sensors Article

A Consistency Evaluation and Calibration Method for Piezoelectric Transmitters Kai Zhang 1, *, Baohai Tan 1 and Xianping Liu 2 1 2

*

School of Geosciences and Technology, China University of Petroleum, Qingdao 266555, China; [email protected] Logging Branch of China National Petroleum Corporation Bohai Drilling Engineering CO. LTD., Tianjin 300280, China; [email protected] Correspondence: [email protected]; Tel.: +86-156-2108-1452

Academic Editor: Vittorio M. N. Passaro Received: 27 March 2017; Accepted: 26 April 2017; Published: 28 April 2017

Abstract: Array transducer and transducer combination technologies are evolving rapidly. While adapting transmitter combination technologies, the parameter consistencies between each transmitter are extremely important because they can determine a combined effort directly. This study presents a consistency evaluation and calibration method for piezoelectric transmitters by using impedance analyzers. Firstly, electronic parameters of transmitters that can be measured by impedance analyzers are introduced. A variety of transmitter acoustic energies that are caused by these parameter differences are then analyzed and certified and, thereafter, transmitter consistency is evaluated. Lastly, based on the evaluations, consistency can be calibrated by changing the corresponding excitation voltage. Acoustic experiments show that this method accurately evaluates and calibrates transducer consistencies, and is easy to realize. Keywords: piezoelectric transducers; transmitter consistency; impedance analyzer; consistency evaluation; consistency calibration

1. Introduction Acoustic transducers are used to realize energy transformations between alternating electric signals and sonic signals, and these devices play an important role in many domains, such as industry [1], agriculture [2], military [3], and medicine [4]. These transducers can be divided into piezoelectric and magnetostrictive transducers according to their materials, and piezoelectric transducer technologies that use piezoelectric effect to detect energy transformation are known to be more mature and useful [2,5,6]. Transducers that transform electric energy into sonic energy are called transmitters, and those that transform sonic energy into electric energy are called receivers. With the development of transducer-applied technologies, array transducer and transducer combination technologies are evolving rapidly [7,8]. For example, in the process of petroleum exploration, acoustic transmitters are well-ordered and phase-arrayed to control the orientation and enhance the energy of sound waves [9–11]; in the process of nondestructive testing, the inspection performances are limited to regular surfaces, and flexible phased-array techniques are used to compensate for surface irregularities and to fit the surface [12]. While adapting transmitter combination technologies, the parameter consistencies between each transmitter are extremely important because they can determine a combined effort directly. On the other hand, transmitter consistencies can influence the consistencies between each tool. Thereafter, transducers must be chosen to obtain good consistencies. However, transducer manufacturing technologies are very complex and have low production pass rates, and the consistencies are influenced by many factors and are difficult to guarantee. Thus, the consistencies between transmitters must be evaluated and calibrated. Sensors 2017, 17, 985; doi:10.3390/s17050985

