Medical Robert
E. Nelson,
BS,
RTR
John
#{149}
G. Stears,
RT
Acceptance Testing Experience in Testing at Two Major Medical The problems encountered in acceptance testing of newly installed imaging equipment at two major medical institutions over a 10-year period are presented. Acceptance tests were conducted in 129 newly installed imaging systems with conventional acceptance testing methods. A total of 1,132 problems were documented. Problems were classified as major or minor; there were 772 major problems and 360 minor problems. An average of six major and three minor problems were documented in each new equipment installation tested. In some instances, final payment was withheld for several months or more to ensure correction of the problems identified. This experience confirms the need of thorough acceptance testing of new imaging equipment before final payment is made to the vendor. Index terms: Acceptance testing #{149}Radiology and radiologists, departmental management Radiology
and
radiologists,
design
of radiologi-
cal facilities 1992;
183:563-567
health
ODERN
standards
production of high
care requires of quality, and
of radiologic technical
Joel
#{149}
inations
requires
optimal
skilled
ment quality maintenance in such ceptance
control program.
a program tests on
achieved
to operate,
with equip-
and
preventive The first step
is to perform newly installed
equipment. That is, verify new equipment operates signed
consisexam-
performance.
The latter can only be the aid of a comprehensive
ac-
that the as it is de-
complies
with
the
baseline
for
future
quality control evaluations. The objective of this article is to review the acceptance testing experience and the problems encountered at two major medical facilities-the Mayo
Clinic,
Rochester,
the University and Clinics, Birmingham. publications tance testing 7); however, type
and
countered We address From the Department of Radiology, Univerof Alabama Hospitals and Clinics, University of Alabama at Birmingham, Birmingham, AL 35294 (R.E.N., G.T.B.), and Department of Diagnostic Radiology, Mayo Clinic and Foundahon. Rochester, Minn (J.G.S., J.E.G.). From the
Minn,
and
of Alabama Hospital University of Alabama at There have been several concerned with accepand quality control (1to our knowledge, the
frequency
of problems
en-
have not been reviewed. those issues herein.
standards
are
met, and high-quality images are obtamed. At both institutions, acceptance tests are performed when the vendor cornpletes the installation and the equipment is ready for clinical use. Thus, problems in equipment function and performance are identified and corrected before clinical use. The results of acceptance tests in 129 new equipment installations over a 10year period were evaluated. Seventy-six pieces of equipment were installed at the Mayo Clinic and 53 at the University of Alabama at Birmingham. The 129 installa-
lions were equipment
reg-
ulatory standards, and produces high-quality images. This in turn mmimizes radiation exposure to patients and staff. Furthermore, when acceptance tests are performed on new equipment, it ensures that all contract (bid) requirements are satisfactorily met between the vendor and institution before clinical use. Also of importance is that data from acceptance provide
with the manufacturer’s specsU.S. Food and Drug Administra-
tion (FDA) performance in
personnel
equipment
PhD
accordance fications,
high the
is essential
objective. The of high-quality
E. Gray,
Systems: Systems
examinations
quality
achieving this tent production and
PhD
T. Barnes,
ofRadiologic 129 Imaging Facilities’
M
tests
Radiology
#{149} Gary
Physics
divided into seven general classifications. Equipment
from
20 different equipment manufacturers were included in the evaluation. These manufacturers
are
listed
in Table
I and
represent a broad cross-section of the medical x-ray imaging industry. Thirteen categories of radiologic compo-
nents
were
developed
for problem
fication.
Problems
were
further
as major
or minor,
depending
identi-
classified on
the
se-
verity of the problem. Major problems (a) interfered with the desired patient examinafions and equipment operation or resulted in a hazard to the patient, (b) noticeably compromised image quality, (c)
resulted
from
FDA
standards, hour meet
in a 10% or greater regulations
or
(d)
and
required
deviation
performance
more
than
I
to repair. Minor problems (a) did not the required bid specifications or
performance
standards
and,
in the
judg-
ment of the authors, did not interfere with patient examinations and equipment operation or result in a hazard to the patient; (b) resulted in less than a 10% deviation from FDA regulations and performance standards; or (c) required less than 1 hour to repair.
