication flow-chart. In this process, low temperature wafer bonding technology is vital. Low temperature processing can effectively avoid problems such as introduction of thermal stress, possible contamination, defect generation and doping profile broadening, and also increases the bond strength [27]. Sealed cavities can prevent the hydrolysis of water under high electric fields, thus reducing the energy losses and effectively averting electric breakdown [28].

Sensors 2015, 15

23210

Figure 3. Final fabricated transducers on the 6 inch silicon wafer. 4. Experimental Results 4.1. C-V Characteristics The Agilent 4284A PRECISION LCR METER is used to analyze the effects of the DC bias voltage on the transducer capacitance. The oscillation frequency range of the LCR meter is from 20 Hz to 1 MHz. The DC bias voltage ranges from 0 V to 45 V and then from 45 V back to 0.0 V in a cycle. When the scan cycle number is two, the voltage step length is 0.5 V and the operating frequency is 1 MHz, the C-V characteristic curve is plotted as shown in Figure 4. The black curve and the green curve represent the first and second cycles, respectively. The red curve and the disperse points represent the average capacitance under different bias voltages. -9

1.25

x 10

1.2

Capacitance/F

1.15

1.1

1.05

1

0.95

0.9

first cycle second cycle 0

5

10

15

20 25 Bias Voltage/V

30

35

40

45

Figure 4. The C-V character. The capacitance basically remains unchanged when the DC bias voltage ranges from 40 V to 45 V. This phenomenon can be accounted for by the reason that the membrane has collapsed. As a result, the collapsed voltage is 40 V. Additionally, the deviation between the output capacitance values is small when the DC bias voltage is increasing and when the DC bias voltage is decreasing. This indicates that the transducer has a fine hysteresis performance. Moreover, the consistency between the output

Sensors 2015, 15

23211

capacitance values of the first and second cycles is also fine. This implies that the transducer has excellent repeatability performance. 4.2. Underwater Experiments 4.2.1. Output Pressure and Bandwidth Testing Under the hypothesis that V (V) is the collected voltage and S (dB) is the sensitivity of the standard hydrophone, the sound pressure level (SPL) can be expressed as: SPL = 20 log ( ) −

(3)

The pressure can then be expressed as: P = 10



(4)

= 1 μPa.

where

The experimental schematic diagram and configuration are shown in Figure 5a,b, respectively. The left CMUT is driven by multiple cyclical bursts of sinusoids from the signal generator (Agilent 33521A) amplified by a high voltage amplifier (HA-205) at 250 kHz. The bias and excitation voltages are 20 V and 20 , respectively.

(a)

(b)

900

0

measurement value throretical value

800

-1 Normalized Magnitude/dB

Pressure/Pa

700

600

500

400

-3

-4

-5

300

200 10

-2

20

30

40 Distance/cm

(c)

50

60

70

-6

200

300

400

500 600 Frequency/kHz

700

800

900

(d)

Figure 5. (a) The experimental schematic diagram (b) The experimental configuration (c) the measurement and theoretical pressures at different distances (d) The −6 dB bandwidth.

Sensors 2015, 15

23212 Table 3. The sound pressure at different distances.

1 2 3 4 5 6 7 8 9 10 11 12 13

Distance/cm 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00 50.00 55.00 60.00 65.00 70.00

Measurement Pressure/Pa 786.7700 769.6664 701.2516 598.6294 513.1109 444.6961 461.7998 427.5924 393.3850 324.9702 307.8665 290.7628 273.6591

Theoretical Pressure/Pa 807.3977 733.4918 666.3509 605.3558 549.9440 499.6043 453.8725 412.3269 374.5841 340.2962 309.14684 280.8488 255.1410

Derivation Pressure/Pa 20.6277 −36.1746 −34.9007 6.7264 36.8331 54.9082 −7.9273 −15.2655 −18.8009 15.3260 1.28034 −9.9140 −18.5181

The −6 dB bandwidth of the hydrophone is 31% and the sensitivity is −199.8 dB @ 250 kHz (0 dB re 1 V/μPa). Based on Equations (3) and (4), the measurement pressures at different distances away from the CMUT are obtained and are plotted in Figure 5c and listed in Table 3. If the pressure at the sound source is P0, the distance between the sound source and a point on the sound source axis is x and the attenuation coefficient is α, the pressure (P) at the point can be expressed as [29]: P=P

(5)

The pressure at the sound source (983.9 Pa) and attenuation coefficient (0.01946) are obtained from the measurement pressure and distance data. Equation (5) can then be written as: P = 978.3

.

