A~~ulyrlca C’hl,nica Acra. 75 ( 1975) 169-l 80 0 Elsevicr Scientific Publishing Compnny. Amstcrdnm

AMPEROMETRIC DETERMINATION CARBOXYLIC ACIDS WITH AN ENZYME ELECTRODE

M. NANJO Departrnerir

(Received

169 - Printed

in The Ncthcrlnnds

OF ALCOHOLS, ALDEHYDES AND IMMOBILIZED ALCOHOL OXIDASE

and G. G. GUILBAULT qf Cliertrl.~try,

29th July

University

of New Orlcrms.

New

0rleatr.s. Ltr. 70122 (U.S.rl.)

1974)

Chemical tests for measurement of blood alcohols are the subject of much interest. particularly in legal cases, and there is a need for quick, simple and reliable assay methods. Likewise. the determinations of carboxylic acids and aldehydes are of importance in various industrial fields, but such assays are difficult to carry out. especially at low concentrations. In the past. alcohols_ have been determined by gas chromatograplry1-3. and rcdox titration”.‘. Selective spectrophotometric enspectrophotometry4*’ zymatic methods utilizing alcohol dehydrdgenase have been reported’-“. but it is not possible to couple these with an electroanalytical device such as an ion-selective electrode; these methods also require a coenzyme. NAD. and deproteinization of the serum sample before U.V. measurement. Guilbault and Lubrano” used a new enzyme for assay of alcohol, alcohol oxidase, by amperometrically detecting the hydrogen peroxide formed during the enzymatic reaction. Because the activity’ of the available enzyme was low. it was not possible to immobilize the enzyme and to build an enzyme electrode in a method similar to that described for assay of g1ucose’2*13. Hence, measurements were conducted only in bulk solution with soluble enzyme. In this study, an alcohol enzyme electrode which was useful for the assay of blood ethanol in concentrations as low as 10 mg ‘%, is described. Additionally, this chemically bound alcohol oxidase electrode was found to be useful for the assay of aldehydes and carboxylic acids based on a measurement of the consumption of dissolved oxygen in the enzymatic reaction, similar to that used in the uricase-uric acid systemi4. EXPERIMENTAL

A Heath polarograph (Model EUA-19-2) combined with a Heath operational amplifier system (Model EUW-19A) was used. A constant potential, -0.G V us. SCE. was applied to the electrode. Reagettts All chemicals were reagent grade. The solid, activity measured by hydrogen peroxide

alcohol oxidase formation rate)

(0.02 was

units/mg from the

170

M. NANJO.

G. G. GUI LBAULT

rnycclium of (t Basitiior~~ycete and was a gift of Wyeth Laboratories, Inc. This enzyme is identical with that used by Guilbault and Lubrano”. Alcohol oxidase was immobilized according to a modification of the method reported by Brown et ul. I5 as follows.

Alcohol oxidase ( 100 mg) and plasma albumin ( 100 mg, Sigma Chemical Co., from Bovirw Scrtrrn) were dissolved in 5.0 ml of phqsphate buffer, 0.1 M. pH 8.2. Then 5 drops of glutaraldehyde (Sigma Chemical Co., SO’;/ aqueous solution) was added. and the mixture was stirred well. The solution was frozen by dry ice-acetone coolant and the frozen mixture was warmed slowly in a refrigerator for a day. The sponge-like copolymer was washed with the buffer solution and stored in a refrigerator until USC. This method has been shown to be effective. and convenient for glucose oxidase”. L-amino acid oxidase”’ and uricase14. Platinum disc electrodes (Beckman Model 39273) were used as the solid base electrode to sense the dissolved oxygen change, the same as used for the uricase electrode14. The procedure for the construction of the enzyme electrode was as follows. The immobilized alcohol oxidase was mounted on the surface of the platinum electrode (diameter 1.0 cm) and secured with a nylon cloth and O-rings; the electrode was then stored in the buffer solution (0.1 M phosphate buffer, pH 8.2) at room temperature. The enzyme electrode together with a calomel reference electrode (SCE) were placed into a stirred buffer solution (10.0 ml) at an applied potential of -0.G V vs. SCE. When the current was at a constant level-a constant limiting current from reduction of dissolved oxygenthe sample solution (0.01410 ml) was pipetted into the buffer solution and the initial rate of change in the dissolved oxygen limiting current, and the final steady-state current, were recorded with a Heath Recorder (Model EU-205-I 1). The total measuring time was only 2 min. The amount of substrates present was calculated from a calibration curve of initial rate or steady-state current vs. substrate concentration. In the case of the assay of blood ethanol, 10.0 ml of 0.1 M phosphate buffer (pH 8.2) was placed in a thermostated water jacket (3O”C), and 0.025 ml of serum was added directly to the stirring buffer solution. After each measurement. the electrode was rinsed with distilled water and dipped into the buffer to recover the dissolved oxygen level around the enzyme and to wash out any remaining reaction products (2-3 min). The dissolved oxygen content in the buffer solution was found to be constant during a series of assays owing to the slow equilibrium of oxygen between buffer solution and air for these experimental conditions. RESULTS