I
RESULTS
sity
MATERIALS
AND
The acceptance
METHODS
tests performed
at both
RSNA scientific assembly. Received September 9, 1991; revision requested October 10; revision received December 13; accepted De-
institutions were similar and comparable in scope. The protocols employed have been previously described elsewhere (1-
cember
5). Newly
installed
to ensure
that
1990
C
23. Address
RSNA,
1992
reprint
requests
to G.T.B.
equipment
it functions
Table problems
and
and
problems
found
number of major
in
and
360 were
minor.
of
for each
the seven types of equipment stalled. A total of 1,132 problems identified. Of these, 772 were
is evaluated
properly
minor
2 shows the total and the number
of
inwere
major In general, the 563
more complex the system, the greater the number of problems. The major problems in the different equipment component categories are listed in Table 3, and the minor problems are listed in Table 4. Lack of compliance with FDA performance standards was responsible
for
at least
one
problem
in
each number in the x-ray beam
of the 13 categories. The largest of major problems occurred calibration and operation of generators. Collimation and alignment and equipment integrity categories were also high in deficiencies. Most minor problems occurred in installation and equipment integrity. For both the major and minor problems,
collimation
accounted lems.
and
beam
for 171 of the
The
x-ray
alignment
1,132
generator
prob-
category
are present during most of the acceptance testing. However, problems are not corrected when identified. Rather, acceptance testing is completed, and a
in the
acceptance
types
of equipment,
list
Classification of the data in major and minor problems in ferent component categories completely indicate the range
of all problems
is provided
as the
initial
list
to the
of problems,
that the dates the rected satisfactorily highlighting
This lems this
problems.
are
that
problems
are
documented and acceptance can be performed efficiently interruption. Disadvantages with
less-effective
philosophies, documented
had
to be replaced.
Of
tubes,
image
intensifiers,
the
grids
pickup tubes, high tension transformers, and tabletops were also replaced. One portable radiography unit was rejected as a result of evaluation. Tables 5 and 6 identify components that were replaced.
The
philosophies differ
slightly
involvement
of the
two
instituto the
of the
vendor’s service Clinic, both inhouse and vendor service engineers are present during acceptance testing, and minor problems are corrected at that time if possible. If problems cannot be corrected immediately, the acceptance testing is continued and completed to the degree possible. Capabilities not tested during this initial phase are tested at a later date when the equipment should be functioning engineers.
correctly.
At Mayo
Advantages
of this
while
repairs
are
being
564
#{149} Radiology
(four
However,
major,
both
were at the
Table Types
Studied
assem-
ifi)
Inc (Chicago)
Imaging
GE Medical
(Rochester, NJ)
Corporation
Systems
Jamieson Film Liebel-Flarsheim nati)
(Denver)
(Milwaukee)
Company Company
(Dallas) (Cincin-
Philips Medical Systems (Shelton, Conn) Picker International, Inc (Highland Heights, Ohio) Schimadzu
X-Ray
Corporation
land Heights, Ohio) Siemens Medical Systems, NJ)
test-
(High-
Inc (Iselin,
Tecnomed USA (Bay Shore, NY) Thomson CGRt Transworld X-Ray Corporation (Charlotte, NC) Varian Associates (Palo Alto, Calif) Xerox Medical Systems*
helped
2 of Equipment
Bucky
and equipment inGrid and Bucky often overlooked dur-
Manufacturers
Fischer
minor)
acceptance
and
terms of the difdoes not of prob-
as illusproblems
Inc* Eastman Kodak Company NY) Elscinct, Inc (Hackensack,
testing
one
is
testing and, therefore, is of practical impor-
Amerisys, Amrad,
for
identify a surprisingly large number of problems with new equipment installations. Ultrasound, computed tomographic, and magnetic resonance imagers were not included in this study because we have been less involved
*
t
and Associated
No longer Purchased
Major
and
in business. by GE in 1988.
Minor
Problems No. of Problems
Type
of
No. of Units
Equipment
ap-
Radiography
and
fluoroscopy*
Radiographyt Mobile radiography
Processors and darkrooms Mammography Special procedures, heart ters
Includes
add-on
7,3
106
169
99
344 268
24 21
93 18
38 11
131 29
12
74
27
101
17
173 7
74 5
247 12
129 general radiography and tomography and trauma, digital
Total
27
2
Total Indudes t Includes
Minor
cathe-
Miscellaneous*
*
Major
and fluoros-
copy
made.