(6)

The theoretical and measured sound pressures at various distances are plotted in Figure 5a and listed in Table 3. The average error between the theoretical and measured values is only 4.5%. Additionally, 143.43 Pa pressure is achieved underwater at a distance of 1 m from the CMUT under 20 V excitation. A frequency domain plot of the signal received by the hydrophone is shown in Figure 5b. The −6 dB center frequency is 540 kHz and the transducer has a bandwidth of 840 kHz for a relative bandwidth of 155%. This bandwidth implies that the designed structure achieves the intended aim of increasing the bandwidth. 4.2.2. Distance Testing The hydrophone is then replaced by a CMUT; an experiment schematic is shown in Figure 6a. When the distance between the two CMUTs is 30 cm, then the filtered signal is as shown in Figure 6a. The first signal is the transmitting signal and the second signal is the receiving signal.

Sensors 2015, 15

23213

(a)

(b)

(c) Figure 6. (a) Schematic diagram of the distance testing experiment; (b) The transmitting and receiving signal when the distance between the two CMUTs; The frequency of the transmitting signal is 250 kHz; (c) The measurement and real distances between the two transducers. Table 4. The real and measurement distances between the two CMUTs. 1 2 3 4 5 6 7 8 9 10 11 12 13 14

Real Distance/cm 5 10 15 20 25 30 35 40 45 50 55 60 65 70

Measurement Distance/cm 5.65 11.46 15.22 21.72 26.45 31.67 36.37 41.25 46.30 51.31 55.85 61.23 65.83 70.67

Deviation/cm 0.65 1.46 0.22 1.72 1.45 1.67 1.37 1.25 1.3 1.31 0.85 1.23 0.83 0.67

Sensors 2015, 15

23214

The measured distance (31.67 cm) between the two CMUTs can be obtained from the speed of ultrasonic propagation underwater (1480 m/s) by multiplying the time that the signal takes to travel from the transmitting CMUT to the receiving CMUT (2.14 × 10−4 s). Similarly, measured distances at actual distances of 10–70 cm at 5 cm step increments are plotted in Figure 6b and listed in Table 4. The results show that the measured distances are greater than the real distances by around 1 cm. The distance between the transducers and the packaging is not considered and the deviation related to the CMUT moving from one location to another can account for this result. In addition, the actual speed of ultrasonic propagation underwater of less than 1480 m/s may also contribute to the result. However, this experiment does demonstrate that the designed CMUT operates well underwater. 4.2.3. Underwater Imaging The experimental underwater imaging configuration is shown in Figure 7a. The imaging target is placed at a specified distance from the transducer. The transducers, of which one is used to transmit ultrasonic waves and the other is used to receive these ultrasonic waves, are moved from the left side to the right side of the imaging target. The initial 2D ultrasonic imaging result is obtained through bandpass filtering, envelope detection, logarithmic compression and image processing, with results as shown in Figure 7b.

Lateral distance [mm]

-150 -100 -50 0 50 100 150 100

(a)

200

300 400 Axial distance [mm]

500

600

(b)

Figure 7. (a) The underwater imaging experiment configuration; (b) The imaging result of the target. While the edge of the target imaging result is fuzzy, we can see that the target is obviously there. Therefore, this experiment has basically demonstrated that the designed transducer can be used for underwater imaging. 1D arrays and 2D arrays will be studied in the future, and a pronounced increase in the imaging results will be obtained. 5. Conclusions A CMUT with an insulation layer appended between its top electrodes and vibration membrane for underwater imaging is designed, fabricated and tested in this paper. The CMUT shows fine hysteresis and repeatability performances. The 155% bandwidth of the transducer proves that the designed

Sensors 2015, 15

23215

structure is beneficial for increasing transducer bandwidth. Distance and imaging experiments demonstrate that the designed transducer shows promise for applications in the underwater field. Acknowledgments This work was supported by National Natural Science Foundation of China under Grant 61127008, and the “863” program of China under Grant 2013AA09A412. Author Contributions The author Jinlong Song designed the structure, organized the experiments and drafted the manuscript. Chenyang Xue provided with the funding support. Wendong Zhang and Changde He designed the fabrication process. Jing Miao completed the processing of the transducer. Rui Zhang and Yuan Liu performed the data collection and analysis. Linfeng Mu and Juan Cui reviewed and edited the manuscript. Conflicts of Interest The authors declare no conflict of interest. References 1.