www.mdpi.com/journal/sensors

Sensors 2017, 17, 985

2 of 9

and are difficult to guarantee. Thus, the consistencies between transmitters must be evaluated and calibrated. Sensors 2017, 17, 985 2 of 9 To date, the reciprocity theorem is mostly used to measure transducer consistencies [13,14]. This method is accurate, but its experiment process is tedious and only suited for reciprocity transducers. To date,electronic the reciprocity theorem usedbytoimpedance measure transducer [13,14]. Transducer parameters can is bemostly measured analyzers consistencies and, in this study, a This method is accurate, but its experiment process is tedious and only suited for reciprocity variety of transmitter acoustic energies caused by electronic parameter differences are analyzed, and transducers. Transducerand electronic parameters can beonmeasured by impedance analyzers and, in a consistency evaluation calibration method based these parameters is proposed. Experiments this study, a variety of transmitter acoustic energies caused by electronic parameter differences are show that this method can efficiently evaluate and calibrate transducer consistencies. analyzed, and a consistency evaluation and calibration method based on these parameters is proposed. Experiments show that this method can efficiently and calibrate transducer consistencies. 2. The Consistency Evaluation Method Based on evaluate an Impedance Analyzer Impedance analyzers deliver low-level current the object that needs to be measured, and 2. The Consistency Evaluation Method Based on an into Impedance Analyzer the object’s resistance and reactant characteristics are then recorded. By changing the current Impedance analyzers deliver low-level currentaccurately into the object that needs to be measured, and the frequency continuously, impedance analyzers perform a wide range of frequency object’s resistance and reactant characteristics are then recorded. By changing the current frequency measurements [15]. Thereafter, many parameters and performances are obtained by calculating the continuously, impedance accurately perform range of frequencyhalf measurements [15]. resistances and reactantanalyzers characteristics, such as thea wide resonance frequency, power points, Thereafter, manyfrequency, parametersmaximum and performances are obtained by calculating the resistances and reactant anti-resonance admittance, mechanical quality factor, free capacitance, static characteristics, such as the resonance frequency, half power points, anti-resonance frequency, maximum capacitance, dynamic capacitance, dynamic impedance, dynamic inductance, and electromechanical admittance, mechanical quality factor, free capacitance, static capacitance, dynamic capacitance, coupling factor. dynamic impedance, dynamic inductance, electromechanical coupling factor. Resonance is a phenomenon in which and a vibrating system or external force drives another system Resonance is a phenomenon in which a vibrating system or external force drives another system to oscillate with greater amplitude at a specific preferential frequency. Frequencies at which the to oscillate with greater amplitude at a specific preferential frequency. Frequencies at which the response amplitude is a relative maximum are known as the resonant (or resonance) frequencies of response amplitude is a relative maximum are known as theforces resonant resonance) of the system. At resonant frequencies, small periodic driving have(or the ability to frequencies produce large the system. At resonant frequencies, small periodic driving forces have the ability to produce large amplitude oscillations. The maximum conductance is the value at the resonance frequency, which is amplitude oscillations. The maximum is the value resonance frequency, which the reciprocal of dynamic resistance. Theconductance mechanical quality factoratisthe a dimensionless parameter that is the reciprocal of dynamic resistance. The mechanical quality factor is a dimensionless parameter describes the energy that a piezoelectric object consumes to overcome the internal friction at the that describes the energy piezoelectric an object consumes to overcome the to internal friction at the resonant frequency, and itthat alsoacharacterizes oscillator’s bandwidth relative its center frequency. resonant frequency,object and ittranslates also characterizes an oscillator’s bandwidth to its center frequency. The piezoelectric mechanical into electrical energy relative or electrical into mechanical The piezoelectric object translates mechanical into electrical energy or electrical into mechanical energy, energy, and the electromechanical coupling factor is used to describe the degree of these energy and the electromechanical coupling factor is used to describe the degree of these energy translations. translations. A A piezoelectric piezoelectric transducer transducer can can be be regarded regarded as as only only aa capacitance capacitance (C (C00)) in in static static state, state, whereas whereas dynamic impedance must be considered while it is vibrating and emitting energy, dynamic impedance must be considered while it is vibrating and emitting energy, which which can can be be described by a capacitance (C ), a resistance (R ), and an inductance (L ) that are connected 1 1 1 described by a capacitance (C1), a resistance (R1), and an inductance (L1) that are connected in in series series [16]. [16]. The The equivalent equivalent circuit circuit of of the the transducer transducer is is shown shown in in Figure Figure 1. 1.

L1 AC

C0

C1 R1

Figure 1. Equivalent circuit of a transmitter; only C0 is considered in static state, whereas C1, R1 and L1 Figure 1. Equivalent circuit of a transmitter; only C0 is considered in static state, whereas C1 , R1 and L1 must be considered connecting in series while vibrating. must be considered connecting in series while vibrating.

According to Ohm’s law, based on the assumption that the input voltage of the transducer is U According Ohm’s law, basedpower on theofassumption that circuit the input voltage of the transducer is U at at one moment,to the instantaneous the equivalent is as follows: one moment, the instantaneous power of the equivalent circuit is as follows: (1) P  UI  U 2(G f  jB f )  U 2G f  jU 2 B f P = U I = U 2 ( G f + jB f ) = U 2 G f + jU 2 B f (1) where Gf and Bf are the total conductance and the total susceptance at a certain frequency, U2Gf is the 2Bf is the reactive power. real power and where Gf and Bf U are the total conductance and the total susceptance at a certain frequency, U2 Gf is the real power and U2 Bf is the reactive power.