At the University of Alabama Hospitals and Clinics, the vendor’s service engineers are not essential to the acceptance testing process. They are invited to be present and, in general,
different
problems
Acoma X-ray Company (Wheeling, Advanced Medical Systems, Inc (Geneva, Ohio) Alpha Imaging (Wffloughby, Ohio)
the
equipment
fewer problems per installation
approaches
proach are that problems are easily communicated and corrected when identified, and, as a result, follow-up measurements are minimized. Disadvantages are that many problems are not documented and that the staff has to wait
in delaying
of the of the
Clinic
ing
in regard
result
grid
acceptance category
than at the University of Alabama Hospitals and Clinics (nine major, six minor). Other factors that may have affected the number of problems documented at the two institutions are the difference in location and service groups involved, and that different personnel performed the acceptance tests.
DISCUSSION
tions
As a result
Mayo
video
can
in the
Table 1 Equipment
communica-
subsystems
This
testing
well
tion about certain types of problems, and additional vendor visits may be required to resolve a problem or probfinal acceptance clinical use.
ing the
testing without may re-
accounted for 152 problems and equipment integrity for 145. Problems in equipment integrity included parts or subsystems that were not delivered or were nonoperational and shipment of the wrong equipment or parts. As a result, a number of components or
lems.
documenta-
found in a category, by reviewing the
bly, mechanical, tegrity categories. assemblies are
until all probadvantages of
of these
and
or incomplete.
found
except
uncorrected
system
lems trated
problems were corare noted, thereby
process is repeated are resolved. The
of acceptance
lacking
vendor’s service organization. When the problems have been corrected, evaluations are performed and a revised list of problems developed. The revised list is essentially the same
sult
operational components tested, were replaced most often. X-ray
tion
testing
fluoroscopy head, and
772
and urologic radiography dedicated chest units.
360 and
1,132
fluoroscopy.
units.
May
1992
Table 3 Major Problems
Found
with Acceptance
Testing
(n
772) No. of
Parameter X-ray source assembly Focal spot size
(n
=
Tested
Problems
37)
12 8 6 11
Alignment
Half-value layer Miscellaneous X-ray generator
calibration
Kilovolt peak Milliampere
and
and operation
kilovoltage
Time, milliampere Line compensator
(it
=
125)
36 28 8 3 22 11 17
waveform
seconds
Hand switch Exposure Output
waveform
Collimation
and
beam
light field
Radiography Radiography Indicators Source-to-image
image cassette
Positive
(n
alignment
Radiography
=
119) 25 10 22 12
intensifier
distance
beam
12
28 10
limitation
Detents
Grid and Grid
Bucky assembly
(n
59)
=
41 18
Bucky Automatic
exposure
Density
control
performance
(n
Tube potential and patient thickness Sensing cell variations Miscellaneous Fluoroscopic
automatic
rates (n
brightness
brightness
Maximum 5-minute
5
control
control
Input
tabletop timer
and
exposure
image
Image
quality
rate
rate (n
=
84)
intensifier
Monitor
31
performance
Video
29
camera (n
Mechnical
24 72)
=
Table
31 11
Tubecrane
Image Image
intensifier intensifier
tower tower
movements alignment
13 6
11
Miscellaneous
(n
Tomography
=
25)
Operation
12
Alignment Fulcrum Speed and
2 7
exposure
2 2
Miscellaneous
and darkroom
Imaging, processing, Processors Multiformat
cameras (n
Installation Corridor
=
0 0 0 0 0 0 0
lights
location
Equipment aesthetics Room aesthetics Painting
Miscellaneous integrity type
Wrong on
order
Volume
183
Number
#{149}
=
104)
25 14 27
24)
Miscellaneous code
(n
38
Nonoperational Not provided =
20)
4
Tie wrapping
Equipment
=
0)
warning
Equipment
(n
5 11
Darkroom
(n
exposure
8 2 19 3 19
tracking
exposure
phosphor
Fluoroscopic
Error
36 10
tracking
Interlock
Other
52)
51)
=
Automatic
Parts
=
settings
problems
2
19 5
tance. Examples of problems include grids with the wrong ratio, grids not centered to the x-ray source, grids with the wrong focal range, Bucky assemblies demonstrating grid lines at more than 10-msec exposure times, and Bucky assemblies in which the grid would slow down or stop when minimal pressure was applied to the assembly cover. As noted
