.,
:
:
AAPM :: Tutorial
Stephen
.
Mammography Lincoln
author
introduces
first
of this
series
physics
in general.
on
principles
Annual the
Symposium
physics
applied
details
is used
and detection
clinical objects that parameters, and the
basic
diographic fit versus
to obtain
clinical
those
differing
of general
about
and the choices
that
need
emphasis is on image clinical application
are also
radiography
by the
exist
to be made
are to
demands.
patients
formation. of the
ma-
equipment
clinical
Interrelationships
In the basic
is intended and radiography
to be visualized, the patient’s anatomy, receptor. In this tutorial, breast imaging
imaging. The risk considerations,
cal development
imaging
information
images.
Physics. the
The analysis in particular
from
of the
x-ray
need image
principles
differ
because
of transmission
on Basic
of mammography,
in radiographic
systems
important
INTRODUCTION Radiography
trate
1 5th
Mammographic
in several
‘
the of tutorials
in terms of a mammographic system. in the understanding of mammography
assist
System1
PhD
The
discussed
PhD
as a Radiographic
B. Hubbard,
diologic
Balter,
the
formation
between
the
x-ray is used
beam to illus-
to obtain
optimal
In addition, technique,
dose,
ra-
beneits histoni-
and
reviewed.
The formation of a radiograph begins with the formation of a shadow image; this is shown schematically in Figure i A uniform x-ray beam is directed toward the patient. As it passes through the patient, it is attenuated. There will be a differential transmission of radiation if the attenuation of an object differs from that of its surroundings. This differentially transmitted beam carries the diagnostic information. After this beam exits the patient, it is detected and captured by an image me.
S
ceptor.
The
captured
patient
and
object.
dered. Our
The image concentration
tance only
makes
sense
the handling In particular,
Index
terms:
I
From
1990;
the
radiologic of the
©RSNA,
Department
IL 605
read,
in the
context
radiography,
quality
of the
assurance
a visible and
representation
a diagnostic
.
Breast
that report. widen
lead from the In addition, clinical
referral, diagnosis
radiography,
of the
impression
for future reference. is not intended to minimize
activities diagnostic
the
men-
impon-
positioning radiologic
process,
of the diagno-
which
includes
treatment, and follow-up. does not stand by itself; it is
radiation
dose
#{149} Images,
quality
Hubbard,
and
#{149} Physics
10:103-113 of Medical
Physics,
Rush
mc, Glencoc, Ill. From the 15th Annual Symposium vember 7, 1989; accepted November 8. Address Grove,
to produce
is ultimately
of other diagnostic information, the ability to obtain a correct
Breast
RadloGraphics
is processed
image
may then be archived on image creation
of the many other to the completion
patient sis
image The
University,
Chicago,
on Basic Physics reprint requests
and
Fields,
Griffith,
at the 1988 RSNA annual to the author, 4 1 1 3 West
meeting. End Rd.
Broadbent, Received Downers
No-
1 5-2307.
1990
103
X-RAY SOURCE
----
Figure
2.
Image
iions
of the
tient.
Variations
sult
formation
x-ray
in a range
beam
requires
por-
to penetrate
in patient
the
thickness
of radiation
pa-
will
intensities
re-
reach-
ing different points on the image receptor (eg, A, B, C, D). The dynamic range (latitude) of the image receptor must be sufficient
to record
these
varied
intensities.
INCIDENT BEAM
dx
OBJECT PATIENT
Figure
3.
Subject
differential
different
IMACER
Figure 1. Projection radiographic systems provide information about structures within patients by means of a shadowgraph of the area of interest produced with a penetrating x-ray beam emanating from a small source.
certainly not to be confused of any radiologic procedure social environments. SYSTEM
REQUIREMENTS
104
U
RadioGraphics
A radiographic
imaging
with will
system
a cure. change
must
contrast
transmission
structures
is formed of x rays
in the patient.
by the
through
In a uni-
form slab of tissue, the subject contrast of a dense object will depend on the thickness of the object (dx) and the difference be-
tween ject
the attenuation and
As time passes, both in response to changes
produce
coefficients
its surrounding
a spatially
uniform
of the ob-
tissue.