Sensors 2017, 17, 985

3 of 9

Based on the assumption that the spring constant in Hooke’s Law of one transducer is k, the average vibration displacement of the transducer in a very short time (∆t) is ∆X, and according to Hooke’s law, the mechanical power is as follows: P0 = k × ∆X 2 /∆t

(2)

Based on the assumption that all the electrical real powers are converted into mechanical power, then, U 2 G f = k∆X 2 /∆t (3) Based on the aforementioned Equation (3), we conclude that the transducer conductance is proportional to the square of its displacement, whereas the input voltage (U), the spring constant (k), and the time (∆t) are certain. To confirm the aforementioned conclusion, analogue simulations were performed to the most common-used laminated transducer and cylindrical transducer by ANSYS software (Ansys Inc., Canonsburg, PA, USA) [17]. In the process of analogue simulations, PZT4 was used for piezo material, whose density is 7500 kg/m3 . For the laminated transducer, the length of the piezoelectric ceramic block was 12 mm, while the width was 9 mm and the height was 1 mm. Meanwhile, the length of the substrate metal block was 24 mm, while the width was 9 mm and the height was 1 mm. The polarization of the piezoelectric ceramic block was along the through-thickness orientation. For the cylindrical transducer, the inner diameter for the piezoelectric ceramic block was 46 mm, the exterior diameter was 53 mm and the height was 48 mm. The polarization of the piezoelectric ceramic block was along the radial orientation. Finite element models for laminated and cylindrical transducers are shown in Figure 2a,b. Normalized simulation results are presented in Figure 2c,d, where all conductance curves coincide with the square of displacement curves (especially at the frequencies near the resonance frequency). Transmitters should always work at nearby resonance frequencies. In the experiments and simulations, almost all the transmitter conductance curves are found to be similar to the Gaussian curve while working at those frequencies. Thereafter, the following empirical expression is proposed: G ( Fx , Gmax , Qm , Fs ) = Gmax eln 0.5×4Qm

2 /F 2 ×( F − F )2 s x s

(4)

where Fx is the working frequency, Fs is the resonance frequency, Gmax is the maximum conductance, and Qm is the mechanical quality factor. Five transducers were selected and measured to confirm the aforementioned empirical expression, including one laminated transducer, one cylindrical transducer, one sandwich lamination structure ultrasonic cleaning transducer, one arc-shaped bending transducer and one much larger arc-shaped bending transducer isolated by rubber. Calculated conductance curves that are based on this empirical expression and actual measurement conductance curve are shown in Figure 3. Therefore, the calculated and actual values of nearby resonance frequencies are in substantial agreement. When Equation (3) is substituted with Equation (4), the following equation can be obtained: r ∆X =

U 2 ∆t 2 2 2 Gmax eln 0.5×4Qm /Fs ×( Fx − Fs ) k

(5)

As is well known, the sound pressure fired by one transmitter is proportional to the vibration displacement of its surface. Because the vibration displacement ∆X is difficult to measure, the peak-to-peak sound pressure value of the emitted acoustic signals is used to instead. Thereafter, transducers of the same material have the same spring constant (k), and their input voltage (U) and time (∆t) are determined values. The following equation is deduced: Psound ∝ U

q

Gmax eln 0.5×4Qm

2 /F 2 ×( F − F )2 s x s

(6)

G ( Fx , Gmax , Qm , Fs )  Gmax eln 0.54Qm

2

/ Fs 2 ( Fx  Fs )2

(4)

where Fx is the working frequency, Fs is the resonance frequency, Gmax is the maximum conductance, and Qm17,is985 the mechanical quality factor. Sensors 2017, 4 of 9 Five transducers were selected and measured to confirm the aforementioned empirical expression, including one laminated transducer, one cylindrical transducer, one sandwich lamination Equation ultrasonic (6) is deduced based on the one assumption that all thetransducer electrical and real one powers structure cleaning transducer, arc-shaped bending muchare larger converted into mechanical power. However, part of the energy is consumed in the process of on arc-shaped bending transducer isolated by rubber. Calculated conductance curves that are based electromechanical and theactual electromechanical coupling factor (K usedshown to describe the 3. P ) is are this empirical coupling, expression and measurement conductance curve in Figure degree of electromechanical coupling. On the other hand, piezoelectric transducers must consume Therefore, the calculated and actual values of nearby resonance frequencies are in substantial energy to overcome internal friction, and the mechanical quality factor is used to describe it. agreement.