aliphatic

AND DISCUSSION

Alcohol oxidase is reported to catalyze alcohols according to the reaction18-20:

the oxidation

of lower

primary

ENZYME

ELECTRODES

FOR

/JLCOHOL.

ALDEHYDES

171

ETC.

Alc0lNrl0*idauc ’ RCHO

RCH,OH+O,

-‘- H LO2

(1)

This reaction has been coupled with amperometric monitoring of the hydrogen the activity of the cnzymc peroxide formed. with a soluble enzyme l ‘al’. Howcvcr, was found to be too low to be useful for a practical assay as an immobilized enzyme electrode’ ’. Clarkl’ also reported the instability of his soluble alcohol oxidase electrode which lost half its activity after standing overnight at pH 8.5 (instability believed to be due to the loss of FAD from the electrode). However. it has proved possible to make a very stable and sensitive alcohol oxidase enzyme electrode with chemically *bound enzyme. and to monitor the decrease in the dissolved oxygen level with a platinum electrode which of peroxide formation functions as an oxygen electrode at -0.6 V US. SCE (instead measured at +O.G V OS. SCE). Bulk retrctiou of’ itlcoltol osititrscf br.itlt etlrartol At first, the enzymatic reaction was studied ampcrometricully hydrogen peroxide in solution with a platinum clcctrode at At this potential, the following oxidation of hydrogen peroxide 8-t 2). H202 + 0,+2H++2cWith careful pretreatment of the platinum electrode, have reported that the accurate quantitative measurement centrations as low as 10 /tM hydrogen peroxide’ ‘.

I

0

I

I

I

I

by monitoring +0.6 V OS. SCE. takes place (pH (2)

Guilbault and Lubrano was possible at con-

-

4 lime, min Fig:. 1. Bulk cnzymutic reaction clctc~_~rl h> ‘J~,IIIIILIII~ ulcctrodc. Ethanol (38 mg ‘x,) added to the cnzymc buNcr. pi-l 8.2, 30°C) at A. (I) 3 Dnysolution (nlcohol oxidasc 10 mg in 1.0 ml. I.0 M phosphate V vs. SCE: (II) fresh cnzymc solution. anodic old enzyme solution. cathodic current chungc 111 -0.6 current chungc at +0.6 V us. SCE: (III) rresh cnzymc solution. cathodic current change ut -0.6 V us. SCE. 2

172

M. NANJO.

G. G. GUILBAULT

In Fig. I, the anodic current increase is shown, indicating the formation of hydrogen peroxide at +0.6 V us. SCE using freshly prepared enzyme solution (curve II). But the activity of the enzyme solution, as. measured at this voltage. appears to become very weak soon after the preparation of this solution. as higher than 6.5 (ref. 20). reported by Clark’ ‘. even on storage in pH solutions However. dissolved oxygen is reduced to water, consuming 4 electrons per molecule of oxygen at -0.6 V us. SCE: 0,+2

H20+4

e--+4

OH-

(3)

As shown in Fig. 1 (curve I), the consumption of dissolved oxygen was detected as a cathodic current decrease at -0.6 V IX. SCE using 3 day-old enzyme solution. The initial rate of decreasing oxygen current, expressed as PA min - i, was faster than that of hydrogen peroxide formation at +0.6 V us. SCE. Hydrogen peroxide is also electroactive at negative potentials: H,O,-t-2

e-+2

OH-

(4) Curve II’1 in Fig. 1 shows the cathodic current change from the formation V us. SCE. This cathodic current increase was of hydrogen peroxide at -0.6 observed only when freshly prepared enzyme solution was added to the ethanol buffer solution,‘indicating an accumulation of hydrogen peroxide on the platinum electrode surface. Therefore, the cathodic current, i,, at -0.6 V us. SCE can be expressed as follows: i, = iilro2 f ino.