the
risk in the
and benefit medical and
primary
x-ray
beam that is simultaneously capable of penetrating the patient and is sensitive to attenuation differences within the patient. Such a beam will develop contrast in passing through the patient. The x-ray beam must also have suitable geometric properties. Spatial resolution is directly influenced by the focal spot size. The source-image distance and anode angulation determine the radiation field size. Motion unsharpness is influenced by exposure time. This in turn is a function of tube loadability and therefore partially anode angulation. The source-image distance and patient position determine the magnification of the image. Magnification geometry reduces the effects of scatter but usually requires smaller x-ray focal spots. Other considerations, for any imaging system and geometry, include patient dose and scattered radiation control. It can be seen that no general solution can be obtained for all radiography; appropriately optimized, dedicated systems are needed for different examinations.
U
Hubbard
Volume
10
Number
1
The choice of an appropriate imaging system is complicated by the need to balance radiographic contrast, noise, and patient dose. In most imaging systems, contrast and dose are both reduced as the voltage applied across the x-ray tube increases. If receptor sensitivity is increased, fewer x-ray quanta are needed to form the image. Quantum noise is therefore increased. Sensitive receptors tend to have an inherent noisy structure. This further increases the total image noise. The pres-
ence there
of noise are real
in the upper
image limits
obscures on tube
the visibility voltages and
of low-contrast detail. receptor sensitivities.
Thus,
The portion of the patient containing the object of interest has a major influence on imaging technique. The thickness of the part sets a lower limit on beam penetrability. A sufficient number of photons must be transmitted through the part so as to form a usable image. This is seen in Figure 2 The differences in transmitted intensity across the image must be limited so as to fit within the dynamic range of the image receptor (too much contrast is not helpful) However, the difference in transmission through a region in the object of interest relative to adjacent tissue must produce enough radiographic contrast so that the object can be distinguished from noise fluctuations (too little contrast is not helpful either) The quality, or penetrating ability, of an x-ray beam is customarily stated in terms of its half-value layer. .
.
.
The formation of contrast in the x-ray beam transmitted through the patient occurs as a result of differential attenuation along different rays in the beam. This is illustrated in Figure 3 In the absence of scatter, the local transmission of a ray passing a distance dx through an object with a linear attenuation coefficient u1 (where
CONTRAST
.
‘0
=
incident
intensity)
is given
by I
dx
11’0e
A nearby uation
nay passing
coefficient
through
u2) will
an identical
have
a local
thickness
transmission
of surrounding given
tissue
(atten-
by
-u,dx
‘2’0e The overall transmission sion and the transmission remainder of the patient
the two
of each through is identical
ray will be the product of this local transmisthe remainder of the patient. Assuming that the for both nays, one can calculate the ratio of
transmissions: u2)dxl
(-(U,
exp
‘1/’2
which
can
be approximated
by 1
‘l/’2
-
[(u1
u2)dx].
-
The term in brackets describes the scatter-free contrast in the transmitted x-ray beam. Contrast depends on the object thickness (dx) and the difference between the two attenuation coefficients (u1 and u2) Contrast will increase linearly with .
the
thickness
coefficients
of the is more
object. complex.
The
influence This
difference
of the
difference
decreases
between as the
quality
attenuation of the
x-ray
beam increases; it increases as the atomic number (Z ) of the object increases, and it increases as the density (p) of the object increases. Thick, dense objects with high atomic numbers are easily seen with a soft x-ray beam. Conversely, thin tissue-density objects with lower atomic numbers are extremely difficult to visualize with a hard x-ray beam.
January
1990
Hubbard
U
Ra4ioGrapbics
U
105
0.05
ATTENUATION
X-RAY
SPECTRA
WATER
.