Sensors 2017, 17, 985 Sensors 2017, 17, 985

0.8 0.8

1.0 1.0

Displacement Displacement Conductance Conductance Square of displacement Square of displacement

0.6 0.6 0.4 0.4 0.2 0.2 0.0 0.0 12000 12000

4 of 9 4 of 9

(b)

Normalized Value Normalized Value

Normalized Value Normalized Value

1.0 1.0

(a)

Displacement Displacement Conductance Conductance Square of displacement Square of displacement

0.8 0.8 0.6 0.6 0.4 0.4 0.2 0.2 0.0 0.0

12500 12500

13000 13000

13500

14000

13500 14000 Frequency(Hz) Frequency(Hz)

14500 14500

15000 15000

2000 2000

4000 4000

6000 6000

8000 8000

(c) (c)

10000

12000

14000

16000

18000

20000

10000 12000 14000 16000 18000 20000 Frequency(Hz) Frequency(Hz)

(d) (d)

Figure 2. Finite element models for (a) laminated and (b) cylindrical transducers. Normalized Figure 2. Finite element element models models for for (a) Figure 2. Finite (a) laminated laminated and and (b) (b) cylindrical cylindrical transducers. transducers. Normalized Normalized simulation results ((c) ((c) for for laminated and (d) for cylindrical transducers) indicate that the conductance simulation results laminated and (d) for cylindrical transducers) indicate that the conductance simulation results ((c) for laminated and (d) for cylindrical transducers) indicate that the conductance curves coincide with with the square square of displacement displacement curves. curves coincide coincide curves with the the square of of displacement curves. curves. 35 35

30 30

G(ms) G(ms)

20 20 15 15 10 10

30 30

20 20 15 15 10 10

5 5

5 5

0 0

0 0

12000 12000

14000 14000

16000 18000 16000 18000 Frequency(Hz)

20000 20000

22000 22000

T4_actual_conductance T4_actual_conductance T4_calculated_conductance T4_calculated_conductance T5_actual_conductance T5_actual_conductance T5_calculated_conductance T5_calculated_conductance

25 25

G(ms) G(ms)

T1_actual_conductance T1_actual_conductance T1_calculated_conductance T1_calculated_conductance T2_calculated_conductance T2_calculated_conductance T2_actual_conductance T2_actual_conductance T3_actual_conductance T3_actual_conductance T3_calculated_conductance T3_calculated_conductance

25 25

25000 25000

30000 30000

35000 35000

40000 40000

Frequency(Hz)

45000 45000

50000 50000

Frequency(Hz) Frequency(Hz)

55000 55000

60000 60000

65000 65000

Figure 3. Calculated conductance curves based on the empirical expression and actual conductance Figure 3. Calculated Calculated conductance curves based on the empirical empirical expression expression and and actual actual conductance conductance Figure measured 3. conductance based on the curves by an impedance curves analyzer for five kinds of transducers: a laminated transducer (T1), curves measured by analyzer for of aalaminated transducer (T1), measured by an an impedance impedance analyzer for five five kinds kinds of transducers: transducers: laminated transducer (T1), acurves cylindrical transducer (T2), a sandwich lamination structure ultrasonic cleaning transducer (T3), an aa cylindrical transducer (T2), a sandwich lamination structure ultrasonic cleaning transducer (T3), an cylindricalbending transducer (T2), a sandwich structure ultrasonic cleaning transducer (T3), by an arc-shaped transducer (T4) and lamination a much larger arc-shaped bending transducer isolated arc-shaped bending transducer (T4) and aa much larger arc-shaped bending bending transducer isolated by arc-shaped bending transducer (T4) and much larger arc-shaped transducer isolated by rubber (T5). rubber rubber (T5). (T5).