(5)

from hydrogen peroxide according to eqn. (4) and iH202 is the current is the current from dissolved oxygen according to eqn. (3). The magnitudes are considered to depend on the enzymatic reaction and of inlOr and i,,o. degree of diffusion of the substances involved in the reaction to the electrode, and reflect the apparent enzymatic reaction rate. From Fig. 1, it becomes apparent that if fresh enzyme solutions are used, hydrogen peroxide can be detected at both potentials, even at -0.6 V us. SCE. The formation of hydrogen peroxide compensates for the decrease of current from the decrease of dissolved oxygen, since hydrogen peroxide accumulates around the electrode, and a net increase in current results. However. an aged enzyme solution (greater than 3 days old) cannot produce sufficient hydrogen peroxide to overcome the consumption of dissolved oxygen and a decrease of the cathodic current results, except at very high substrate concentrations (above 0.05 M).

where iD,o.

Enzyme

electrode

response

In Fig. 2, the responses of an enzyme electrode are shown. When ethanol is added to the buffer solution, it diffuses into the layer of chemically bound ’ alcohol oxidase where it is oxidized at the expense of dissolved oxygen and is converted to hydrogen peroxide and acetaldehyde. Curve I, in Fig. 2, shows the response of a 7 day-old electrode at +0.6 V US. SCE. The formation of hydrogen peroxide is thus proven also in the case of

ENZYME

ELECTRODES

FOR

ALCOHOL.

ALDEHYDES

173

ETC.

immobilized alcohol oxidase. However, an enzyme electrode 12 days old showed no response to the same ethanol solution at -1-0.6 V US. SCE. Yet. the response at -0,6 V US. SCE became more sensitive. indicating that the enzyme consumed dissolved oxygen faster than hydrogen peroxide accumulated at the enzyme electrode. The initial rate of dissolved oxygen decrease, i.e. the apparent initial rate of the enzymatic reaction, was found to be proportional to ethanol added, using an electrode that had aged over 2 weeks. L

I

_e-/’

I J_

_-

-

)I4

__--

d

-2,

4 -1

l

I

I

I

0

2

4

6

Time,

I 8

Ethanol,

min

I 20 mg%

Fig. 2. Rcsponsc of an cnzymc clcctrodc (glutoraldchydc bound alcohol oxidasc) to ethanol (added at A). T=30”C. ethanol 38 mg ‘%, in 0.1 M phosphate buffer, pH 8.2. (I) 7 Day-old clcctrodc, at +0.6 V us. SCE: (II) I2 day-old clcctrodc. at -0.6 V us. SCE: (111) 3 day-old elcctrodc, ut -0.6 V us. SCE. Fig. 3. Rclntionship between enzyme clcctrodc at -0.6 (II) stcndy-state method.

elcctrodc rcsponsc und ethanol V us. SCE. 0.1 M Phosphntc

concentration bufl’cr. pH

mcrlsurcd with II 14 day-old 8.2. 30°C. (I) Rate method:

A typical relation curve based on the initial rate of reaction as a measure of ethanol is shown in Fig. 3. The steady-state current. which shows the amount of consumed dissolved oxygen for the enzymatic reaction, was also proportional to the ethanol concentration. The sensitivity at -0.6 V us. SCE was found to increase with increasing time and number of measurements, reached a maximum after 2 weeks, and then kept a constant value for more than 4 months. For good analytical results, an electrode should be prepared and used only after 2 weeks.