E >I-
I-
U)
z
z
LU
LU I-
U
z
U. U.. 0. LU
LU
50 KVP
>
unfiltered
I-
10
0 U
z 0
R
LU
50
filtered b; 1.5 mm, Al -‘
z
LU
,
10
I-
I-
20
4.
produced
The when
ENERGY
shape
target
In
(keV)
of the x-ray
a thick
(I)
50
140
30
PHOTON
Figure
spectrum
is struck
0.01
by an
electron beam can be approximated by a tnangular shape. In this illustration, 50 and 60 kVp spectra are represented by heavy and light solid lines. Filtering of the x-ray beam will remove many of the low-energy x-ray photons. The 50.kVp spectrum, filtered by 1 .5 mm of aluminum, is shown by the
dashed
0.05
KVP
line.
ic
20
110
PHOTON
60
ENERGY
100
150
(keV)
Figure 5. The attenuation of x rays between 1 0 and 1 50 keY is the result of three independent processes: the photoelectric process, coherent scattering, and the Compton process. The total attenuation coefficient of the beam at any energy is the sum of the three individual attenuation coefficients.
X-RAY SPECTRA
The previous analysis, nect for a monoenengetic lion have trates the
The
first
a spectrum behavior
which used a single beam of radiation. of photon of an x-ray
approximation
energies. The system in which
of the spectrum
angular distribution of intensity ure 4. The initial spectrum (50 the tube voltage can be changed. initial spectrum produced at 60 number of photons (intensity)
.
kVp
the
line
relative
to that
under
attenuation Almost
coefficient all practical
u, is only
sources
con-
of radia-
following simplified model illusthe x rays have a range of energies.
produced
by a thick
x-ray
target
is a tn-
versus energy. This is shown schematically in FigkVp) can be modified in two principal ways. First, In Figure 4 , the lighter straight line shows the kVp. Increasing the voltage increased the total This is seen by the greater total area under the 60-
50-kVp
line.
The
average
photon
energy
has
also been shifted toward a higher value as the voltage was increased. Second, the shape of the spectrum can be modified by allowing the beam to pass through an absorber on filter on its way to the patient. Filtration usually removes more of the lower-energy photons than the higher-enengy photons. The result of filtration en with an increased average beam some of the filtration is accomplished
lope
of the x-ray
tube
or portions
is a reduction energy (beam by means
of the
light
of x-ray beam intensity togethquality) In an imaging situation, of structures, such as the enve-
localizer.
.
Additional
aluminum
fil-
tration is usually added to bring the beam quality to an acceptable value. The genenal appearance of the filtered spectrum shown in Figure 4 is typical of most diagnostic beams. The actual beam parameters are close to those used in clinical xeromammography.
106
U
RadioGraphics
U
Hubbard
Volume
10
Number
1
PHOTON U. LU
TRANSMISSION
IN
0
WATER
U
z
LU
I-. IU) U)
.c z 0 U) U) U)
Figure
6.
thickness
properties
and
x rays properties change rapidly with energy, transmission will also vary considerably. Since x-ray photons of very low energy
-I
2
5 U.
The attenuation
of tissue determine the fraction that is transmitted. Since attenuation
negligible to the transmissibility, tribute information
0 01
PHOTON
(keV)
ENERGY
X-ray beam filtration used for screen-film complicated than that previously described tration is done properly, the beam quality should be noted here that mammographic mography
and
Two
for
reference
xenomammography
energies
are
are shown
of
have
they do ofnotthe concontent ra-
diograph.
Most of the x-ray
that
is not
transmitted
utes
to the patient’s
beam
energy
eventually
Integral
contnib-
dose.
mammography is considerably more and will not be discussed here. If filis lowered for this special situation. It beams optimized for screen-film mamnot
interchangeable.
on the horizontal
axis of Figure 4 They are shown because most of the photons that are useful for breast imaging lie between these two energies. The physical explanation for these particular values follows. For the purpose of discussing x-ray interactions, one may simplify the descniption of the patient to that of a uniform cient for water as a function of energy tions of x rays with matter at diagnostic ton interactions are the most important. attenuations
caused
by
each
.
thickness of water. The attenuation coeffiis shown in Figure 5 Of the three interacenergies, the photoelectric and the CompTotal attenuation is the sum of the
individual
.
mechanism.