When Equation (3) is substituted with Equation (4), the following equation can be obtained: When Equation (3) is substituted with Equation (4), the following equation can be obtained:

U 22 t 0.5 4 Qm 2 / Fs 2 ( Fx  Fs )2 X  U t Gmax elnln 0.5  4 Qm 2 / Fs 2 ( Fx  Fs )2 X  k Gmax e k

(5) (5)

As is well known, the sound pressure fired by one transmitter is proportional to the vibration

Sensors 2017, 17, 985

5 of 9

When the electromechanical coupling and mechanical quality factors are considered, Equation (6) can be described as follows: q 2 2 2 (7) Psound ∝ U × K P (1 − 2π/Qm ) Gmax eln 0.5×4Qm /Fs ×( Fx − Fs ) Equation (7) describes the relationship between sound pressures that are caused by working transmitters and the electronic parameters that are measured by impedance analyzers. That is, parameter consistencies of piezoelectric transmitters can be evaluated by measurement based on the impedance analyzers and the calculation based on Equation (7). As the calculation endings based on Equation (7) are proportionality values, the calculation results are normalized by dividing each result by the maximum result. After the influences of the electronic parameters have been calculated successfully, calibrations can be performed by adjusting the input voltages. 3. Experiments and Results To validate the presented theory, eight laminated transmitters with resonant frequencies of about 15 kHz were chosen, and their electronic parameters were measured by an impedance analyzer. Thereafter, their normalized sound pressures were calculated by using Equation (7) (Table 1), where KP , Qm , Gmax , and Fs denote electromechanical coupling factor, mechanical quality factor, maximum conductance, and resonance frequency, respectively. Table 1. Electronic parameters of the eight laminated transmitters and their normalized sound pressure calculation results based on Equation (7). KP , Qm , Gmax , and Fs denote electromechanical coupling factor, mechanical quality factor, maximum conductance, and resonance frequency, respectively. NO.

KP

Qm

Gmax (ms)

FS (Hz)

Normalized Calculation Results

1 2 3 4 5 6 7 8

0.432 0.428 0.431 0.401 0.400 0.380 0.371 0.399

36.941 40.474 32.116 17.817 29.354 39.903 27.454 34.420

21.242 23.018 17.715 10.727 15.584 22.739 17.331 20.204

15368 15496 15420 15268 15500 15686 15840 15560

0.919 0.832 0.745 0.516 0.617 1.000 0.967 0.732

To measure their real sound pressure, the eight transmitters were emitted by the same electrical excitation source with a frequency of 15.5 kHz and an amplitude of 600 V, and the acoustic signals were received by the broadband hydrophone. The acoustic experiments were conducted in a water tank with the dimensions of 5.0 m × 5.0 m × 4.0 m. The fluid density and velocity were 1000 kg/m3 and 1500 m/s, respectively. As shown in Figure 4, the experimental apparatus included an electrical excitation resource, cylindrical transmitter, host computer, positioning control system, multichannel acquisition system, gain-controlled amplifier, and hydrophone [18,19]. The excitation resource produced high-voltage pulses to emit the transducer and synchronize signals to start the high-accuracy synchronous data acquisition system. The hydrophone and transmitters were accurately positioned with the positioning control system. The diagram of a laminated transmitter and the positioning control system are shown in Figure 5. The distance between the hydrophone and laminated transmitters was two meters. Eight transmitters were fixed in one octagonal mounting bracket, and the angle between neighboring transmitters was 45 degrees. In the process of the experiments, only one of these eight transmitters was opposite the hydrophone. While the selected transmitter was being excited by the excitation resource, the hydrophone collected the sound pressure curves. Then, the mounting bracket rotated 45 degrees, causing another transmitter to become excited, and the sound pressure curves were subsequently

Sensors 2017, 17, 985

6 of 9

Figure 5. Lastly, the normalized calculated and measured sound pressure are shown in Figure 6 and Table 2. Sensors17, 2017, Sensors 2017, 98517, 985

6 of69of 9

Host computer Figurecollected. 5. Lastly,The thefiring normalized calculated and measured sound pressure are shown in Figure 6 and voltage curve and eight sound pressure curves are shown in Figure 5. Lastly, the Table 2. normalized calculated and measured sound pressure are shown in Figure 6 and Table 2. Router

Host Positioning Acquisition Excitation computer control Signal resource system system Router Positioning control system

Hydrophone

Transmitter

Acquisition Excitation Signal resource system Water rank

Figure 4. Experimental system for acoustic waveform measurements in a water tank with dimensions of 5.0 m × 5.0 m × 4.0 m. The positioning system allowed the precise adjustment of the relative position Hydrophone between the transmitter and the hydrophone. Positioning Transmitter control system, acquisition system and host computer are connected by router and ethernet. Water rank