2. 3.

4.

5.

6.

7.

8.

Oralkan, O.; Ergun, A.S.; Cheng, C.H.; Johnson, J.A.; Karaman, M.; Khuri-Yakub, B.T. Underwater acoustic imaging using capacitive micromachined ultrasonic transducer arrays. In Proceedings of the OCEANS’02 MTS/IEEE, Biloxi, MI, USA, 29–31 October 2002; pp. 2354–2360. Sutton, J.L. Underwater acoustic imaging. IEEE Proc. 1979, 67, 554–566. Oralkan, O.; Ergun, A.S.; Johnson, J.A.; Karaman, M.; Demirci, U.; Kaviani, K.; Lee, T.H.; Khuri-Yakub, B.T. Capacitive micromachined ultrasonic transducers: Next-generation arrays for acoustic imaging? IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2002, 49, 1596–1610. Ito, S.; Suzuki, T.; Ito, T.; Katoh, O.; Ojio, S.; Sato, H.; Ehara, M.; Suzuki, T.; Kawase, Y.; Myoishi, M.; et al. Novel Technique Using Intravascular Ultrasound-Guided Guidewire Cross in Coronary Intervention for Uncrossable Chronic Total Occlusions. Circ. J. 2004, 68, 1088–1092. Courtney, B.K.; Munce, N.R.; Anderson, K.J.; Thind, A.S.; Leung, G.; Radau, P.E.; Foster, F.S.; Vitkin, I.A.; Schwartz, R.S.; Dick, A.J.; et al. Innovations in imaging for chronic total occlusions: A glimpse into the future of angiography’s blind-spot. Eur. Heart J. 2008, 29, 583–593. Park, K.K.; Lee, H.; Kupnik, M.; Khuri-Yakub, B.T. Fabrication of Capacitive Micromachined Ultrasonic Transducers via Local Oxidation and Direct Wafer Bonding. J. Microelectromech. Syst. 2011, 20, 95–103. Ladabaum, I.; Khuri-Yakub, B.T.; Spoliansky, D.; Haller, M.I. Micromachined ultrasonic transducers (MUTs). In Proceedings of the 1995 IEEE Ultrasonics Symposium, Seattle, WA, USA, 7–10 November 1995; Volume 1, pp. 501–504. Jeong, B.; Kim, D.; Hong, S.; Chung, S.; Shin, H. Performance and reliability of new CMUT design with improved efficiency. Sens. Actuators A Phys. 2013, 199, 325–333.

Sensors 2015, 15 9. 10. 11. 12. 13.

14. 15. 16. 17.

18.

19.

20.

21.

22.

23. 24.