Contrast
in mammography
is
almost totally attributable to the photoelectric interaction because it is a function of the atomic number Z. The quantity of Compton attenuation is mainly a function of density; this varies slightly in the breast. The upper reference energy in Figure is that at which less than 25% of the total interactions are due to the photoelectric effect. Above this energy, most of the x-ray interactions in the patient will create radiation dose without creating much tissue contrast. At a given energy, if scatter is disregarded, the attenuation coefficient and the thickness
determine
the
fraction
of the
incident
energy
that
will
5
be transmitted
6 shows two transmission curves: one for 5 cm of water (an approximation ofthe breast) and one for 20 cm ofwater (an approximation of the abdomen) The lower reference energy in Figure 6 corresponds to a i % transmission through 5 cm of water. Photons with lower energy are almost totally absorbed in tissue. They contribute to patient dose without yielding diagnostic inforthrough
an
object.
Figure
.
mation.
These
photons
should
be excluded
from
incident
to produce
in thick
obtained
January
As shown in Figure 6, the low-energy higher. Therefore, it is not possible patient parts that is comparable to that
the
as possible. ten is much
1990
reference
x-ray
point
beam
radiographic
in breast
as much
for 20 cm of wacontrast
images.
Hubbard
U
Ra4ioGrapbics
U
107
>.
0.14
SPECTRAL THROUGH
I-
Figure
7.
The filtered
50-kVp
x-ray
trum
shown in Figure 4 is further by its transmission through 5 cm
spec-
the effect
.
a “filtered
WATER
transmitted
?
beam”
0.2
o.
“
i
0
10
20
(50
medium,
the
through
kVp
the
+ i-mm
30
40
ENERGY
PHOTON
beam is incident on an absorbing are modified by its transmission
of passing
5 cm
transmissIon factor
modified ofwater.
The high-energy portion of the transmitted spectrum is mainly determined by the shape of the corresponding portion of the incident spectrum. The shape of the low-energy portion of the transmitted spectrum is mainly determined by the attenuation properties of 5 cm of water.
If an x-ray its spectrum
TRANSMISSION
50
(keV)
intensity
and
medium.
shape of 7 shows
Figure
aluminum
filtration)
through 5 cm ofwater. The incident spectrum has a peaked shape. The transmitted on exit spectrum has a narrower peaked shape. The intensity of the transmitted spectrum is normalized to match the intensity of the incident spectrum. This normalization allows us to compare the shapes of the two spectra. It may be seen that the lower-energy portion of the transmitted spectrum is primarily shaped by the transmission characteristics of water, while the high-energy sides of the two spectra are similar.
The filtered
shapes of the spectra of the transmitted beams, 50-kVp beams are incident on 5 cm of water,
when
both
filtered
and
are shown in Figure intensity of the filtered
un-
8. The beam
two beams have been normalized by increasing the by approximately 50%. The spectra of these two exit beams are strikingly similar. The principal difference is that the filtered input beam produces a spectrum that is shifted
slightly
toward
higher
energy.
Thus,
there
will
be
a small
reduction
in
contrast in the images produced by the filtered beam. Figure 9 shows the spectra of the corresponding incident beam. The total input intensity of the unfiltered beam is much higher than that of the filtered beam. Thus, the filtered beam will produce much less patient dose. This is true even though it was necessary to increase the quantity of x-ray production by about 50% for the case of the filtered beam. This latter requirement increases heat loading on the x-ray tube and may mequine an increase in exposure time.
BREAST IMAGING
Requirements anatomy and signs
of cancer.
cause tween
contrast fat and
is generally soft tissue.
erable
contrast.
Since
is usually
tial
108
U
Ra4ioGrapbks
for breast imaging are dominated by the of objects such as masses and calcifications,
modest.
resolution
U
Both
Hubbard
The
as well
normal
and
abnormal
structures
nature are
created by the small differences Small amounts of calcium in the
the typical visualization
as adequate
breast
calcification
of small
calcifications
of normal
which
are
difficult
to image
in atomic breast can
is small,
breast
diagnostic
the absolute
requires
be-
number beyield consid-
contrast
excellent
spa-
contrast.
Volume
10
Number
1
c\
>..
INCIDENT
X-RAY
I-. U)
TRANSMITTED
z
SPECTRA
SPECTRA
LU
I-.
I-
filtered
z
U)
z
LU I-
unfiltered
LU
z
>
unfiltered
-
LU
I-
>
I-
-I LU
LU
0
10
20
30
PHOTON
40
50
ENERGY
0
(keV)
8, 9.