As different transducers were excited by different excitation resources, the excitation voltages Figure 4. Experimental system for acoustic waveform measurements in water dimensions for each transmitter could befor adjusted bywaveform the logging tools to makein upaa for thetank lackwith of consistency. The Figure 4. Experimental system acoustic measurements water tank with dimensions of 5.0 m × 5.0 m × 4.0 m. The positioning system allowed the precise adjustment of the relative position adjustment of input excitation voltages can be calculated according to Equation (7). Then, the acoustic of 5.0 m × 5.0 m × 4.0 m. The positioning system allowed the precise adjustment of the relative position between the transmitter and the hydrophone. Positioning control system, system and host The experiments were repeated each transducer was excited by itsacquisition own excitation between the transmitter and thewhile hydrophone. Positioning control system, acquisition systemvoltage. and host computer are connected by router and ethernet.

adjustment of excitation andethernet. calibrated sound pressures are shown in Table 2 and Figure 6. computer are connected byvoltages router and

As different transducers were excited by different excitation resources, the excitation voltages for each transmitter could be adjusted by the logging tools to make up for the lack of consistency. The adjustment of input excitation voltages can be calculated according to Equation (7). Then, the acoustic experiments were repeated while each transducer was excited by its own excitation voltage. The adjustment of excitation voltages and calibrated sound pressures are shown in Table 2 and Figure 6.

(a)

(a)

(b)

(b)

(c)

(d)

Figure 5. The diagram of one laminated transmitter (a), the diagram of the positioning control system Figure 5. The diagram of one laminated transmitter (a), the diagram of the positioning control system (b), the firing voltage curve excited by the excitation resource (c), and eight sound pressure curves (b), the firing voltage curve excited by the excitation resource (c), and eight sound pressure curves collectedby bythe thehydrophone hydrophone(d) (d)while whileeight eighttransmitters transmitterswere werebeing beingexcited, excited,respectively. respectively. collected

(c)

(d)

Figure 5. The diagram of one laminated transmitter (a), the diagram of the positioning control system (b), the firing voltage curve excited by the excitation resource (c), and eight sound pressure curves collected by the hydrophone (d) while eight transmitters were being excited, respectively.

1 0.919 2 0.832 3 0.745 4 0.516 5 0.617 Sensors 2017, 17, 985 6 1.000 7 0.967

1386.4 1312.6 1135.6 1004.8 997.1 1538.8 1457.2

0.901 0.853 0.738 0.653 0.648 1.000 0.947

665.9 703.4 813.0 918.8 925.9 600 633.6

1539.9 1449.1 1563.8 1592.5 1538.3 1582.9 1565.4

0.967 0.910 0.982 1.000 0.966 0.994 0.983

7 of 9

1.0 0.9

Normalized value

0.8 0.7 0.6 0.5 0.4

Calculated sound pressure Measured sound pressure Calibrated sound pressure

0.3 0.2 0.1 0

1

2

3

4

5

6

7

8

9

Transmitter number

Figure 6. The calculated sound pressure curve of eight laminated transmitters based on on Equation (7) (7) Figure 6. The calculated sound pressure curve of eight laminated transmitters based Equation (see Table 1), the measured sound pressure curve obtained by the measurement system in a water (see Table 1), the measured sound pressure curve obtained by the measurement system in a water tank, tank, calibrated sound pressure curve measured the samesystem systemafter afterexcitation excitationvoltages voltagesfor andand thethe calibrated sound pressure curve measured byby the same foreach eachtransmitter transmitterwere wereadjusted adjustedaccording accordingtotoEquation Equation(7), (7),respectively. respectively.