in Tables
3 and
4, 73 prob-
lems in the grid and Bucky assembly category were found at the two institutions during the past 10 years. Six grids
were
replaced
because
of radio-
graphically demonstrated defects, one fluoroscopic grid with a radiographically demonstrated defect at its periphery was accepted because the defect was outside the television and cine field of view, four grids were replaced because the wrong grid ratios were supplied, five grids with a 3640-inch (90-100-cm) focus were installed on units with a 44-inch (112cm) source-to-image distance, and three grid tunnels were found to have the incorrect focal length. The remaining 29 problems were associated with Bucky assemblies. In six cases, the moving grid was not centered to the x-ray source during an exposure (lateral decentering), and in seven cases the grid was not orthogonal to the x-ray beam (off-level decentenng). Other examples of Bucky assembly problems included grid lines evident at more than 10-msec exposure times, Bucky assemblies that were difficult to move and position, and wall Bucky assemblies that were not properly counterbalanced. The most common mechanical radiography table, tube crane, and image intensifier tower problems encountered were rough or difficult movement (due to binding, defective, or misaligned bearings), disturbing noise during
movement,
and
would not release release. Additional intensifier tower nonfunctional
locks
that
or only partially table and image problems included
limit
switches
that
per-
mitted the table to continue to drive or tilt beyond normal limits, nonfunctioning drives and power assist, dragging brakes, noisy drives, and the omission of or nonoperational interlocks. Additional tube crane problems were cables catching other equipment, cables and/or tie wraps that limited movement, and the omission of cushions, stops, and detents. The latter allowed unsatisfactory and destructive metal-to-metal collisions and collisions with walls, control booths, or delicate imaging components. The number of major problems asRadiology
565
#{149}
sociated with equipment integrity is large, approximately 13% of all major problems, which is somewhat surprising for new equipment installations. Included in this category are (a) yendor notification of parts or components on order at the time of acceptance testing, (b) missing parts or components that the vendor did not indicate, (c) incorrect parts or components supplied, and (d) nonoperational parts or components that were subsequently replaced. Many minor installation integrity problems (eg, room aesthetics, room painting) were associated
with
institution
rather
than
vendor responsibility. It is advantageous to specify in the purchase agreement that a substantial portion of the purchase price will be withheld until the equipment meets specifications. In several cases, final payment was withheld for several months. In two cases (one at each institution), final payment was withheld for more than 1 year before the problems were resolved. In one of those cases, the equipment manufacturer made several modifications to the installed equipment and, later, to production systems. The amount of money that should be withheld is negotiable but should be about 20% of the purchase price. As a result of acceptance testing, a number of expensive major and minor components were found to be defective or did not meet specifications and were subsequently replaced. Examples are listed in Tables 5 and 6 and include x-ray tubes, an x-ray generator (the wrong type was supplied), x-ray generator high-tension transformers, tabletops (bent), a digital image
processing
tubes, (those
intensifiers, television supplied
desired
degree
control
module,
video pickup monitors and stands would not allow the of movement
and
viewing flexibility), a mobile radiography unit, grids with the problems mentioned earlier, the grid-cassette mounting bracket and the lower section of a radiographic U-arm (because of poor design and/or manufacturing practices, the grid could not be aligned to the x-ray beam), and a large number of miscellaneous circuit boards. Many of these defective components (eg, high tension transformer, digital processing module, bent tabletop, and malfunctioning printed circuit boards) would have been detected during routine clinical use if the equipment was not acceptance tested. However, equipment failure or malfunction is likely to result in additional patient positioning, 566