(8)
ted through
The
final
spectra
5 cm of water
low-energy
been
absorbed
scale)
that
are of similar
photons
from
the
by the aluminum.
would
yield
of the
the
The normal shape ties if compression used to accommodate ceptable.
makes
or unfiltered
and intensity.
unfiltered
spectrum,
These
are the incident
(9)
spectra
shown
entire
was
ENERGY
50
(keV)
of the possible
such
breast the
as ultrasonography
requirements system.
recognized
early
which
would
spectra
have
(plotted
special
need
in the history
thickness
for
low
almost otherwise
on the same photon much
results
meet
equalizes
the
demanding
in breast
x-ray
for
of mammography.
is
trans-
image
These imaging
me-
and
res-
are specialtechnol-
imaging.
led to progressive
kilovoltage
intensidetector be unac-
contrast
mammography. systems. Other
enless
in a much
(high-contrast)
roles
have
transmit-
variation in transmitted If a wide-latitude loss of contrast may
of narrow-latitude
have The
beams
absorbs
8. At higher energies, it has
to a uniform
use
of mammography
imaging
50-kVp
The water
more intensity; at lower intensity of the filtered spectrum
ceptoms. Theme are two image receptor systems that olution requirements of general and screening ized screen-film systems and xeroradiographic The
110
in Figure
of the breast produces a wide is not used during mammography. the beam, the corresponding
Compression and
filtered
shape
transmitted
ergies, the filtered beam has a little intensity. The markedly lower total lower dose to the patient.
ogies
30
9.
Figures
mission
20
PHOTON
8#{149}
all the
10
specialization
adequate
Direct
of the
subject
exposure
contrast
of industri-
al x-ray film was shown to produce usable images at doses of about i 00 mGy. Today, the use of specialized screens and films has reduced dose to around i mGy, while producing diagnostically useful images. Some of this dose reduction is ob-
tamed filter,
by use of a dedicated as well
as with
In xeroradiography,
x-ray
an appropriate
a charged
unit
equipped compression
selenium
plate
with
a molybdenum
target
and
device.
is used
as the detector,
along
with
a unique processing mechanism. It overcomes many of the problems associated with varying tissue thickness. The sensitivity of the plate is not as great as that of current screen-film systems. Even with separately optimized target and filter combinations (molybdenum-aluminum or tungsten-aluminum), xeromadiography presently requires doses of about 3 mGy.
January
1990
Hubbard
U
RadioGrapbics
U
109
0.7
-
BREAST CANCER INCIDENCE MASSACHUSETTS
0 6 -
>-
.
0 5
LOW
DOSE
BREAST CANCER INCIDENCE
FLUOROSCOPY
-
.-
-
I-
&o.i
-
0.4
0.2-
-
0. 3
r15
-
.‘
Exams
‘-NaturaI
--.-#{149}-.--
J
__
FIT
I
IncIdence
05
BREAST
DOSE,
Grays
BREAST
Grays
DOSE.
10. 11. Figures 10, 11. (10) High doses of radiation have been correlated with an increased incidence of cancer. In this illustration, high fluoroscopic doses correlate with an increased incidence of breast cancer. The large errors (vertical bars) make it impossible to define fully the functional form of the relationship between dose and risk. The linear, no-threshold hypothesis is shown by the dotted line. (11) The graph is an expansion of the low-dose portion of Figure 1 0 so as to analyze the dose delivered in modern mammography. Under the linear no-threshold hypothesis, the total incidence of breast cancer in a population
be seen grams
DOSE
that were
of women
this is very
undergoing
close
1 5 mammographic
to the incidence
examinations
in the population
is shown.
in which
It can
no mammo-
obtained.
The dose delivered to a patient during irradiation has an associated risk. Figure iO demonstrates the risk of breast cancer induction due to high radiation doses. The uncertainties, as shown by the error bars, are consistent with either the usual linear dose-risk fit on other functional models. However, the doses required for modcnn mammography are much less than those used in the study upon which Figure i 0 is based. Figure i 1 , an enlarged detail from Figure 1 0, illustrates the risk associated with modern screen-film mammographic examinations. If the linear doserisk hypotheses are correct, the total risk associated with i 5 mammograms is the value plotted in Figure i i This is a small value in comparison to the natural risk of breast cancer. Even though the risk may be small for an individual patient, it .
could
be considered
significant
when
millions
of patients
are irradiated.