30”

174

M. NANJO.

Consrrrnption

of H,O,

thrirlg

emymt

G. G. GUILBAULT

ic reactiou

Figure 4 shows the consumption of hydrogen peroxide during the enzymatic reaction at kO.6 V c’s, SCE. The currents, proportional to hydrogen peroxide concentration (points A and C). were found to decrease when ethanol was added to the hydrogen peroxide solution as monitored with an enzyme electrode. This consumption of hydrogen peroxide could not be observed with only the base platinum electrode. which shows that there is no bulk chemical reaction between ethanol and hydrogen peroxide under these experimental conditions. Rather. this peroxide consumption is catalyzed by the alcohol oxidase present in the enzyme electrode. Also, the hydrogen peroxide current at both fO.6 V and -0.6 V US. SCE decreased when acetaldehyde was added to the solution as monitored by the enzyme electrode. The addition of semicarbazide. which reacts with aldehydes forming semicarbazones. decreased the rate of oxygen uptake. These results can only be explained by some subsequent reaction following reaction (I), and the observed change of the property of enzyme electrode during the experiment is _ considered to be due mainly to-an activation of this subsequent reaction which consumes hydrogen peroxide. Resporwe to acetaldeltyde and acetic crcitl The enzyme electrode was found also to aldehydes and carboxylic acids,

0

0

2

4-e Time,

10 min.

to respond not only showing a response

OO

to alcohols. but such as that in

1 Substrate,

2 10”

M

4. Consumption or hydrogen pcroxidc during the enzymatic reaction mcrtsurcd by an ugcd cnzymc clcctrodc (14 days old). 0.1 M Phosphate buffer. pH 8.2. 30°C. (A) Addition or 3.8 mg ‘7; cthunol to I(O.45 mM H202) ut +0.6 V DS. SCE: (B) addition or 0.45 m&f HrOz to II (phosphntc bulTcr solution) at .- 0.6 V us. SCE: (C) addition of 3.8 mg I;/, of ethanol to (II + HIO1). Fig.

Fig. 5. Comparison ol’ the clcctrode response to cthunol, ucctaldchydc ilnd ucctic acid. Chemically bound alcohol oxidasc clcctrodc. 0.1 M phosphlttc bulk. pH X.2. 30°C. (I) Ethanol; (II) ncctaldchyde; (111) sodium ncctutc.

3

ENZYME

ELECTRODES

FOR

ALCOHOL.

ALDEHYDES

175

ETC.

Fig. 2. curve II. The sensitivity of the electrode for formaldehyde, acetaldehyde. acetic acid and formic acid was as good as lhat of ethanol. and thus should be useful for the assay of low levels of aldehydes and carboxylic acids. In Fig. 5, the relation of the initial rate to the concentrations of ethanol, acetaldehyde and acetic acid. is shown.

The effect of pH on the initial rate of change in current. at -0.6 V LX SCE. of 3.8 mg ‘%; ethanol solution measured with an alcohol oxidase electrode. was studied. As indicated in Fig. ‘6. the optimal pH for the enzymatic reaction was found to be approximately 8.0. which is almost identical with the value of 7.8 for hydrogen peroxide formations reported by Guilbault and Lubrano’ *.

1.2

00

6

l. II

‘A+ 7

6

9

7. El&t

11

or ummonium

I

2--

sulkttc

rcsponsc:

(0)

conccntrution

phosphate on

the

-

6

I 10

1

N9,)2SO, 0 Q%

buflix. 0.1 M: (a) rcsponsc

I 6

I

4 (

PH

Fig. 6. Eflixt or pH on the clcctrodc Fig.

10

glycinc

or nn ngcd

alcohol oxidasc clectrodc: (I) dissolved oxygen mcnsurcmcnt at -0.6 V cs. SCE. (II) hydrogen pcroxidc mcnsurcmcnt nt +0.6 V vs. SCE 38 mg I>,;cthnnol.

bull+.

0.1 M.

(one 38

month old) mg ‘2; ethanol:

Figure 7 shows the effect of ionic strength (ammonium sulfate) on the response of an enzyme electrode at both +0.6 V and -0.6 V us. SCE. Both rates decreased with increasing concentrations of ammbnium sulfate. However, the rate of peroxide formation measured at +0.6 V US. SCE was quenched faster than V us. SCE. Thus. high conthe rate of dissolved oxygen consumption at -0.6 centrations of salt solutions are one of the interferences in an assay, and for good results, the ionic strength should be kept as low as possible and constant. Selectivity

contrast

Alcohol oxidases to the sensitive

are specific for lower primary aliphatic alcohols17-19. In ethanol response of the electrode, methanol gives a very