4. Discussion Table 2.and TheConclusions calculated sound pressures of eight laminated transmitters based on Equation (7); thecalculated measured and and normalized peak-to-peak pressures measured by the acoustic waveform The measured results show sound that the consistencies between transducers are not measurement system; the adjustment excitation voltages for each transmitter depending on the satisfactory. Table 2 and Figure 6 show that the calculated results are in accord with the experimental measured sound pressures and Equation (7); the measured peak-to-peak (*) and normalized sound measurement results. Thereafter, by adjusting the excitation voltages for each transmitter, the pressures (*) measured by the same system after voltage adjustment. consistencies between transducers were greatly improved. There are some discrepancies in the measured pressures after the voltage adjustment, because the acoustic experiment method had many Calculation Measured Sound Normalized Adjustment Measured Sound Normalized NO. influencing factors the endings uncertain, such asVoltage the fixed of *the the ResultsmakingPressure (Pa) Pressure (V) strength Pressure (Pa)transducer, Pressure * acoustic condition, and the gas and oil0.901 bubbles outside sheath, which change 1 coupling 0.919 1386.4 665.9of the rubber 1539.9 0.967 2 0.832 1312.6 voltage was0.853 703.4 (a voltage1449.1 0.910not with time. Also, as the excitation adjusted by hand regulator), it was 3 0.745 1135.6 0.738 813.0 1563.8 0.982 perfectly accurate, either. 4 0.516 1004.8 0.653 918.8 1592.5 1.000 This study0.617 introduces the of piezoelectric transmitters that can 5 997.1electronic parameters 0.648 925.9 1538.3 0.966 be 6 1.000 1538.8 Thereafter,1.000 600 between these 1582.9 0.994the measured by impedance analyzers. the relationship parameters and 7 0.967 1457.2 0.947 633.6 1565.4 0.983 transmitter acoustic performances is analyzed, and a consistency evaluation and calibration method for these transmitters is proposed. Acoustic experiments show that this method can evaluate and Astransducer different transducers were excited with by different excitation resources, the excitation voltages calibrate consistencies. Compared the traditional reciprocity theorem, this method is for each transmitter could be adjusted by the logging tools to make up for the lack of consistency. easier to realize and is suited for all transducers (not only reciprocity transducers). The adjustment of input excitation voltages can be calculated according to Equation (7). Then, the acoustic experiments were repeated while each transducer was excited by its own excitation voltage. The adjustment of excitation voltages and calibrated sound pressures are shown in Table 2 and Figure 6.

4. Discussion and Conclusions The calculated and measured results show that the consistencies between transducers are not satisfactory. Table 2 and Figure 6 show that the calculated results are in accord with the experimental measurement results. Thereafter, by adjusting the excitation voltages for each transmitter, the consistencies between transducers were greatly improved. There are some discrepancies in the measured pressures after the voltage adjustment, because the acoustic experiment method had many influencing factors making the endings uncertain, such as the fixed strength of the transducer, the acoustic coupling condition, and the gas and oil bubbles outside of the rubber sheath, which change

Sensors 2017, 17, 985

8 of 9

with time. Also, as the excitation voltage was adjusted by hand (a voltage regulator), it was not perfectly accurate, either. This study introduces the electronic parameters of piezoelectric transmitters that can be measured by impedance analyzers. Thereafter, the relationship between these parameters and the transmitter acoustic performances is analyzed, and a consistency evaluation and calibration method for these transmitters is proposed. Acoustic experiments show that this method can evaluate and calibrate transducer consistencies. Compared with the traditional reciprocity theorem, this method is easier to realize and is suited for all transducers (not only reciprocity transducers). Acknowledgments: This research is financially supported by China Postdoctoral Science Foundation (F1501072B). The authors would like to thank Xiaolong Hao at China University of Petroleum for his help in experiments. Author Contributions: Kai Zhang conceived and performed the theoretical derivation; Kai Zhang and Baohai Tan designed and performed the experiments; Xianping Liu designed and performed the analogue simulations; Kai Zhang and Xianping Liu wrote the paper. Conflicts of Interest: The authors declare no conflict of interest.

References 1. 2. 3.

4. 5. 6. 7.

8. 9. 10. 11. 12. 13. 14. 15. 16.