Many different doses could be defined. For example, one might consider the dose at the point on the patient’s skin at which the beam enters (skin dose) , the dose at the midline of the breast, or the average dose to the glandular tissue of the breast. These three kinds of doses vary in their accuracy, utility, and ease of measumement.
Figure three that ical model backs skin
i 2 schematically
associated dose
shows
these
can be measured directly. The on physical phantom in order with
is delivered
skin
dose.
to different
As shown points
concepts.
Skin
dose
other two require to be evaluated. in Figure in the
patient.
is the only
one
of the
some sort of mathematThere are several draw-
i 3 , for Skin
different dose
views, is not
the
directly
additive to obtain a sensible total dose for a multiview study. Patient dose should be minimized to a value as low as is consistent with clinically adequate imaging. This can be accomplished by a variety of means. We have seen how the selection of an appropriate beam quality and filter is of major importance. The choice of image receptor and processing will also effect the dose. For
110
U
RadioGrapbics
U
Hubbard
Volume
10
Number
1
SKIN
or
ENTRANCE
AVERAGE
()
GLANDULAR
MIDLINE
12. Figures central at which
the
defined
central
shown
have
doses
example,
(13)
skin
for
maximum
fraction
of the
We must
If two
beams
enter
by different
film
the
practical
can
processing.
compression,
beam
from
will
dose
is not
directions,
together.
each
Average
the dose
by 1 5%-30%
examination
reduce
dose,
relative
to
technique,
since
the
effective
tissue
thickness
benefits
of mammography.
such
transmitted
is reduced.
PUBLIC HEALTH
death
from
breast
cancer
can
be
and
different
values
Proper
as the
the meal risks
breast
glandular
to all of the glandular
risk
against How tribute
identify
increases
average
delivered
be added.
can reduce
film
The
to add these
views
processing
breast.
of the dose
at which the as the point
of mammography is the induction of cancer due to the examining x-ray beam. The primary benefit is the detection of cancers at stages for which there will be increased mates of survival and cure compared with those for unexamined populalions. It is to be noted that epidemiologic studies of benefit defined in this way automatically include the risks of irradiation. In addition, real patient benefits are often associated with true-negative examinations. Efforts are also made to reduce benefits and risks to monetary terms. Assuming that the cost to society of an “early”
clearly
the
It is inappropriate
1 .5-minute
x-ray
through
It is the average
dose.
produced
3-minute
required
as the
point.
in gray.
its own
glandular
that
ray is halfway
at any given
tissue
will
13. The skin or entrance dose is measured at the point beam enters the breast. The midline dose is defined
12, 13. (12) ray of the x-ray
obtained,
this
information
the cost of a mammogmaphic screening program. can the benefits of mammography be improved? by developing techniques for improving image
Medical quality
can
The
be
real
balanced
physicists and reducing
conpa-
tient dose. Quality assurance programs can optimize and stabilize the quality of existing systems. The accreditation of facilities that meet certain operating and imaging standards becomes one way in which the public is told where quality examinations can be obtained. Quality assurance methods used for mammography are similar to those applicable to general radiography. These include control of the mechanical aspects of the equipment (including patient positioning devices) , x-ray beam calibration, the evaluation of focal spot characteristics, and the performance of automatic exposure control systems. The quality of image processing is important for all of radiography but is critical for mammography. Retake and artifact analysis are tools that often give insight into overall system problems. Because mammography makes special demands on the imaging system, it mequines
January
a specialized
1990
quality
assurance
program.
For example,
a 2#{176} star
pattern
Hubbard
is
U
RadioGrapbics
U
111
-.--.---S
-
14. Figures 14, size on many
15.
(14)
When
mammographic
ter to be seen. This pattern (15) For many mammographic measure the effective focal
15. a 2#{176} star pattern is used
to measure
machines,
blur
the
is, however, systems, spot size.
resulting
pattern
the effective is too
close
focal
spot
to the
cen-
adequate for most general radiographic systems. the 0.5#{176} star pattern can be used to successfully
17.