Radiology
#{149}
Table
4
Minor
Found
Problems
Acceptance
with
Testing
(n
=
360)
No. of Parameter X-ray source assembly Focal spot size Alignment Half-value Miscellaneous
(n
Tested
Problems
30)
=
1 1 27 1
layer
X-ray
calibration and and kilovoltage
generator Kilovolt peak Milliampere
Time, milliampere Line compensator
operation waveform
(n
=
17)
I 2 2
seconds
1
5 1
Hand switch Exposure Output waveform
Collimation
5
and beam
alignment
(n
=
52)
field
Radiography
light
Radiography
image
Radiography Indicators
cassette
Source-to-image
15
intensifier
2 0 9 9 2
distance
limitation
Positive beam Detents
15
Grid and Bucky assembly
(ii
14)
=
Grid
5 9
Bucky
Automatic exposure control performance Density settings Tube potential and patient thickness
(n =4) 3 0 0 1
tracking
Sensing cell variations Miscellaneous Fluoroscopic
automatic
(n
rates
Automatic
brightness
control
and
brightness
control
0
tracking
Interlock
0
Maximum
tabletop
5-minute Input
exposure
phosphor
0 0 0
rate
exposure
image
Image Video
rate
timer
Fluoroscopic
(n
quality
=
12) 3 4 5
intensifier camera
Monitor
performance
(n
Mechnical
39)
=
Table Tubecrane Image intensifier
Image
intensifier
14 17 3 0 5
tower movements tower alignment
Miscellaneous
Tomography
(n =5)
Operation Alignment Fulcrum Speed and
0 0 2 1 2
exposure
Miscellaneous
Iinagmg, processing, Processors Darkroom Multiformat Installation Corridor
and darkroom
cameras
(n
warning
lights
=
15 5 14
aesthetics
19
4 10 5
Miscellaneous Equipment
integrity type Parts on order Nonoperational Wrong
Not provided =
code
(n
=
41) 3 1 16
21
55)
Miscellaneous Error
19)
location
Room aesthetics Painting
(n
=
72)
Tie wrapping Equipment
(n
11 3 5
integrity
Equipment
Other
exposure
0)
=
53 problems
2
May
1992
Our experience clearly demonstrates that acceptance testing should be routinely performed in newly installed equipment before it is used for clinical imaging and before final payment is made to the vendor. #{149} Acknowledgments: the assistance Chakraborty,
The authors Wu, PhD, Dewey Narkates,
acknowledge Dev P. PhD, Rich-
of Xizeng
PhD, ard L. Morin, PhD, CarolJ. Mount, RTR, Jerome P. Taubel, RTR, and Timothy R. Daly, RTR, in acceptance testing, and the assistance of Phyllis J. Roybal in preparing the manuscript.
References 1.
creased contrast grids,
from
patient discomfort, problems, additional injections,
or repeat
More
subtle
tube focal generous
tional ation, contrast
Vnliime
patient scheduling contrast media imaging.
problems
include
spots being specifications
Electrical Manufacturers image intensifiers with resolution,
is:;
#{149} Number
x-ray
larger than the of the Na-
grid
2
artifacts,
Associpoor in-
radiation associated decreased
and
with contrast
decreased
tected would ment
use
would
not
have
been
de-
during routine clinical use but result in suboptimal equipperformance, suboptimal image
quality,
and
to the
patient.
increased
2.
misaligned resulting
of a lower than commonly employed grid ratio, increased radiation dose resulting from the use of a high-ratio grid in low-scatter conditions, and poor image quality and repeat imaging because of improper calibration and performance in accordance with automatic exposure control standards. In most cases, these problems
the
dose
radiation
3.
4.
Lin PP,
Kriz
JK, Rauch
ceptance
testing
equipment. Physics,
New 1982.
PL, et al, eds.
of radiological York:
American
Institute
6.
ology, 1983. Gray JE, Winkler
Stears
JG, Frank
Quality control in diagnostic ville, Md: Aspen, 1983.
imaging.
NT,
Hendee
WR, ed. The selection of radiologic equipment. Williams & Wilkins, 1985. Dixon RL, ed. MRI acceptance quality control. Madison, Wis: Physics, 1988. Knight KR. Is your equipment
Second
7.
of
Acceptance testing protocols: a systematic approach to evaluating radiologic equipment. Reston, Va: American College of Radi-
Source
Imag,
Gray JE, Morin RL. imaging equipment. 16.
October Purchasing Radiology
ED.
Rock-
and perforBaltimore:
mance 5.
Ac-
imaging
testing Medical
and
acceptable? 1990; 30-34. medical 1989; 171:9-
dose
Radinlntv
#{149} S67