176

M. NANJO. I 0.6

G. G. GUILBAIJLT

I

i

M

Methanol,

Fig. 8. Elcctrodc rcsponsc to methanol measured by an aged phosphate bulk. pH 8.2, 30°C: (I) dissolved oxygen at -0.6 at +0.6 V vs. SCE.

clectrodc (one month old) V us. SCE: (II) hydrogen

in 0.1 M peroxide

V US. SCE. In Fig. 8. the relationship of the rates at poor response at -0.6 both potentials is shown for methanol; an aged enzyme electrode was used. As c&n be seen from Fig. 8, methanol gives a response at both potentials in the same concentration range (higher than 0.1 M), but the response is much lower than that for ethanol shown in Fig. 5. The rate at -0.6 V us. SCE shows a maximum around 0.5 M of methanol, then decreases. This indicates that the formation of hydrogen peroxide becomes predominant at this concentration rather than a consumption of hydrogen peroxide. In Table I. the substrate selectivity of the alcohol oxidase electrode is shown at both potentials. The apparent selectivity at -0.6 V vs. SCE with the chemically bound alcohol oxidase is very different from the reported values for soluble alcohol oxidase11*18*1g. The source of the alcohol oxidase was the same as that of TABLE

I

SUBSTRATE

SELECTIVITY

OF ALCOHOL

aXIDASE -

A Icohols

Relaf iae reucriait~

Methanol Ethunol Ally1 alcohol n-Propanol n-Butanol

100 28 17 5.3 2.1

o Rclativc reactivity mcasurcd by the cnzymc curves and cxprcssed us initial rate per molt

-

3 12000 190 1300 2500 elcctrodc was obtained of alcohols (/iA min-’

0.98 0.46 1.14 0.36 0.21 from the slopes mole-‘).

of the calibration

ENZYME

ELECTRODES

FOR

ALCOHOL.

ALDEHYDES

177

ETC.

Janssen et ~l.‘e*19. although the niethod of assay for enzyme activity is quite different. Jansscn et al. measured the activity spectrophotometrically by reacting this prevents the subsequent reactions the peroxide formed with o-dianisidine; which remove hydrogen peroxide from the system. The present procedure involves total consumption of dissolved oxygen at -0.G V US. SCE. Thus, the apparent selectivity may result from the rate-determining reaction in the enzymatic oxidation. In Table II, a comparison of the reactivity of several .alcohols. together with their aldehydes and carboxylic acids is shown. In the case of ethanol and butanol. the rate-determining step was found to be in the last reaction, i.e., the oxidation of the carboxylic acid produced from the oxidation of the alcohol. However, itself gives a very slow rate, in the case of the methanol system, methanol whereas formaldehyde and formic acid consume dissolved oxygen much faster than methanol. Also, the rate of formaldehyde oxidation is faster than that of acetaldehyde. On the other hand. the relative selectivity obtained at +O.G V U.S. SCE has the same trend as that observed by Janssen Is* l‘-‘. TABLE

II

COMPARISON ALDEHYDE

OF THE REACTIVITY AND CARBOXYLIC ACID

CHJOH HCOH HCOONu CzH,OH d-i&OH CH .COONa n-C;HloOH II-C,H,&OONa 0 Rclntivc

rate of change

OF

ALCOHOLS

WITH

THEIR

CORRESPONDING

1.00 17600 5640 1.00 0.89 0.88 1.00 0.94 in current

ut - 0.6 V vs. SCE.

phosphutc

bulkr.

0.

I M. pH 8.2.

These selectivity data demonstrate the possibility of assay of ethanol in V US. SCE. Also, Table II shows the presence of excess of methanol at -0.G that it is possible to assay aldehydes or carboxylic acids if they are present alone in solutions, or in mixtures after a separation of each component. Assny

of blood

alcohol An assay of blood ethanol was attempted to demonstrate a typical application of the alcohol oxidase electrode. Since blood contains several hundreds of compounds, it is necessary to consider the possible interference of some of them. Such interferences were considered in two categories: the direct reaction of other substrates with enzyme; and the consumption of dissolved oxygen by non-enzymatic reactions, such as by ascorbic acid or cysteine. In the first category, some acids and hydroxy acids were found to be oxidized by alcohol oxidase (see Table III), and thus interfered with the assay of ethanol in blood (i.e., lactic acid which is present in blood at concentrations of 8-17 mg “A). However, the .response of lactic acid to the electrode is 37’j/,

178 TABLE

G. G. GUILBAULT

111

RELATIVE REACTIVITY OF ENZYME ELECTRODE __--.___ --.----_..--Acids ---

NANJO.