-
Figures
16, 17 (16) The processing of a charged, unexposed xeroradiographic plate is called dark dusting. In the resulting image, the numerous white spots are indicative of a plate that has exceeded its useful life. The dank spots demonstrate that toner powder is not being dispersed properly by the processor. (17) The phantom used as part of the American College of Radiology mammography accreditation program contains fibers, specks, and circular disks. This radiograph shows some of the most prominent objects in the phantom as well as some system artifacts.
suitable for evaluation of the effective image shown in Figure 14 was made small focal spot and imaging geometry
focal spot size for general radiography. The with such a star on a mammographic unit; the results in a blur pattern too close to the
center for evaluation. Figure i 5 shows the the same machine. Even though this image pattern is none too Quality assurance
mographic
units
large. methods
as well
for
image
as to screen-film
image
covers
processing
systems.
of a 0.5#{176}star
pattern
an 8 X i 0-inch should
Figure
be applied
U
RadioGrapbics
U
Hubbard
Volume
on
the blur
to xeromam-
i 6 is an example
“dark dusting,” a useful test of the xeromadiographic system. A prepared has not been exposed to radiation is processed, and the resulting image clues to the proper performance of the plate and processor.
112
made
cassette,
of plate provides
10
that
Number
1
Table 1 Typical Parameters
for Two
Radiographic Chest Radiography
Parameter Kilovoltage (kVp) Fi!ter Skin dose (mGy)
Focal
Procedures Mammography
140 0.1 Cu + 1 Al 0. 1
30 0.03 Mo 2
1.0
0.5
1 83 22
65 4.5
spot
size (mm) Source-image distance (cm) Thickness (cm)
Note-Al
aluminum,
Cu
copper,
Mo
molybdenum.
Image quality is one of the major components of current accreditation programs for mammogmaphic facilities. A resolution “standard” developed by Dr Harold Lasky (i) is used to test the performance of mammognaphic systems in accreditation programs such as that of the Illinois Chapter of the American Cancer Society. This standard consists of several small specks of zinc tellunide and rock salt, which
are placed
on a patient’s
breast
and
imaged
during
mammography.
The visualiza-
tion of the standard is a major criterion for approval in this accreditation program. The national mammography accreditation program of the American College of Radiology includes testing the performance of the imaging system on a phantom containing fibers, specks, and circular “masses.” A radiograph of this phantom is shown in Figure 1 7. The standard requires visualization of an adequate number of each of these objects.
The
materials
lion
radiography.
require with
discussed
different those
in this
However,
the
optimizations.
article
can
special
Table
i compares
of screen-film
mammography. tween the requirements for these two 1 The chest has a greater thickness; quined for chest radiography. 2.
Objects
soft
large that radiographs. with that
in the
tissue
have
bone
much
and
greater
the difference
a much lower radiographic contrast Film with lower contrast is used used in mammography. In addition,
kVp are used 3
chest
and
to minimize
subject
to other of other
chest
There are examinations. thus, a more
.
tween
be extended
requirements
forms
main
contrast.
x-ray The
soft tissue
OThER RADIOGRAPHIC
will
parameters
differences
penetrating
between
tasks
radiographic
three
inherent
of projec-
imaging
beam
is me-
difference
and
SYSTEMS
be-
lung
be-
are so
is required to obtain optimal chest for chest radiography compared x-ray tube voltages in excess of 100
contrast.
The spatial resolution that is adequate for chest radiography is lower than that required for mammography. Since the sensitivity of radiographic detectors and their spatial resolutions are trade offs with each other, this permits the use of a much more sensitive screen-film system for chest radiography. Among other things, this reduces the exposure time, which in turn minimizes motion unsharpness. The use of a more sensitive image receptor and a higher kilovoltage peak in chest radiography results in lower patient dose than that required for mammography. As shown in Table i , the use of a composite copper-aluminum filter can me.
duce the dose even general radiography.
1
January
.
Lasky
HJ.
Quality
bly
and
ber
30-December
1990
Annual
further
than
standard Meeting
that
needed
for mammograms. of the
Radiological
with
the all-aluminum
Presented Society
at the 72nd
of North
America,
filter
Scientific Chicago,
used
for
Assem-
REFERENCE
Novem-
5, 1986.
Hubbard
U
RadioGrapbks
U
113