23216

Emadi, T.A.; Buchanan, D.A. Multiple moving membrane CMUT with enlarged membrane displacement and low pull-down voltage. IEEE Electron Device Lett. 2013, 34, 1578–1580. Ergun, A.S.; Yaralioglu, G.G.; Khuri-Yakub, B.T. Capacitive micromachined ultrasonic transducers: Theory and technology. Am. Soc. Civ. Eng. 2014, 16, 76–84. Jin, X.; Ladabaum, I.; Khuri-Yakub, B.T. The microfabrication of capacitive ultrasonic transducers. J. Microelectromech. Syst. 1998, 7, 295–302. Salim, M.S.; Malek, M.F.A.; Heng, R.B.W.; Juni, K.M.; Sabri, N. Capacitive Micromachined Ultrasonic Transducers: Technology and Application. J. Med. Ultrasound 2012, 20, 8–31. Cianci, E.; Foglietti, V.; Caliano, G.; Pappalardo, M. Micromachined capacitive ultrasonic transducers fabricated using silicon on insulator wafers. Microelectron. Eng. 2002, 61–62, 1025–1029. Suo, X. Simulation and Deformation Analysis of CMUT Membrane; South China University of Technology: Guangzhou, China, 2013. Chen, J.; Cheng, X.; Chen, C.; Li, P. A Capacitive Micromachined Ultrasonic Transducer Array for Minimally Invasive Medical Diagnosis. J. Microelectromech. Syst. 2008, 17, 599–610. Liu, J.; Oakley, C.; Shandas, R. Capacitive micromachined ultrasonic transducers using commercial multi-user MUMPs process: Capability and limitations. Ultrasonics 2009, 49, 765–773. Doody, B.C.; Cheng, X.; Rich, A.C.; Lemmerhirt, D.F.; White, R.D. Modeling and Characterization of CMOS-Fabricated Capacitive Micromachined Ultrasound Transducers. J. Microelectromech. Syst. 2011, 20, 104–118. Helin, P.; Czarnecki, P.; Verbist, A.; Bryce, G.; Rottenberg, X. Severi S: Poly-SiGe-based CMUT array with high acoustical pressure. In Proceedings of the 2012 IEEE 25th International Conference on Micro Electro Mechanical Systems (MEMS), Paris, France, 29 January–2 February 2012; pp. 305–308. Olcum, S.; Oguz, K.; Senlik, M.N.; Yamaner, F.Y.; Bozkurt, A.; Atalar, A.; Koymen, H. Wafer bonded capacitive micromachined underwater transducers. In Proceedings of the 2009 IEEE International Ultrasonics Symposium (IUS), Rome, Italy, 20–23 September 2009; pp. 976–979. Jin, X.; Ladabaum, I.; Degertekin, F.; Calmes, S.; Khuri-Yakub, B.T. Fabrication and characterization of surface micromachined capacitive ultrasonic immersion transducers. J. Microelectromech. Syst. 1999, 8, 100–114. Wong, S.H.; Kupnik, M.; Watkins, R.D.; Butts-Pauly, K.; Khuri-Yakub, B.T. Capacitive Micromachined Ultrasonic Transducers for Therapeutic Ultrasound Applications. IEEE Trans. Biom. Eng. 2010, 57, 114–123. Logan, A.S. The Design, Fabrication and Characterization of Capacitive Micromachined Ultrasonic Transducers for Imaing Applications. Ph.D. Thesis, The University of Waterloo, Waterloo, ON, Canada, 2010. Li, Y.; He, C.; Zhang, J.; Zhang, H.; Song, J.; Xue, C. Design and analysis of MEMS capacitive ultrasonic transducer. Transducer Microsyst. Technol. 2014, 33, 73–75. Zhuang, X. Capacitive Micromachined Ultrasonic Transducers with Through-Wafer Interconnects. Ph.D. Thesis, Stanford University, Stanford, CA, USA, 2008.

Sensors 2015, 15

23217

25. Bozkurt, A.; Ladabaum, I.; Atalar, A.; Khuri-Yakub, B.T. Theory and analysis of electrode size optimization for capacitive microfabricated ultrasonic transducers. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 1999, 46, 1364–1374. 26. Emadi, T.A.; Buchanan, D.A. Development of a novel configuration for a MEMS transducer for low bias and high resolution imaging applications. Proc. SPIE 2014, 8976, doi:10.1117/12.2037840. 27. Tong, Q.Y.; Cha, G.; Gafiteanu, R.; Gosele, U. Low temperature wafer direct bonding. J. Microelectromech. Syst. 1994, 3, 29–35. 28. Tsuji, Y.; Kupnik, M.; Khuri-Yakub, B.T. Low temperature process for CMUT fabrication with wafer bonding technique. In Proceedings of the Ultrasonics Symposium (IUS), San Diego, CA, USA, 11–14 October 2010; pp. 551–554. 29. Feng, N. Ultrasonics Handbook; Nanjing University Press: Nanjing, China, 2006; pp. 16–31. © 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/4.0/).

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Underwater Imaging Applications.

A capacitive micromachined ultrasonic transducer structure for use in underwater imaging is designed, fabricated and tested in this paper. In this str...
NAN Sizes 1 Downloads 12 Views