Na, W.S.; Baek, J. Impedance-based non-destructive testing method combined with unmanned aerial vehicle for structural health monitoring of civil infrastructures. Appl. Sci. 2017, 7, 15. [CrossRef] Lipper, A.; Borden, J. Urethane transducer encapsulation versus oil filled boot encapsulation of piezoelectric transducers. In Proceedings of the 2012 OCEANS, Hampton Roads, VA, USA, 14–19 October 2012. Lawry, T.J.; Wilt, K.R.; Roa-Prada, S.; Ashdown, J.D.; Saulnier, G.J.; Scarton, H.A.; Das, P.K.; Pinezich, J.D. Electrical optimization of power delivery through thick steel barriers using piezoelectric transducers. In Proceedings of the Conference on Energy Harvesting and Storage-Materials, Devices, and Applications, Orlando, FL, USA, 5–6 April 2010. Lee, H.S.; Kundu, J.; Kim, R.N.; Shin, Y.K. Transducer of erbb2.1 (tob1) as a tumor suppressor: A mechanistic perspective. Int. J. Mol. Sci. 2015, 16, 29815–29828. [CrossRef] [PubMed] Li, D.H.; Wu, M.; Oyang, P.X.; Xu, X.F. Cymbal piezoelectric composite underwater acoustic transducer. Ultrasonics 2006, 44, E685–E687. [CrossRef] [PubMed] Chi, F.L.; Wu, Y.Z.; Yan, Q.B.; Shen, Y.H.; Jiang, Y.; Fan, B.H. Sensitivity and fidelity of a novel piezoelectric middle ear transducer. ORL J. Otorhinolaryngol. Relat. Spec. 2009, 71, 216–220. [CrossRef] [PubMed] Qiu, Y.; Gigliotti, J.V.; Wallace, M.; Griggio, F.; Demore, C.E.M.; Cochran, S.; Trolier-McKinstry, S. Piezoelectric micromachined ultrasound transducer (PMUT) arrays for integrated sensing, actuation and imaging. Sensors 2015, 15, 8020–8041. [CrossRef] [PubMed] Wang, Z.; Xue, Q.-T.; Chen, Y.-Q.; Shu, Y.; Tian, H.; Yang, Y.; Xie, D.; Luo, J.-W.; Ren, T.-L. A flexible ultrasound transducer array with micro-machined bulk PZT. Sensors 2015, 15, 2538–2547. [CrossRef] [PubMed] Che, X.H.; Qiao, W.X.; Wang, R.J.; Zhao, Y.H. Numerical simulation of an acoustic field generated by a phased arc array in a fluid-filled cased borehole. Pet. Sci. 2014, 11, 385–390. [CrossRef] Che, X.H.; Qiao, W.X.; Ju, X.D.; Lu, J.Q.; Wu, J.P. An experimental study on azimuthal reception characteristics of acoustic well-logging transducers based on phased-arc arrays. Geophysics 2014, 79, D197–D204. [CrossRef] Lu, J.Q.; Ju, X.D.; Qiao, W.X.; Men, B.Y.; Wang, R.J.; Wu, J.P. Azimuthally acoustic logging tool to evaluate cementing quality. J. Geophys. Eng. 2014, 11, 045006. [CrossRef] Casula, O.; Toullelan, G.; Dumas, P. Inspection of complex components with flexible phased array transducer. J. Acoust. Soc. Am. 2008, 123, 3118. [CrossRef] Primakoff, H.; Foldy, L.L. A general theory of passive linear electroacoustic transducers and the electroacoustic reciprocity theorem. II. J. Acoust. Soc. Am. 1947, 50. [CrossRef] Foldy, L.L.; Primakoff, H. A general theory of passive linear electroacoustic transducers and the electroacoustic reciprocity theorem. I. J. Acoust. Soc. Am. 1945, 109. [CrossRef] Hoja, J.; Lentka, G. A family of new generation miniaturized impedance analyzers for technical object diagnostics. Metrol. Meas. Syst. 2013, 20, 43–52. [CrossRef] Lin, S.; Xu, J. Effect of the matching circuit on the electromechanical characteristics of sandwiched piezoelectric transducers. Sensors 2017, 17, 329. [CrossRef] [PubMed]

Sensors 2017, 17, 985

17. 18. 19.

9 of 9

Li, L.; Wang, L.; Qin, L. Simulation of 1-3 piezoelectric composite cylindrical transducer. Ferroelectrics 2014, 471, 17–22. [CrossRef] Wu, J.; Qiao, W.; Che, X.; Ju, X.; Lu, J.; Wu, W. Experimental study on the radiation characteristics of downhole acoustic phased combined arc array transmitter. Geophysics 2013, 78, D1–D9. [CrossRef] Zhang, K.; Ju, X.D.; Lu, J.Q.; Men, B.Y. Design of acoustic logging signal source of imitation based on field programmable gate array. J. Geophys. Eng. 2014, 11, 045008. [CrossRef] © 2017 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).

A Consistency Evaluation and Calibration Method for Piezoelectric Transmitters.

Array transducer and transducer combination technologies are evolving rapidly. While adapting transmitter combination technologies, the parameter cons...