M.

----

-.-------

(EtllilllOl)

Acetic acid Formic acid Lactic acid Butyric acid Pyruvic acid Monochloroucctic acid ----__-.--.----------..-----.-

SOME

ACID

SUBSTRATES -.-.--.-

-.--.----.--.

Reltrriw wwliril --.----.-..-

I” --------_

---.-_--

- .--.-----..

AT

THE

ALCOI-101,

.--.__~----_.-_---

OXIDASE

- _._.__ __ _____

- __.__.---_-..-__---.-.--.-

._._---

____ __

I .ot.l

0.88 0.51

0.37

0.2 I 0.03 0.001

n Rclutivc rate or chnngc in current at No rcsponsc to acctonc and trichloroacctic

--..-

__..---.--.-.-.-..-.-_.

-0.6 V acid.

rs. SCE,

----. .--_-

-.-------.

phosphate

hufkr.

0.1

M,

_.__ .-__---___ pl-I 8.2. 3O’C.

of ethanol response. Thus, in the assay of high levels of ethanol. the effect of lactic acid becomes less serious. Furthermore. the interference of lactic acid in serum may be eliminated by the addition of lactate dehydrogenase to the sample before assay. In the second category. various substances were tested under the same conditions of assay, with a platinum electrode (without enzyme). *None of the compounds commonly present in blood (i.e.. ascorbic acid. cysteine. phenylalanine. glucose. uric acid), were found to reuct at the electrode. Also the effect of the content of oxygen in the sample solution was tested: the small amount of serum (0.025 ml) was found not to give any change in the dissolved oxygen level of the large excess of buffer solution (10.0 ml). The calibration curves prepared from aqueous ethanol solutions or ethanol control solutions (Sigma Chemical Co.) did not give good agreement with a spectrophotometric method based on alcohol dehydrogenase (ADH). probably owing to interference from acids in serum. Next. a calibration curve W&ISprepared by adding an aqueous ethanol solution to a reconstituted freeze-dried control prepared from pooled human blood (Moni-Trol I. Dade, Miami). This curve is shown in Fig. 9. The calibration curves did not pass the origin, but gave a good agreement with the spectrophotometric method as shown in Table IV. Moni-Trol II gave’ the same calibration curve. indicating there were no serious interferences from abnormal serum. The rate and steady-state methods were found to be useful for the assay of blood ethanol especially in the assay of high ethanol levels, which is the most common case of assay in the clinical laboratory. Thus. the enzyme electrode utilizing dissolved oxygen consumption is concluded to be very sensitive (1 mg ‘x, of ethanol), reproducible (standard deviation by rate method 2.9’%,. by steady-state method 4.9X, when 0.05 ml of a 0.08 mg “/, ethanol control was added to 10.0 ml buffer solution). very rapid (complete analysis takes less than 5 min), and simple (as little as 25 ccl of strum sample can be added directly without any pretreatment). Additionally, this enzyme electrode is also useful for the assay of carboxylic acids. such as acetic acid and formic

5NZYME

ELECTRODES

I

OO

FOR ALCOHOL.

I

I

5

10

I 15

Ethanol

I 20

ALDEHYDES



25

179

ETC.

10

mg%

added.

Fig. 9. C:llibration cwvcs for blood ethanol ussay at -0.G V rs. SCE. Aged (one month old) alcohol oxidilsc clcctrode, 0.1 M pll~~sphatc bufkr. pH 8 2. 3O’C. (I) Row method: (II) steady-state method. TABLE

IV

RESULTS

----

OF BLOOD

----

0.4 4.2 8.2 12.2 16.2 20.3

ETHANOL

Rtiw 0.0 2.5 6.0 Il.9 17.6

tmhotl

SIw/_v-srcire

t~rethrl

-__--

--

0.4 2.7 5.8

IO.9

21.9

* According to Sigma Technical L’Avcritgc of 3 determinations.

ASSAY“

18.5 20.4

Bulletin. No. 33-UV.

acid, and aldehydes such as acetaldehyde and formaldehyde as can be seen from Tables II and III. from normal It may be possible to measure lactic acid in serum samples individuals who have not ingested alcohol (below 10 mg ‘j/U).This enzyme electrode should be applicable not only to the clinical field, but also to various industrial areas, such as the food and brewing industries.

Science

The authors Foundation.

wish to acknowledge the linancial assistance of the National Grant No. GP-3 1518A. in carrying out this research. We

180 also wish to thank Dr. S. S. Laboratories for a gift of alcohol

M. NANJO.

Kuan for oxidase.

his technical

G. G. GUILBAULT

assistance

and

Wyeth

SUMMARY

A new enzyme electrode has been developed for the rapid and simple measurements of blood ethanol. A platinum electrode was coupled with immobilized alcohol oxidase, sensing the dissolved oxygen consumption amperometrically. Blood ethanol can be measured within several minutes without any pretreatments. The only reagent required is phosphate buffer. The enzyme electrode was found to respond not only to alcohols but also to aldehydes and carboxylic acids. REFERENCES 1 2 3 4 5 6 7 8 9 10 1I 12 13 14 15

J. T. Kung and J. E. Whitney. A~~crl. Clrern.. 33 (1961) 1505. C. Blucstcin and l-1. N. Posmantcr. rl~rol. C/W~I.. 38 (1966) 1856. H. Lymons and J. Bard, C/in. Chon.. 10 (1964) 429. V. W. Reid und D. G. Sulmon. n!~tr/_vsr (Lowlu~). 80 (1955) 704. M. Mantel and M. Anbar. AIIN/. Cl~eol.. 36 (1964) 936. B. Jusciskis and J. P. Wurrincr. Awf. C/rem. 38 (1966) 563. T. Takahnshi nnd C. Hnyusc. J. Clwrn. Sot. Jup.. Irul. C/renr. Sect.. 55 (1952) 205. R. K. Bonnichscn nnd H. Thcorll. Scam/. J. Clirl. Lab. Inuesr.. 3 (1951) 58. R. Bonnichscn and C. T. Lundgren. Actct P/wwwco/. Tmicol.. 13 (1957) 256. M. Browcr and J. E. Woodbrigc. C/in C/WIII.. 16 (1970) 539. G. G. Guilbault and G. J. Lubruno. And. C/rim. Acttr. 69 (1974) 189. G. G. Guilbault und G. J. Lubrano. And. C/rim Actcc. 60 ( 1972) 254. G. G. Guilbault and G. J. Lubruno, Amt/. C/~/nr. Acttr, 64 (1973) 439. M. Nanjo und G. G. Guilbault. Arrd. C/rer~. 46 (1974) 1769. G. Brown, C. T. Thomas. G. Gcllbc. D. Domurndo. A. M. Bcryonncrtu und G. Guillon. Biorecht~o/. Biow~~.. 105 ( 1973) 359. 16 M. Nunjo and G. G. Guilbault. Artctl. Chim. Actu. 73 (1974) 367. 17 L. C. Clark. Jr. in L. Wingard (Ed.), E,~z~vne E!@,tecr/~lg. Wiley-lntcrscicncc, New York, 1972. p. 377. 18 F. W. Jnnsscn. H. W. Ruclius. i3ioc/1ent. Biopllp. Amt. 151 (1968) 330. 19 F. W. Jansscn. R. M. Kcrwin and H. W. Ruclius. Bicdwtw Biophyss. Res. Cowwtrr~.. 20 (1965) 630. 20 Y. Tuni. T. Miya. I-I. Nishiknwn and K. Ogatu, Apr. Bid. C/W~I., 36 (1972) 68.

Amperometric determination of alcohols, aldehydes and carboxylic acids with an immobilized alcohol oxidase enzyme electrode.

A~~ulyrlca C’hl,nica Acra. 75 ( 1975) 169-l 80 0 Elsevicr Scientific Publishing Compnny. Amstcrdnm AMPEROMETRIC DETERMINATION CARBOXYLIC ACIDS WITH A...
822KB Sizes 0 Downloads 0 Views